ORTHOPAT ADVANCE PERIOPERATIVE AUTOTRANSFUSION SYSTEM
Applicant
Haemonetics Corp.
Product Code
CAC · Anesthesiology
Decision Date
Oct 25, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5830
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Haemonetics OrthoPAT advance (Orthopedic Perioperative Autotransfusion) System is indicated for use to salvage Red Blood Cells (RBCs) from blood lost intraoperatively and postoperatively during surgical procedures where the expected rate of processing of salvaged blood and fluid aspirated from the surgical site is less than or equal to 2 liters per hour. Autotransfusion is indicated for patients who meet at least one of the following criteria: • The patient is expected to lose sufficient blood in the period so as to require RBC transfusion, and autotransfusion will likely reduce or eliminate the need for allogeneic blood transfusion. • Religious beliefs cause the patient to refuse allogeneic transfusion, but accept autologous transfusion. • Compatible allogeneic blood is not available. • The patient is unable to donate sufficient quantities of autologous blood prior to surgery to adequately cover the anticipated transfusion requirement. • The patient or physician prefers perioperative autotransfusion rather than preoperative autologous donation or transfusion of allogeneic blood.
Device Story
OrthoPAT advance is an electro-mechanical perioperative autotransfusion system. It collects shed blood from surgical sites or wound drains via a disposable reservoir; processes blood using a Dynamic Disk® centrifuge to pack RBCs; washes cells to remove plasma, cell fragments, and irrigation fluid; and transfers processed RBCs to a bag for gravity reinfusion. Operated by clinical staff in surgical settings. The system includes a display unit for monitoring, a base unit with centrifuge and pneumatic controls, and disposable sets. New features include updated software for CCD camera calibration and detection, increased battery capacity (Lithium Ion), and modified reservoir materials. The system provides audible/visual alarms for process triggers (volume/time). Benefits include reduced reliance on allogeneic blood transfusion and support for patients with specific transfusion requirements.
Clinical Evidence
Bench testing only. Studies validated software UI updates, reservoir biocompatibility, package integrity/shelf life, electrical safety of the battery, and EMC compliance. No clinical data presented.
Technological Characteristics
Electro-mechanical autotransfusion system. Materials: Polycarbonate (SABIC HP4R, Lexan HP4NR), Polypropylene (TOTAL 3620 WZ), Polyester, ePTFE. Sensing: CCD camera for float tracking/fluid level, light bar, RBC sensor. Energy: AC power (100-240V) and Lithium Ion battery (8.4V 4600mAH). Connectivity: Standalone. Sterilization: EtO. Software: Embedded control system for pneumatics, centrifuge, and UI.
Indications for Use
Indicated for patients undergoing surgery with expected blood loss requiring RBC transfusion, including those refusing allogeneic blood, lacking compatible allogeneic blood, unable to donate autologous blood pre-surgery, or preferring autotransfusion. Applicable where processing rate is ≤ 2 liters/hour.
Regulatory Classification
Identification
An autotransfusion apparatus is a device used to collect and reinfuse the blood lost by a patient due to surgery or trauma.
Predicate Devices
Haemonetics OrthoPAT Perioperative Autotransfusion System (K992723)
Submission Summary (Full Text)
{0}------------------------------------------------
K122262;
### 510(k) Summary
Date: July 26, 2012
OCT 25 2012
Submitter: Haemonetics Corporation 400 Wood Road. Braintree MA 02184
Contact: Erica Diaz Regulatory Affairs Specialist Phone: 781-356-9798 Fax: 781-356-3558 Email: erica.gasca@haemonetics.com Alternate Contact: Greg Calder Regulatory Affairs Manager Phone: 781-356-9538 Fax: 781-356-3558 Email: gcalder@haemonetics.com
Device Information: Trade Name: Haemonetics OrthoPAT advance Perioperative Autotransfusion System Regulation Number: 21 CFR 868.5830 Regulation Name: Autotransfusion Apparatus Regulatory Class: Class II (two) Product Code: CAC
Predicate Device Information: Trade Name: Haemonetics OrthoPAT Perioperative Autotransfusion System Regulation Number: 21 CFR 868.5830 Regulation Name; Autotransfusion Apparatus Regulatory Class: Class II (two) Product Code: CAC
#### Device Characteristics Summary:
The OrthoPAT advance Perioperative Autotransfusion System is an evolution of the previously cleared OrthoPAT Perioperative Autotransfusion System. The OrthoPAT Perioperative Autotransfusion System was most recently cleared via 510(k) K992723 on October 18, 1999.
The OrthoPAT advance System is designed to salvage Red Blood Cells (RBCs) from blood lost intraoperatively and postoperatively during surgical procedures where the expected rate of processing of salvaged blood and fluid aspirated from the surgical site is less than or equal to two liters per hour.
Blood shed from the surgical site or wound drain is collected in the reservoir; processed in the Dynamic Disk® separation chamber to pack the RBCs; washed to remove the cell fragments, irrigation fluid, plasma, and other undesirable components that are found in the fluid portion of the shed blood; then transferred to a bag for gravity reinfusion to the patient.
The OrthoPAT advance System consists of the following three parts that work together to collect and process the RBCs lost during and after surgery:
- OrthoPAT advance device: the electro-mechanical device and display screen. .
- Disposable set: the single-use collection material including reservoir, processing set, A&A line, post-op . line, and vacuum line.
- Solutions: solutions for collecting and processing salvaged blood (examples: anticoagulant and wash . solutions).
The OrthoPAT advance device, together with the disposables to be used with the device, is the subject of this 510(k) submission.
{1}------------------------------------------------
## Non-Clinical Testing Summary:
ﻢ
The following non-clinical performance testing was submitted in support of a determination of substantial equivalence between the subject and predicate device. A summary of the performance testing is presented below in Table 1: Summary of Performance Studies. Test data demonstrates that the device met all clinical and performance requirements, and that the subject device is as safe, as effective, and performs as well as or better than the predicate device.
| Test Name | Test Report # | Test Intent | Test Result |
|----------------------------------------------------|----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Bench Testing for<br>Software | TR-SOF-100367<br>TR-SOF-100377<br>TR-SOF-100378 | The intent of these studies was<br>to validate an update to the<br>software user interface with<br>display of Estimated Blood Loss<br>(EBL) volume. | Passed |
| Biocompatibility<br>Testing for Reservoir | TR-BIO-100023-A | The intent of this study was to<br>demonstrate the biocompatibility<br>of the new reservoir, containing<br>materials different from the<br>predicate device. | Passed |
| Transportation &<br>Aging Testing for<br>Reservoir | TR-ACG-100094AA<br>TR-DIS-101222A<br>TR-DIS-101338 | The intent of these studies was<br>to demonstrate the package<br>integrity and device stability of<br>the new reservoir, up to a six<br>month shelf life. | Passed |
| Bench Testing for<br>Electrical Hardware | TR-ELE-100445 | The intent of this study was to<br>demonstrate the electrical safety<br>of the increased-capacity battery. | Passed |
| EMC | TR-ELE-100445A | The intent of this testing was to<br>demonstrate the EMC<br>compliance of the modified<br>device. | Passed |
#### Table 1: Summary of Performance Studies
{2}------------------------------------------------
### Comparison to Predicate Summary:
The OrthoPAT advance Perioperative Autotransfusion System is an evolution of the OrthoPAT Perioperative Autotransfusion System. The OrthoPAT Perioperative Autotransfusion System was most recently cleared via 510(k) K992723 on October 18, 1999. The OrthoPAT advance System is designed to perform the same types of procedures as the previously-cleared OrthoPAT system, utilizing identical disposable sets as well as additional disposables designed for the new system. The previously-cleared OrthoPAT System to the subject OrthoPAT advance System include a software user interface update, a change in reservoir material and orientation of the reservoir filter, and an increase in battery capacity.
A summary of the OrthoPAT advance System comparison to the predicate OrthoPAT System is presented in Table 2: Comparison of the OrthoPAT advance System to the Previously-Cleared Predicate OrthoPAT System.
| | Predicate OrthoPAT Device (K992723) | Subject OrthoPAT advance Device |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------|
| Indications for Use | The Haemonetics OrthoPAT<br>(Perioperative AutoTransfusion) System<br>is indicated for use to salvage Red Blood<br>Cells (RBCs) from blood lost<br>intraoperatively and postoperatively<br>during surgical procedures where the<br>expected rate of processing of salvaged<br>blood and fluid aspirated from the<br>surgical site is less than or equal to 2<br>liters per hour. Autotransfusion is<br>indicated for patients who meet at least<br>one of the following criteria:<br>• The patient is expected to lose<br>sufficient blood in the perioperative<br>period so as to require RBC transfusion,<br>and autotransfusion will likely reduce<br>or eliminate the need for allogeneic<br>blood transfusion.<br>• Religious beliefs cause the patient.to<br>refuse allogeneic transfusion, but accept<br>autologous transfusion.<br>• Compatible allogeneic blood is not<br>available.<br>• The patient is unable to donate<br>sufficient quantities of autologous<br>blood prior to surgery to adequately<br>cover the anticipated transfusion<br>requirement.<br>• The patient or physician prefers<br>perioperative autotransfusion rather<br>than preoperative autologous donation | Same |
| Table 2: Comparison of the OrthoPAT advance System to the Previously-Cleared Predicate OrthoPAT | | | | | | |
|-------------------------------------------------------------------------------------------------|----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------|---------------------------------|------|
| System | | | | | | |
| Table 2 (cont.): | | | | | | |
| | Hardware Device: | Predicate OrthoPAT Device (K992723) | Subject OrthoPAT advance Device | | | |
| Display unit | • Keypad | Contains seven membrane switches and corresponding LED indicator lights | Contains thirteen buttons that provide pressed status to the display processor. Seven of the buttons shall provide associated key lights that can be turned on and off by software The SELECT button shall provide three key lights associated with it that can be turned off by the Display Processor | | | |
| | • Display | LCD (480x640 pixel) | Same | | | |
| Base Unit | • Centrifuge | 0 to 5500 rpm<br>Pneumatic control of disposable disk diaphragm | Same | | | |
| | • Header Arm | The header arm holds the header portion of the disk stationary during operation | Same | | | |
| | • Red blood cell sensor | Light source and sensor to differentiate between air, waste, fluid, and red blood cells | Same | | | |
| | • Spill containment system | Spill containment well and spill containment ring | Spill containment well drain and spill collection bag | | | |
| Rotary Valves | • Inlet Valve | Controls the flow of fluid into the disk | Same | | | |
| | • Effluent Valve | Controls the flow of fluid out of the disk | Same | | | |
| | Pneumatic System | | | | | |
| | • Vacuum | 0 to -350mmHG<br>[0 to -39.99 kPA;<br>0 to -399.9 mbar] | Same | | | |
| | • Pressure | 0 to 250mmHG<br>[0 to 33 kPA;<br>0 to 333 mbar] | Same | | | |
| | • Pump | Maintains selected vacuum range | Same | | | |
| | Reservoir Optics System | | | | | |
| | • Light Bar | Emits light towards the light reflector located in the reservoir | Same | | | |
| | • CCD Camera | Tracks the movement of the float and monitors the fluid level of the reservoir | Same | | | |
| Power Supply | | | | | | |
| | • AC Power | AC power accommodates worldwide voltage and frequencies from 100V to 240V, 50/60Hz | Same | | | |
| | • Battery power | Nickel Metal hydride 7.2 volts and provides 1800mAH | Lithium Ion 8.4V 4600mAH | | | |
| Rear panel | | | Predicate OrthoPAT Device (K992723) | Provides mechanism for mounting an IV | Subject OrthoPAT advance Device | Same |
| Software | | | | | | |
| | Base Unit | Controls device subsystems such as<br>pneumatics, centrifuge, reservoir, etc. | New software for the CD camera<br>calibration and detection algorithm;<br>Added memory storage area for<br>calibration data, machine recovery<br>information data, and procedure<br>summary and event data | | | |
| | • Display Unit | Control LCD display, keypad, key lights,<br>and tone generator | Additional functionality utilizing a new<br>keypad to add AC volume, process<br>volume trigger, process time trigger,<br>interval timer reset and display toggling<br>between graphic and text messages;<br>A new audible alarm will be generated<br>and a visual text message will be<br>displayed when the reservoir volume<br>reaches the process volume trigger<br>volume or process time trigger set by<br>operator;<br>Additional icons will be generated and<br>displayed for use with the battery and<br>charging system | | | |
| Disposables: | | | | | | |
| | Integrated Processing Set | Collects blood in reservoir, processes<br>blood from collection reservoir and<br>separates into RBC's and waste | Designed to use Quickconnect collection<br>reservoir and processing set. | | | |
| • | Anticoagulant line | Delivers anticoagulant to shed blood | Component of Aspiration and<br>Anticoagulation line | | | |
| • | Anticoagulant<br>Roller Clamp | Allows the operator to adjust the flow rate<br>of the anticoagulant | Component of Aspiration and<br>Anticoagulation line | | | |
| • | Anticoagulant<br>Spike | Incorporates a drip chamber to allow the<br>operator to visually monitor the<br>anticoagulant flow rate | Component of Aspiration and<br>Anticoagulation line | | | |
| • | Anticoagulant Y<br>Adapter | Allows the anticoagulant to flow into the<br>aspiration line and mix with the blood as<br>it flows toward the reservoir | Component of Aspiration and<br>Anticoagulation line | | | |
| • | Aspiration<br>Connector | Connector to either the wound drain set or<br>the intraoperative suction set | Component of Aspiration and<br>Anticoagulation line | | | |
| • | Aspiration Line | Connects the suction tip or wound drain<br>to the reservoir | Same | | | |
| • | Effluent Stopcock | Diverts fluid from the rotor into either the<br>waste bag or the RBC bag | Same | | | |
| • | Inlet Stopcock | Diverts fluid into the rotor from either the<br>reservoir or the saline bag | Same | | | |
| • | Saline line | Runs between inlet stopcock and saline<br>spike | Same | | | |
| • | Separation<br>Chamber (rotor,<br>disk) | Spins in the centrifuge to separate the<br>shed blood and process the RBCs | Same | | | |
| | Aspiration and<br>Anticoagulant line<br>HAR-A-1003 | N/A | Connects suction tip to reservoir,<br>anticoagulates shed blood | | | |
| • | Anticoagulant | N/A | Allows the operator to adjust the flow | | | |
{3}------------------------------------------------
.
:
and the comments of the comments of the comments of
{4}------------------------------------------------
.
{5}------------------------------------------------
Table 2 (cont.):
・・
·
.
.
.
| | Predicate OrthoPAT Device (K992723) | Subject OrthoPAT advance Device | |
|------------------------|---------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| • | Anticoagulant<br>Spike | N/A | Incorporates a drip chamber to allow the<br>operator to visually monitor the<br>anticoagulant flow rate |
| • | Anticoagulant Y<br>Adapter | N/A | Allows the anticoagulant to flow into the<br>aspiration line and mix with the blood as<br>it flows toward the reservoir<br>Material: PVC |
| | Aspiration and<br>Anticoagulant line<br>1400T | Connects suction tip to reservoir,<br>anticoagulates shed blood | Same |
| • | Anticoagulant<br>Roller Clamp | Allows the operator to adjust the flow rate<br>of the anticoagulant | Same |
| • | Anticoagulant<br>Spike | Incorporates a drip chamber to allow the<br>operator to visually monitor the<br>anticoagulant flow rate | Same |
| • | Anticoagulant Y<br>Adapter | Allows the anticoagulant to flow into the<br>aspiration line and mix with the blood as<br>it flows toward the reservoir<br>Material: PVC | Same |
| | Post-op Suction set | Connects to wound drain tubing and<br>collects blood postoperatively from<br>wound drain | Same |
| | Reservoir | Serves four functions: Transmission of<br>vacuum, storage of salvaged blood,<br>filtration of salvaged blood, and fluid<br>volume measurement | Same |
| • | Body | Material: ABS Plastic | Material: Polycarbonate<br>SABIC HP4R |
| • | Cover | Material: ABS Plastic | Material: Polycarbonate<br>Lexan HP4NR with 8H9D266 White |
| • | Mesh filter | Removes clots and large debris from the<br>shed blood<br>Material: ABS frame frame/copolyester<br>mesh | Material: Polypropylene<br>Frame: TOTAL 3620 WZ<br>Screen: Polyester, Saati<br>PES 200/43 with Plasma Treatment |
| • | Hydrophobic<br>Filter | Material: PVDF sheet | Frame: HP4R<br>Filter: 1um ePTFE, CVL-HC1 with Pre-<br>Filter 4.3 um HOVOGLAS HC4011 |
| • | Grommet | Material: Silicone rubber | OrthoPAT advance does not use a<br>grommet |
| • | Float ball | Floats on the blood as it fills the reservoir,<br>monitored by reservoir light bar.<br>Material: Polypropylene GR2 | Material: Polypropylene<br>Plasma treated then pad printed with<br>MARIBU TPU 980 Black |
| • | Light reflector | Material: ABS with white clariant | Same<br>Material: ABS Terlux 2802 HD with<br>white clariant UN0001 |
| • | Tubing | Material: PVC | Same |
| • | Pre-filter | Capture and retain debris as fluid enters<br>the reservoir<br>Material: Thermally reticulated<br>polyester/polyurethane foam | OrthoPAT advance does not use a Pre-<br>Filter |
| | Predicate OrthoPAT Device<br>(K992723) | Subject OrthoPAT advance<br>Device | |
| • Straw | Pulls fluid from the bottom of the<br>reservoir<br>Material: PVC | Same | |
| • Reservoir Ports | Shorts line which end in a connector<br>Material: PVC | Same | |
| • Aspiration connector | Polypropylene | Component of A&A Line HAR-A-<br>1003 Anticoagulant Y Adapter | |
| • Relief Valve | Not a part of the system | Additional part assembled to cover<br>for connection to outlet line or<br>reservoir<br>Silicone<br>Vernay PN VL1001M12 | |
| Waste bag | Stores the waste fluid and saline<br>wash that are separated from the<br>concentrated RBCs<br>PVC | Same | |
| RBC bag | Stores the concentrated RBCs for<br>transfusion back to the patient<br>PVC | Same | |
| Sterilization | Gamma | EtO | |
| Shelf Life | | | |
| • Processing Set | | | |
| 1150H (Integrated) | 3 years | N/A - See below for OPT-P-1000 | |
| OPT-P-1000 | N/A | 3 years | |
| • HAR-A-10003 | N/A | 3 years | |
| • 1400T | 3 years | 3 years | |
| • Post-op Suction Set | 3 years | Same | |
| • Reservoir | N/A | 6 months | |
| • Waste Bag | 3 years | Same | |
| • RBC Bag | 3 years | Same | |
.
:
{6}------------------------------------------------
Table 2 (cont.):
, '
Erica Diaz
Regulatory Affairs Specialist
Haemonetics Corporation
7-26-12
Date
7
{7}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
OCT 25 2012
Haemonetics Corporation c/o Ms. Erica Diaz Regulatory Affairs Specialist 355 Wood Road Braintree, MA 20184
Re: K122262
OrthoPAT advance system Perioperative Autotransfusion System Regulation Number: 21 CFR 868.5830 Regulation Name: Autotransfusion Apparatus Regulatory Class: Class II Product Code: CAC Dated: July 26, 2012 Received: July 27, 2012
### Dear Ms. Diaz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{8}------------------------------------------------
Page 2 - Ms. Erica Diaz
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
{9}------------------------------------------------
## Indications for Use
510(k) Number (if known):
Device Name: Haemonetics OrthoPAT advance (Orthopedic Perioperative Autotransfusion) System
Indications for Use:
The Haemonetics OrthoPAT advance (Orthopedic Perioperative Autotransfusion) System is indicated for use to salvage Red Blood Cells (RBCs) from blood lost intraoperatively and postoperatively during surgical procedures where the expected rate of processing of salvaged blood and fluid aspirated from the surgical site is less than or equal to 2 liters per hour. Autotransfusion is indicated for patients who meet at least one of the following criteria:
- The patient is expected to lose sufficient blood in the period so as to require RBC ● transfusion, and autotransfusion will likely reduce or eliminate the need for allogeneic blood transfusion.
- Religious beliefs cause the patient to refuse allogeneic transfusion, but accept autologous . transfusion.
- . Compatible allogeneic blood is not available.
- The patient is unable to donate sufficient quantities of autologous blood prior to surgery to . adequately cover the anticipated transfusion requirement.
- The patient or physician prefers perioperative autotransfusion rather than preoperative . autologous donation or transfusion of allogeneic blood.
Prescription Use _ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--------------------------------------------------------|-----------------|
| | Page _1_ of _1_ |
| (Division Sign-Off) | |
| Division of Cardiovascular Devices | |
| 510(k) Number | K122 262 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.