THEYFIT MALE CONDOM

K122219 · Theyfit · HIS · Sep 27, 2013 · Obstetrics/Gynecology

Device Facts

Record IDK122219
Device NameTHEYFIT MALE CONDOM
ApplicantTheyfit
Product CodeHIS · Obstetrics/Gynecology
Decision DateSep 27, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.5300
Device ClassClass 2
AttributesReal-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K122219 · Sep 27, 2013THEYFIT MALE CONDOMTheyfitPublished clinical study (Reece M, Herbenick D, Sanders SA et al., Sex Transm Infect 2008; 84(2):143-149)The sponsor utilized a published clinical study to provide evidence of clinical performance, specifically regarding breakage, slippage, and acceptability outcomes for the device sizes.Clinical performance; Breakage rates; Slippage rates; Acceptability

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Breakage, slippage and acceptability outcomes of a condom fitted to penile dimensions; Published clinical studyUsers of fitted condomsNot applicable for this studyBreakage, slippage, and acceptability

Indications for Use

The TheyFit Male Condom is used for contraceptive and for prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted infections).

Device Story

TheyFit Male Condom is a natural rubber latex sheath designed to cover the penis; features reservoir end; cylindrical shape. Available in 22 distinct size combinations of length and width. Provided pre-lubricated with silicone-based lubricant; non-spermicidal. Users employ a fitting kit (FitKit) to select appropriate size. Intended for over-the-counter use to prevent pregnancy and sexually transmitted infections.

Clinical Evidence

Clinical performance data for sizes O77-Z22 provided via published study (Reece M, et al. Sex Transm Infect 2008; 84(2):143-149) evaluating breakage, slippage, and acceptability outcomes. Bench testing performed per ASTM D3492-08 and ISO 4074:2002, including airburst, water leak, freedom from holes, and dimensional analysis; all acceptance criteria met.

Technological Characteristics

Natural rubber latex sheath; pre-lubricated with silicone-based lubricant. Available in 22 size variations (lengths 163-208mm; widths 49-64mm). Testing conducted per ASTM D3492-08 and ISO 4074:2002 standards.

Indications for Use

Indicated for individuals seeking contraception and protection against sexually transmitted infections.

Regulatory Classification

Identification

A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.

Special Controls

*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K122219 # 510(k) Summarv SEP 2 7 2013 Image /page/0/Picture/3 description: The image shows the logo for TheyFit condoms. The logo is in black and white, with the word "TheyFit" in a bold, sans-serif font. The "I" in "Fit" is replaced by a condom shape. There is a registered trademark symbol to the right of the "T". This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. | DATE: | September 24, 2013 | |--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | APPLICANT: | TheyFit<br>Michael Cecil, MD<br>4140 Tate Street<br>Covington, GA 30014<br>Tel: 404-441-4435<br>Email: mpcecil@earthlink.net | | OFFICIAL CORRESPONDENT: | Penny Northcutt, RAC, FRAPS, CQA<br>Regulatory Consultant for TheyFit<br>REGSolutions, LLC<br>Tel: 678-428-6978<br>Fax: 678-513-0937<br>Email: pennynorthcutt@theregsolutions.com | | TRADE NAME: | TheyFit Male Condom | | CLASSIFICATION NAME: | Condom | | COMMON OR USUAL NAME: | Condom | | DEVICE CLASSIFICATION<br>AND PRODUCT CODE: | Class II per 21 CFR §884.5300<br>Obstetrics/Gynecology<br>Product Code: HIS | | PREDICATE DEVICE NAME: | Karex Large and Extra Large Natural Rubber Latex<br>Condoms, K113061 | ## DESCRIPTION OF THE DEVICE: The TheyFit Male Condom is made of a natural rubber latex sheath, which completely covers the penis with a closely fitted membrane. The condom is shaped with a reservoir end and is flat with a cylindrical shape. TheyFit condoms are available in 22 sizes, of different length/width combinations. TheyFit Condoms are provided pre-lubricated with a silicone-based lubricant. TheyFit condoms are not provided with spermicide. The user must use a fitting kit (FitKit) to select the appropriate size TheyFit Condom. · The sizes available are as follows: {1}------------------------------------------------ # 510(k) Summary | TheyFit Male Condom | | | |---------------------|-----|-----| | Model Number Chart | | | | O77 | N17 | Z22 | | O88 | N21 | Z21 | | O99 | N22 | | | O11 | N77 | | | O17 | D11 | | | O21 | D17 | | | O22 | D21 | | | N88 | D22 | | | N99 | Z11 | | | N11 | Z17 | | #### Legend | Condom Length | Condom Lay Flat Width | |---------------|-----------------------| | O = 163mm | 77 = 49mm | | N = 178mm | 88 = 51mm | | D = 193mm | 99 = 53mm | | Z = 208mm | 11 = 55mm | | | 17 = 57mm | | | 21 = 60mm | | | 22 = 64mm | ## INTENDED USE/INDICATIONS FOR USE: The subject and predicate device have the same Indications for Use statement, which is, "The TheyFit Male Condom is used for contraceptive and for prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted infections)." ## TECHNOLOGICAL CHARACTERISTICS: The subject and predicate devices have different technological characteristics. The different technological characteristics of the subject device include its available size range compared to the predicate device and the inclusion of a fitting kit to help the user select his condom size. These different characteristics of the subject device could affect safety and effectiveness, (e.g., clinical slippage and breakage rates and selection of an appropriate condom size). However, the different technological characteristics of the subject device do not raise new types of safety and effectiveness questions because FDA has cleared condoms with different characteristics than previously cleared male condoms that could affect condom failure rates, including new condom material, size, and performance claims (e.g., extra strength). {2}------------------------------------------------ # 510(k) Summary ## PERFORMANCE DATA: | Description | Summary and Conclusion | |----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Biocompatibility | The TheyFit Condoms are identical with regard to<br>material composition and manufacturing process as the<br>predicate device. Therefore, biocompatibility testing<br>completed on the predicate device was leveraged to<br>support the biocompatibility of the TheyFit Condoms. | | Airburst | ASTM D3492-08 and ISO 4074:2002<br>Acceptance criteria met | | Water Leak | ASTM D3492-08 and ISO 4074:2002<br>Acceptance criteria met | | Freedom from Holes | ASTM D3492-08 and ISO 4074:2002<br>Acceptance criteria met | | Dimensional Analysis | ASTM D3492-08 and ISO 4074:2002<br>Acceptance criteria met | | Clinical Performance | Clinical performance data was provided for sizes 077-<br>Z22 via a published clinical study (Reece M, Herbenick D,<br>Sanders SA et al, Breakage, slippage and acceptability<br>outcomes of a condom fitted to penile dimensions. Sex<br>Transm Infect 2008; 84(2):143-149.) | ## CONCLUSION: The TheyFit Male Condoms (sizes O77-Z22) are substantially equivalent to their proposed predicate devices. . {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines. Food and Drug Administration 0903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 September 27, 2013 TheyFit % Penny Northcutt, RAC, FRAPS, COA Executive Director REGSolutions, LLC 717 Lakeglen Drive Suwanee, GA 30024 Re: K122219 > Trade/Device Name: TheyFit Male Condom (sizes 077-222) Regulation Number: 21 CFR 884.5300 Regulation Name: Condom Regulatory Class: Class II Product Code: HIS Dated: September 19, 2013 Received: September 20, 2013 Dear Penny Northcutt: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may, therefore, manns of the Act include requirements for annual registration, listing of general oonline profice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {4}------------------------------------------------ Page 2 - Penny Northcutt device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, # Benjamin R. Fisher -S Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation . Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ 510(k) Number (if known): K122219 #### Device Name: TheyFit Male Condom Indications For Use: - The TheyFit male condom is used for contraceptive and for prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted infections). | TheyFit Male Condom<br>Model Number Chart | | | |-------------------------------------------|-----|-----| | O77 | N17 | Z22 | | O88 | N21 | Z21 | | O99 | N22 | | | O11 | N77 | | | O17 | D11 | | | O21 | D17 | | | O22 | D21 | | | N88 | D22 | | | N99 | Z11 | | | N11 | Z17 | | #### Legend | Condom Length | Condom Lay Flat Width | |---------------|-----------------------| | O = 165mm | 77 = 49mm | | N = 180mm | 88 = 51mm | | D = 195mm | 99 = 53mm | | Z = 210mm | 11 = 55mm | | 17 = 57mm | | | 21 = 60mm | | | 22 = 64mm | | AND/OR Prescription Use _ --- (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Benjamin R. Fisher -5 2013.09.27 13:29:57 -04'00' Page 1 of 1 Traditional 510(k) TheyFit TheyFit Male Condoms
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...