← Product Code [PCE](/productcode/PCE) · K122206

# GARAVENTA X3 INCLINED PLATFORM LIFT (K122206)

_Garaventa (Canada) Ltd [Dba Garaventa Lift] · PCE · Mar 21, 2013 · Physical Medicine · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K122206

## Device Facts

- **Applicant:** Garaventa (Canada) Ltd [Dba Garaventa Lift]
- **Product Code:** [PCE](/productcode/PCE.md)
- **Decision Date:** Mar 21, 2013
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 890.3930
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine
- **Attributes:** Therapeutic

## Indications for Use

The Garaventa X3 inclined platform lift is intended to mechanically transport one person in a wheelchair or in a fold-down seat up and down stairs in a private or public facility either indoors or outdoors.

## Device Story

Garaventa X3 is a motorized inclined platform lift for transporting individuals in wheelchairs or fold-down seats on straight stairways. System uses extruded aluminum guide rails; upper rail houses drive rack and control wiring; lower rail provides stabilization. On-board rack and pinion drive system is powered by 24V DC batteries. Device is computer-controlled via an on-board microprocessor. Wireless call stations at each landing communicate via 2.4 GHz ISM band to send/call the unit or fold/unfold the platform. Safety network monitors switches, landing cams, battery voltage, and motor current independently of the microprocessor to ensure safe stops. Operated by the user or caregiver at landings or on the platform. Provides mobility assistance for disabled persons in public or private settings.

## Clinical Evidence

No clinical data. Bench testing only. Functionality and safety verified through design-phase testing in accordance with ASME 18.1-2008 and CSA B335-09 standards.

## Technological Characteristics

Extruded aluminum guide rails; rack and pinion drive; 24V DC battery power; wireless 2.4 GHz ISM band communication; microprocessor-based control system; electromechanical safety monitoring network; indoor/outdoor use; straight stairway configuration.

## Regulatory Identification

A permanently mounted wheelchair platform lift is a motorized vertical or inclined platform lift device permanently installed in one location that is intended for use in mitigating mobility impairment caused by injury or other disease by providing a guided platform to move a person from one level to another, with or without a wheelchair. A portable wheelchair elevator is a motorized lift device that is not permanently mounted in one location and that is intended for use in mitigating mobility impairment caused by injury or other disease by providing a means to move a person, with or without a wheelchair, from one level to another (e.g., portable platform lifts, attendant-operated stair climbing devices for wheelchairs).

## Special Controls

(i) Appropriate analysis and nonclinical testing (such as that outlined in the currently FDA-recognized edition of ASME A18.1 “Safety Standard for Platform Lifts and Stairway Chair Lifts”) must demonstrate that the safety controls are adequate to prevent a free fall of the platform in the event of a device failure; 
(ii) Appropriate analysis and nonclinical testing (such as that outlined in the currently FDA-recognized edition of ASME A18.1 “Safety Standard for Platform Lifts and Stairway Chair Lifts”) must demonstrate the ability of the device to withstand the rated load with an appropriate factor of safety; 
(iii) Appropriate analysis and nonclinical testing (such as that outlined in the currently FDA-recognized edition of ASME A18.1 “Safety Standard for Platform Lifts and Stairway Chair Lifts”) must demonstrate the ability of the enclosures to prevent the user from falling from the device; and 
(iv) Appropriate analysis and nonclinical testing (such as that outlined in the currently FDA-recognized editions of AAMI/ANSI/IEC 60601-1-2, “Medical Electrical Equipment—Part 1-2: General Requirements for Safety—Collateral Standard: Electromagnetic Compatibility—Requirements and Tests,” and ASME A18.1 “Safety Standard for Platform Lifts and Stairway Chair Lifts”) must validate electromagnetic compatibility and electrical safety.

## Predicate Devices

- STAIR-LIFT (Models GSL-1, GSL-2 & GSL-3) ([K981486](/device/K981486.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

### 510k Summary

As required by 21 CFR 807.92 (c)

# MAR 2 1 2013

1- Date summary prepared: March 19, 2013

#### 2 - Owner/submitter/applicant/sponsor information:

GARAVENTA (CANADA) LTD 7505 134A Street, Surrey BC, Canada V3W 7B3 1-604-594-0422 / 1-800-663-6556 Fax 1-604-594-9915

#### 3- Submission official correspondent:

Jay Mansour Mansour Consulting LLC 845 Aronson Lake Court. Roswell, GA 30075 USA 678-908-8180 Fax 678-623-3765

#### 4- Device information:

Common/usual/classification name: ELEVATOR, WHEELCHAIR

Device name: Garaventa X3 inclined platform lift.

Device classification:

| FDA 3 letter code              | PCE               |
|--------------------------------|-------------------|
| FDA regulation number: 21 CFR: | 890.3930          |
| Regulation medical specialty   | Physical Medicine |
| Review panel                   | Physical Medicine |
| Class                          | 2                 |

- 5- Substantial equivalency is claimed against the following predicate device(s):

| 510k number | Trade or Proprietary or Model Name       | Manufacturer           |
|-------------|------------------------------------------|------------------------|
| K981486     | STAIR-LIFT (Models GSL-1, GSL-2 & GSL-3) | GARAVENTA (CANADA) LTD |

#### 6- Description of the device:

The Garaventa X3 inclined platform lift is intended to mechanically transport one person in a wheelchair or in a fold down seat equipped with a seat belt, up and down stairs in a private or public facility either indoors or outdoors, on straight stairways.

Page F-1

{1}------------------------------------------------

The Garaventa X3 inclined platform lift uses two custom-cut extruded aluminum guide rails to guide and support the platform. An upper rail houses the drive rack that is engaged by the pinion gear providing platform movement as well as acting as the raceway for the control wiring. A lower rail stabilizes the platform keeping it in a level traveling position. The Garaventa X3 inclined platform lift has steel guide rails and haul ropes or cables are used for those products instead of extruded aluminum.

The Garaventa X3 inclined platform lift does not require a special cabinet to house the drive system. The drive system of the Garaventa X3 inclined platform lift is an on-board rack and pinion drive system. The drive system is protected from accidental contact by the user by a conveyance "hard" cover.

The Garaventa X3 inclined platform lift is a computer-controlled device equipped with a microprocessor and the software resides inside the conveyance controller.

The safety network monitors several switches, landing cam, the battery voltage and motor current limits. These are independent of the microprocessor system and provide a positive electromechanical means to reliably and safely stop the lift in any possible hazardous situation.

The controller is located in the conveyance unit. It receives the operating commands from the call station or from the conveyance controller. The conveyance control module always has a priority over the call stations.

The conveyance controller is powered by 24V DC batteries that are attached to the conveyance. The conveyance controller also controls the "barrier arms" motor, platform side-load motor and the platform motor in either of the landings.

The status of the safety switches and the emergency stop button are monitored by the conveyance controller.

There is a call station located at each landing. The call station transmits commands using a wireless simplex channel while the conveyance continuously scans all RF signals (decoded) from both call stations.

Call stations allow the user to call and send the unit into the desired landing and also fold and unfold the platform. The call stations are linked with the conveyance through the wireless link. The system uses a 2.4 GHz ISM band using encoded pulse for each call station does not process the button / command but simply transfers the state of each button to the conveyance.

The Garaventa X3 inclined platform lift is provided in four models: one fully automatic, two semi automatic, and one fully manual.

#### 7- Intended use:

The Garaventa X3 inclined platform lift is a motorized lift device intended for medical purposes to provide a means for a disabled person to move a wheelchair from one level to another

8- Indications for use:

Page F-2

{2}------------------------------------------------

The Garaventa X3 inclined platform lift is indicated to mechanically transport one person in a wheelchair or in a fold-down seat up and down stairs in a private or public facility either indoors or outdoors.

### 9- Basis for a determination of substantial equivalency:

- (a) Indications for use:
The indications for use are the same as the predicate device(s).

- (b) Technological characteristics:
The main differences between this device and predicate device are:

- i. X3 uses battery only in the control system.
- ii. X3 uses remote control call station using specified frequency.
- The table below reflects how the change in design is controlled and tested against, providing the basis of the determination of substantial equivalency in a summarized format.

| #  | Subject  | Main design            | Reason of change    | Test conducted                                                                  |
|----|----------|------------------------|---------------------|---------------------------------------------------------------------------------|
|    |          | item/feature change    |                     |                                                                                 |
| 1  | Rail     | X3 Uses the Rail       | Ease in production  | Use existing GSL 2 Rail System for straight lifts applications. Tested with "X3 |
|    | System   | System of GSL2, not    | and installation of | Safety Circuit Test" attached in this report in tab J Item 26 to comply with    |
|    |          | the tube system of     | the product.        | ASME 18.1-2008 and CSA B335-09.                                                 |
|    |          | GSL 1/3.               |                     |                                                                                 |
| 2  | Hanger   | X3 uses the Hanger of  | Ease in production  | Use the existing hanger of GSL 1 & 3 for complete control system. Safety        |
|    | System   | GSL 1/3, not the bulky | and installation of | tests done and documented in "X3 Safety Circuit Test" attached in this          |
|    |          | Hanger of GSL 2.       | the product.        | report in Tab J Item 26 to comply with ASME 18.1-2008 and CSA B335-09.          |
| 3  | Platform | X3 uses the Platform   | Align control       | Use the existing platform of GSL 1 & 3 for conveyance system. Safety tests      |
|    | System   | of GSL 1/3, not the    | system with the     | done and documented in "X3 Safety Circuit Test" attached in this report in      |
|    |          | GSL 2 Platform.        | hanger system       | Tab J Item 26 to comply with ASME 18.1-2008 and CSA B335-09.                    |
| 4  | Call     | Remote Control Call    | Ease in control and | Remote control Call station designed for wireless operation to improve the      |
|    | Station  | Station with specified | portability of call | wired operations of GSL 2 and GSL 1 & 3 for portability and ease of             |
|    |          | frequency.             | station. Cost-      | operation. Allocated band frequency has no conflict with outside radio          |
|    |          |                        | effective.          | frequencies. Design compliance to ASME 18.1 - 2008 and CSA 8 335-09.            |
| ട് | Power    | The whole control      | Cost-effectiveness  | Instead of using the Main AC power of 120 V, X3 uses battery only to            |
|    | Source   | system is by battery   | and ease in         | minimize power consumption and ease in installations. Design compliance         |
|    |          | only.                  | installation.       | to ASME 18.1 - 2008 and CSA B 335-09.                                           |

- (c) Non clinical tests- brief discussion: Tests on functionality and safeties has been done in the design phase in accordance to applicable standards. See Tab J3 Item 26 : Test Report X3 Safety Circuit.
- (d) Clinical tests- brief discussion: Not applicable

Page F-3

{3}------------------------------------------------

- (e) Non clinical and clinical tests- conclusions drawn demonstrating that the device is as safe and as effective, and performs as well as or better than the predicate device(s): As shown in the safety tests and functionality of the equipment, X3 performs effectively as the predicate devices. , .
:

the submit the state of the subject of the states of the states

:

.

.

:

.

.

-

and and the comments of the country

.

:

:

.

{4}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Image /page/4/Picture/2 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. In the center of the seal is a stylized caduceus, a symbol often associated with medicine and healthcare. The caduceus consists of a staff with two snakes coiled around it, topped by a pair of wings.

March 21, 2013

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Garaventa (Canada) Ltd. % Jay Mansour, MSQA, BE, RAC, LA Mansour Consulting 845 Aronson Lake Court Roswell, GA 30075

Re: K122206

Trade/Device Name: Garaventa X3 Inclined Platform Lift Regulation Number: 21 CFR 890.3930 Regulation Name: Wheelchair elevator Regulatory Class: Class II Product Code: PCE Dated: February 11, 2013 Received: February 22, 2013

Dear Mr. Jay Mansour:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

We note that your device exceeded the Limitations of exemptions from section 510(k) of the Federal Food, Drug, and Cosmetic Act (21 CFR Part 890.9), and therefore required the submission and clearance of a premarket notification prior to commercial distribution in the United States. Future devices of this same type, that meet the exemption criteria and do not exceed the limitations of exemptions found in 21 CFR Part 890.9, will be exempt from the premarket notification requirements of the Act.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 – Dear Mr. Jay Mansour

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Victor Krauthamer -S

Victor Krauthamer, Ph.D. Acting Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

## Indications for Use

510(k) Number (if known): K122206

Device Name: Garaventa X3 inclined platform lift

Indications For Use:

The Garaventa X3 inclined platform lift is intended to mechanically transport one person in a wheelchair or in a fold-down seat up and down stairs in a private or public facility either indoors or outdoors.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Brian D. Pullin - S

(Division Sign-off) Division of Neurological and Physical Medicine Device 510(k) Number K122206

Page 1 of

---

**Source:** [https://fda-staging.innolitics.com/device/K122206](https://fda-staging.innolitics.com/device/K122206)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
