K122154 · Fig, LLC · IRP · Nov 21, 2012 · Physical Medicine
Device Facts
Record ID
K122154
Device Name
POWERPLAY MUSCLE MASSAGER
Applicant
Fig, LLC
Product Code
IRP · Physical Medicine
Decision Date
Nov 21, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5650
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PowerPlay model PPRT-01 is intended to be an over-the-counter portable inflatable tube massage system which simulates kneading and stroking of tissue with the hands by use of inflatable pressure wraps. This device can be used to: - Temporarily increase blood circulation in the treated areas; - Temporary relief of minor muscle aches and pains.
Device Story
PowerPlay Muscle Massager is a portable, battery-operated OTC device for muscle pain relief and circulation enhancement. System uses an electronically controlled air pump and solenoid valves to inflate single-bladder compression wraps (calf, ankle, knee, hip, or shoulder). User selects up to 3 output ports and pressure levels (30, 50, or 70 mmHg). Device operates via microprocessor-controlled cycles: inflation to preset pressure, followed by a 30-second rest period for deflation. If multiple ports are active, system sequences inflation to ensure only one wrap inflates at a time. Treatment is preset to 20 minutes per output. Device includes LED display for pressure monitoring and safety alarms (high/low pressure, low battery, overpressure safety vent). Operated by the user in home or clinic settings. Benefits include simulated manual kneading/stroking to alleviate minor aches and improve circulation.
Clinical Evidence
No clinical testing performed. Substantial equivalence supported by non-clinical bench testing, including electrical safety, EMC, mechanical integrity, environmental, and life cycle testing. Bench testing verified equivalent pressure delivery and bladder fill times compared to predicate devices.
Technological Characteristics
Powered inflatable tube massager. Materials: PVC air bladders, non-woven medical fabric (Polyester/Softesse). Energy: Rechargeable battery or AC adapter. Connectivity: Standalone unit with quick-disconnect air ports. Control: Microprocessor-based with tactile switches and LED display. Safety: Integrated pressure transducer, internal safety vent (release at 2 psi/100 mmHg). Dimensions: Portable case. Sterilization: Supplied clean, non-sterile.
Indications for Use
Indicated for individuals seeking temporary relief of minor muscle aches and pains and temporary increase in blood circulation. Contraindicated for persons with suspected/active/untreated neuropathy, DVT, ischemic vascular disease, severe arteriosclerosis, pulmonary edema, congestive heart failure, thrombophlebitis, or active infection; or on legs with vein ligation, gangrene, dermatitis, open wounds, recent skin grafts, massive edema, extreme deformity, or extremities insensitive to pain.
Regulatory Classification
Identification
A powered inflatable tube massager is a powered device intended for medical purposes, such as to relieve minor muscle aches and pains and to increase circulation. It simulates kneading and stroking of tissues with the hands by use of an inflatable pressure cuff.
Predicate Devices
King Tai Holdings, Ltd., Verseo Leg Massager (K092435)
Nihon Seimitsu Sokki Co., Ltd., Portable Air Massager (K071596)
Salton, Inc., Relaxor Perfect Touch Air Massaging System (K030437)
Submission Summary (Full Text)
{0}------------------------------------------------
# Fig, LLC
1638 South Main Street Tulsa, OK 74119 Telephone: 800-994-0464
Section 5: 510(k) Summary
Date Prepared: July 18, 2012
510(k) Number: K 22154
# Sponsor
.
- Fig, LLC 1638 South Main Street Tulsa, OK 74119 Phone: 918-556-6939 Fax: 918-747-1078
#### Contact Person
Joe Adkins, Senior Project Manager Phone: 216-849-7847 Email: safeguardmd@aol.com
Device Name
----
Trade Name: PowerPlay Model Number: PPRT-01 Common or Usual Name: Inflatable Leg Massager Proprietary Name: PowerPlay Muscle Massager Classification Name: Powered Inflatable Tube Massager Class II Product Code: IRP Regulation Number: 21 CFR 890.5650
> Model PPRT-01 Powered Inflatable Tube Massager 510(k) Submission for Fig, LLC July 18, 2012
l 1
{1}------------------------------------------------
### Identification of Predicate Devices
| King Tai Holdings, Ltd., Verseo Leg Massager | K092435 |
|----------------------------------------------------------|---------|
| Nihon Seimitsu Sokki Co., Ltd., Portable Air Massager | K071596 |
| Salton, Inc., Relaxor Perfect Touch Air Massaging System | K030437 |
#### Device Description
The PowerPlay Muscle Massager is a lightweight, portable, rechargeable battery powered OTC device that is intended to be used by individuals to relieve minor muscular aches and pains and to temporarily increase blood circulation in the treated arcas.
The system utilizes pneumatically controlled, single bladder compression wraps actuated by an electronically controlled air pump unit and solenoid valves. The intermittent compression provided by the system simulates stroking and kneading of the hands. All pump, battery and control components are protectively housed in a plastic case. A mylar control panel overlays 4 tactile touch control switches, an LED display for monitoring pressure, LED "selected channel", "low battery" and "charging" indicators. There is also a port for connecting the battery charger/AC adapter plug.
The compression wraps consist of a Polyvinyl Chloride (PVC) air bladder encased inside soft, non-woven medical fabric (made from Polyester or Softesse - a fabric composed of common kitchen sponge material), which is adhered to the PVC air bladder. The wraps are supplied clean, non-sterile, packaged individually. A calf, ankle, knee, hip or shoulder massage wrap may be supplied with the PPRT -01 system or available as accessory items.
In operation, the user selects any number of 3 different outputs (Port 1, Port 2 and/or Port 3) and sets the pressure desired for each port (30, 50 or 70 mmHg). An LED corresponding to each output indicates the selected ports, with the digital display indicating the pressure setting for each (see instructions for use). User determined compression wraps containing air bladders are connected to the unit via externally accessible plastic quick-disconnect air ports. The control unit then inflates the wraps to the preset pressure. The LED that corresponds to the activated output channel flashes slowly while that wrap is being filled with air. Wrap pressure is monitored by an integrated pressure transducer and system software. Once the pressure reaches the proper level, the pump is turned OFF for a predetermined "rest" period, and the wrap deflates to ambient pressure through a valve inside the plastic case. After the "rest" period, the next wrap is sequenced, and so on. This cycle repeats for 20 minutes on each output (preset treatment period) and then that output turns off. When all selected outputs have completed their 20 minute treatment period the unit powers off. The unit power can be turned off at any time by pressing and holding the I/O button for approximately 2 seconds.
The "rest" period is internally preset to allow each wrap approximately 30 seconds between inflations in order to ensure adequate deflation time between compressions. When multiple outputs are selected, only one wrap is inflated at a time.
> Model PPRT-01 Powered Inflatable Tube Massager 510(k) Submission for Fig, LLC July 18, 2012
{2}------------------------------------------------
# Intended Use
The PowerPlay model PPRT-01 is intended to be an over-the-counter portable inflatable tube massage system which simulates kneading and stroking of tissue with the hands by use of inflatable pressure wraps. This device can be used to:
- . Temporarily increase blood circulation in the treated areas;
- . Temporary relief of minor muscle aches and pains.
#### Contraindications
The PowerPlay model PPRT-01 must not be used by persons with the following conditions:
- Suspected, active or untreated: Neuropathy, deep vein thrombosis, ischemic vascular . disease, severe arteriosclerosis, pulmonary edema, congestive heart failure, thrombophlebitis or an active infection;
- . On a leg where wraps would interfere with the following conditions: vein ligation, gangrene, dermatitis, open wounds, a recent skin graft, massive edema or extreme deformity of the leg;
- On extremities that are insensitive to pain; .
- Where increased circulation is undesirable; ◆
# Comparison of Indications
The Indications for Use for the PowerPlay Muscle Massager are the same as those for the Verseo Leg Massager (K092435), Portable Air Massager (K071596) and Relaxor Perfect Touch (K030437) predicate devices listed.
#### Substantial Equivalence of Technological Characteristics
· The PowerPlay Muscle Massager is equivalent to the predicate devices listed in operating principals to achieve identical results. That is, to provide massage by inflating and deflating massage garments wrapped around a part of the body. All systems, including the subject device, utilize microprocessor controlled pumps to deliver pressurized air to bladders that are attached to the user, using a cycle times of approximately 30 - 60 seconds. Each cycle consists of inflation of a bladder followed by a rest period during which the bladder deflates and the treated area relaxes without any compression.
The PowerPlay system uses similar means for pressure delivery to the wraps as the predicate devices. Pressurized air is delivered by the pump to the wraps. Although the Verseo unit (K092435) and the Portable Air Massager unit (K071596) incorporate the pump and wrap
> Model PPRT-01 Powered Inflatable Tube Massager 510(k) Submission for Fig, LLC July 18, 2012
13
{3}------------------------------------------------
into a single assembly, the PowerPlay delivers air to the wrap via flexible, plastic air tubes connected to the plastic pump / control unit by means of locking, quick disconnect couplings (like the Relaxor Perfect Touch (K030437). As with the Relaxor Perfect Touch, the PowerPlay Muscle Massager can be used with a variety of available compression wraps (i.e., shoulder, leg, etc.). All the predicate wraps, as well as those available for use with the PowerPlay system, are comprised of single bladder PVC chambers encased in a covering of soft, non-latex, non-woven medical fabric for increased patient comfort and biocompatibility compliance.
A difference in the subject device from all the predicates is that the PowerPlay unit has provisions for up to 3 separate massage wraps to be used at the same time. If more than one wrap is being used, the additional outputs are cycled during the rest period of the first output, keeping a constant cycle time of approximately 30 seconds for each. Each output port is independently limited to a 20 minute treatment time.
Cycle and treatment session times are factory preset and cannot be changed (predicate devices vary cycle and session times with selection of preset programs, except that the Perfect Touch system has a preset 15 minute treatment session time). Pressure settings on the predicate devices are adjusted automatically (dependent on the selected program) at levels between 45 and 320 mmHg, whereas the PowerPlay pressure is selectable to one of 3 preset levels (30, 50 and 70 mmHg).
Unlike the predicate systems, the microprocessor and pump unit is powered by internal rechargeable batteries, and can be connected to the main AC power line (through the battery charger / AC adaptor) while in use, allowing uninterrupted prolonged service. The Relaxor Perfect Touch must be connected to an AC adaptor during use, while the Verseo and Portable Air Massager units are powered from commonly available disposable consumer batteries.
A subtle difference between the predicate devices and the PowerPlay is in their respective alarm functions. Although all units are equipped with safety alarms, their functions differ somewhat. The subject device includes multiple and visual alarms including: High and low pressure alarms, low battery alarm and system malfunction overpressure safety via an internal safety vent with a release pressure of 2 psi (approximately 100 mmHg).
The management of FIG, LLC has reviewed these differences between the PowerPlay and those predicates cited and concluded that they pose no safety or efficacy concerns.
A detailed comparison table can be found in Section 10, Executive Summary, of this submission,
#### Non-Clinical Testing
Non-clinical validation, including electrical safety, EMC, mechanical integrity, environmental and life cycle testing have shown that the PowerPlay Muscle Massager has performance characteristics substantially equivalent to or surpassing those of the listed predicate devices. In-house bench testing has verified equivalent pressure delivery and wrap (bladder) fill time to the low pressure settings of the Relaxor Perfect Touch and Nihon Portable Air Massager, with overall system operation substantially equivalent to all predicate devices listed.
> Model PPRT-01 Powered Inflatable Tube Massager 510(k) Submission for Fig, LLC July 18, 2012
{4}------------------------------------------------
An informal focus group study was conducted by the sponsor in which the proposed Instructions for Use, a PowerPlay unit and various available site wraps were supplied to the participants for the purposes of determining the adequacy of the IFU. Most respondents reported the ability to use the device without referencing the IFU, with none of the participants reporting any difficulty in use with the aid of the proposed Instructions for Use.
# Clinical Testing
No formal clinical testing was performed on the PowerPlay Muscle Massager.
#### Summary Conclusion
Per the requirements of 21 CFR 807, non-clinical validation testing and the information provided in the accompanying 510(k) submission, Fig, LLC concludes that the PowerPlay model PPRT-01 (with accessory compression wraps) is safe, effective and performs in a manner that is substantially equivalent to the predicate devices listed.
> Model PPRT-01 Powered Inflatable Tube Massager 510(k) Submission for Fig, LLC July 18, 2012
ાર
{5}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
# Letter Dated: November 21, 2012
Fig. LLC % Mr. Joe Adkins Safeguard Manufacturing & Development, Inc. 2839 Harvest Moon Drive Orange Park, FL 32073
Re: K122154
Trade/Device Name: PowerPlay Muscle Massager, PPRT-01 Regulation Number: 21 CFR 890.5650 Regulation Name: Powered inflatable tube massager Regulatory Class: Class II Product Code: IRP Dated: October 10, 2012 Received: October 12, 2012
Dear Mr. Joe Adkins:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{6}------------------------------------------------
Page 2 - Mr. Joe Adkins
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson
Victor Krauthamer, Ph.D. Acting Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Section 4: Indications for Use Statement
510(k) # K |2 ZJ 5 Y
#### Indications for Use:
The PowerPlay model PPRT-01 is intended to be an over-the-counter portable inflatable tube massage system which simulates kneading and stroking of tissue with the hands by use of inflatable pressure cuffs. This device can be used to:
- Temporarily increase blood circulation in the treated areas; .
- Temporary relief of minor muscle aches and pains. �
AND/OR Over-The-Counter Use: _ X Prescription Use: (Part 21 CFR 801 Subpart D) (Part 21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mark A. Mullan
(Division Sign-Off) (Division Sign-Off)
Division of Surgical, Orthopedic,
Division of Surgical, Orthopedic, Division of Surgious.
and Restorative Devices
510(k) Number -
Model PPRT-01 Powered Inflatable Tube Massager 510(k) Submission for Fig, LLC July 18, 2012
10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.