← Product Code [ILY](/productcode/ILY) · K122125

# ASA LASER SH1 (K122125)

_El.En Electronic Engineering Spa · ILY · Jan 25, 2013 · Physical Medicine · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K122125

## Device Facts

- **Applicant:** El.En Electronic Engineering Spa
- **Product Code:** [ILY](/productcode/ILY.md)
- **Decision Date:** Jan 25, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.5500
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine
- **Attributes:** Therapeutic

## Indications for Use

The ASA laser SH1 is intended to provide topical heating for the purpose of elevating tissue temperature for the temporary relief of muscle and joint pain and stiffness, arthritis pain, or muscle spasm, the temporary increase in local blood circulation and/or promoting relaxation of muscle.

## Device Story

ASA laser SH1 is a medical laser system utilizing a Nd:YAG laser source (1064nm wavelength). Laser radiation is delivered to the treatment area via fiber optics and a handpiece. Device features a warning light on the top cover to indicate system readiness. Operated by healthcare professionals in a clinical setting. The device provides topical heating to elevate tissue temperature, which aids in pain relief, muscle relaxation, and increased local blood circulation. The system is compact (38x22x42 cm) and weighs 13 kg.

## Clinical Evidence

Clinical study performed on three human subjects with varying skin phototypes. Results demonstrated the device successfully induced and maintained skin temperatures between 40°C and 45°C for at least 10 minutes. Data indicates the device is as safe and effective as the predicate.

## Technological Characteristics

Nd:YAG laser source; 1064nm wavelength; fiber optic delivery system; 115Vac 50/60Hz power requirement; 750VA; dimensions 38cm x 22cm x 42cm; weight 13kg; warning light indicator.

## Regulatory Identification

An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.

## Special Controls

*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.

## Predicate Devices

- HILT Family Laser ([K051537](/device/K051537.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K122/25

## 510(K) Summary

JAN 2 5 2013

#### Submitter:

El.En. S.p.A.

via Baldanzese, 17

50041 Calenzano (FI), Italy

### Contact:

Paolo Peruzzi

Regulatory Affairs Manager

Phone: +39.055.8826807

E-mail: standards@elen.it

Date Summary Prepared:

Device Trade Name:

Common Name:

Classification Name:

Equivalent Devices:

Device Description:

November 26, 2012

ASA laser SH1

Nd:YAG Laser System

Infrared Lamp

ILY, 21 CFR 890.5500

HILT Family Laser (K051537)

The ASA laser SH1 is a medical laser system equipped with a Nd: Y AG laser source.

The Nd:YAG laser radiation has a wavelength of 1064nm and is delivered to the treatment area through a fiber optics and a delivery handpiece connected to its distal end.

A warning light is located on the top cover, close to the control panel. Light ON state indicates that the system is enabled and ready.

Overall weight of the device is 13 kg, and the size is 38 cm x 22 cm x 42 cm (H x W x D).

Electrical requirement is 115Vac 50/60Hz, 750VA.

{1}------------------------------------------------

Indications for Use:

Comparison:

Nonclinical Performance Data:

Clinical Performance Data:

.

The ASA laser SH1 is intended to provide topical heating for the purpose of elevating tissue temperature for the temporary relief of muscle and joint pain and stiffness, arthritis pain, or muscle spasm, the temporary increase in local blood circulation and/or promoting relaxation of muscle.

The ASA laser SH1 is substantially equivalent to its predicate device.

It shares same indication for use, same principle of operation essentially same technological and characteristics and performances.

None

A clinical study has been performed on three live human subjects with three different skin phototypes. The study has shown that ASA laser SH1 is able to induce a skin temperature between 40℃ and 45°C and maintain that temperature for at least 10 minutes for all patients.

The results of clinical performance tests show that ASA laser SH1 system is as safe, as effective and performs as well as its predicate device for the indications for use mentioned above.

Additional Information:

Conclusion:

None

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a person embracing or supporting another, represented by flowing lines.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

EL.EN. Electronic Engineering S.p.A % Mr. Paolo Peruzzi · Regulatory Affairs Manager 17 Via Baldanzese Calenzano, Italy 50041

Re: K122125

Trade/Device Name: ASA laser SH1 Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared lamp Regulatory Class: Class II Product Code: IL Y Dated: November 27, 2012 Received: November 29, 2012

Dear Mr. Peruzzi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be t found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

January 25, 2013

{3}------------------------------------------------

Page 2 – Mr. Paolo Peruzzi

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

# Peter D. Rumm -S

Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(K) Number (if known):

Device Name: ASA laser SHI

Indications for Use:

The ASA laser SHI is intended to provide topical heating for the purpose of elevating tissue temperature for the temporary relief of muscle and joint pain and stiffness, arthritis pain, or muscle spasm, the temporary increase in local blood circulation and/or promoting relaxation of muscle.

Prescriptive Use _ X (Part 21 CFR 801 Subpart D) OR

Over-the-Counter Use (Part 21 CFR 801 Subpart C) ti

场

: খু-

(PLEASE NO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence ofCDRH, Office of Device Evaluation (ODE)

Neil R Ogden 2013.01.25 08:10:41 -05'00'

(Division Sign-Off)

Division of Surgical Devices

510(k) Number ________________________________________________________________________________________________________________________________________________________________ K122125

---

**Source:** [https://fda-staging.innolitics.com/device/K122125](https://fda-staging.innolitics.com/device/K122125)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
