K122112 · Bio Compression Systems, Inc. · IRP · Apr 9, 2013 · Physical Medicine
Device Facts
Record ID
K122112
Device Name
2004-OC MASSAGE SYSTEM, 2008-OC MASSAGE SYSTEM
Applicant
Bio Compression Systems, Inc.
Product Code
IRP · Physical Medicine
Decision Date
Apr 9, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5650
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The 2004-0C and 2008-OC Massage Systems are powered inflatable tube massagers intended to temporarily relieve minor muscle aches and/or pains, and to temporarily increase circulation to the treated areas. The devices are intended for home use by people who are in good health.
Device Story
Device consists of air compressor and garment with 4 (2004-OC) or 8 (2008-OC) inflatable chambers; applied to limbs. Compressor provides sequential pneumatic compression (distal to proximal) to increase circulation and relieve muscle aches. User manually adjusts pressure (30-80 mmHg) via dial; in-line check valve limits output to 80 mmHg. System inflates chambers sequentially, holds pressure to prevent reverse flow, then releases all simultaneously. Fixed cycle times controlled by motorized valve. Used in home setting by healthy individuals. Output is mechanical compression; provider/user monitors pressure via calibrated gauge. Benefits include temporary relief of minor aches and improved circulation.
Clinical Evidence
No clinical data. Bench testing only: electrical leakage, inflation pressure per segment, pressure adjustment, pressure gradient, gauge accuracy, and cycle times.
Technological Characteristics
Powered inflatable tube massager; 4 or 8 segmented chambers. Max pressure 150 mmHg; operating range 30-80 mmHg. Motorized valve for fixed cycle timing. Calibrated dial gauge (0-125 mmHg). In-line check valve (80 mmHg limit). 115VAC power. Bilateral treatment adapter available.
Indications for Use
Indicated for healthy individuals for temporary relief of minor muscle aches/pains and temporary increase of local circulation.
Regulatory Classification
Identification
A powered inflatable tube massager is a powered device intended for medical purposes, such as to relieve minor muscle aches and pains and to increase circulation. It simulates kneading and stroking of tissues with the hands by use of an inflatable pressure cuff.
{0}------------------------------------------------
# 11. 510(k) Summary
## APPLICANT'S INFORMATION
Robert Freidenrich, CEO Bio Compression Systems, Inc 120 West Commercial Avenue Moonachie, NJ, 07074 201 939-0716 PH: 201 939-4503 FX: E-mail: Internet: http://www.biocompression.com Medical Establishment Registration No .:
#### 2424387
#### SUBMITTER'S INFORMATION
Maureen Garner, President New World Regulatory Solutions, Inc. P.O. Box 5374 Toms River, NJ 08754 732-779-7422 PH: 732-270-4829 Fax: E-mail: Internet: www.newworldreg.com
DATE PREPARED: March 29, 2013
#### DEVICE INFORMATION
| DEVICE NAME: | 2008-OC and 2004-OC Massage Systems |
|-------------------------|--------------------------------------------------|
| Classification Panel: | Surgical, Orthopedic and Restorative Devices |
| Classification Number: | 890.5650 |
| Product Nomenclature: | Powered Inflatable Tube Massager |
| Product Code(s): | IRP |
| Trade/Proprietary Name: | 2008-OC Massage System<br>2004-OC Massage System |
| Common Name: | 2008-OC Massage System<br>2004-OC Massage System |
#### DEVICE CLASSIFICATION
Powered Inflatable Tube Massager Devices are classified as Class II devices, and reviewed by the Division of Surgical, Orthopedic and Restorative Devices.
#### PREDICATE DEVICE
DJS Massager (K112479; product code IRP) manufactured by Mego Afec AC LTD.
#### DEVICE DESCRIPTION
The device consists of an air compressor with manually adjustable pressure settings (i.e., the user manually adjusts the pressure, while the inflate/deflate cycle times remain constant) and a sleeve or garment containing four (2004-OC) or eight (2008-OC) discrete, segmented inflatable chambers externally applied to the limb.
The pump consists of a compressor capable of a maximum pressure of 150 mmHg, and provides graduated, or gradient, pressurization to the chambers (e.g., sequential inflation of distal to proximal, with distal chambers inflated to a greater pressure than the proximal ones).
# APR 0 9 2013
{1}------------------------------------------------
An in-line check valve limits the output pressure to 80 mmHg, and a calibrated dial gauge displays pressure in the range of 0-125 mmHq.
As each chamber is inflated, the pressure is held constant until all chambers are inflated, in order to prevent reverse gradient flow. Once all chambers are inflated, they are then all released simultaneously, and the cycle repeats. Pressures within the distal chambers have a default value set at the factory (60 mmHg distal), and can be adjusted by the user to distal pressures between 30 mmHg and 80 mmHg. Pressures decrease by 5 mmHg in each proximally sequential chamber.
Garments are available in sizes to accommodate varying limb lengths. An adapter is available to support bilateral treatment.
#### INDICATIONS FOR USE
The Model 2004-OC and 2008-OC Massage Systems are powered inflatable tube massagers intended to temporarily relieve minor muscle aches and to temporarily increase circulation to the treated areas. The device is intended for home use by people who are in good health.
#### TECHNOLOGICAL CHARACTERISTICS
The manufacturer believes that the technological characteristics of the 2004-OC Massage Systems are substantially similar to those of the predicate device.
#### PERFORMANCE DATA
Before being released, every device is tested and must meet all performance specifications. In addition to aesthetic acceptance criteria, functional testing includes electrical leakage, inflation pressure in each segment, pressure adjustment, pressure gradient across the segments, air pressure gauge accuracy, and inflation/deflation cycle times.
### STATEMENT OF SUBSTANTIAL EQUIVALENCE
The Bio Compression Systems Model 2004-OC and 2008-OC Massage Systems and the predicate device all provide sequential inflation pressure from distal to proximal segments. Both the Bio Compression and predicate devices offer adjustable pressure ranges.
The Bio Compression devices have decreasing gradient pressures from distal to proximal segments, which prevent flow of lymph back into the limb, while predicate device does not have decreasing gradient pressures. The Bio Compression devices provide continuously adjustable pressure in the range between 30 to 80 mmHg. while the predicate device provides three pressure adjustments options in increments of low, medium and high pressure ranges within the 20 to 80 mmHg range. The Bio Compression devices have fixed cycle times that are fixed by a motorized valve, and are independent of garment size and number (one or two garments). The predicate device does not have a motorized valve and uses back pressure to trigger the cycle to the next chamber of the garment.
Both the Bio Compression devices and the applicant device operate within clinically-established parameters. The differences between the predicate and the applicant devices do not impact safety or effectiveness. A table illustrating the similarities and differences is provided below.
{2}------------------------------------------------
## Similarities and Differences with the Predicate Device
| Parameter | 2004-OC | 2008-OC | K112479<br>DJS Massager |
|----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|-----------------------------------------------------------------------|
| Intended Use | Temporarily relieve minor muscle aches and/or pains, and to temporarily increase<br>circulation to the treated areas. The devices are intended for home use by people<br>who are in good health. | | |
| Principal of<br>Operation | Sequential pneumatic<br>compression | Same | Same |
| Weight | 8 pounds | 8 pounds | 5.1 pounds |
| Dimensions, inches | 5.5H X 8 L X 12 W | 5.5H X 8 L X 12 W | 3.9 H X 10.2 L X 5.1 W |
| # of Segments in<br>garment | 4 | 8 | 4 |
| Sequential segment<br>inflation | Yes | Yes | Yes |
| Distal to Proximal<br>gradient | Yes | Yes | No |
| Inflation Time, each<br>segment | 18 seconds | 5.5 seconds | Approx. 10 (one garment)<br>or 18 (two garments)<br>seconds |
| Inflation Pressure | 30 - 80 mmHg distal,<br>adjustable in 1 mmHg<br>increments | 30 - 80 mmHg distal,<br>adjustable in 1 mmHg<br>increments | 20 - 80 mmHg, adjustable<br>in 3 ranges: 20-30, 40-60<br>and 70-80 |
| Locking pressure<br>adjustment knob | Yes | Yes | Yes |
| Pressure Gauge | Yes, 0 - 125 mmHg | Yes, 0 - 125 mmHg | No |
| Deflation Time | 18 seconds | 5.5 seconds | Approx. 10 (one garment)<br>or 18 (two garments)<br>seconds |
| Pause time between<br>inflation cycles | None (deflation time is<br>pause time) | None (deflation time is<br>pause time) | Adjustable in 3 levels:<br>10-20, 30-50 and 60-70<br>seconds |
| Total Cycle Time | 90 seconds, fixed<br>(one or two garments). | 50 seconds, fixed<br>(one or two garments) | Variable: Approx. 50<br>(one garment) or 90<br>seconds (two garments) |
| Garments Available | Sized leg and<br>arm/shoulder | Same | Same |
| Fail-safe hose<br>connectors | Yes | Yes | Yes |
| Bilateral Treatment<br>Option | Yes | Yes | Yes |
| Power Requirements | 115VAC, 50-60Hz | Same | Same |
Based upon safety and performance testing, compliance with voluntary standards, and comparison to the predicate devices, the manufacturer believes that the 2004-OC and 2008-OC Massage Systems are substantially equivalent to the predicate device, and do not raise any new questions of safety or effectiveness.
{3}------------------------------------------------
Image /page/3/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring three curved shapes, resembling a stylized caduceus or a representation of human figures.
## April 9, 2013
Bio Compression Systems, Inc. c/o Ms. Maureen N. Garner President 1983 Hazelwood Road Toms River, NJ 08754
Re: K122112/S001
Trade/Device Name: 2004-OC Massage System, 2008-OC Massage System Regulation Number: 21 CFR 890.5650 Regulation Name: Powered Inflatable Tube Massager Regulatory Class: Class II Product Code: IRP Dated: April 1, 2013 Received: April 2, 2013
Dear Ms. Garner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G60 Silver Spring, MD 20993-002
{4}------------------------------------------------
Page 2 - Ms. Maureen N. Garner
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Joyce M. Whang -S
for Victor Krauthamer, Ph.D. Acting Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K122112
Device Name: 2004-0C Massage System, 2008-0C Massage System
Indications For Use:
The 2004-0C and 2008-0C Massage Systems are powered inflatable tube massagers intended to temporarily relieve minor muscle aches and/or pains, and to temporarily increase circulation to the treated areas. The devices are intended for home use by people who are in qood health.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use X (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
# Joyce M Whang -
(Division Sign Off) Division of Neurological and Physical Medicine Devices (DNPMD)
510(k) Number _K122112
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.