K122105 · Solinas Medical, Inc. · DXZ · Dec 7, 2012 · Cardiovascular
Device Facts
Record ID
K122105
Device Name
SMI CARDIOVASCULAR PATCH
Applicant
Solinas Medical, Inc.
Product Code
DXZ · Cardiovascular
Decision Date
Dec 7, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.3470
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SMI Cardiovascular Patch is intended for cardiovascular patching.
Device Story
SMI Cardiovascular Patch is a cardiovascular patch used for surgical repair. Device consists of knitted polyethylene terephthalate (polyester) fabric, silicone sheet, and nickel-titanium (nitinol) alloy mesh. Available in flat (2 cm x 7.5 cm) and curved configurations (6, 7, 8, 9, 10 mm diameters; 4.5 cm length; 360° arc angle). Manufacturing process involves bonding polyester fabric to silicone rather than stitching. Used by surgeons in cardiovascular procedures to patch tissue. Bench testing confirms mechanical integrity including tensile strength, burst strength, suture pullout, and water permeability.
Clinical Evidence
No clinical data. Bench testing only. Testing included tensile strength, burst strength, suture pullout, and water permeability to demonstrate performance specifications are met and equivalent to the predicate device.
Technological Characteristics
Materials: knitted polyethylene terephthalate (polyester) fabric, silicone sheet, nickel-titanium (nitinol) alloy mesh. Configurations: flat and curved. Manufacturing: bonded fabric-to-silicone interface. Sterilization, packaging, and labeling are equivalent to predicate.
Indications for Use
Indicated for cardiovascular patching in patients requiring cardiovascular repair.
Regulatory Classification
Identification
An intracardiac patch or pledget made of polypropylene, polyethylene terephthalate, or polytetrafluoroethylene is a fabric device placed in the heart that is used to repair septal defects, for patch grafting, to repair tissue, and to buttress sutures.
{0}------------------------------------------------
## Appendix 4: 510(k) Summary per (21CFR807.92)
:
DEC 7 2012
| 510(k) Number | K122105 | |
|--------------------------------------------------------------------------------------------|------------------------------------------------------|---------------|
| Submitter Name and Address | | |
| Name: | Solinas Medical, Inc. | |
| Contact: | Michael Kolber | |
| | Consultant, Regulatory Affairs | |
| Address: | 443 Costa Mesa Terrace, Unit A | |
| | Sunnyvale, CA 94085 | |
| Telephone: | 650-793-5015 | |
| Fax: | 408-720-9466 | |
| Date Prepared: | November 7, 2012 | |
| General Device Information | | |
| Product Name: | SMI Cardiovascular Patch | |
| Common Name: | Cardiovascular patch | |
| Classification: | 21CFR870.3470 | |
| Classification Name | Intracardiac patch or pledget made of polypropylene, | |
| | polyethylene terephthalate, or | |
| | polytetrafluoroethylene | |
| Device Class: | Class II | |
| Product Code: | DXZ | |
| Predicate Device | | |
| Manufacturer | Device Name | 510(k) Number |
| Solinas Medical Inc. | SMI Cardiovascular Patch | K112683 |
| Description | | |
| The SMI Cardiovascular Patch is designed for cardiovascular patching. The SMI | | |
| Cardiovascular Patch is comprised of a knitted polyethylene terephthalate (polyester) | | |
| fabric covered silicone sheet and with a thin nickel-titanium (nitinol) alloy mesh. It is | | |
| provided in various sizes and as flat and curved shapes. The SMI Cardiovascular Patch in | | |
| the flat configuration is 2 cm x 7.5 cm and the following curved sizes: 6, 7, 8, 9, and 10 | | |
| mm diameters, 4.5 cm length, and 360° arc angle. A manufacturing change was also | | |
| made to bond the polyester fabric to the silicone. | | |
| Intended Use (Indications) | | |
| The SMI Cardiovascular Patch is intended for cardiovascular patching. | | |
| Comparison to the Predicate Device | | |
| The predicate SMI Cardiovascular Patch includes a curved configuration of 6mm | | |
| diameter, 3, 7.5, and 15 cm lengths, and 120° arc angle and flat configuration of 1 and 2 | | |
| cm width by 3, 7.5, and 15 cm lengths. This submission increases the number sizes | | |
| available and implements a manufacturing change to securely bond the polyester fabric | | |
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a stylized symbol that resembles two curved arrows facing opposite directions. The arrows are arranged vertically, with the top arrow curving to the right and the bottom arrow curving to the left. The arrows are thick and black, and there are small dots on either side of the symbol.
K122105 for the SMI Cardiovascular November 7.
rather than stitching the fabric to the silicone.
The SMI Cardiovascular Patch has the same intended use and fundamental scientific technology as the predicate device. The technological characteristics of the SM1 Cardiovascular Patch are substantially equivalent to the predicate device including biocompatibility, sterilization, packaging, and labeling. Through bench performance testing it was demonstrated that the additional sizes, dimension changes and manufacturing change does not adversely affect safety and effectiveness.
Summary of Non-clinical Testing
The non-clinical test data provided in this submission demonstrated that the SMI Cardiovascular Patch meets the performance specifications. The submission includes bench mechanical testing, including: Tensile Strength, Burst Strength, Suture Pullout, and Water Permeability Tests. These tests demonstrate the SMI Cardiovascular Patch has results that are substantially equivalent to the predicate device.
Statement of Equivalence
The SMI Cardiovascular Patch has the same indications for use and the same technological characteristics as the predicate device. Based on this and the data provided in this pre-market notification, the subject device and the predicate device have been shown to be substantially equivalent.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a symbol. The symbol is a stylized representation of a human figure, with three overlapping profiles suggesting a sense of community and support. The logo is printed in black and white.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
DEC 7 2012
Solinas Medical, Inc. C/O James Hong 443 Costa Mesa Terrace Sunnyvale, CA 94089
Re: K122105
Trade/Device Name: SMI Cardiovascular Patch Regulation Number: 21 CFR 870.3470 Regulation Name: Intracardiac patch or pledget made of polypropylene, polyethylene terephthalate, or polytetrafluoroethylene Regulatory Class: Class II Product Code: DXZ Dated: November 07, 2012 Received: November 08, 2012
Dear Mr. Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.goy/MedicalDevices/Resourcesfor You/Industry/default.htm.
for
Sincerely vours.
Kenneth J. Cavanaugh
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) for the Solinas Medical, Inc. SMI Cardiovascular Patch July 16, 2012
## Indications for Use
510(k) Number (if known): _
Device Name: Solinas Medical, Inc. SMI Cardiovascular Patch
Indications for Use: The Solinas Medical, Inc. SMI Cardiovascular Patch is indicated for cardiovascular patching.
Prescription Use _ × (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|-----------------------------------------------------------|---------|
| (Division Sign-Off)<br>Division of Cardiovascular Devices | |
| 510(k) Number | K122105 |
Page 26
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.