K122098 · Andon Health Co, Ltd. · DXN · Mar 21, 2013 · Cardiovascular
Device Facts
Record ID
K122098
Device Name
APPS-HEALTH01
Applicant
Andon Health Co, Ltd.
Product Code
DXN · Cardiovascular
Decision Date
Mar 21, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The Apps-Health01 is intended for use in home settings as an aid for people to review, analyze, and evaluate test results which are measured from a blood pressure monitor or scale.
Device Story
Apps-Health01 is an iOS-based data management software application. It interfaces with external health devices (blood pressure monitors and scales) via Bluetooth or Wi-Fi to retrieve, store, and display measurement data. Users operate the app on mobile devices (iPhone, iPad, iTouch) to view trends in blood pressure, weight, and calories; log food and activity; and set reminders for medication or measurements. The software provides a historical view of data to assist users in monitoring their health metrics. It does not perform measurements itself but acts as a repository and visualization tool for data collected by peripheral hardware. The output is presented to the user on the mobile device screen to facilitate self-management of health goals.
Clinical Evidence
No clinical data. Performance validation was limited to software unit and system testing, wireless coexistence testing, and compliance with IEC 60601-1, IEC 60601-1-2, and FCC standards.
Technological Characteristics
iOS software application; operates on iPhone, iPad, iTouch; connectivity via Bluetooth and Wi-Fi; data management and visualization functionality; software validation performed.
Indications for Use
Indicated for home users to review, analyze, and evaluate blood pressure and weight measurements obtained from compatible blood pressure monitors or scales.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{0}------------------------------------------------
# MAR 2 1 2013
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of SMDA and 21 CFR 807.92.
## 1.0 submitter's information
Name: Andon Health Co., Ltd. Address: No 3, Jinping Road, Ya' an street TIANJIN, 300193 Phone number: 86-22-60526161 Fax number: 86-22-6052 6162 Contact: Yi Liu Date of Application: 7/11/2012
#### 2.0 Device information
| Trade name: | Apps-Health01 |
|----------------------|--------------------------|
| Common name: | Data management software |
| Classification name: | Data management software |
## 3.0 Classification
Production code: DXN, MNW Regulation number: 21 CFR 870.2770, 21 CFR 870.1130 Classification: II Panel: 870 Cardiovascular
#### 4.0 Predict device information
Manufacturer: Andon Health Co., Ltd. Device: KD-931 510(k) number: K102939
#### 5.0 Device description
Apps-Health01 iHealth App is an iOS software for using with health devices, such as Blood Pressure Monitor and Scale. When used with the health devices, iHealth App could transfer data from the device's memory and count, manage, share the data.
#### 6.0 Intended use
{1}------------------------------------------------
#### Apps-Health01 FDA 510(k) Files
.
:
.
:
The Apps-Health01 is intended for use in home settings as an aid for people to review, analyze, and evaluate test results which are measured from a blood pressure monitor or scale. .
| item | Apps-Health01 | KD-931 (K102939) |
|-----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product type | software | System(include meter<br>and software) |
| Indications for<br>use | The Apps-Health01 is intended<br>for use in home settings as an<br>aid for people to review,<br>analyze, and evaluate test<br>results which are measured<br>from a blood pressure monitor<br>or scale. | KD-931 Fully Automatic<br>Electronic Blood<br>Pressure Monitor are for<br>use by medical<br>professionals or at home<br>and are non-invasive<br>blood pressure<br>measurement system<br>intended to measure the<br>diastolic and systolic<br>blood pressures and<br>pulse rate of an adult<br>individual by using a<br>non-invasive technique<br>in which an inflatable<br>cuff is wrapped around<br>the upper arm. The cuff<br>circumference is limited<br>to 22cm-48cm. |
| Software<br>platform | iOS | iOS |
| Hardware<br>requirements | iPhone, iTouch, iPad | iPhone,iTouch,iPad |
| Function | Blood pressure: Test or manually input blood pressure data. | Blood pressure: Test blood pressure data. |
| | Weight: Test or manually input body composition data. | My BP trends: View your blood pressure trends. |
| | My calorie Trends: View your calories consumed vs. calories burned. | History: A customizable view of your historical data. |
| | Reminders: Set reminders (Medicine/Measure blood pressure /Measure weight/Sport) to stay focused on your goals. | |
| | My plan: Add and modify your weight goals. | |
| | My BP trends: View your blood pressure trends. | |
| | My weight Trends: View your weight data. | |
| | History: A customizable view of your historical data. | |
| | My Diary: Log food and activities to stay on track. | |
| Meter<br>Compatibility | Scales, Blood Pressure Monitor | Blood Pressure Monitor |
| Connection with peripheral<br>device by | Bluetooth (HS5 with WIFI additionally, KD-931 by apple 30PIN) | Apple 30PIN |
| Performance<br>test provided | Software validation | Software validation |
## 7.0 Summary comparing technological characteristics with predicate device
## Page 2 of 4
{2}------------------------------------------------
## Apps-Health01 FDA 510(k) Files
.
:
.
. . . . . . . . . .
5- 3
. :
:
:
:
.
{3}------------------------------------------------
## 8.0 Performance summary
Software validation of Apps-Health01 included system test and unit test. We also performed the Wireless coexistence test, as well as test according to IEC 60601-1 and 60601-1-2, FCC test.
### 9.0 Comparison to the predict device and the conclusion
Apps-Health01 is similar with the predicted device FULLY AUTOMATIC ELECTRONIC BLOOD PRESSURE MONITOR MODEL(K102939), However, Apps-Health01 is only a software, it doesn't include meter. And it can used together with Scales, Blood Pressure Monitor instead of Blood Pressure Monitor only.
However, the test in this submission provides demonstration that these small differences do not raise any new questions of safety and effectiveness.
### Page 4 of 4
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle. The logo is black and white and appears to be a scanned image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 21, 2013
Andon Health Co., Ltd. c/o Ms. Yi Liu No 3, Jinping Road, Ya' an Street Tianjin China 300190 CN
Re: K122098
Trade/Device Name: Apps-Health01 Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive blood pressure measurement system Regulatory Class: Class II Product Code: DXN, MNW Dated: January 3, 2013 Received: January 11, 2013
Dear Ms. Yi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{5}------------------------------------------------
Page 2 - Ms. Liu Yi
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Owen Faris -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Apps-Health01 FDA 510(k) Files
# Indication for Use
510(k) Number (if known): K122098
Device Name: Apps-Health01
Indication For Use:
The Apps-Health01 is intended for use in home settings as an aid for people to review, analyze, and evaluate test results which are measured from a blood pressure monitor or scale.
Prescription Use Yes (21 CFR Part 801 Subpart D)
And/Or
Over the Counter Use Yes (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
Oliver L. Owen P. Faris - S
2013.03.21 15:27:35
-04'00'
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k)
Page 1 of _ 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.