K122089 · Merit Medical Systems, Inc. · DQX · Sep 21, 2012 · Cardiovascular
Device Facts
Record ID
K122089
Device Name
OSTIAL PRO STENT POSITIONING SYSTEM
Applicant
Merit Medical Systems, Inc.
Product Code
DQX · Cardiovascular
Decision Date
Sep 21, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1330
Device Class
Class 2
Indications for Use
The Ostial Pro Stent Positioning System is intended for use in aorta-ostial procedures to introduce and position stents and other interventional devices within the coronary and peripheral vasculature. In addition, the Ostial Pro Stent Positioning System is intended to facilitate the alignment of interventional devices and function as an alignment tool.
Device Story
Disposable medical-grade guidewire system; used by interventional cardiologists and radiologists in aorta-ostial procedures (coronary and renal stenting). Facilitates precise stent implantation and alignment of interventional devices. Compatible with 6, 7, and 8 French catheters. Provided sterile for single use. Device functions as mechanical alignment tool to assist physician during interventional procedures; improves accuracy of stent placement in ostial lesions. Benefits include reduced procedural complexity and enhanced precision in stent positioning.
Clinical Evidence
No clinical data. Bench testing only. Testing included dimensions, torque strength, torque-ability, feet deflection force, austenite finish temperature, wire Af, coating integrity, catheter compatibility, radio-opacity, feet force deflection/compression, fatigue loading, and tensile strength (linear/angular).
Technological Characteristics
Disposable guidewire system; medical-grade materials. Compatible with 6, 7, 8 French catheters. Sterilized via ethylene oxide (AAMI/ANSI/ISO 11135-1). Packaging compliant with AAMI/ANSI/ISO 11607-1 and ISO 2233. Mechanical design features feet for positioning/alignment. No software or electronic components.
Indications for Use
Indicated for use in aorta-ostial procedures to introduce and position stents and other interventional devices within the coronary and peripheral vasculature, and to facilitate alignment of interventional devices. Used by interventional cardiologists and radiologists for coronary and renal stenting procedures.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
K122089
SEP 21 2012
### 510(k) Summary Submitter Name: Merit Medical Systems, Inc. Address: 1600 West Merit Parkway South Jordan, UT 84095 General Telephone Number: (801) 316-4956 Provisions Fax Number: (801) 253-6982 Contact Person: David Thomas Date of Preparation: . July 11, 2012 Registration Number: 1721504 Ostial Pro Stent Positioning System Trade Name: Subject Common/Usual Name:Stent Positioning System Device Classification Name: Catheter Guidewire Trade Name: Ostial Pro Stent Positioning System Common/Usual Name: Stent Positioning System Classification Name: Catheter Guidewire Predicate Device Premarket Notification:K062192 Manufacturer: Merit Medical Systems, Inc. Class II Classification 21 CFR § 870.1330 Cardiovascular The Ostial Pro Stent Positioning System is intended for use in aorta-ostial procedures to introduce and position stents and other interventional devices within the coronary and peripheral Intended Use vasculature. In addition, the Ostial Pro Stent Positioning System is intended to facilitate the alignment of interventional devices and function as an alignment tool.
Section 5
20
{1}------------------------------------------------
| Device Description | The Ostial Pro Stent Positioning System is a medical grade, disposable guidewire system. The Ostial Pro Stent Positioning System will be used by interventional cardiologists and interventional radiologists to ensure precise stent implantation in aorta-ostial procedures. The product will be used in coronary and renal stenting procedures. This product is provided sterile and intended for single use. This finished product will be compatible with 6, 7 and 8 French catheters. |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technological Characteristics | The technological characteristics of the subject Ostial Pro Stent Position System are substantially equivalent to those of the predicate device, the Ostial Pro Stent Positioning System, 510(k) K062192. |
:
- Production Comments of Children
and and the comments of
1. The state of the state of the states of the states of the states of the states of the states of the states of the states of the states of the states of the states of the s
.
.
and the contraction of the control of the country of
.
.
.
:
K122089 Page 2 of 4
{2}------------------------------------------------
No performance standards have been established under Section 514 of the Food, Drug and Cosmetic Act for these devices. Performance testing of the subject Ostial Pro Stent Positioning System was conducted based on the risk analysis and based on the requirements of the following FDA guidance document and international standards:
- . AAMI/ANSI/ISO 11135-1: 2007, Sterilization of health care products - Ethylene oxide - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices
- AAMI/ANSI/ISO 11607-1: 2006, Packaging for terminally . sterilized medical devices Part 1: Requirements for materials, sterile barrier systems and packaging systems
- ISO 2233:2000, Packaging complete, filled transport . packages and unit loads - conditioning and testing
## Safety & Performance Tests
- . ASTM D4169-09, Standard practice for performance testing of shipping containers and systems
- AAMI/ANSI/ISO 10993-7: 2008 Biological Evaluation of . Medical Devices - Part 7: Ethylene Oxide Sterilization Residuals
- ASTM F1980:2007 Standard Guide for Accelerated Aging of . Sterile Barrier Systems for Medical Devices.
- Coronary and Cerebrovascular Guidewire Guidance Jan ● 1995
- Non-Clinical Engineering Tests and Recommended Labeling . for Intravascular Stents and Associated Delivery Systems -April 2010.
The following is a list of all significant testing that was successfully completed:
{3}------------------------------------------------
| Safety &<br>Performance<br>Tests cont. | Design Verification<br>Dimensions<br>Torque Strength<br>Torque-Ability<br>Feet Deflection Force Test<br>Body Af (Austenite Finish Temperature) Test<br>Wire Af Test<br>Coating Adherence/Integrity<br>Catheter Compatibility<br>Radio-opacity<br>Feet Force Deflection/Compression Test<br>Fatigue Loading of Feet<br>Linear Tensile Strength<br>Angular Tensile Strength |
|------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Safety &<br>Performance<br>Tests cont. | The results of the testing demonstrated that the subject Ostial Pro<br>Stent Positioning System met the pre-determined acceptance<br>criteria applicable to the safety and efficacy of the device. |
| Summary of<br>Substantial<br>Equivalence | Based on the indications for use, design, and safety and<br>performance testing, the subject Ostial Pre Stent Positioning<br>System meets the requirements that are considered essential for<br>its intended use and is substantially equivalent to the predicate<br>device, the Ostial Pro Stent Positioning System, manufactured by<br>Ostial Solutions LLC. |
and the comments of the comments of
23
.
. K122089
Page 4
{4}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing and tail. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
2 1 2012
Merit Medial Systems, Inc.
c/o David Thomas
1600 West Merit Parkway
South Jordan, Utah 84095
Re: K122089
Trade/Device Name: Ostial Pro Stent Positioning System Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guidewire Regulatory Class: Class II Product Code: DQX Dated: August 23, 2012 Received: August 24, 2012
Dear Mr. Thomas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{5}------------------------------------------------
## Page 2 – Mr. David Thomas
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
W. A. Hillen
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Section 4
Indications for Use
510(k) Number (if known):
Device Name: Ostial Pro Stent Positioning System
Indications for Use:
The Ostial Pro Stent Positioning System is intended for use in aorta-ostial procedures to introduce and position stents and other interventional devices within the coronary and peripheral vasculature. In addition, the Ostial Pro Stent Positioning System is intended to facilitate the alignment of interventional devices and function as an alignment tool.
× Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use -(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
U.A. Miller
(Division Sign-Off) Division of Cardiovascular Devices
510(k) Number_ 11 122089
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.