← Product Code [DPS](/productcode/DPS) · K122073

# ELI 280 ELECTROCARDIOGRAPH (K122073)

_Mortara Instrument, Inc. · DPS · Aug 2, 2012 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K122073

## Device Facts

- **Applicant:** Mortara Instrument, Inc.
- **Product Code:** [DPS](/productcode/DPS.md)
- **Decision Date:** Aug 2, 2012
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.2340
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Pediatric

## Indications for Use

The ELI 280 is intended to be a high-performance, multi-channel, interpretive electrocardiograph. As a resting electrocardiograph, ELI 280 simultaneously acquires data from multiple channels. Once the data is acquired, it can be reviewed and/or stored, and/or printed with optional resting ECG interpretation provided for physician over-read. It will be a device primarily intended for use in hospitals, but may be used in medical clinics and offices of any size.

## Device Story

Multi-channel resting electrocardiograph; acquires 12-lead ECG data (up to 15 leads) via WAM or AM12 acquisition modules; utilizes color touch-screen LCD for waveform display, menu navigation, and demographic entry; includes thermal writer for various print formats (6+6, Cabrera, real-time rhythm). Used in hospitals, clinics, and offices by physicians or trained staff. Provides optional automated adult/pediatric ECG interpretation to assist physician over-read; identifies best 10-second sample based on noise. Supports data storage and transmission via LAN, WLAN, USB, or modem. Output allows clinicians to review, store, and print ECGs to support diagnostic decision-making.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Multi-channel resting electrocardiograph; color touch-screen LCD; thermal writer; connectivity via LAN, WLAN, USB, and modem. Compatible with Mortara WAM and AM12 acquisition modules. Software provides interpretive analysis for adult and pediatric ECGs.

## Regulatory Identification

An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.

## Predicate Devices

- ELI 250(c) Electrocardiograph ([K101403](/device/K101403.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

510k Summary

## Special 510(k) Notification

AUG 2 2012

## 510(k): ELI 280 Electrocardiograph Device Summary

#### Submitter:

Date: July 10, 2012

Chien Hui (Amy) Yang, Regulatory Affairs Engineer Mortara Instrument, Inc. 7865 N. 86" Street Milwaukee, WI 53224 (414) 354-4760 FAX: (414) 354-1600 Phone: Contact: Amy Yang (see above)

| Trade Name:          | ELI 280 Electrocardiograph                  |
|----------------------|---------------------------------------------|
| Common Name:         | Electrocardiograph                          |
| Classification Name: | Electrocardiograph<br>(Per 21 CFR 870.2340) |

#### Legally marketed devices to which S.E. is claimed:

The ELI 280 is an update to the ELI Series Electrocardiographs and is substantially equivalent to ELI 250(c) Electrocardiograph (K101403) and other Mortara devices presently in distribution

#### Description:

The proposed Mortara Instrument ELI 280 is multi-channel, resting electrocardiograph utilizing a color LCD with touch screen for display of ECG waveforms, custom keyboard, menu options and status information. The proposed modification to the ELI 280 will simultaneously acquire data from 12 lead patient cables, up to 15 leads when available. Once the data is acquired, it can be reviewed, and/or stored, and/or printed with optional resting ECG interpretation provided for physician over-read.

The ELI 280 is able to acquiré, analyze, display and print electrocardiograms acquired through the Mortara front-end amplifier. The size of the screen will allow preview of the record for technician to assess the quality of the acquired ECG.

The ELI 280 utilizes a color touch screen LCD for display of ECG waveforms, custom keyboard, menu options and status information. The touch screen overlay on the LCD display allows the user to interact with the cardiograph by touching areas on the screen for select functions, input demographic parameters and navigate through menus using the touch screen interface. The keyboard is part of the ELI 280 touch screen design and allows patient data entry as well as control of the functions and options available for the unit.

The ELI 280 incorporates a thermal writer that allows printouts using several formats available to the user, from the 6+6 channels to the Cabrera formats. The writer is also used by the unit for real time, continuous rhythm printouts. The ELI 280 offers, adult and pediatic interpretation capability to assist the physician over-read of the electrocardiogram, and review the electrocardiogram to identify the best 10 second sample based on noise.

The ELI 280 is intended to be used with the Mortara Wireless Acquisition Module (WAM) and Mortara Acquisition Module (AM12) patient cables. The ELI 280 acquires ECG waveforms from the AM12 patient cable. The ELI 280 also offers storage capability in order to retrieve or transmit stored records. Transmission can be achieved using one of the options communication media designed in the unit: LAN, WLAN, USB port and/or Modem.

{1}------------------------------------------------

Image /page/1/Picture/16 description: The image shows a white heart shape inside of a black square. The heart is centered in the square, and the top of the heart is slightly rounded. The black square provides a stark contrast to the white heart, making it stand out. The image is simple, but the contrast between the colors makes it visually appealing.

## Special 510(k) Notification

### Intended Use:

The ELI 280 is intended to be a high-performance, multi-channel, interpretive electrocardiograph. As a resting electrocardiograph, ELI 280 simultaneously acquires data from multiple channels. Once the data is acquired, it can be reviewed and/or stored, and/or printed with optional resting ECG interpretation provided for physician over-read. It will be a device primarily intended for use in hospitals, but may be used in medical clinics and offices of any size.

## Indications for Use:

The proposed Mortara ELI 280 Electrocardiograph is a non-invasive prescription device.

- The device is indicated for use to acquire, analyze, display and print electrocardiograms. .
- The device is indicated for use to provide interpretation of the data for consideration by a physician. .
- The device is indicated for use in a clinical setting, by a physician or by trained personnel who are � acting on the orders of a licensed physician. It is not intended as a sole means of diagnosis.
- The interpretations of ECG offered by the device are only significant when used in conjunction with a . physician over-read as well as consideration of all other relevant patient data.
- The device is indicated for use on adult and pediatric populations.
- The device is not intended to be used as a vital signs physiological monitor.

{2}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo consists of a stylized eagle with three stripes forming its body and wing. The eagle faces right. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular fashion around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

AUG 2 2012

Mortara Instrument, Inc. c/o Ms. Any Yang Regulatory Affairs Engineer 7865 N 86th Street Milwaukee, WI 53224

Re: K122073

> Trade/Device Name: ELI 280 Electrocardiograph Regulatory Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: II (two) Product Code: 74 DPS Dated: July 11, 2012 Received: July 16, 2012

Dear Ms. Yang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

## Page 2 - Ms. Any Yang

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21. CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known):

Mortara ELI 280 Electrocardiograph Device Name:

Indications for Use:

The proposed Mortara ELI 280 Electrocardiograph is a non-invasive prescription device.

- The device is indicated for use to acquire, analyze, display and print electrocardiograms.
- The device is indicated for use to provide interpretation of the data for consideration by a physician.
- The device is indicated for use in a clinical setting, by a physician or by trained . personnel who are acting on the orders of a licensed physician. It is not intended as a sole means of diagnosis.
- The interpretations of ECG offered by the device are only significant when used in . coniunction with a physician over-read as well as consideration of all other relevant patient data.
- The device is indicated for use on adult and pediatric populations.
- The device is not intended to be used as a vital signs physiological monitor.

Prescription Use × AND/OR (21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Vision Sign-Off) (Division Sign-Off)
Division of Cardiovascular Devices

510(k) Number K122023

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of ..

---

**Source:** [https://fda-staging.innolitics.com/device/K122073](https://fda-staging.innolitics.com/device/K122073)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
