K122020 · Zolar Technology & Mfg Co., Inc. · GEX · Jun 5, 2013 · General, Plastic Surgery
Device Facts
Record ID
K122020
Device Name
PHOTON PHOTON PLUS
Applicant
Zolar Technology & Mfg Co., Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jun 5, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Photon is intended for use by dentists for excision, vaporization, ablation and coagulation of oral soft tissue procedures. The specific applications are as follows: SOFT TISSUE: The metal handpiece with cutting fiber does not have any direct contact to the soft tissue, skin or cell during the operation. - Exposure of Unerupted teeth - Fibroma removal - Frenectomy - Gingival Troughing for crown impressions - Gingivectomy - Gingivoplasty - Hemostasis and coagulation - Gingival incision and excision hemostasis - Implant recovery - Incision and drainage of abscess - Leukoplakia - Operculectomy - Oral papillectomies - Pulpotomy as an adjunct to root canal therapy - Reduction of gingival hypertrophy - Soft tissue crown lengthening - Gingival bleeding index - Excisional and incisional biopsies - Treatment of aphthous ulcer, canker sores and herpetic - Vestibuloplasty. LASER PERIODONTAL PROCEDURES: - Sulcular debridement (removal of diseased or inflamed soft tissue in the periodontal pocket to improve clinical indices including:, probe depth, attachment loss and tooth mobility) - Removal of highly inflamed edematous tissue affected by bacteria penetration of the pocket lining and junctional epithelium. The Photon Plus is intended for use by dentists for excision, vaporization, ablation and coagulation of oral soft tissue procedures. The specific applications are as follows: SOFT TISSUE: The metal handpiece with cutting fiber does not have any direct contact to the soft tissue, skin or cell during the operation. - Exposure of Unerupted teeth - Excision of lesion or removal of Granulation tissue - Frenectomy - Gingival Troughing for crown impressions - Gingivectomy - Gingivoplasty - Hemostasis and coagulation - Gingival incision and excision hemostasis - Implant recovery - Incision and drainage of abscess - Leukoplakia - Operculectomy - Oral papillectomies - Pulpotomy as an adjunct to root canal therapy - Reduction of gingival hypertrophy - Soft tissue crown lengthening - Gingival bleeding index - Excisional and incisional biopsies - Treatment of aphthous ulcer, canker sores and herpetic - Vestibuloplasty. LASER PERIODONTAL PROCEDURES: - Sulcular debridement (removal of diseased or inflamed soft tissue in the periodontal pocket to improve clinical indices including:, probe depth, attachment loss and tooth mobility) - Debridement of diseased epithelial lining. TOOTH WHITENING INDICATIONS - Laser assisted whitening/bleaching of teeth - Light activation for bleaching materials for teeth whitening.
Device Story
Photon/Photon Plus are dental diode lasers (980nm wavelength; 100mW-10W power) used by dentists for soft tissue surgery and tooth whitening. Device utilizes solid-state diode laser energy delivered via flexible fiber to a handpiece; activated by footswitch. Provides surgical precision for procedures like gingivectomy, frenectomy, and periodontal debridement. Output is laser energy; healthcare providers use visual feedback and clinical judgment to perform procedures. Benefits include precise tissue management, hemostasis, and efficient whitening. Operated in dental clinics.
Clinical Evidence
No clinical data; bench testing only. Safety and effectiveness established via electrical, mechanical, and thermal testing, and verification of design specifications against predicates.
Technological Characteristics
Solid-state diode laser; 980nm wavelength; 100mW-10W power; 650nm aiming beam. Pulsed or CW modes. Color touch screen interface. Handpiece is autoclavable. Powered by 120V/240V or battery. Standards: IEC 60825-1, IEC 60601-2-22, ANSI/AAMI ST79, ISO 17665-1.
Indications for Use
Indicated for dental patients requiring oral soft tissue surgery (excision, vaporization, ablation, coagulation), periodontal procedures (sulcular debridement, epithelial lining debridement), and tooth whitening. Contraindications not specified.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Picasso/Picasso Lite Dental Diode Lasers (K102359)
{0}------------------------------------------------
# 510(K) SUMMARY
ﻠﺴﻠﺴﻠﺔ
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is K122020.
#### 1. Submitter's Identification:
ZOLAR TECHNOLGY & MFG CO. INC. 6033 SHAWSON DRIVE, UNIT #19, MISSISSAUGA, ON, L5T 1H8, CANADA
Contact:
Joginder (Joe) Pal General Manager
TEL : 905-593-3605/ 905-581-5235 FAX: 1-888-426-7728 EMAIL: SALES@ZOLARTEK.COM
Date Summary Prepared: June 3, 2013
### 2. Name of the Device:
Photon/Photon Plus Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX
### 3. Common or Usual Name:
Dental Diode Laser
### 4. Predicate Device Information:
Picasso/Picasso Lite Dental Diode Lasers, K102359
stLase Dental Laser, K111689
### ડ. Device Description:
The Photon/Photon Plus is intended for use by dentists in oral soft tissue and tooth whitening procedures.
{1}------------------------------------------------
Photon/Photon Plus Diode Laser is a surgical device at the cutting edge of technology, designed for a wide variety of oral soft tissue and tooth whitening procedures with the wavelength of 980nm and the power range of 100mW to 10W.
Photon Series Diode Laser utilizes a solid state diode a laser energy source. The energy is delivered to the operating area by means of a delivery system consisting of a flexible fiber connecting the laser source and the handpiece. The device is activated by means of a footswitch.
#### 6. Intended Use:
The Photon is intended for use by dentists for excision, vaporization, ablation and coagulation of oral soft tissue procedures. The specific applications are as follows:
SOFT TISSUE: The metal handpiece with cutting fiber does not have any direct contact to the soft tissue, skin or cell during the operation.
- Exposure of Unerupted teeth .
- . Fibroma removal
- . Frenectomy
- . Gingival Troughing for crown impressions
- . Gingivectomy
- . Gingivoplasty
- . Hemostasis and coagulation
- . Gingival incision and excision hemostasis
- . Implant recovery
- . Incision and drainage of abscess
- . Leukoplakia
- . Operculectomy
- . Oral papillectomies
- . Pulpotomy as an adjunct to root canal therapy
- ' Reduction of gingival hypertrophy
- . Soft tissue crown lengthening
- . Gingival bleeding index
- . Excisional and incisional biopsies
- . Treatment of aphthous ulcer, canker sores and herpetic
- . Vestibuloplasty.
## LASER PERIODONTAL PROCEDURES:
- Sulcular debridement (removal of diseased or inflamed soft tissue in the periodontal pocket to . improve clinical indices including:, probe depth, attachment loss and tooth mobility)
- Removal of highly inflamed edematous tissue affected by bacteria penetration of the . pocket lining and junctional epithelium
## Indications for Use for Photon Plus:
The Photon Plus is intended for use by dentists for excision, vaporization, ablation and coagulation of oral soft tissue procedures. The specific applications are as follows:
SOFT TISSUE: The metal handpiece with cutting fiber does not have any direct contact to the soft tissue, skin or cell during the operation.
{2}------------------------------------------------
- . Exposure of Unerupted teeth
- Excision of lesion or removal of Granulation tissue .
- . Frenectomy
- . Gingival Troughing for crown impressions
- . Gingivectomy
- . Gingivoplasty
- . Hemostasis and coagulation
- . Gingival incision and excision hemostasis
- . Implant recovery
- . Incision and drainage of abscess
- . Leukoplakia
- . Operculectomy
- . Oral papillectomies
- . Pulpotomy as an adjunct to root canal therapy
- . Reduction of gingival hypertrophy
- . Soft tissue crown lengthening
- . Gingival bleeding index
- Excisional and incisional biopsies .
- . Treatment of aphthous ulcer, canker sores and herpetic
- . Vestibuloplasty.
# LASER PERIODONTAL PROCEDURES:
- Sulcular debridement (removal of diseased or inflamed soft tissue in the periodontal pocket to improve clinical indices including;, probe depth, attachment loss and tooth mobility)
- Debridement of diseased epithelial lining .
# TOOTH WHITENING INDICATIONS
- Laser assisted whitening/bleaching of teeth .
- Light activation for bleaching materials for teeth whitening. .
## 7. Comparison to Predicate Devices:
The subject device is of a comparable type and substantially equivalent to the previously licensed Picasso and Picasso Lite marketed by AMD Lasers LLC and the stLase by Dental Photonics, Inc. To facilitate the comparison, the principal features of the subject devices are listed side-by-side with the predicate devices. Refer to the comparison table shown below:
{3}------------------------------------------------
# Zolar Photon Plus and Stlase IFU Comparison Chart
| Zolar Photon Plus (subject) | St Lase Predicate (K111689) |
|-----------------------------------------------------|---------------------------------------------------------|
| Exposure of Unerupted teeth | incision, excision, vaporization, ablation, |
| Excision of lesion or removal of Granulation tissue | removal of granulation tissue, |
| Frenectomy | |
| Gingival Troughing for crown impressions | Frenectomy, Frenotomy, |
| Gingivectomy | gingival troughing, crown lengthening, |
| Gingivoplasty | gingivectomy, |
| Hemostasis and coagulation | gingivoplasty, |
| Gingival incision and excision hemostasis | hemostasis, or coagulation of intraoral and |
| | extraoral soft tissue (including marginal and |
| | interdental gin giva and epithelial lining of free gin |
| Implant recovery | giva) |
| Incision and drainage of abscess | implant recovesry/uncovery, |
| Leukoplakia | incisions and draining of abscesses |
| Operculectomy | Leukoplakia |
| Oral papillectomies | operculectomy |
| Pulpotomy as an adjunct to root canal therapy | papillectomy |
| Reduction of gingival hypertrophy | pulpotomy as an adjunct to root canal therapy |
| | papillectomy |
| Soft tissue crown lengthening | |
| Gingival bleeding index | crown lengthening |
| Excisional and incisional biopsies | gingival troughing |
| | biopsy |
| Treatment of aphthous ulcer, canker sores and | |
| herpetic | |
| Vestibuloplasty. | Treatment of aphthous ulcers |
| LASER PERIODONTAL PROCEDURES: | vestibuloplasy, |
| Sulcular debridement (removal of diseased | |
| or<br>inflamed soft tissue in the periodontal | dental soft tissue indications including: incision, |
| pocket to to improve clinical<br>indices | exci sion, vaporization, ablation, |
| including:, probe depth, attachment loss | biopsy, |
| and tooth mobility) | removal of granulation tissue, |
| | laser assisted flap surgery, debridement of diseased |
| Debridement of diseased epithelial lining | epithelial lining, , tissue retraction for impressions, |
| | , excision of lesions, leukoplakia, removal of |
| TOOTH WHITENING INDICATIONS | hyperplastic tissues, and removal of hyperplastic |
| | tissues, , sulcular debridement (removal of diseased |
| Laser assisted whitening/bleaching of teeth | or inflamed soft tissue in the periodontal pocket),. |
| Light activation for bleaching materials for teeth | and light activation of bleaching materials for teeth |
| whitening. | whitening |
| | |
:
-
{4}------------------------------------------------
# Zolar Photon and Picasso IFU Comparison Chart
| Zolar Photon(subject) | Picasso Predicate(K102359) |
|-------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| Exposure of Unerupted teeth | Exposure of unerupted teeth |
| Fibroma removal | Fibroma removal |
| Frenectomy | Frenectomy and frenotomy |
| Gingival Troughing for crown impressions | Gingival troughing for crown impression |
| Gingivectomy | Gingivectomy |
| Gingivoplasty | Gingivoplasty |
| Hemostasis and coagulation | Hemostasis and coagulation |
| Gingival incision and excision hemostasis | Gingival incision and excision |
| Implant recovery | Implant recovery |
| Incision and drainage of abscess | Incision and drainage of abscess |
| Leukoplakia | Leukoplakia |
| Operculectomy | Operculectomy |
| Oral papillectomies | Oral papillectomnies |
| Pulpotomy as an adjunct to root canal therapy | Pulpotomy as an adjunct to root canal therapy |
| Reduction of gingival hypertrophy | Reduction of gingival hypertrophy |
| Soft tissue crown lengthening | Soft tissue crown lengthening |
| Gingival bleeding index | Gingival bleeding index,<br>probe depth, attachment loss,<br>and tooth mobility |
| Excisional and incisional biopsies | |
| | |
| Treatment of aphthous ulcer, canker sores and | Treatment of canker sores, herpetic and aphthous |
| herpetic | ulcers of the oral mucosa |
| Vestibuloplasty. | Vestibuloplasty |
| | Excisional and incisional biopsies |
| LASER PERIODONTAL PROCEDURES: | |
| | |
| Sulcular debridement (removal of diseased or | Sulcular debridement (curettage, removal of<br>diseased, infected, inflamed and necrosed soft |
| inflamed soft tissue in the periodontal<br>pocket to improve clinical indices | tissue in the periodontal pocket to improve clinical |
| including:, probe depth, attachment loss | indices including: gingival index, gingival bleeding |
| and tooth mobility) | index, probe depth, attachment loss and tooth |
| | mobility). |
| Removal of highly inflamed edematous tissue | Removal of highly inflamed edematous tissue |
| affected by bacteria penetration of the | affected by bacteria penetration of |
| pocket lining and junctional epithelium | the pocket lining and junctional epithelium. |
| | Picasso assisted new attachment procedure |
| | (cementum-mediated periodontal ligament |
| | new-attachment to the root surface in the absence |
| | of long junctional epithelium). |
| | |
.
:
{5}------------------------------------------------
# Comparison of Stlase and Zolar Photon Plus (Technical Specifications)
| Specifications | Dental Photonics StLase<br>(K11689) | Zolar--Photon Plus<br>(Subject Device) |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| Wavelength | | |
| | 980 nm | 980nm |
| Power | | |
| | 25 w | 10 W |
| Power range | 0.5-25w | 100mW to 10W |
| Increments | | |
| | 0.1-0.5w | 0.1 w |
| Operating modes | | |
| | Pulsed or CW | Pulsed or CW |
| Amining beam | 650 nm, 5mW | 650nm, 2 mW |
| Cooling | | |
| | Air cooled, closed control water | Air cooled |
| Weight | | |
| | around 1.8 KG | around 1.2 KG |
| Power requirements | | 120v/60 HZ or 240 V to charge- |
| | 120v/60 HZ or 240 V/50HZ | Battery |
| Sterilization method | Steam autoclave | Steam autoclave |
| Physical Media Source | Diode | Diode |
| Wattage | 0.5 to 25 watts | 0.1 to 10 watts |
| Control Panel | Color Touch screen | Colour Touch screen |
| Settings | | 22 Treatable & 3 customizable |
| Frequency | 20,000Hz | 1-5000Hz |
| Pulse Interval | 0.025ms - 3ms | 0.1ms-9.9s |
| Pulse Energy | Continuouse & Pulse Modes | Continuouse & Pulse Modes |
| Laser Clasification | Class 4 | Class 4 |
| Aiming Beam | 650nm, 5mW | 3 levels of Intensity |
| Foot Switch Operation | | Yes |
| Dental Applications | Deatal Soft Tissue indications;<br>General surgey for<br>incision/excision, vaporazation;<br>abalation and coagulation of<br>soft tissues;<br>Dental laser operation incl. light<br>activation of bleathing materils<br>for tooth whitenting. | Soft Tissue Surgery Periodontal<br>& Hygiene Endodontic,<br>Whitening / Bleaching. |
| Autoclave | Hand Piece Autoclavable | Hand Piece Autoclavable |
.
.
.
{6}------------------------------------------------
| Specifications | Picasso Lite/Picasso<br>(K102359) | Zolar-Photon/Photon Plus<br>(Subject Device) |
|--------------------------------|-----------------------------------------------------------------|----------------------------------------------|
| Physical Media Source | Diode | Diode |
| Wattage | 0.1 to 2.5 watts / 0.1 to 7<br>watts | 0.1 to 3.0W / 0.1 to 10 watts |
| Wavelength nm | 810 nm | 810 nm/980 nm |
| Control Panel | Color Touch Screen /<br>Universal Icon Membrane<br>Touch screen | Color Touch screen |
| Settings | 8 Treatable (prest values) /<br>Manual | 22/23 Treatable & customizable |
| Pulse Energy | Continuous & Pulse Modes | Continuous & Pulse Modes |
| Delivery Fiber Diameter | 200-1000um | 200-1000um |
| Laser Classification | Class 4 | Class 4 |
| Aiming Beam | Variables (adjustable) | Variables ( 3 settings) |
| Foot Switch Operation | Yes | Yes |
| Audible/Visual<br>Notification | Yes | Yes |
| Autoclave | Hand Piece Autoclavable | Hand Piece Autoclavable |
| Applicable standards | IEC 60825-1<br>IEC 60601-2-22 | IEC 60825-1<br>IEC 60601-2-22 |
Comparison of Picasso/Picasso Lite and Zolar Photon (Technical Specifications)
The Photon Plus / Photon Dental Diode Lasers are of a comparable type and substantially equivalent to the currently marketed Picasso / Picasso Lite (K102359) and stLase Dental Lasers (K111689). Although the subject devices are somewhat smaller and lighter than Picasso devices, the overall construction, materials and performance as well as key features are equivalent. The subject devices have the same intended uses as the predicate devices and the Photon Plus subject device includes features which are well established on the stLase Dental Laser systems.
## Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence 8. are as follows:
The subject devices have been tested for thermal, electrical and mechanical safety, and conform to applicable medical device safety standards, IEC 60825-1 and IEC 60601-2-22.
Testing was conducted to validate and verify that the Photon/Photon Plus devices met all design specifications and were substantially equivalent to the predicate devices.
Photon/Photon Plus have also been tested to assure compliance to the requirements of various published standards, including ANSI/AAMI ST79, AAMI TIR 12, ANSI/AAMI/ISO 17665-1, ANSI/AAMI ST77, ANSI/AAMI ST81, ISO 17664, ANSI/AAMI ST67, ANSI/AAMI ST8, EN 556.
## 9. Discussion of Clinical Tests Performed:
Not Applicable
{7}------------------------------------------------
### 10. Conclusion:
We have demonstrated that there are no significant differences between the subject devices, and the predicate devices, Photon/Photon Plus and the predicate devices, Picasso/Picasso Lite Dental Diode Lasers, K102359, and stLase Dental Laser, K111689, in terms of safety and effectiveness based on electrical, mechanical and environmental test results. Based on this information, the 8subject devices are substantially equivalent to the predicate devices.
{8}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/8/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Public Health Service
Letter dated: June 5, 2013
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
Zolar Technology & Mfg. Co., Inc. % MDI Consultants, Inc. Ms. Susan D. Goldstein-Falk 55 Northern Boulevard, Suite 200 Great Neck, New York 11021
Re: K122020 -
Trade/Device Name: Photon/Photon Plus Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general surgery and Plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: May 17, 2013 Received: May 20, 2013
Dear Ms. Goldstein-Falk:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{9}------------------------------------------------
Page 2 - Ms. Susan D. Goldstein-Falk
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Mark Nielkerson -S
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health-
Enclosure ·
{10}------------------------------------------------
## Page 1 of 3
# Indications for Use
510(k) Number (if known): K122020
# Device Name: Photon/Photon Plus
## Indications for Use for Photon:
The Photon is intended for use by dentists for excision, vaporization, ablation and coagulation of oral soft tissue procedures. The specific applications are as follows:
SOFT TISSUE: The metal handpiece with cutting fiber does not have any direct contact to the soft tissue, skin or cell during the operation.
- . Exposure of Unerupted teeth
- Fibroma removal .
- Frenectomy .
- . Gingival Troughing for crown impressions
- Gingivectomy .
- Gingivoplasty .
- Hemostasis and coagulation .
- Gingival incision and excision hemostasis .
- Implant recovery
- Incision and drainage of abscess
- Leukoplakia
- Operculectomy
- Oral papillectomies
## Pulpotomy as an adjunct to root canal therapy_
- Reduction of gingival hypertrophy
- Soft tissue crown lengthening
- Gingival bleeding index
- Excisional and incisional biopsies
- Treatment of aphthous ulcer, canker sores and herpetic .
- Vestibuloplasty.
## LASER PERIODONTAL PROCEDURES:
- · Sulcular debridement (removal of diseased or inflamed soft tissue in the periodontal pocket to improve clinical indices including:, probe depth, attachment loss and tooth mobility)
- Removal of highly inflamed edematous tissue affected by bacteria penetration of . the pocket lining and junctional epithelium
Neil R Ogden Nell R Ogden
2013.06.04 09:27:24:04 09:27:24 04:00
(Division Sign-Off) for MXM
Division of Surgical Devices
510(k) Number K122020
{11}------------------------------------------------
## Indications for Use for Photon Plus:
The Photon Plus is intended for use by dentists for excision, incision, vaporization, ablation and coagulation of oral soft tissue procedures. The specific applications are as follows:
SOFT TISSUE: The metal handpiece with cutting fiber does not have any direct contact to the soft tissue, skin or cell during the operation.
- Exposure of Unerupted teeth .
- Excision of lesion or removal of Granulation tissue .
- t Frenectomy
- ◆ Gingival Troughing for crown impressions
- . Gingivectomy
- Gingivoplasty �
- 0 Hemostasis and coagulation
- Gingival incision and excision hemostasis .
- Implant recovery .
- Incision and drainage of abscess .
- Leukoplakia ●
- Operculectomy .
- Oral papillectomies .
- Pulpotomy as an adjunct to root canal therapy
- Reduction of gingival hypertrophy
- Soft tissue crown lengthening
- Gingival bleeding index
- Excisional and incisional biopsies
- Treatment of aphthous ulcer, canker sores and herpetic
- Vestibuloplasty ...
Neil R Ogdens 2013.06.04 09:27-52 -04'00"
(Division Sign-Off) for MXM Division of Surgical Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________
{12}------------------------------------------------
# LASER PERIODONTAL PROCEDURES:
- Sulcular debridement (removal of diseased or inflamed soft tissue in the periodontal pocket to improve clinical indices including:, probe depth, attachment loss and tooth mobility)
- Debridement of diseased epithelial lining
## TOOTH WHITENING INDICATIONS
- Laser assisted whitening/bleaching of teeth .
- . Light activation for bleaching materials for teeth whitening.
Prescription Use X (Per 21 CFR 801 Subpart D)
OR
Over-The Counter Use (21 CFR 807 Subpart C)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Neil R Ogdena Ma
2013.06.04 09:28:26-04'00'_
(Division Sign-Off) for MXM Division of Surgical Devices 510(k) Number ___K122020_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.