Histoacryl Flexible topical skin adhesive is intended for topical application to hold closed easily approximated skin edges of minimum-tension wounds from clean surgical incisions and simple, thoroughly cleansed, trauma-induced lacerations. Histoacryl Flexible may be used in conjunction with, but not in place of, deep dermal sutures.
Device Story
Histoacryl Flexible is a sterile, liquid topical skin adhesive composed of n-butyl-2-cyanoacrylate monomer, softener, stabilizer, and colorant. Provided in single-patient-use plastic ampoules. Upon contact with water or water-containing substances (e.g., tissue), the monomer polymerizes to form a flexible film that bonds skin edges together. Used by clinicians for wound closure; acts as a microbial barrier while the film remains intact. Performance is equivalent to predicate cyanoacrylate adhesives. Benefits include maintaining wound approximation for minimum-tension wounds.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing, including lap shear strength (ASTM F2255-05), peel adhesion (ASTM F2256-05), impact strength (ASTM F2458-05), polymerization time, heat of polymerization, viscosity, and hydrolytic degradation. Microbial barrier properties evaluated in vitro using strike-through testing against gram-positive, gram-negative, motile, and non-motile organisms and fungi.
Technological Characteristics
Liquid n-butyl-2-cyanoacrylate monomer with softener, stabilizer, and D&C Violet #2. Sterile single-use ampoule. Biocompatibility per ISO 10993-5, -6, -10, -11. Performance standards: ASTM F2255-05, F2256-05, F2458-05. Sterilization: ETO (ampoule), gamma radiation (pouch), membrane filtration/aseptic filling (adhesive).
Indications for Use
Indicated for topical closure of minimum-tension wounds, including clean surgical incisions and simple, thoroughly cleansed, trauma-induced lacerations. May be used with deep dermal sutures.
Regulatory Classification
Identification
A tissue adhesive for the topical approximation of skin is a device intended for topical closure of surgical incisions, including laparoscopic incisions, and simple traumatic lacerations that have easily approximated skin edges. Tissue adhesives for the topical approximation of skin may be used in conjunction with, but not in place of, deep dermal stitches.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: “Tissue Adhesive for the Topical Approximation of Skin.” See § 878.1(e) of this chapter for the availability of this guidance document.(b)
*Tissue adhesive for non-topical use* —(1)*Identification.* A tissue adhesive for non-topical use, including adhesives intended for use in the embolization of brain arteriovenous malformation or for use in ophthalmic surgery, is a device used for adhesion of internal tissues and vessels.(2)
*Classification.* Class III (premarket approval). As of May 28, 1976, an approval under section 515 of the act is required before this device may be commercially distributed. See § 878.3 of this chapter.
Predicate Devices
Histoacryl and Histoacryl Blue Topical Skin Adhesive (K111959)
{0}------------------------------------------------
K121976 Page 1 of 3
#### 510(k) SUMMARY (as required by 21 CFR 807.92)
Histoacryl Flexible Topical Skin Adhesive April 15, 2013
COMPANY:
B.
Aesculap®, Inc. 3773 Corporate Parkway Center Valley, PA 18034 Establishment Registration Number: 2916714
CONTACT:
Kathy A. Racosky 610-984-9291 (phone) 610-791-6882 (fax) kathy racosky@aesculap.com
COMMON NAME: Topical Skin Adhesive
CLASSIFICATION NAME: Tissue Adhesive
REGULATION NUMBER: 878.4010
PRODUCT CODE: MPN
#### DEVICE DESCRIPTION
Histoacryl Flexible Topical Skin Adhesive is a sterile liquid topical skin adhesive composed of n-butyl-2-cyanoacrylate monomer, softener, stabilizer, and colorant (D&C Violet #2). It is provided in 0:5 ml single patient use plastic ampoules. Each ampoule is sealed within a foil pouch so the exterior of the ampoule can remain sterile. The tissue adhesives remain liquid until exposed to water or water-containing substances including tissue, after which it cures (polymerizes) and forms a film that bonds to the underlying surface.
In vitro studies have shown that Histoacryl Flexible acts as a barrier to microbial penetration as long as the adhesive film remains intact. Clinical studies were not conducted to demonstrate microbial barrier properties and a correlation between microbial barrier properties and a reduction in infection have not been established.
#### INDICATIONS FOR USE
Histoacryl Flexible topical skin adhesive is intended for topical application to hold closed easily approximated skin edges of minimum-tension wounds from clean surgical incisions and simple, thoroughly cleansed, trauma-induced lacerations. Histoacryl Flexible may be used in conjunction with, but not in place of, deep dermal sutures.
{1}------------------------------------------------
#### SUBSTANTIAL EQUIVALENCE
Aesculap. Inc. believes that the Histoacryl Flexible Topical Skin Adhesive is substantially equivalent to the Aesculap Histoacryl and Histoacryl Blue Topical Skin Adhesive (K111959) and Dermabond (P960052).
### PURPOSE FOR PREMARKET NOTIFICATION
The purpose for this submission is to gain marketing clearance for the new Histoacry Flexible Topical Skin Adhesive.
### TECHNOLIGICAL CHARACTERISTICS (compared to Predicate(s))
The technological characteristics of the Aesculap Histoacryl Flexible Topical Skin Adhesive are equivalent in performance to the predicate devices Histoacryl and Histoacryl Blue Topical Skin Adhesive, Aesculap Inc. (K111959) and Dermabond (P960052). The subject device is shown to be substantially equivalent and has the same performance characteristic to the predicate devices through comparison in technology, Indication For Use, mechanism of action, intended application and performance. Both Histoacryl and Histoacryl Blue Topical Skin Adhesive and Histoacryl Flexible Topical Skin Adhesive devices use n-butyl-2-cyanoacrylate to facilitate wound closure. Histoacryl Flexible is designed to bond to the skin to provide flexible wound closure maintaining wound approximation.
In vitro studies have shown that Histoacryl acts as a barrier to microbial penetration as long as the adhesive film remains intact. Clinical studies were not conducted to demonstrate microbial barrier properties and a correlation between microbial barrier properties and a reduction in infection have not been established.
The main differences between Histoacryl Flexible and Histoacryl are
- · a change in formulation to allow flexible wound closure
- · storage conditions at or below 77°F (25°C) versus 72°F (22°C)
#### Biocompatibility:
The biological evaluation of Histoacryl Flexible Topical Skin Adhesive device has been performed in accordance with ISO 10993, "Biological Evaluation of Medical Devices Part -1: Evaluation and Testing, for breached or compromised surfaces with blood contact for the wound closure and subsequent layer adhesives".
Biocompatibility tests were conducted for a "breached or compromised surface with blood contact device with prolonged contact duration of greater than 24 hours but less than 30 days". Tests include MEM cytotoxicity (ISO 10993-5), intracutaneous reactivity (ISO 10993-10), maximization sensitization (ISO 10993-10), systemic toxicity (ISO 10993-11), and muscle implantation (ISO 10993-6).
The results provide evidence that Histoacryl Flexible is safe and biocompatible for its intended use and therefore substantially equivalent to the predicate devices.
{2}------------------------------------------------
### Performance testing:
Testing was performed in accordance to FDA's Class II Special Control Guidance Document for Tissue Adhesive for the Topical Approximation of Skin to demonstrate that the Aesculap Histoacryl Flexible Topical Skin Adhesive is substantially equivalent to other predicate devices. The following comparative testing demonstrated substantially equivalent performance to Histoacryl and Histoacryl Blue Topical Skin Adhesive and Dermabond:
- Lap shear strength (ASTM F2255-05) .
- Peel adhesion strength (ASTM F 2256-05) .
- Impact strength (ASTM F2458-05) .
- Set (polymerization) time .
- Heat of polymerization .
- Viscosity .
- . GC Chemical Analysis
- · Hydrolytic degradation
- Ease of expression
Microbial barner testing was conducted using Histoacryl and Histoacryl Flexible. The method was a strike through test that was conducted with common organisms known to cause infections and represent gram positive, gram negative, motile and non-motile as well as fungi. The challenge was at a minimum concentration of 1 x 10° cfu.
#### Sterilization and Shelf-Life:
Sterilization of Histoacryl Flexible Topical Skin Adhesive is the same as the predicate device. Histoacryl and Histoacryl Blue Topical Skin Adhesive. The sterilization process consists of 1) ETO sterilization of the ampoule, 2) gamma radiation sterilization of the aluminum pouch; both are sterilized before 3) the sterility of the liquid topical skin adhesive which is guaranteed by membrane filtration and aseptic filling.
Real-time testing data has been generated to support this submission. Current data supports a 24 month shelf life.
#### Conclusion:
Based on the nonclinical testing Histoacryl Flexible Topical Skin Adhesive has been demonstrated to be substantially equivalent to Histoacryl and Histoacryl Blue Topical Skin Adhesive and Dermabond.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
Letter dated: May 17, 2013
Aesculap, Inc. % Ms. Kathy A. Racosky Senior Regulatory Affairs Specialist 3773 Corporate Parkway Center Valley, Pennsylvania 18034
Re: K121976
Trade/Device Name: Aesculap Histoacryl Flexible Topical Skin Adhesive Regulation Number: 21 CFR 878.4010 Regulation Name: Tissue adhesive Regulatory Class: Class II Product Code: MPN Dated: April 15, 2013 Received: April 18, 2013
Dear Ms. Racosky:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
Page 2 - Ms. Kathy A. Racosky
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm11.5809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## Mark N. Melkerson -S
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page 1 of 1
#### INDICATIONS FOR USE STATEMENT A.
**510(k) Number:** K121976
Device Name: Aesculap Histoacryl Flexible Topical Skin Adhesive
#### Indications for Use:
Histoacryl Flexible topical skin adhesive is intended for topical application to hold closed easily approximated skin edges of minimum-tension wounds from clean surgical incisions and simple, thoroughly cleansed, trauma-induced lacerations. Histoacryl Flexible may be used in conjunction with, but not in place of, deep dermal sutures.
Prescription Use _____________________________________________________________________________________________________________________________________________________________
and/or Over-the-Counter Use (21 CFR 801 Subpart C)
(Part 21 CFR 801 Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF
NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
# Jiyoung Dang -S
(Division Sign-Off) Division of Surgical Devices 510(k) Number: K121976
178
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.