K121941 · Arrow International (Subsidiary of Teleflex Inc.) · LJS · Sep 7, 2012 · General Hospital
Device Facts
Record ID
K121941
Device Name
ARROW PICC POWERED BY ARROW VPS STYLET
Applicant
Arrow International (Subsidiary of Teleflex Inc.)
Product Code
LJS · General Hospital
Decision Date
Sep 7, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5970
Device Class
Class 2
Indications for Use
The Pressure Injectable PICC is indicated for short or long term peripheral access to the central venous system for intravenous therapy, blood sampling, infusion, power injection of contrast media and allows for central venous pressure monitoring. The maximum pressure of power injectors used with the pressure injectable PICC catheter may not exceed 300 psi. The maximum pressure injection flow rate ranges from 4 ml/sec to 6 ml/sec. Refer to the product specific labeling for the maximum pressure injection flow rate for the specific lumen being used for pressure injection. The VPS Stylet and Console are indicated for guidance and tip positioning for central venous catheters. The Stylet provides stiffness for use in placement of the catheter, intravascular capability for ECG detection and recording and intravascular ultrasound for catheter guiding and positioning. The VPS Stylet, when used with the VPS Console, provides real-time catheter tip location information by using the patient's physiological (cardiac electrical activity and blood flow) information. When the VPS System guidance indicator shows a Blue Bullseye, the catheter tip is in the desired location. The VPS System is indicated for use as alternative method to fluoroscopy or chest x-ray for central venous catheter tip placement confirmation in adult patients when a steady Blue Bullseye is obtained. NOTE: If a steady Blue Bullseye is not obtained, standard hospital practice should be followed to confirm catheter tip location. Limiting but not contraindicated situations for this technique are in patients where alterations of cardiac rhythm change the presentation of the P-wave as in atrial flutter, severe tachycardia and pacemaker-driven rhythm, and in central venous catheterization procedures performed through femoral or saphenous vein access which change the presentation of the P-wave. In such patients, who are easily identifiable prior to central venous catheter insertion, the use of an additional method is required to confirm catheter tip location.
Device Story
Device combines pressure-injectable PICC with VPS Stylet for real-time catheter tip positioning. PICC: polyurethane catheter with Blue Flex Tip for central venous access. VPS Stylet: polyimide tube containing distal Doppler sensor and ivECG sensing wire. Stylet inserts into catheter lumen; connects to VPS Console. Console processes physiological signals (cardiac electrical activity and blood flow) to provide real-time tip location. Output: 'Blue Bullseye' indicator on console display confirms desired tip location. Used in clinical settings by healthcare providers as alternative to fluoroscopy/chest X-ray for tip placement confirmation. Benefits: reduces radiation exposure; provides real-time guidance during insertion.
Clinical Evidence
Bench testing only. Performance testing included catheter tensile strength and burst testing, and stylet tensile, Hi-pot, continuity, and removal testing. No clinical data presented.
Technological Characteristics
Materials: radiopaque polyurethane catheter, polyimide stylet, stainless steel wire. Sensing: Doppler sensor and ivECG sensing wire. Form factor: 4-6 Fr, 1-3 lumen, 40-55 cm catheters. Connectivity: wired connection to VPS Console. Sterilization: sterile kit configuration.
Indications for Use
Indicated for adult patients requiring short or long-term central venous access for IV therapy, blood sampling, infusion, power injection of contrast media, and CVP monitoring. Contraindicated/limited in patients with cardiac rhythm alterations (atrial flutter, severe tachycardia, pacemaker-driven rhythm) or femoral/saphenous access, where P-wave presentation is altered.
Regulatory Classification
Identification
A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”
Predicate Devices
Vascular Positioning System (VPS System) Stylet (K103255)
ArrowADVANTAGE Pressure Injectable Peripherally Inserted Central Catheter (PICC) (K113277)
{0}------------------------------------------------
K121941
pg 1 of 3
## 510(K) SUMMARY
7 2012
SEP
### Submitter Information
6
| Name: | Arrow International, Inc (subsidiary of Teleflex Inc.) |
|-------------------|--------------------------------------------------------|
| Address: | 2400 Bernville Road<br>Reading, PA 19605-9607 USA |
| Contact Person: | Elizabeth Duncan<br>Sr. Regulatory Affairs Specialist |
| Telephone Number: | (610) 378-0131 Extension 603220 |
| Fax Number: | (610) 478-3179 |
| Email: | elizabeth.duncan@teleflex.com |
| Date Prepared: | June 28, 2012 |
#### Device Name
Device Trade Name: Arrow PICC powered by Arrow VPS Stylet Common Name, Catheter: Peripherally Inserted Central Catheter (PICC) Common Name, Stylet: Catheter, Ultrasound, Intravascular
Classification Name, Catheter: Percutaneous, implanted, long-term intravascular catheter per 21 CFR: 880.5970
Classification Name, Stylet: Diagnostic Intravascular Catheter per 21 CFR 870.1200
### Predicate Devices
- K103255: Vascular Positioning System (VPS System) Stylet �
- K 113277: ArrowADVANTAGE5 Pressure Injectable Peripherally Inserted Central . Catheter (PICC)
- K080604: Pressure Injectable PICC .
#### Device Description
The Arrow PICC powered by Arrow VPS Stylet has the following characteristics:
- . 4 Fr, 1-Lumen, 40-55 cm catheter preloaded with VPS Stylet
- 5 Fr, 2-Lumen 40-55 cm catheter preloaded with VPS Stylet ●
- 6 Fr, 3-Lumen 40-55 cm catheter preloaded with VPS Stylet .
The Arrow PICC is pre-loaded with the Arrow VPS Stylet and will be provided in sterile kit configurations.
The Arrow Pressure Injectable PICC is a short-term or long-term, single use catheter designed to provide access to the central venous system. It consists of a non-tapered, radiopaque
{1}------------------------------------------------
K121941
p3 2 of 3
polyurethane extruded catheter body with a softer, contoured Blue Flex Tip. The catheter can be used for the injection of contrast media.
The Arrow VPS Stylet is a polyimide tube containing a Doppler sensor on a coax cable and an intravascular electrocardiogram (ivECG) signal sensing stainless steel wire. The Doppler sensor and the exposed portion of the ivECG are located at the distal end of the stylet and are used to detect and transmit physiological information to the VPS Console for analysis. The proximal end contains a connector to the VPS Console or to an extension cable that in turn connects to the VPS Console. The stylet was designed to be inserted and removed from any catheter with a luminal diameter of at least 0.021 inch. For user convenience, Arrow has created the Arrow PICC powered by Arrow VPS Stylet in which the Arrow PICC is provided pre-loaded with the Arrow VPS Stylet.
## Indications for Use and Intended Use
The Pressure Injectable PICC is indicated for short or long term peripheral access to the central venous system for.intravenous therapy, blood sampling, infusion, power injection of contrast media and allows for central venous pressure monitoring. The maximum pressure of power injectors used with the pressure injectable PICC catheter may not exceed 300 psi. The maximum pressure injection flow rate ranges from 4 ml/sec to 6 ml/sec. Refer to the product specific labeling for the maximum pressure injection flow rate for the specific lumen being used for pressure injection.
The VPS Stylet and Console are indicated for guidance and tip positioning for central venous catheters. The Stylet provides stiffness for use in placement of the catheter, intravascular capability for ECG detection and recording and intravascular ultrasound for catheter guiding and positioning. The VPS Stylet, when used with the VPS Console, provides real-time catheter tip location information by using the patient's physiological (cardiac electrical activity and blood flow) information. When the VPS System guidance indicator shows a blue bullseve, the catheter tip is in the desired location.
The VPS System is indicated for use as an alternative method to fluoroscopy or chest x-ray for central venous catheter tip placement confirmation in adult patients when a steady blue bullseve is obtained. NOTE: If a steady Blue Bullseye is not obtained, standard hospital practice should be followed to confirm catheter tip location.
Limiting but not contraindicated situations for this technique are in patients where alterations of cardiac rhythm change the presentation of the P-wave as in atrial flutter, severe tachycardia and pacemaker-driven rhythm, and in central venous catheterization procedures performed through femoral or saphenous vein access which change the presentation of the Pwave. In such patients, who are easily identifiable prior to central venous catheter insertion, the use of an additional method is required to confirm catheter tip location.
{2}------------------------------------------------
### Technological Characteristics and Substantial Equivalence
The Arrow PICC powered by Arrow VPS Stylet is substantially equivalent to the Vascular Positioning System (VPS System) Stylet (K103255), ArrowADVANTAGE Pressure Injectable Peripherally Inserted Central Catheter (PICC) (K113277) and the Pressure Injectable PICC (K080604) in terms of indications for use, design, manufacturing process, functional performance, and materials of construction. The subject device combines the predicate Arrow VPS Stylet and the Arrow Pressure Injectable PICCs; there is no change to the previously cleared devices or their indications for use.
#### Nonclinical Testing
The results of the catheter performance testing: tensile strength, and burst and the stylet performance testing: tensile, Hi-pot, Continuity and removal testing demonstrate that the Arrow PICC powered by Arrow VPS Stylet is as safe, as effective and performs comparably to the predicate VPS stylet and Pressure Injectable PICC.
#### Conclusions
The predicate and the subject devices have the same indications for use, intended use, design, materials, and are manufactured using the same processes. The results of the testing performed have demonstrated that combining the two previously cleared devices does not raise new issues of safety or effectiveness and therefore the combination is considered substantially equivalent to the cited predicate devices.
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
SEP 7 2012
Arrow International, Inc. c/o Ms. Elizabeth Duncan Sr. Regulatory Affairs Specialist 2400 Bernville Road Reading, PA 19605
Re: K121941
Trade/Device Name: Arrow PICC powered by Arrow VPS Stylet Regulation Number: 21 CFR 880.5970 Regulation Name: Percutaneous Implanted Long-Term Intravascular Catheter Regulatory Class: Class II Product Code: LJS, OBJ Dated: August 29, 2012 Received: August 30, 2012
Dear Ms. Duncan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
## Page 2 - Ms. Elizabeth Duncan
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): _
## Device Name: Arrow PICC powered by Arrow VPS Stylet
### Indications for Use:
The Pressure Injectable PICC is indicated for short or long term peripheral access to the central venous system for intravenous therapy, blood sampling, infusion, power injection of contrast media and allows for central venous pressure monitoring. The maximum pressure of power injectors used with the pressure injectable PICC catheter may not exceed 300 psi. The maximum pressure injection flow rate ranges from 4 ml/sec to 6 ml/sec. Refer to the product specific labeling for the maximum pressure injection flow rate for the specific lumen being used for pressure injection.
The VPS Stylet and Console are indicated for guidance and tip positioning for central venous catheters. The Stylet provides stiffness for use in placement of the catheter, intravascular capability for ECG detection and recording and intravascular ultrasound for catheter guiding and positioning. The VPS Stylet, when used with the VPS Console, provides real-time catheter tip location information by using the patient's physiological (cardiac electrical activity and blood flow) information. When the VPS System guidance indicator shows a Blue Bullseye, the catheter tip is in the desired location.
The VPS System is indicated for use as alternative method to fluoroscopy or chest x-ray for central venous catheter tip placement confirmation in adult patients when a steady Blue Bullseye is obtained. NOTE: If a steady Blue Bullseye is not obtained, standard hospital practice should be followed to confirm catheter tip location.
Limiting but not contraindicated situations for this technique are in patients where alterations of cardiac rhythm change the presentation of the P-wave as in atrial flutter, severe tachycardia and pacemaker-driven rhythm, and in central venous catheterization procedures performed through femoral or saphenous vein access which change the presentation of the P-wave. In such patients, who are easily identifiable prior to central venous catheter insertion, the use of an additional method is required to confirm catheter tip location.
Prescription Use X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number k121941
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.