BLUE, LATEX POWDER FREE EXAMINATION GLOVES, TESTED FOR USE WITH CHEMOTHERAPY DRUGS WITH PROTEIN LABLING CLAIM (50UG/DM
Applicant
Northstar Healthcare Holdings
Product Code
LZC · General Hospital
Decision Date
Sep 20, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Attributes
3rd-Party Reviewed
Indications for Use
The examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation of Chemotherapy Drugs:
Device Story
Disposable latex examination glove; textured surface; blue color. Designed for use by medical/dental professionals to provide barrier protection against contaminants. Tested for permeation resistance against specific chemotherapy drugs per ASTM D6978-05. Provides protection during handling of hazardous drugs; breakthrough times vary by agent (e.g., Carmustine and Thiotepa show low breakthrough times). Used in clinical/medical settings. Benefits include reduced risk of cross-contamination and protection from chemical exposure during chemotherapy administration.
Clinical Evidence
Bench testing only. Compliance demonstrated with ASTM D3578-05 (dimensions, physical properties), ASTM D5151-06 (pinholes), ASTM D6124-06 (powder residual), ASTM D5712-10 (protein levels), and ISO 10993-1/10 (biocompatibility, skin irritation, sensitization). Chemotherapy drug permeation tested per ASTM D6978-05.
Indicated for use as a disposable patient examination glove worn by medical personnel to prevent cross-contamination between patient and examiner. Tested for use with specific chemotherapy drugs.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Latex Powder Free Examination Gloves (Blue) Tested For Use With Chemotherapy Drugs (K083409)
Submission Summary (Full Text)
{0}------------------------------------------------
Abbreviated 510(k) Premarket Notification Latex Powder Free Examination Gloves
510(k) SUMMARY
5.
Section 5, 510(k) Summary Page 1 of 3
? �。
K121926/.
| DATE: | February 17, 2010 | | SEP 2 0 2012 |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| OWNER: | McKesson Medical Surgical International<br>70 Sir John Rogerson's Quay<br>Dublin 2, Ireland | | |
| OFFICIAL CORRESPONDENT: | Anthony L. Giaccio<br>Manager, Quality Systems and Regulatory Affairs<br>Telephone: 815-385-0100<br>Fax: 815-385-0114<br>Email: Anthony.Giaccio@mckesson.com | | |
| DEVICE NAME: | Trade Name: | Textured, Blue, Latex Powder Free Examination<br>Gloves, Tested For Use With Chemotherapy<br>Drugs With Protein Labeling Claim (50μg/dm2<br>Or Less Of Water Soluble Protein) | |
| | Common Name: | Patient Examination Gloves | |
| | Classification: | Patient Examination Gloves | |
| | Class: | Class I | |
| | Product Code: | LZC | |
# PREDICATE DEVICE(S):
| Predicate<br>510(k) | Device Name | Indication | Clearance<br>Date | Company |
|---------------------|------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|---------------------------------|
| K083409 | Latex Powder Free<br>Examination Gloves<br>(Blue) Tested For<br>Use With<br>Chemotherapy Drugs | The latex examination glove (Tested for<br>Use with Chemotherapy Drugs) is a<br>disposable device intended for medical<br>and dental purposes that is worn on the<br>examiner's hand to prevent<br>contamination between patient and<br>examiner. | 29 Jul 2009 | WRP ASIA<br>PACIFIC SDN,<br>BHD |
DEVICE DESCRIPTION: Textured, Blue, Latex Powder Free Examination Gloves, Tested For Use With Chemotherapy Drugs With Protein Labeling Claim (50µg/dm² Or Less Of Water Soluble Protein).
{1}------------------------------------------------
# STATEMENT OF INTENDED USE:
The examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
i
In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation of Chemotherapy Drugs:
### Chemotherapy Drug Permeation (Breakthrough Detection Time in Minutes)
| Chemotherapy Drug | Average Breakthrough<br>Detection Time (Min) |
|-------------------------------------------|----------------------------------------------|
| Carmustine (BCNU) (3.3 mg/mL) | 15.4 |
| Cisplatin (1.0 mg/mL) | No breakthrough up to 240 min |
| Cyclophosphamide (Cytoxan)<br>(20.0 g/mL) | No breakthrough up to 240 min |
| Dacarbazine (DTIC) (10.0<br>mg/mL) | No breakthrough up to 240 min |
| Doxorubicin Hydrochloride (2.0<br>mg/mL) | No breakthrough up to 240 min |
| Etoposide (20.0 mg/mL) | No breakthrough up to 240 min |
| Fluorouracil (Adrucil) (50.0<br>mg/mL) | No breakthrough up to 240 min |
| Methotrexate (25.0 mg/mL) | No breakthrough up to 240 min |
| Mitomycin C (0.5mg/mL) | No breakthrough up to 240 min |
| Paclitaxel (Taxol) (6.0mg/mL) | No breakthrough up to 240 min |
| Thiotepa (10.0 mg/mL) | 1.6 |
| Vincristine Sulfate (1.0 mg/mL) | No breakthrough up to 240 min |
The maximum testing time is 240 minutes. Please note that the following drugs, Carmustine and Thiotepa, have extremely low permeation time of less than 30 minutes.
The Latex Powder Free Examination Glove is substantially equivalent to the predicate device with regard to physical characteristics, design, product features, and intended use. Both gloves are made with latex using similar manufacturing processes. In addition, both gloves have been tested for use with chemotherapy drugs.
## TECHNOLOGICAL CHARACTERISTICS:
{2}------------------------------------------------
Abbreviated 510(k) Premarket Notification Latex Powder Free Examination Gloves
Section 5, 510(k) Summary Page 3 of 3
:
## ASSESSMENT OF NONCLINICAL DATA:
| Characteristic | Standard | Device<br>Performance |
|--------------------------|--------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| Dimension | ASTM Standard D3578-05 | Meets |
| Physical Properties | ASTM Standard D3578-05 | Meets |
| Freedom from<br>Pinholes | 21 CFR 800.20; ASTM D5151-<br>06 | Meets |
| Powder Residual | ASTM Standard D6124-06 | Meets<br>Results generated<br>values below 2mg of<br>residual powder |
| Protein Level | ASTM Standard D5712-10 | Meets<br>Results generated<br>values below 50<br>mcg/mg of protein |
| Biocompatibility | Biological Evaluation of Medical<br>Devices part 1: Evaluation and<br>Testing (ISO 10993-1:2009) | Meets |
| | Primary Skin Irritation in rabbits<br>(ISO 10993-10:2010) | Gloves are non-<br>irritating |
| | Dermal Sensitization in the<br>guinea pig (ISO 10993-10:2010) | Gloves do not display<br>any potential for<br>sensitization |
### CONCLUSIONS:
The Latex Powder Free Examination Gloves meet the requirements of established standards ASTM D3578-05, ASTM D5712-10, ASTM D5151-06, ASTM D6124-06 ISO 10993-1:2009 and ISO 10993-10:2010.
Based on the comparison of intended use, design, materials and performance, the Latex Powder Free Examination Gloves Tested for Use With Chemotherapy Drugs are substantially equivalent to the predicate device.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
SEP 20 2012
Northstar Healthcare Holdings C/O Mr. Ned Devine Senior Staff Engineer Underwriters Laboratories. Incorporated 333 Pfingsten Road Northbrook, Illinois 60062
Re: K121926
Trade/Device Name: Textured, Blue, Latex Powder Free Examination Gloves, Tested For Use With Chemotherapy Drugs With Protein Labeling Claim (50 ug/dm2 Or Less Of Water Soluble Protein)
Regulation Number: 21 CFR 880.6250
Regulation Name: Patient Examination Gloves
Regulatory Class: Class I
Product Code: LZC
Dated: September 4, 2012
Received: September 5, 2012
Dear Mr. Devine:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2- Mr. Devine
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
For Scott Robertson
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{5}------------------------------------------------
### INDICATIONS FOR USE
#### 510(k) Number (if known): K121926
#### Device Name: Textured, Blue, Latex Powder Free Examination Gloves, Tested For Use With Chemotherapy Drugs With Protein Labeling Claim (50ug/dm2 Or Less Of Water Soluble Protein).
Indications for Use: The examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation of Chemotherapy Drugs:
| Chemotherapy Drug and Concentration | Minimum Breakthrough Detection Time in<br>Minutes |
|-----------------------------------------|---------------------------------------------------|
| Carmustine (BCNU)(3.3 mg/mL) | 15.4 |
| Cisplatin (1.0 mg/mL) | >240 |
| Cyclophosphamide (Cytoxan) (20.0 mg/mL) | >240 |
| Dacarbazine (DTIC) (10.0 mg/mL) | >240 |
| Doxorubicin Hydrochloride (2.0 mg/mL) | >240 |
| Etoposide (20.0 mg/mL) | >240 |
| Fluorouracil (Adrucil) (50.0 mg/mL) | >240 |
| Methotrexate (25 mg/mL) | >240 |
| Mitomycin C (0.5mg/mL) | >240 |
| Paclitaxel (taxol) (6.0 mg/mL) | >240 |
| Thiotepa (10.0 mg/mL) | 1.6 |
| Vincristine Sulfate (1.0 mg/mL) | >240 |
The maximum testing time is 240 minutes. Please note that the following drugs, Carmustine (BCNU) and Thiotepa have extremely low permeation time of less than 30 minutes.
AND/OR
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use X (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
ion of Anestheslology, General Hospital Infection Control, Dental Devices
510(k) Number;
K121926
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.