K121878 · Medtronic, Inc. · OCL · Nov 2, 2012 · General, Plastic Surgery
Device Facts
Record ID
K121878
Device Name
CARDIOBLATE CRYOFLEX SURGICAL ABLATION PROBES
Applicant
Medtronic, Inc.
Product Code
OCL · General, Plastic Surgery
Decision Date
Nov 2, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Cardioblate CryoFlex Surgical Ablation System is intended for minimally invasive cardiac surgical procedures, including the treatment of cardiac arrhythmias. The Cardioblate CryoFlex 7 cm, 10 cm, and 10-S probes plus the Clamp and Surgical Ablation Console freeze target tissue and block the electrical conduction pathways by creating an inflammatory response and cryonecrosis.
Device Story
System uses argon-based cryogen for controlled freezing of cardiac tissue; intended for minimally invasive cardiac surgery. Components include console, tank regulator, and single-use disposable probes (7cm, 10cm, 10-S). Probes feature heat-treated stainless steel shafts, malleable ablation segments with bellows for kink resistance, and integrated thermocouples for temperature monitoring. Surgeon shapes probe shaft for stability in operating field; movable insulative sleeve adjusts ablation zone size. Probe connects to console via hose; system creates cryonecrosis to block electrical conduction pathways. Used by surgeons in operating room; output is thermal energy (freezing) to treat arrhythmias. Benefits include fast, controlled tissue ablation.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by verification and validation bench testing.
Technological Characteristics
Argon-based cryosurgical system. Probes constructed of heat-treated stainless steel. Features include integrated thermocouples, bellows-configured ablation segments, and movable insulative sleeves. System includes console, tank regulator, and pressure transducer cables. Sterile, single-use disposable probes.
Indications for Use
Indicated for patients undergoing minimally invasive cardiac surgical procedures requiring treatment of cardiac arrhythmias. Contraindications not specified.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
K 121878
### NOV 2 2012
# 510(k) Summary
#### General Information l.
#### Submitter Information a.
| Submitter's Name and Address | |
|------------------------------|--------------------------|
| | Medtronic, Inc. |
| | Medtronic Cardiovascular |
| | 8200 Coral Sea Street NE |
| | Mounds View, MN 55112 |
Contact Person
Kari Christianson Senior Regulatory Affairs Specialist Tel: 763-514-9529 FAX: 763-526-8377 Email: kari.j.christianson@medtronic.com
Date of Summary
June 27, 2012
Cardioblate CryoFlex Surgical Ablation System Proprietary Name of Device
Common/Usual Name Cryosurgical System
Classification Name
Classification
Class II, per 21 CFR 878.4350
Product Code:
Predicate Device
OCL
SurgiFrost 7cm, 10cm Cryosurgical Devices (K053436)
Surgical Device, for Ablation of Cardiac Tissue
Page 1 of 3
Page 15 of 95
{1}------------------------------------------------
#### Device Description II.
#### System and Components a.
The Medtronic Cardioblate® CryoFlex® Surgical Ablation Probes, Console, and accessories are used together as a system. The system is composed of the following components:
| Components | Model number |
|-------------------------------------------|--------------|
| Cardioblate CryoFlex Console | 65CS1 |
| Cardioblate CryoFlex Console, Refurbished | R65CS1 |
| Cardioblate CryoFlex 7cm Probe | 60SF7 |
| Cardioblate CryoFlex 10cm Probe | 60SF2 |
| Cardioblate CryoFlex 10S Probe | 60SF3 |
| Tank Carrier | 65TC1 |
| Tank Regulator | 65RAXNA |
| Tank Regulator, Refurbished | R65RAXNA |
| Pressure Transducer Cable, 6ft. (1.8m) | 67PS6 |
| Gas Hose, 8ft (2.4m) | 67H08 |
| Power Cord, North American. | 671PCNA |
## b. System Overview
The Medtronic Cardioblate CryoFlex Surgical Ablation System is a crvoablative surgical device that is used in the cryosurgical treatment of cardiac arrhythmias. The system utilizes an argon-based cryogen for fast, controlled freezing.
The Cardioblate CryoFlex Probes are single use, disposable cryoprobes that are designed for use with the system. Each probe has an integrated thermocouple for monitoring temperature at is ablation segment. The probes are supplied sterile, cannot be reused or re-sterilized, and are available in 7cm, 10cm, and 10-S (slightly stiffer) versions.
The probes shaft and ablation segment are made of specially heat-treated stainless steel. The shaft was designed with enough malleability that the surgeon can easily shape it while still maintaining enough stiffness to ensure its stability in the operating field.
The malleable ablation segment has a bellows configuration, which provides kink resistance and thermal performance. A movable insulative sleeve on the shaft allows the surgeon to vary the size of the ablation zone.
Page 2 of 3
Page 16 of 95
Special 510(k): Modification to K053436 Medtronic . -
{2}------------------------------------------------
The probe has a 3m (10ft) connection hose which is to be handed out of the sterile field to an operating room nurse for connection to the control panel. Note that once the probe is connected to the control panel, it should not be disconnected until the end of the procedure because it cannot be reconnected. The probe is disabled once it is disconnected.
### Indications for Use 111.
The Cardioblate CryoFlex Surgical Ablation System is intended for minimally invasive cardiac surgical procedures, including the treatment of cardiac arrhythmias. The Cardioblate CryoFlex 7 cm, 10 cm, and 10-S probes plus the Clamp and Surgical Ablation Console freeze target tissue and block the electrical conduction pathways by creating an inflammatory response and cryonecrosis.
There is no change to the indications for use as compared to the predicate device.
#### IV. Comparison to Predicate Devices
The Medtronic Cardioblate CryoFlex Surgical Ablation system is substantially equivalent to the predicate surgical ablation system, has the same intended use and the same design with the exception of an alternate insulative sheath component on the 7cm, 10cm and 10S models (60SF7, 60SF2, and 60SF3, respectively), and a modified heat exchange component on the 10S model (60SF3).
#### Summary of Performance Data V.
Verification and validation testing has demonstrated that the Medtronic CryoFlex Surgical Ablation System is substantially equivalent to the predicates.
#### VI. Conclusion
Based on the accumulated technical information, intended use, verification tests and performance data provided, the Cardioblate CryoFlex Surgical Ablation disposable probes and console are substantially equivalent to the currently marketed predicate devices.
Page 3 of 3
Special 510(k): Modification to K053436 Medtronic ( . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
{3}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring three abstract human figures, possibly representing individuals or families, with flowing lines suggesting movement or progress.
### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Medtronic Inc. c/o Mr. Kari Christianson Senior Regulatory Affairs Specialist 8200 Coral Sea Street NE, Mail Stop MVS83 Mounds View, MN 55112
NOV. 2 2012
Re: K121878
Trade Name: Cardioblate CryoFlex Surgical Ablation Probes Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II (two) Product Code: OCL Dated: June 27, 2012 Received: June 28, 2012
Dear Mr. Christianson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 – Mr. Kari Christianson
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
15 Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K 121878
# Indications for Use
510(k) Number (if known):
Device Name: Cardioblate CryoFlex Surgical Ablation System
Indications For Use: The Cardioblate CryoFlex Surgical Ablation System is intended for minimally invasive cardiac surgical procedures, including the treatment of cardiac arrhythmias. The Cardioblate CryoFlex 7 cm, 10 cm, and 10-S probes plus the Clamp and Surgical Ablation Console freeze target tissue and block the electrical conduction pathways by creating an inflammatory response and cryonecrosis.
Prescription Use _ X
AND/OR
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of *_*_
Division Sign-Off) vision of Cardiovascular Devices
510(k) Number K121878
Page 1 of 1
Page 6 of 95
Special 510(k): Modification to K053436 Medtronic Confidential
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.