The Vertecem II Bone Cement is used for the fixation of pathological fractures of the vertebral body using Vertebroplasty or Kyphoplasty procedures. Painful vertebral compression fractures of the vertebral body may result from osteoporosis, benign lesions (hemangioma), or malignant lesions (metastatic cancers, myeloma).
Device Story
Vertecem II Mixing Kit is a two-component, self-polymerizing PMMA bone cement system. Components consist of a powder (polymethyl methacrylate/acrylate, radio-opacifier, benzoyl peroxide initiator) and a liquid (methyl methacrylate, DmpT activator, stabilizer). Upon mixing, DmpT activates the initiator, triggering exothermic polymerization to form a viscous, radio-opaque, injectable paste. Used by physicians in clinical settings (OR) during Vertebroplasty or Kyphoplasty procedures. The cement is injected into the vertebral body to stabilize lesions and fractures. Curing occurs at room temperature; heat is generated during the setting process. Benefits include stabilization of vertebral compression fractures and pain relief.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing only, including physical, chemical, and mechanical comparative performance testing against predicate devices.
Technological Characteristics
Two-component PMMA bone cement system. Powder: polymethyl methacrylate/acrylate, radio-opacifier, benzoyl peroxide. Liquid: methyl methacrylate, DmpT activator, stabilizer. Self-polymerizing at room temperature via exothermic reaction. Radio-opaque. Sterile packed components.
Indications for Use
Indicated for fixation of pathological vertebral body fractures via Vertebroplasty or Kyphoplasty. Patient population includes individuals with painful vertebral compression fractures secondary to osteoporosis, benign lesions (hemangioma), or malignant lesions (metastatic cancers, myeloma).
Regulatory Classification
Identification
Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”
{0}------------------------------------------------
| aap Biomaterials GmbH<br>Lagerstraße 11 – 15<br>64807 Dieburg<br>Germany | Vertecem II Mixing Kit | K121876 |
|--------------------------------------------------------------------------|-------------------------------------------------|---------|
| 5. 510(k) Summary | 164-0008-01<br>Date of issue:<br>2 2. Juni 2012 | |
| 510(k) Premarket Notification PO-32 | page 1 of 3 | |
.
# 5. 510(k) summary
:
OCT 1 9 2012
:
:
| Preparation date: | June 22, 2012 |
|-------------------|------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | aap Biomaterials GmbH<br>Lagerstraße 11-15<br>64807 Dieburg<br>Germany<br>Phone: +49 6071 / 929-0<br>Fax: +49 6071 / 929-100 |
| Contact person: | Volker Stirnal |
| Trade name: | Vertecem II Mixing Kit |
| Common name: | PMMA Bone Cement (For Vertebroplasty) |
| Classification: | Polymethylmethacrylate (PMMA) Bone Cement<br>21 CFR 888.3027, Class II |
| Product Code: | LOD, NDN |
| Panel: | Orthopedics |
# Predicate device to which substantial equivalence is claimed:
| Manufacturer | Device Name | 510(k) # |
|------------------|--------------|-----------|
| Teknimed S.A.S. | Vertecem | (K090435) |
| Manufacturer | Device Name | 510(k) # |
| aap Biomaterials | BonOs Inject | (K090460) |
{1}------------------------------------------------
| | | K 121876 |
|--------------------------------------------------------------------------|----------------------------------------------------------------------------------------|---------------------------------------------------------------|
| aap Biomaterials GmbH<br>Lagerstraße 11 – 15<br>64807 Dieburg<br>Germany | Vertecem II Mixing Kit<br><br>5. 510(k) Summary<br>510(k) Premarket Notification PO-32 | 164-0008-01<br>Date of issue:<br>22. Juni 2012<br>page 2 of 3 |
### Device description:
Vertecem II is a radio-opaque, injectable acrylic bone cement used for the treatment of pathological fractures of the vertebral body using a Vertebroplasty or Kyphoplasty procedure.
The Vertecem II Mixing Kit consists of sterile packed components. When the cement components are mixed together, they become a self hardening, radioopaque bone cement.
## Scientific concepts, significant physical and performance characteristics:
Bone cements in general are self-polymerizing two-component systems comprising a powder and a liquid which polymerize at room temperature immediately after they are mixed together.
The major powder component is polymethyl methacrylate / acrylate. Furthermore a radio-opacifier and benzoyl peroxide as initiator is included.
The liquid mainly consists of methyl methacrylate. It is furthermore comprised of an activator and a stabilizer to prevent premature polymerization.
When the powder and liquid components are mixed together, the activator DmpT, contained in the liquid activates the initiator in the powder component. This reaction starts the polymerization of the MMA, which is bonded with the polymer powder during ongoing polymerization. A description of polymerization technology is depictured in section 10- Executive summary, annex 10 - D.
As a result, a viscous injectable paste is obtained which can be introduced into a vertebral body using a suitable application system. Heat is generated during setting as a result of the progressive polymerization and exothermic reaction respectively. After curing the bone cement is able to stabilize the vertebral lesions and vertebral compression fractures. The setting or curing time is greatly influenced by the temperature of the components and environment, which is common for all acrylic bone cements.
#### Statement of the intended use:
The Vertecem II Bone Cement is used for the fixation of pathological fractures of the vertebral body using Vertebroplasty or Kyphoplasty procedures. Painful vertebral compression fractures of the vertebral body may result from osteoporosis, benign lesions (hemangioma), or malignant lesions (metastatic cancers, myeloma).
{2}------------------------------------------------
| aap Biomaterials GmbH<br>Lagerstraße 11 – 15<br>64807 Dieburg<br>Germany | Vertecem II Mixing Kit<br><br>5. 510(k) Summary<br>510(k) Premarket Notification PO-32 | 164-0008-01<br><br>Date of issue:<br>2 2. Juni 2012<br>page 3 of 3 |
|--------------------------------------------------------------------------|----------------------------------------------------------------------------------------|--------------------------------------------------------------------|
|--------------------------------------------------------------------------|----------------------------------------------------------------------------------------|--------------------------------------------------------------------|
VI91471
## Summary of technological characteristics of the new device in comparison to the predicate devices:
Vertecem II bone cement comprises the same materials, mechanical safety and performance as the legally marketed devices BonOs Inject and Vertecem.
| Trade Name | Vertecem II Mixing Kit | Vertecem | BonOs Inject |
|----------------------------------|---------------------------------|---------------------------------|---------------------------------|
| Common name | PMMA Bone<br>Cement | PMMA Bone<br>Cement | PMMA Bone<br>Cement |
| Responsible<br>manufacturer | aap Biomaterials | Teknimed S.A.S. | aap Biomaterials |
| 510(k) Number | - | K090435 | K090460 |
| Device<br>Classification<br>Name | Cement, Bone,<br>Vertebroplasty | Cement, Bone,<br>Vertebroplasty | Cement, Bone,<br>Vertebroplasty |
| Product Code | NDN | NDN | NDN |
| Classification | Class II | Class II | Class II |
| Regulation no. | 21 CFR 888.3027 | 21 CFR 888.3027 | 21 CFR 888.3027 |
The effectiveness and substantial equivalence of Vertecem II Bone Cement was determined by physical, chemical and mechanical comparative tests to Vertecem and BonOs Inject by comparing the results of the relevant data.
In summary, Vertecem II Bone Cement is as safe and effective for the declared indications as the predicate devices BonOs Inject and Vertecem.
In conclusion substantially equivalence has been demonstrated for Vertecem II Mixing Kit in comparison to Vertecem (K090435) and BonOs Inject (K090460).
{3}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines forming its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
aap Biomaterials GmbH % Mr. Volker Stirnal Director, Quality Assurance and Regulatory Affairs Lagerstrasse 11-15 Dieburg, Germany D-64807
Re: K121876
Trade/Device Name: Vertecem II Mixing Kit Regulation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) bone cement Regulatory Class: Class II Product Code: MDN, LOD Dated: October 15, 2012 Received: October 16, 2012
Dear Mr. Stirnal:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
OCT 1 9 2012
{4}------------------------------------------------
# Page 2 - Mr. Volker Stirnal
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part. 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Sincerely yours,
for
Mark N. Melkerson
Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
| aap Biomaterials GmbH<br>Lagerstraße 11 – 15 | Vertecem II Mixing Kit | 164-0007-01 |
|----------------------------------------------|-------------------------------------|---------------------------------|
| 64807 Dieburg<br>Germany | 4. Indications for Use Statement | Date of issue:<br>22. Juni 2012 |
| | 510(k) Premarket Notification PO-32 | |
# 4. Indications for Use
## 510(k) Number:
Device Name:
Vertecem II Mixing Kit
#### Indications for Use:
The Vertecem II Bone Cement is used for the fixation of pathological fractures of the vertebral body using Vertebroplasty or Kyphoplasty procedures. Painful vertebral compression fractures of the vertebral body may result from osteoporosis, benign lesions (hemangioma), or malignant lesions (metastatic cancers, myeloma).
Samuel D. Payne
(Division Sign-Off) (Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K121876
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.