Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis. 2. Rheumatoid arthritis. 3. Correction of functional deformity. 4. Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable by other techniques. 5. Revision procedures where other treatment or devices have failed. Acetabular shells and femoral stems with porous coatings are indicated for uncemented biological fixation. Non-coated or polyethylene components may be used with mating components that are indicated for either cemented or uncemented use. The Biomet G7™ Freedom™ Constrained Liner is indicated for use as a component of a total hip prosthesis in primary and revision patients at high risk of dislocation due to a history of prior dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease, or intraoperative instability, and for whom all other options to constrained acetabular components have been considered.
Device Story
Modular acetabular system for total hip arthroplasty; provides surgeons with multiple shell, liner, and head options. Components include titanium alloy shells with porous plasma spray coating, UHMWPE liners (neutral, high wall, or face-changing), and cobalt chrome femoral heads. System includes constrained liner/head options for high-dislocation-risk patients. Used in clinical/OR settings by orthopedic surgeons. Device replaces damaged hip joint surfaces to restore function and reduce pain. System compatibility extends to existing Biomet ceramic heads and femoral stems. Output is a reconstructed hip joint assembly; clinical decision-making relies on surgeon assessment of patient anatomy, bone quality, and stability requirements.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by extensive bench testing, including poly push-out, lever-out, torque-out, rim impingement, fatigue testing, range-of-motion studies, screw pull-through/torsional properties, and metal head fretting/corrosion analysis.
Technological Characteristics
Modular hip prosthesis. Materials: Titanium alloy (ASTM F136) shell, Porous Plasma Spray coating (ASTM F1580), UHMWPE liners (ASTM F648, ArComXL® or E1®), Cobalt chrome femoral heads (ASTM F1537). Features: Solid/limited hole shell designs, neutral/high wall/face-changing liners, constrained liner options. Mechanical fixation via uncemented porous coating or cemented techniques. No software or electronic components.
Indications for Use
Indicated for patients with non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, functional deformity, proximal femur fractures (non-union, femoral neck, trochanteric), or revision needs. Constrained liners indicated for primary/revision patients at high risk of dislocation (prior dislocation, bone loss, laxity, neuromuscular disease, intraoperative instability).
Regulatory Classification
Identification
A hip joint metal/polymer constrained cemented or uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device prevents dislocation in more than one anatomic plane and has components that are linked together. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and an acetabular component made of ultra-high-molecular-weight polyethylene with or without a metal shell, made of alloys, such as cobalt-chromium-molybdenum and titanium alloys. This generic type of device is intended for use with or without bone cement (§ 888.3027).
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Hip Joint Metal/Polymer Constrained Cemented or Uncemented Prosthesis.”
Predicate Devices
Porous Plasma Spray (PPS®) RingLoc®+ Acetabular System (K093235)
{0}------------------------------------------------
K121874
Image /page/0/Picture/1 description: The image shows the logo for Biomet Inc. The logo is in black and white and features a stylized font. The letters are bold and outlined, giving them a three-dimensional appearance.
### 510(k) Summary
# NOV 2 3-2012
Preparation Date:
Contact Person:
Proprietary Name:
Classification Name:
Common Name:
Applicant/Sponsor:
August 23, 2012
Biomet Manufacturing Corp. 56 East Bell Drive P.O. Box 587 Warsaw, IN 46581-0587 Establishment Registration Number: 1825034
Becky Earl Regulatory Specialist
G7™ Acetabular System
Acetabular System
LPH- Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented (21 CFR 888.3358)
LZO-Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented or Non-Porous, Uncemented (21 CFR 888.3353)
OQG-Prosthesis, hip, semi-constrained, metal/polymer + additive, porous uncemented (21 CFR 888.3358)
KWZ-Prosthesis, hip, constrained, cemented or uncemented, metal/polymer (21 CFR 888.3310) 0
jDI- Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented (21 CFR 888.3350)
OQH-Hip, semi-constrained, cemented, metal/polymer + additive, cemented (21 CFR 888.3350)
OQI-Hip, semi-constrained, cemented, metal/ceramic/polymer + additive, porous uncemented (21 CFR 888.3353)
PBI-Prosthesis, hip, constrained, cemented or uncemented, metal/polymer, + additive (21 CFR 888.3310)
Mailing Address: P.O. Box 587 Warsaw, IN 46581-0587 Toll Free: 800.348.9500 Office: 574.267.6639 Main Fax: 574.267.8137 ww.biomet.com
Shipping Address: 56 East Bell Drive Warsaw, IN 46582
p. 1 of 3
{1}------------------------------------------------
# K121874
#### Legally Marketed Devices To Which Substantial Equivalence Is Claimed:
#### Acetabular Shells:
Porous Plasma Spray (PPS®) RingLoc®+ Acetabular System, K093235 (Biomet) Reflection Acetabular Cup System, K033442 (Smith and Nephew) Continuum Acetabular System, K103662 (Zimmer)
#### Acetabular Liners:
Ringloc® E1® Acetabular Liners, K093549 (Biomet) Reflection Acetabular Cup System, K033442 (Smith and Nephew) Continuum Acetabular System, K103662 (Zimmer)
Constrained Heads and Liners: Freedom Constrained Liners, K043537 (Biomet) Cobalt Chrome Femoral Heads: Cobalt Chrome Femoral Components, K911684 (Biomet)
#### Device Description:
The G7™ Acetabular System is a modular system, designed to provide numerous options for surgeons and patients in one compatible system. The titanium alloy (ASTM F136) shell is available in both a solid shell and limited hole design, with an apical hole and plug. The outer diameter of the shells is coated with Biomet's Porous Plasma Spray (ASTM F1580). The acetabular liners are manufactured from UHMWPE (ASTM F648) as ArComXL® or E1® (UHMWPE infused with vitamin E) and available in a Neutral, High Wall, or Ten Degree Face Changing design. Additionally, the system has provided for a constrained liner and head when such an option is required. The Biomet Freedom" product line has added a 32mm Freedom" head (cobalt chrome, ASTM F1537) to its offerings, as well as a new series of constrained liners (E1®, ASTM F648) designed for compatibility with the G7™ shells. A new line of cobalt chrome (ASTM F1537) femoral heads in varying sizes and neck lengths, with either a Type 1 or 12/14 taper, has been added to the system. The system is compatible with Biomet® ceramic heads and femoral stems.
#### Indications for Use:
Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis. 1.
- 2. Rheumatoid arthritis.
- 3. Correction of functional deformity.
- Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with 4. head involvement, unmanageable by other techniques.
- ഗ് Revision procedures where other treatment or devices have failed.
Acetabular shells and femoral stems with porous coatings are indicated for uncemented biological fixation. Non-coated or polyethylene components may be used with mating components that are indicated for either cemented or uncemented use.
#### Indications for Biomet G7™ Freedom™ Constrained Liners:
The Biomet G7™ Freedom™ Constrained Liner is indicated for use as a component of a total hip prosthesis in primary and revision patients at high risk of dislocation due to a history of prior dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease, or intraoperative instability, and for whom all other options to constrained acetabular components have been considered.
{2}------------------------------------------------
Klzs74
#### Summary of Technologies:
The technological characteristics of the proposed device are the same as or similar to the predicates. The G7™ acetabular shells are similar in design, dimensions, and intended use to their Biomet predicate, K093235, but have incorporated a different locking mechanism, which is also similar to legally marketed predicates, K091508 and K103662. The acetabular liners are made of the same material as their predicates, and use previously cleared liner types, sizes and designs. The additional constrained head is a line extension to the Biomet Freedom™ constrained heads, and the constrained liners use the same design principles as the Freedom liner predicate, K030047. The cobalt chrome femoral heads are very similar to the Biomet predicate, K911684.
#### Non-Clinical Testing:
The following testing has been completed to support substantial equivalence: v
- Poly Push Out .
- Poly Lever Out .
- Poly Torque Out .
- Poly Rim Impingement .
- Poly Deformed Cup Push Out .
- Poly Deformed Cup Lever Out
- Poly Fatigue, Lever Out �
- Poly Constrained Liner Rim Impingement and Lever Out Testing .
- Range-of-Motion Study (Poly) .
- Poly Deformed Cup Push In .
- . Freedom Head Pull Out of Constrained Liner
- Screw Pull Through �
- Screw Torsional Properties .
- Torque Curve for MIM Screw Hole Plugs .
- Verification of Insertion Torque and Measurement of Torque Out .
#### Justifications:
- Poly Wear Simulator .
- Metal Head Justification .
- . MRI Justification
- Metal Head Axial Head Pull Off .
- Metal Head Fretting/Corrosion ●
#### Clinical Testing:
None provided as a basis for substantial equivalence.
Testing demonstrates that the modifications made to the G7™ Acetabular System do not introduce any new risks of safety or efficacy, and that the G7™ Acetabular System is substantially equivalent to the predicates.
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three wing-like extensions. The bird is positioned within a circle that contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference of the circle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
November 23, 2012
Biomet Manufacturing Corporation % Ms. Becky Earl 56 East Bell Drive P.O. Box 587 Warsaw, Indiana 46581-0587 US
Re: K121874
Trade/Device Name: G7™ acetabular system Regulation Number: 21 CFR 888.3310 Regulation Name: Hip Joint Metal/Polymer Constrained Cemented or Uncemented Prosthesis Regulatory Class: Class II . Product Code: PBI, LPH, LZO, OQG, KWZ, JDI, OQH, OQI
- Dated: October 19, 2012
Received: October 22, 2012
Dear Ms. Earl:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
:
{4}------------------------------------------------
## Page 2 - Ms. Becky Earl
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Erin I. Keith
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known):_K121894
Device Name: G7™ Acetabular System
Indications For Use:
1. Non-inflammatory degenerative joint disease including osteoarthritis and avascular
- necrosis. ..
- 2. Rheumatoid arthritis.
- 3. Correction of functional deformity.
- 4. Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable by other techniques.
- 5. Revision procedures where other treatment or devices have failed.
Acetabular shells and femoral stems with porous coatings are indicated for uncemented biological fixation. Non-coated or polyethylene components may be used with mating components that are indicated for either cemented or uncemented use.
## Indications for Biomet G7 Freedom™ Constrained Liners:
The Biomet G7 Freedom™ Constrained Liner is indicated for use as a component of a total hip prosthesis in primary and revision patients at high risk of dislocation due to a history of prior dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease, or intraoperative instability, and for whom all other options to constrained acetabular components have been considered.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use NO (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
K. Asindi
Jor (Division Sign-Off) . Division of Orthopedic Devices 510(k) Number K121874
Page 1 of 1
42
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.