← Product Code [NHA](/productcode/NHA) · K121843

# NP-CAST ABUTMENT SYSTEM (K121843)

_Osstem Implant Co., Ltd. · NHA · Dec 11, 2012 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K121843

## Device Facts

- **Applicant:** Osstem Implant Co., Ltd.
- **Product Code:** [NHA](/productcode/NHA.md)
- **Decision Date:** Dec 11, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3630
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

NP-Cast Abutment System is intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or overdentures.

## Device Story

NP-Cast Abutment System consists of abutments and screws used for prosthetic restoration in dental implant procedures. Device addresses aesthetic, spatial, and path constraints. Clinicians customize abutments via casting using nonprecious Co-Cr-Mo alloy; post height must exceed 4mm; maximum angulation limited to 30°. System provides support for crowns, bridges, or overdentures. Used in dental clinics by licensed dentists. Benefits include customized prosthetic fit and structural support for dental restorations.

## Clinical Evidence

No clinical studies are submitted. Safety and effectiveness supported by bench testing, including biocompatibility and fatigue testing conducted according to FDA guidance for Class II Special Controls for Root-form Endosseous Dental Implants and Endosseous Dental Abutment.

## Technological Characteristics

Materials: Co-Cr-Mo alloy. Components: Abutment and Abutment Screws. Configurations: Mini and Standard connections, Hex/Non-Hex and Octa/Non-Octa interfaces. Dimensions: Diameters ranging from 4.0mm to 6.3mm. Mechanical device; no software or energy source.

## Regulatory Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

## Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

## Predicate Devices

- SimpleLine II Abutment System ([K112045](/device/K112045.md))
- Prosthetic System ([K110308](/device/K110308.md))
- HU.HS.HG Prosthetic System ([K081575](/device/K081575.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K121843

**OSSTEM**
IMPLANT
OSST

# 507-8 Geoje

SSTEM Implant Co., Ltd 07-8 Geoje3-Dong Yeonje-Gu Busan, 611-804 Republic of Korea

1: +82 51 850-2500 Fax: +82 51 850-4341 www.osstem.com

DEC 1 1 2012

# 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.

Date: May 30, 2012

1. Company and Correspondent making the submission:

- Submitter's Name :
- Address :
- Contact :
- Phone:
- Correspondent's Name:
- Address:
- Contact:
- Phone:
- 2. Device :

Trade or (Proprietary) Name : Common or usual name :

Classification Name :

OSSTEM Implant Co., Ltd. #507-8 Geoje3-Dong Yeonje-Gu Busan, 611-804, Republic of Korea Mr. Hee Kwon Son +82 51 850 2575

HIOSSEN Inc. 85 Ben Fairless Dr. Fairless Hills, PA 19030 Patrick Lim 888 678 0001

NP-Cast Abutment System Dental Device Endosseous dental implant abutment 21CFR872.3630 Class II NHA

- 3. Predicate Device :
SimpleLine II Abutment System, Dentium Co., Ltd., K112045 Prosthetic System, OSSTEM Implant Co., Ltd., K110308 HU.HS.HG Prosthetic System, OSSTEM Implant Co., Ltd., K081575

# 4. Description :

NP-Cast Abutment System is used for prosthetic restoration. It is used for cases with path and aesthetic and spatial constraints. After customization, be sure to use only dental nonprecious metal (Co-Cr-Mo alloy) for casting to make the prosthesis When cast a prosthesis with NP-Cast Abutment System, Post height above the transmucosal collar of NP-Cast Abutment System has to be taller than 4mm and

Maxmimum angulation has to be less than 30°

QS-QI-505-3(Rev.0)

Letter(8.5 X 11in)

{1}------------------------------------------------

- 1) The NP-Cast Abutment System consists of Abutment and Abutment Screws. And Abutment in the NP-Cast Abutment System is distinguished such as below

| Model Name                    | Product name        | Connection | Diameter |          |
|-------------------------------|---------------------|------------|----------|----------|
| NP-Cast<br>Abutment<br>System | NP Cast Abutment    | Mini       | Hex,     | Ø4.0mm   |
|                               |                     |            | Non Hex  | Ø 4.0mm  |
|                               |                     | Standard   | Hex,     | Ø 4.5mm  |
|                               |                     |            | Non Hex  | Ø 4.5mm  |
|                               | HS NP Cast Abutment | Mini       | Octa,    | Ø 5.05mm |
|                               |                     |            | Non Octa | Ø 6.3mm  |
|                               |                     | Standard   | Octa,    | Ø 5.05mm |
|                               |                     |            | Non Octa | Ø 6.3mm  |

- 2) The NP-Cast Abutment System is similar to other commercially available products based on the intended use, the technology used, the claims, the material composition employed and performance characteristics.
- 3) The NP-Cast Abutment System is substantially equivalent in design, function and intended use to The SimpleLine II Abutment System(K112045) of Dentium Co., Ltd. and Prosthetic System (K110308) of OSSTEM Implant Co., Ltd., HU.HS.HG Prosthetic System (K081575) of OSSTEM Implant Co., Ltd.

# - Substantial Equivalence Matrix

|                  | NP-Cast Abutment                                                                                                                                                             | SimpleLine II Abutment System                                  | Prosthetic System                                                                                                                                                      |
|------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer     | Osstem Implant Co., Ltd                                                                                                                                                      | Dentium Co., Ltd.                                              | Osstem Implant Co., Ltd                                                                                                                                                |
| 510(k)<br>Number | New                                                                                                                                                                          | K112045                                                        | K110308                                                                                                                                                                |
| Intended use     | NP-Cast Abutment<br>System is intended for<br>use with a dental implant<br>to provide support for<br>prosthetic restorations<br>such as crowns, bridges,<br>or overdentures. | Intended for use as an<br>aid in prosthetic<br>rehabilitation. | Prosthetic System is<br>intended for use with a<br>dental implant to provide<br>support for prosthetic<br>restorations such as<br>crowns, bridges, or<br>overdentures. |
| Material         | Co-Cr-Mo Alloy                                                                                                                                                               | Co-Cr-Mo Alloy                                                 | Gold Alloy                                                                                                                                                             |

# 5. Indication for use :

The NP-Cast Abutment System is intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or overdentures.

{2}------------------------------------------------

- 6. Review :
The NP-Cast Abutment System has similar material, indication for use, design and technological characteristics as the predicate device.

- 7. Summary of nonclinical testing
Safety tests including biocompatibility have been performed to ensure the devices comply with the applicable international and US regulations.

Fatigue testing was conducted according to the "Guidance for industry and FDA staff Class II Special Controls Guidance Document Root-form Endosseous Dental Implants and Endosseous Dental Abutment" with the worst case scenario of the NP-Cast Abutment System

- 8. Summary of clinical testing No clinical studies are submitted
- 9. Conclusions

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification Osstem Implant Co., Ltd. concludes that the NP-Cast Abutment System is substantially equivalent to the predicate devices as described herein.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular pattern around the symbol. The caduceus is black, and the text is also black.

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

#### December 11, 2012

OSSTEM Implant Company, Limited C/O Mr. Patrick Lim Manager HiOSSEN Incorporated 85 Ben Fairless Drive FAIRLESS HILLS PA 19030

Re: K121843

Trade/Device Name: NP-Cast Abutment System Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: November 28, 2012 Received: December 5, 2012

#### Dear Mr. Lim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

**Kwame** O. Ulmer

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

**OSSTEM Implant Co., Ltd.**

#507-8 Geoje3-Dong Yeonje-Gu Busan, 611-804 Republic of Ko Tel: +82 51 850-2500 Fax: +82 51 850-4341 www.osstem.co

Indications for Use

510(k) Number K_121843

Device Name : NP-Cast Abutment System

Indication for use : NP-Cast Abutment System is intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or overdentures.

Prescription Use X (Per 21CFR801 Subpart D) OR Over-The-Counter Use (Per 21CFR807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

.

Concurrence of CDRH, Office of Device Evaluation (ODE)

2012.12.11 Susan Runner DDS, MA 15:23:21 -05'00'

(Division Sign-Off) (Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number:

QS-QI-505-2(Rev.0)

Letter(8.5 X 1 in)

- 1 -

---

**Source:** [https://fda-staging.innolitics.com/device/K121843](https://fda-staging.innolitics.com/device/K121843)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).