Point 500 HD/500C HD which is the dental panoramic/panoramic and cephalometric device is intended to produce diagnostic quality radiographic images by acquiring digital 2D X-ray images. The cephalometric system produces cephalomertric radiographic images that maybe used in orthodontic diagnosis procedures.
Device Story
Dental digital X-ray imaging system (panoramic and cephalometric); captures 2D X-ray images of teeth, jaw, and TMJ areas. Device consists of X-ray generator, sensor, rotation unit, main frame, and column. Rotation unit orbits patient; sensor detects X-ray data; data transferred to PC. Installed software combines frames to display panoramic/cephalometric images on PC screen. Used in dental clinics by dentists/qualified professionals. Output provides diagnostic quality radiographic images for clinical assessment and orthodontic diagnosis.
Clinical Evidence
No clinical data. Bench testing only, including electrical, mechanical, environmental safety, and performance testing per EN/IEC 60601 series standards.
Technological Characteristics
Extraoral digital X-ray system. Tube voltage: 50-90kV; Tube current: 4-16mA; Focal spot size: 0.5 x 0.5 mm. Connectivity: PC-based image acquisition and display. Standards: EN/IEC 60601-1, 60601-1-3, 60601-2-7, 60601-2-28, 60601-2-32, and 60601-1-2 (EMC).
Indications for Use
Indicated for dental radiographic examinations of teeth, jaw, and TMJ areas, including cephalometric imaging for orthodontic diagnosis. For use by dentists and qualified professionals.
Regulatory Classification
Identification
An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.
{0}------------------------------------------------
1.
Image /page/0/Picture/1 description: The image shows the logo for PointNix. The logo includes a stylized flower-like graphic to the left of the word "SMART". To the right of "SMART" is the phrase "Digital Leader" above the word "PointNix". The word "PointNix" is in a bold, sans-serif font.
Twin Tower 1. 212-1. Kuro-Dong, Kuro-Ku, Seoul, Korea -854-8315 / Fax: 82-2-839-4973
## 510(K) SUMMARY
## JAN 2 4 2013
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 5l0(k) number is K121800. Date: January 17, 2013
#### 1. Submitter's Identification:
PointNix Co., Ltd. #302~303, 3F, Ace Twin Tower 1, 212-1, Kuro-Dong, Kuro-Ku, Seoul, Korea Tel: +82-2-854-8315(ext. 309) Fax: +82-2-839-4973 E-mail: support@pointnix.com info@pointnix.com
#### 2. Name of the Device:
Dental Digital X-ray Imaging Device models Point 500HD and Point 500C HD.
#### 3. Common or Usual Name:
Usual Name : Dental Digital X-ray Imaging System Classification : Class II 21 CFR 872.1800 System, x-ray, extraoral source, digital Class Code : MUH
#### 4. Predicate Device Information:
The Yoshida Dental MFG. Co., LTD.: Panoura 18S - K111231
#### 5. Device Description:
The Point 500 HD/500C HD is a dental panoramic and cephalometric device which is digital X-ray imaging device. It is intended for dental radiographic examinations of teeth, jaw and TMJ areas by exposing X-ray and then converts this information to the digital signals to acquire images. High-defined images are provided by controlling acquired image with computer and this is used to obtain panoramic and cephalometric images for dental purpose. This equipment is
{1}------------------------------------------------
K121800
Image /page/1/Picture/1 description: The image shows the logo for PointNix. The logo includes a stylized graphic to the left of the text. Above the word "PointNix" is the phrase "Digital Leader" in a smaller font.
3F, Ace Twin Tower 1, 212-1, Kuro-Dong, Kuro-Ku, Seoul, Korea Tel: 82-2-854-8315 / Fax: 82-2-839-4973
equipped with an X-ray generator and Sensor for panoramic and cephalometric at Rotation unit which are supported by Main frame and Column. When rotation unit is rotating around the patient's teeth and jaw for acquiring panoramic image, X-ray exposes and sensor detects the data from X-ray generator. Detected data from each frame is transferred to PC, and combined together to display panoramic image on PC screen by installed software on PC.
#### 6. Intended Use:
Point 500 HD/500C HD which is the dental panoramic/panoramic and cephalometric device is intended to produce diagnostic quality radiographic images by acquiring digital 2D X-ray images. The cephalometric system produces cephalomertric radiographic images that maybe used in orthodontic diagnosis procedures. .
#### 7. Comparison to Predicate Devices:
| Characteristic | Model Proposed :<br>Point 500HD, Point 500C<br>HD | Predicated :<br>Panoura 18S |
|-----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | - | K111231 |
| Intended Use | Point 500 HD/500C HD<br>which is the dental<br>panoramic/panoramic and<br>cephalometric device is<br>intended to produce<br>diagnostic quality<br>radiographic images by<br>acquiring digital 2D X-ray<br>images. The cephalometric<br>system produces<br>cephalomertric radiographic<br>images that maybe used in<br>orthodontic diagnosis<br>procedures. | The Panoura 18S dental<br>panoramic and<br>cephalometric device is<br>intended for dental<br>radiographic examinations of<br>teeth, jaw and TMJ areas by<br>producing conventional 2D<br>X-ray images as well as X-<br>ray projection images. The<br>device must only be<br>operated and used by<br>dentist and other legally<br>qualified professionals. |
| Tube Voltage | 50-90kV | 58-82kV |
| Tube Current | 4-16mA | 2-10mA |
| Focal Spot Size | 0.5 x 0.5 mm | 0.5 x 0.5 mm |
| Scan Time | Only Panorama Mode(Point<br>500HD) : 16sec,<br>Added cephalo Mode(Point<br>500C HD) : 0.8sec | Panorama Mode: 16sec<br>Cephalo Mode: 10sec |
{2}------------------------------------------------
K121800
Image /page/2/Picture/1 description: The image shows the logo for PointNix. The logo includes a stylized flower-like graphic to the left of the word "SMART". To the right of "SMART" is the word "PointNix", with the words "Digital Leader" above it. The logo is black and white.
3F, Ace Twin Tower 1, 212-1, Kuro-Dong, Kuro-Ku, Seoul, Korea
Tel: 82-2-854-8315 / Fax: 82-2-839-4973
#### Statement of substantial equivalence 8.
The characteristics of the Point 500HD/500C HD are similar to those of the Panoura 18S (K111231) listed below:
- Intended use
- · Operating characteristics
- · Radiation
Difference between Point 500HD/500C HD and Panoura 18S (K111231) is listed below:
- · Detector type
This difference of detector does not raise any new safety or effectiveness concerns compared to the Panoura 18S (K111231).
#### 9. Discussion of Non-Clinical Tests Performed:
Electrical, mechanical, environmental safety and performance testing according to standards EN/IEC 60601-1, EN/IEC 60601-1-3, EN/IEC 60601-2-7, EN/IEC 60601-2-28, and EN/IEC 60601-2-32 were performed, and EMC testing was conducted in accordance with a standard EN/IEC 60601-1-2. All test results were satisfactory.
#### Discussion of Clinical Tests Performed: 10.
Not Required
#### 11. Conclusions:
The PointNix Co., model Point 500HD and Point 500C HD have been developed and validated according to all applicable standards. Evaluations have proven that the system is safe and effective for the intended use. Risk analysis and safety certifications reveal that there is no new safety issues associated with this system as compared with the predicate devices.
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features the department's name encircling a stylized eagle with its wings spread. The eagle is composed of three curved lines, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
## January 24, 2013
PointNix Company, Limited C/O Mr. Jigar Shah MDI Consultants, Incorporated #302~303, 3F, Ace Twin Tower 1, 212-1, Kuro-Dong Kuro-Ku, Seoul, Korea
Re: K121800
Trade/Device Name: Dental Digital X-ray imaging system point 500 HD and point 500C HD Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral Source X-Ray System Regulatory Class: II Product Code: MUH Dated: January 10, 2013 Received: January 15, 2013
Dear Mr. Shah:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Shah
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Kwame Q. Ulmer
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains a logo for a company called "PointNix". The logo features a stylized graphic to the left of the text, with the word "SMART" underneath. Above the company name, the words "Digital Leader" are printed in a smaller font. The logo is simple and modern, with a focus on the company name.
Twin Tower 1, 212-1, Kuro-Dong, Kuro-Ku, Seoul, Korea 82-2-854-8316 / Fax: 82-2-839-
# Section 2: Indications for Use
**Page** 1 of 1
K 21 800 510(k) Number (if known):
Device Name: PointNix Co., Ltd. Dental Digital X-ray Imaging System Point 500 HD and Point 500C HD
## Indications For Use:
Point 500 HD/500C HD which is the dental panoramic/panoramic and cephalometric device is intended to produce diagnostic quality radiographic images by acquiring digital 2D X-ray images. The cephalometric system produces cephalomertric radiographic images that maybe used in orthodontic diagnosis procedures.
Prescription Use × (Per 21 CFR 801 Subpart D) Over-The Counter Use (21 CFR 807 Subpart C)
(Please do not write below this Line-Continue on another PAGE IF NEEDED)
OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Runner DDS, MA-115:22:50 -05'00'
ﻟﺴﻤﺮ
Division Sign-Off) 'ivision of Anesthesiology, General Hospital ifection Control, Dental Devices
510(k) Number:
2-1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.