WECK REUSABLE OBTURATOR

K121796 · Teleflexmedical, Inc. · GCJ · Oct 11, 2012 · Gastroenterology, Urology

Device Facts

Record IDK121796
Device NameWECK REUSABLE OBTURATOR
ApplicantTeleflexmedical, Inc.
Product CodeGCJ · Gastroenterology, Urology
Decision DateOct 11, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The Weck Reusable Obturator is indicated for use with appropriate disposable Weck Vista cannulas in thoracic, abdominal, and gynecologic minimally invasive surgical procedures to provide a pathway for the introduction of endoscopic surgical devices.

Device Story

Weck Reusable Obturator is a stainless steel surgical instrument used to establish abdominal cavity access during laparoscopic procedures. Operated by trained physicians, the device is inserted into the peritoneum as a guide for a corresponding disposable cannula to create a port of entry. Once the port is established, the obturator is removed, allowing for the insertion and removal of various endoscopic surgical devices while maintaining insufflation. The device is designed for reuse, requiring cleaning and sterilization by the end user. It functions as a mechanical guide to facilitate minimally invasive surgical access.

Clinical Evidence

Bench testing only. Performance testing verified substantial equivalence to predicate obturators and confirmed seamless interchangeability. Cleaning and sterilization instructions were validated per ISO 17664:2004.

Technological Characteristics

Stainless steel construction; reusable design; compatible with existing disposable Weck Vista cannulas; patient-contacting materials compliant with ISO 10993-1; sterilization required by end user per ISO 17664:2004.

Indications for Use

Indicated for use with disposable Weck Vista cannulas in thoracic, abdominal, and gynecologic minimally invasive surgeries to create a pathway for endoscopic devices. Contraindicated where minimally invasive techniques are not appropriate.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Traditional 510(k) Section 8 - Summary of Safety and Effectiveness # 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS OCT 11 1 2012 Reusable Obturator # Weck® Reusable Obturator ## A. Name, Address, Phone, and Fax Number of Applicant Teleflex Medical. Incorporated 2917 Weck Drive Research Triangle Park, NC 27709 USA Phone: 919-433-4918 919-433-4996 Fax: # B. Contact Person Holly Kornegay Regulatory Affairs Associate #### C. Date Prepared June 14, 2012 #### D. Device Name Trade Name: Common Name: Regulatory Classification: Classification Name: Regulation Number: Regulation Name: Product Code: Weck® Reusable Obturator Surgical Trocar . Class II . Endoscope and Accessories 21 CFR 876.1500 Arthroscope and Accessories GCJ #### E. Device Description The Weck® Reusable Obturator is used to establish penetration into the abdominal cavity during laparoscopic surgical procedures. When used with the appropriate corresponding disposable cannula, the system creates a port of entry into the patient, facilitating the access of various diameter devices, while maintaining insufflation at the surgical site. The obturator is positioned into the peritoneum as a guide to the corresponding cannula during minimally invasive surgical procedures, and then is removed in order to provide a pathway for the insertion and removal of various sized surgical devices. The Weck® Reusable Obturator is intended to be used by trained physicians. {1}------------------------------------------------ ## F. Indications for Use The Weck Reusable Obturator is indicated for use with appropriate disposable Weck Vista cannulas in thoracic, abdominal, and gynecologic minimally invasive surgical procedures to provide a pathway for the introduction of endoscopic surgical devices. # G. Contraindications Where minimally invasive techniques are contraindicated, other methods and instrumentation should be employed. ## H. Substantial Equivalence The proposed Weck® Reusable Obturator is substantially equivalent to the predicate disposable obturator used with the following devices: | Predicate Device | Manufacturer | 510(k) No. | Date Cleared | |---------------------------------------------------------|----------------------------------------|------------|--------------| | ADAPT TM Laparoscopic Port<br>and Accessory | Teleflex Medical, Inc. /<br>Taut, Inc. | K010007 | 02/22/01 | | ADAPTTM Universal<br>Laparoscopic Port | Teleflex Medical, Inc. | K082156 | 09/10/2008 | # I. Comparison to Predicate Devices This proposed product line extension to K010007, ADAPTM Laparoscopic Port and Accessory, and K082156, ADAP1M Universal Laparascopic Port, affords a new product code for a reusable stainless steci obturator. . The incorporation of the stainless steel material will allow trained physicians the ability to reuse the obturator, without any modification to material or dimensions of the compatible systems noted above. # J. Materials All patient contacting materials are in compliance with ISO 10993-1. # K. Technological Characteristics A comparison of the technological characteristics of the proposed Weck® Reusable Obturator and the predicates has been performed. The results of this comparison demonstrate that the Weck® Reusable Obturator is compatible with the cannulas of the systems noted above and additionally offers the feature of "reusability". Teleflex Medical, Inc. {2}------------------------------------------------ # L. Performance Data The bench testing has been performed to verify that the performance of the proposed Weck® Resuable Obturator is substantially equivalent to the predicate obturators and that the Weck® Reusable Obturator is seamlessly interchangeable with the predicate obturator. # L. Conclusion Based upon the comparative test results, the proposed Weck® Reusable Obturator is substantially equivalent in performance to the predicate devices cleared to market via 510(k) K010007 and K082156. The modifications made to the Weck® Reusable Obturator do not introduce any new issues of safety and effectiveness. IFU updates have been integrated according to ISO 17664:2004, Sterilization of Medical Devices, to ensure appropriate cleaning and sterilization instructions for reusable devices, by the end user. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the caduceus symbol. Food and Drug Administration 1 0903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Teleflex Medical, Incorporated % Ms. Holly Kornegay Regulatory Affairs Associate 2917 Weck Drive Research Triangle Park, North Carolina 27709 Re: K121796 Trade/Device Name: Weck® Reusable Obturator Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: GCJ Dated: August 1, 2012 Received: September 17, 2012 DEC - 4 2012 Dear Ms. Kornegay: This letter corrects our substantially equivalent letter of October 11, 2012. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ Page 2- Ms. Holly Komegay If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. **Sincerely yours,** Mark N Millman Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure: New Indication For Use {5}------------------------------------------------ # Indications for Use Page I of I 510(k) Number: K121796 Device Name: Weck® Reusable Obturator Indications for Use: The Weck Reusable Obturator is indicated for use with appropriate disposable Weck Vista cannulas in thoracic, abdominal, and gynecologic minimally invasive surgical procedures to provide a pathway for the introduction of endoscopic surgical devices. Prescription Use XX (Part 21 CFR 801 Subpart D) AND/OR Over-the-counter use __ (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Mark H. Miller (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices -10(k) Number K-121796 Teleflex Medical, Inc.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...