K121736 · Waters Medical Systems, LLC · KDL · Aug 16, 2013 · Gastroenterology, Urology
Device Facts
Record ID
K121736
Device Name
PERF-GEN PULSATILE PERFUSION SOLUTION
Applicant
Waters Medical Systems, LLC
Product Code
KDL · Gastroenterology, Urology
Decision Date
Aug 16, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5880
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PERF-GEN® Pulsatile Perfusion is intended to be used for flushing and continuous hypothermic machine perfusion of kidneys at the time of their removal from the donor in preparation for storage, transportation, and eventual transplantation into a recipient.
Device Story
PERF-GEN® is a sterile, non-pyrogenic, non-toxic solution used for in-vitro flushing and continuous hypothermic machine perfusion of explanted kidneys. It acts as a preservation medium during the interval between donor removal and recipient transplantation. The solution is supplied in sterile dispensing bags. It is used in clinical settings by transplant professionals. The solution maintains organ viability through controlled osmolarity (approx. 300 mOsm/kg), sodium (100 mEq/L), and potassium (25 mEq/L) concentrations at a pH of 7.4. By providing a stable environment for the kidney, it supports successful transplantation outcomes.
Clinical Evidence
Bench testing only. Biocompatibility testing (ISO 10993) confirmed non-cytotoxicity, non-toxicity, hemocompatibility, non-genotoxicity, non-sensitization, and non-irritation. Stability testing confirmed 1-year shelf life. Chemical equivalence to predicate confirmed via infrared spectra, elution profiles, and ionic strength measurements.
Technological Characteristics
Sterile, non-pyrogenic, non-toxic aqueous solution. Osmolarity ~300 mOsm/kg; Na 100 mEq/L; K 25 mEq/L; pH ~7.4. Dispensed in sterile bags. Sterilized by filtration. Biocompatibility per ISO 10993.
Indications for Use
Indicated for flushing and continuous hypothermic machine perfusion of explanted kidneys from donors to facilitate storage, transport, and transplantation into recipients.
Regulatory Classification
Identification
An isolated kidney perfusion and transport system and accessories is a device that is used to support a donated or a cadaver kidney and to maintain the organ in a near-normal physiologic state until it is transplanted into a recipient patient. This generic type of device may include tubing, catheters, connectors, an ice storage or freezing container with or without bag or preservatives, pulsatile or nonpulsatile hypothermic isolated organ perfusion apparatus with or without oxygenator, and disposable perfusion set.
{0}------------------------------------------------
Traditional 510(k) Premarket Notification PERF-GEN® Pulsatile Perfusion Solution
# 510(k) Summarv
Date Prepared: December 20th, 2012
#### Submitter's Name / Contact Person
| Submitter | Contact Person |
|-----------------------------|----------------------|
| Waters Medical Systems, LLC | Robert Warren |
| 2112 - 15th Street NW | Phone: 507-288-7777 |
| Rochester, Minnesota 55901 | Fax: 507-252-3700 |
| | Email: bobw@wtrs.com |
### General Information
| Trade Name | PERF-GEN® Pulsatile Perfusion Solution |
|----------------------------|---------------------------------------------------------------------------------------------|
| Common / Usual Name | Cold Storage Solution |
| Product Code | KDL |
| Classification Name | Set, perfusion, kidney, disposable |
| Classification Information | 21 CFR 876.5880<br>Isolated kidney perfusion and transport system and accessories, Class II |
| Predicate Device | BELZER-MPSTM, Trans-Med Corporation (K972066) |
# Device Description
The PERF-GEN® Pulsatile Perfusion (PERF-GEN Solution) is a clear, sterile, non-pyrogenic, non-toxic solution for the in-vitro flushing and temporary continuous perfusion of explanted kidneys. This solution has an approximate calculated osmolarity of 300 mOsm/kg, a sodium concentration of 100 mEq/L, a potassium concentration of 25 mEq/L, and a pH of approximately 7.4 at room temperature.
# Intended Use / Indications
The PERF-GEN® Pulsatile Perfusion is intended to be used for flushing and continuous hypothermic machine perfusion of kidneys at the time of their removal from the donor in preparation for storage, transportation, and eventual transplantation into a recipient.
#### Substantial Equivalence and Summary of Studies
The PERF-GEN Pulsatile Perfusion device is substantially equivalent to the BELZER-MPS UW Machine Perfusion Solution. The PERF-GEN and BELZER-MPS solutions have an identical intended use, chemical composition and principle of operation. Both the subject and predicate device are sterile, non-toxic, transparent solutions dispensed from a bag. Both solutions are sterilized by filtration and aseptically filled in sterile dispensing bags.
The technological difference between PERF-GEN Pulsatile Perfusion compared to the predicate is related to the dispensing bag. The material used for the BELZER-MPS dispensing bags being not specified, it is likely that the bags of PERF-GEN and BELZER-MPS may differ in material. This technological difference has been evaluated
{1}------------------------------------------------
through biocompatibility, stability and chemical testing to ensure the possible bag material difference does not impact the specifications of the perfusion solution itself.
The biomaterial safety of the PERF-GEN Solution has been evaluated through 1SO 10993 compliant testing, which included cytotoxicity test, acute systemic toxicity, hemocompatibility, skin sensitization test in guinea pigs, and primary skin irritation. Results of this testing showed the PERF-GEN Solution is safe for the intended biocontact. Test results confirmed that the PERF-GEN Solution is non-cytotoxic, non-toxic, hemocompatible, nongenotoxic, non-sensitizing, and non-irritating.
Results of this stability study confirm that up to 1 year real-time aging at the recommended storage conditions and several months worst-case storage conditions do not affect the product specifications. The stability testing has showed that aging of test articles at the recommended storage conditions does not affect the product specifications for the PERF-GEN labeled with 1-year shelf life.
Chemical composition and properties of the PERF-GEN Solution and the predicate BELZER-MPS solution were compared by infrared, chromatography and conductivity measurements. The infrared spectra, the elution profiles and the ionic strengths of PERF-GEN and BELZER-MPS solutions were compared. The results of those chemical testing confirm that both the PERF-GEN solution and the predicate BELZER-MPS solution are chemically equivalent.
Results of evaluations did not raise any new questions of safety or effectiveness when compared to the predicate device and therefore the PERF-GEN Solution is substantially equivalent to the BELZER-MPS predicate.
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name arranged in a circular fashion around a symbol. The symbol is a stylized representation of an eagle or bird, with three curved lines forming its body and wings. The text is in all caps and reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 16, 2013
Waters Medical Systems, LLC % Robert Warren General Manager 2112 15th Street NW Rochester, MN 55901
Re: K121736
Trade/Device Name: PERF-GEN® Pulsatile Perfusion Solution Regulation Number: 21 CFR§ 876.5880 Regulation Name: Isolated kidney perfusion and transport system and accessories Regulatory Class: II Product Code: KDL Dated: August 1, 2013 Received: August 5, 2013
Dear Robert Warren,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. or any Fourth all the Act's requirements, including, but not limited to: registration and I ou must occipij marting (21 CFR Part 801); medical device reporting (reporting of
{3}------------------------------------------------
Page 2 - Robert Warren
medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): K 121736
PERF-GEN® Pulsatile Perfusion Solution Device Name:
Indications for Use:
The PERF-GEN® Pulsatile Perfusion Solution is intended to be used for flushing and continuous hypothermic machine perfusion of kidneys at the time of their removal from the donor in preparation for storage, transportation, and eventual transplantation into a recipient.
Prescription Use >< (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
# Benjamin R. Fisher -S 2013.08.16 17:54:34 -04'00'
(Division Sign-Off) Division of Reproductive, Gastro-Renal, and Urological Devices K121736 510(k) Number _
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.