← Product Code [IYN](/productcode/IYN) · K121729

# E-CUBE 7 (K121729)

_Alpinion Medical Systems Co., Ltd. · IYN · Sep 20, 2012 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K121729

## Device Facts

- **Applicant:** Alpinion Medical Systems Co., Ltd.
- **Product Code:** [IYN](/productcode/IYN.md)
- **Decision Date:** Sep 20, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1550
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

The device is intended for use by a qualified physician for the evaluation of soft tissue and blood flow in the clinical applications; Fetal; Abdominal (renal & GYN/pelvic); Pediatric; Small Organ (breast, testes, thyroid); Trans-rectal(TR); Trans-vaginal(TV); Musculo-skeletal(Conventional); Musculo-skeletal (Superficial); Cardiac (adult); Peripheral Vascular (PV); and Urology (including prostate).

## Device Story

E-CUBE 7 is an ultrasound imaging system for medical diagnosis; provides soft tissue and blood flow evaluation. System inputs ultrasound signals via transducers (models GI-Q, L3-12, SP1-5, EN3-10, E3-10); processes signals to generate images; outputs displayed on 18.5" wide flat panel. Operated by qualified physicians in clinical settings. Features ergonomic control panel and user interface. Benefits include optimized diagnostic workflow and image quality. No clinical studies required for clearance; substantial equivalence established via bench testing and comparison to predicate.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including biocompatibility, acoustic output, and safety (thermal, electrical, electromagnetic, mechanical) testing.

## Technological Characteristics

Ultrasound imaging system; 18.5" flat panel display; ergonomic control panel. Transducers: GI-Q, L3-12, SP1-5, EN3-10, E3-10. Modes: B, M, PWD, Color Doppler, Power Doppler, Tissue Harmonic Imaging, 3D/4D. Biocompatible patient-contact materials. Complies with medical device safety standards.

## Regulatory Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- E-CUBE 9 Diagnostic Ultrasound System ([K120060](/device/K120060.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

<12/729

#### 510(k) E-CUBE 7

# 510(k) Summary

In accordance with 21CFR807.92, the following summary of information is provided;

Date June 8tt 2012

Submitter: ALPINION MEDICAL SYSTEMS Co., Ltd. Address: 1, 6 and 7FL Verdi Tower, 72, Digital-ro(St) 26-gil(Rd), Guro-gu, Seoul, Republic of Korea 152-848,

Primary Contact Person

Donghwan Kim QARA Manager Address: 1, 6 and 7FL Verdi Tower, 72, Digital-ro(St) 26-gil(Rd), Guro-gu, Seoul, Republic of Korea 152-848, Phone: +82 70 7465 2068 Fax: +82 2 851 5594 Email: donghwan.kim@alpinion.com

Secondary Contact Person

Yuchi Chu Address: Suite 229, 10604 NE 38th Place, Kirkland, WA 98033, United States Phone: 425 949 4907 Fax: 425 949 4908 Email: ychu@alpinionus.com

E-CUBE 7 Device Trade Name:

Ultrasonic Pulsed Doppler Imaging System Common/Usual Name:

System, Imaging, Pulsed Doppler Ultrasonic · Classification Names

> Ultrasonic Pulsed Doppler Imaging System, 21CFR 892.1550 90-Product Code: IYN Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYO

Predicate Device(s):

Predicate Device(s) K120060 E-CUBE 9 Diagnostic Ultrasound System

Diagnostic Ultrasound Transducer, 21CFR 892.1570, 90-ITX

{1}------------------------------------------------

510(k) E-CUBE 7

Device Description:

E-CUBE 7 product is an ultrasound imaging system for medical diagnosis. The system platform provides optimal patient diagnosis workflow with the 18.5" wide flat panel display, ergonomic control panel with easy user interface, optimal image quality.

Indications For Use:

The device is intended for use by a qualified physician for the evaluation of soft tissue and blood flow in the clinical applications; Fetal; Abdominal (renal & GYN/pelvic); Pediatric; Small Organ thyroid): Trans-rectal(TR); Trans-vaginal(TV); (breast. testes, Musculo-skeletal(Conventional); Musculo-skeletal (Superficial); Cardiac (adult); Peripheral Vascular (PV); and Urology (including prostate).

E-CUBE 7 employs the same fundamental scientific technology as Technology: its predicate device.

Determination of Substantial Equivalence:

Summary of Non-Clinical Tests:

E-CUBE 7 has been evaluated for biocompatibility, acoustic output as well as thermal, electrical, electromagnetic, and mechanical safety, and has been found to conform to applicable medical device safety standards. E-CUBE 7 and its application comply with voluntary standards as detailed in this premarket submission. The following quality management system measures were applied to the development of E-CUBE 7:

- � Medical Device Risk Management
- � Requirements Reviews
- Design Reviews �
- Component Verification ◆
- Integration Review (System Verification) �
- Performance Testing (System Verification) �
- Safety Testing (Compliance Test) ◆
- Design Validation

Transducer materials and other patient contact materials are biocompatible.

#### Summary of Clinical Tests:

The subject of this premarket submission, E-CUBE 7, did not require clinical studies to support substantial equivalence.

Conclusion:

Alpinion Medical Systems Co., Ltd. Considers E-CUBE 7 to be as safe, as effective, and performance is substantially equivalent to the predicate device.

ALPINION MEDICAL SYSTEMS Co., Ltd. will update and include in this summary any other information deemed reasonably necessary by the FDA or the requirements will be published in guidance documents.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal features the department's name encircling a stylized emblem. The emblem is a stylized image of an eagle.

#### Food and Drug Administration

10903 New Hampshire Avenue Silver Spring, MD 20993

SEP 2 0 2012

Mr. Donghwan Kim QARA Manager Alpinion Medical Systems Co., Ltd. 1. 6 and 7FL. Verdi Tower, 72 Digital-ro (St) 26-gil (Rd), Guro-gu SEOUL 152-848 REPUBLIC OF KOREA

Re: K121729

Trade/Device Name: E-CUBE 7 Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, and ITX Dated: July 20, 2012 . Received: July 20, 2012

Dear Mr. Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

This determination of substantial equivalence applies to the following transducers intended for use with the E-CUBE 7. as described in your premarket notification:

#### Transducer Model Number

GI-Q L3-12 SP1-5 EN3-10 E3-10

{3}------------------------------------------------

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

If you have any questions regarding the content of this letter, please contact Shahram Vaezy, PhD at (301) 796-6242.

Sincerely Yours,

Michal O'Kane for

Janine M. Morris Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure(s)

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known):

Device Name: E-CUBE 7

Indications for Use:

The device is intended for use by a qualified physician for the evaluation of soft tissue and blood flow in the clinical applications; Fetal; Abdominal (renal & GYN/pelvic); Pediatric; Small Organ (breast, testes, thyroid); Trans-rectal(TR); Trans-vaginal(TV); Musculo-skeletal(Conventional); Musculo-skeletal (Superficial); Cardiac (adult); Peripheral Vascular (PV); and Urology (including prostate).

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use_ (Part 21 CFR 807 Subpart C)

(Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)

(Division Sign-Off)

(Division Sign-Off)
Division of Radiological Devices
310k 51279

{5}------------------------------------------------

### Diagnostic Ultrasound Indications for Use Form

#### E-CUBE 7 Ultrasound System

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application                     | Mode of Operation |   |     |     |                  |                  |                               |                        |                      |
|------------------------------------------|-------------------|---|-----|-----|------------------|------------------|-------------------------------|------------------------|----------------------|
|                                          | B                 | M | PWD | CWD | Color<br>Doppler | Power<br>Doppler | Tissue<br>Harmonic<br>Imaging | Combined*<br>(Specify) | Other**<br>(Specify) |
| Ophthalmic                               |                   |   |     |     |                  |                  |                               |                        |                      |
| Fetal                                    | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| Abdominal                                | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| Intra-operative (Specify)                |                   |   |     |     |                  |                  |                               |                        |                      |
| Intra-operative (Neuro)                  |                   |   |     |     |                  |                  |                               |                        |                      |
| Laparoscopic                             |                   |   |     |     |                  |                  |                               |                        |                      |
| Pediatric                                | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| Small Organ<br>(breast, testes, thyroid) | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| Neonatal Cephalic                        |                   |   |     |     |                  |                  |                               |                        |                      |
| Adult Cephalic                           |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-rectal                             | P                 | P | P   |     | P                | P                |                               | P                      |                      |
| Trans-vaginal                            | P                 | P | P   |     | P                | P                |                               | P                      |                      |
| Trans-urethral                           |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-esoph. (non-Card.)                 |                   |   |     |     |                  |                  |                               |                        |                      |
| Musculo-skeletal<br>(Conventional)       | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| Musculo-skeletal<br>(Superficial)        | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| Intravascular                            |                   |   |     |     |                  |                  |                               |                        |                      |
| Cardiac Adult                            | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| Cardiac Pediatric                        |                   |   |     |     |                  |                  |                               |                        |                      |
| Intravascular (Cardiac)                  |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-esoph. (Cardiac)                   |                   |   |     |     |                  |                  |                               |                        |                      |
| Intra-cardiac                            |                   |   |     |     |                  |                  |                               |                        |                      |
| Peripheral vessel                        | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| Urology (including prostate)             | P                 | P | P   |     | P                | P                | P                             | P                      |                      |

N = new indication; P = previously cleared by FDA; E = added under appendix

* Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In-Vitro Diaggostic Devices

# #

**(Division Sign-Off)**
Division of Radiological Devices
K121729
QIVD

Prescription User (Per 21 CFR 801.109)

E-2

{6}------------------------------------------------

Diagnostic Ultrasound Indications for Use Form

# E-CUBE 7 with C1-6 Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application                     | Mode of Operation |   |     |     |                  |                  |                               |                        |                      |
|------------------------------------------|-------------------|---|-----|-----|------------------|------------------|-------------------------------|------------------------|----------------------|
|                                          | B                 | M | PWD | CWD | Color<br>Doppler | Power<br>Doppler | Tissue<br>Harmonic<br>Imaging | Combined*<br>(Specify) | Other**<br>(Specify) |
| Ophthalmic                               |                   |   |     |     |                  |                  |                               |                        |                      |
| Fetal                                    | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| Abdominal                                | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| Intra-operative (Specify)                |                   |   |     |     |                  |                  |                               |                        |                      |
| Intra-operative (Neuro)                  |                   |   |     |     |                  |                  |                               |                        |                      |
| Laparoscopic                             |                   |   |     |     |                  |                  |                               |                        |                      |
| Pediatric                                | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| Small Organ<br>(breast, testes, thyroid) |                   |   |     |     |                  |                  |                               |                        |                      |
| Neonatal Cephalic                        |                   |   |     |     |                  |                  |                               |                        |                      |
| Adult Cephalic                           |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-rectal                             |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-vaginal                            |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-urethral                           |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-esoph. (non-Card.)                 |                   |   |     |     |                  |                  |                               |                        |                      |
| Musculo-skeletal<br>(Conventional)       |                   |   |     |     |                  |                  |                               |                        |                      |
| Musculo-skeletal<br>(Superficial)        |                   |   |     |     |                  |                  |                               |                        |                      |
| Intravascular                            |                   |   |     |     |                  |                  |                               |                        |                      |
| Cardiac Adult                            |                   |   |     |     |                  |                  |                               |                        |                      |
| Cardiac Pediatric                        |                   |   |     |     |                  |                  |                               |                        |                      |
| Intravascular (Cardiac)                  |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-esoph. (Cardiac)                   |                   |   |     |     |                  |                  |                               |                        |                      |
| Intra-cardiac                            |                   |   |     |     |                  |                  |                               |                        |                      |
| Peripheral vessel                        |                   |   |     |     |                  |                  |                               |                        |                      |
| Urology (including prostate)             | P                 | P | P   |     | P                | P                | P                             | P                      |                      |

N = new indication; P = previously cleared by FDA; E = added under appendix

* Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

| Concurrence of CDRH, Office of In-Vitro | <i>Diagnostic Devices</i> |
|-----------------------------------------|---------------------------|
|-----------------------------------------|---------------------------|

510

Prescription User (Per 21 CFR 801.109)

{7}------------------------------------------------

Diagnostic Ultrasound Indications for Use Form

### E-CUBE 7 with L3-12 Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application                     | Mode of Operation |   |     |     |                  |                  |                               |                        |                      |  |
|------------------------------------------|-------------------|---|-----|-----|------------------|------------------|-------------------------------|------------------------|----------------------|--|
|                                          | B                 | M | PWD | CWD | Color<br>Doppler | Power<br>Doppler | Tissue<br>Harmonic<br>Imaging | Combined*<br>(Specify) | Other**<br>(Specify) |  |
| Ophthalmic                               |                   |   |     |     |                  |                  |                               |                        |                      |  |
| Fetal                                    |                   |   |     |     |                  |                  |                               |                        |                      |  |
| Abdominal                                |                   |   |     |     |                  |                  |                               |                        |                      |  |
| Intra-operative (Specify)                |                   |   |     |     |                  |                  |                               |                        |                      |  |
| Intra-operative (Neuro)                  |                   |   |     |     |                  |                  |                               |                        |                      |  |
| Laparoscopic                             |                   |   |     |     |                  |                  |                               |                        |                      |  |
| Pediatric                                | P                 | P | P   |     | P                | P                | P                             | P                      |                      |  |
| Small Organ<br>(breast, testes, thyroid) | P                 | P | P   |     | P                | P                | P                             | P                      |                      |  |
| Neonatal Cephalic                        |                   |   |     |     |                  |                  |                               |                        |                      |  |
| Adult Cephalic                           |                   |   |     |     |                  |                  |                               |                        |                      |  |
| Trans-rectal                             |                   |   |     |     |                  |                  |                               |                        |                      |  |
| Trans-vaginal                            |                   |   |     |     |                  |                  |                               |                        |                      |  |
| Trans-urethral                           |                   |   |     |     |                  |                  |                               |                        |                      |  |
| Trans-esoph. (non-Card.)                 |                   |   |     |     |                  |                  |                               |                        |                      |  |
| Musculo-skeletal<br>(Conventional)       | P                 | P | P   |     | P                | P                | P                             | P                      |                      |  |
| Musculo-skeletal<br>(Superficial)        | P                 | P | P   |     | P                | P                | P                             | P                      |                      |  |
| Intravascular                            |                   |   |     |     |                  |                  |                               |                        |                      |  |
| Cardiac Adult                            |                   |   |     |     |                  |                  |                               |                        |                      |  |
| Cardiac Pediatric                        |                   |   |     |     |                  |                  |                               |                        |                      |  |
| Intravascular (Cardiac)                  |                   |   |     |     |                  |                  |                               |                        |                      |  |
| Trans-esoph. (Cardiac)                   |                   |   |     |     |                  |                  |                               |                        |                      |  |
| Intra-cardiac                            |                   |   |     |     |                  |                  |                               |                        |                      |  |
| Peripheral vessel                        | P                 | P | P   |     | P                | P                | P                             | P                      |                      |  |
| Urology (including prostate)             |                   |   |     |     |                  |                  |                               |                        |                      |  |

N = new indication; P = previously cleared by FDA; E = added under appendix

* Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D

:

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUEZON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In-Vitre Diagnostic Devices (IV
(Division Sign-Of)
Division of Radiological Devices
510k K1217890

Prescription User (Per 21 CFR 801.109)

ALPINION MEDICAL SYSTEMS Co., Ltd.

E-4

{8}------------------------------------------------

Diagnostic Ultrasound Indications for Use Form

## E-CUBE 7 with SP1-5 Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application                     | Mode of Operation |   |     |     |                  |                  |                               |                        |                      |
|------------------------------------------|-------------------|---|-----|-----|------------------|------------------|-------------------------------|------------------------|----------------------|
|                                          | B                 | M | PWD | CWD | Color<br>Doppler | Power<br>Doppler | Tissue<br>Harmonic<br>Imaging | Combined*<br>(Specify) | Other**<br>(Specify) |
| Ophthalmic                               |                   |   |     |     |                  |                  |                               |                        |                      |
| Fetal                                    |                   |   |     |     |                  |                  |                               |                        |                      |
| Abdominal                                | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| Intra-operative (Specify)                |                   |   |     |     |                  |                  |                               |                        |                      |
| Intra-operative (Neuro)                  |                   |   |     |     |                  |                  |                               |                        |                      |
| Laparoscopic                             |                   |   |     |     |                  |                  |                               |                        |                      |
| Pediatric                                | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| Small Organ<br>(breast, testes, thyroid) |                   |   |     |     |                  |                  |                               |                        |                      |
| Neonatal Cephalic                        |                   |   |     |     |                  |                  |                               |                        |                      |
| Adult Cephalic                           |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-rectal                             |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-vaginal                            |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-urethral                           |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-esoph. (non-Card.)                 |                   |   |     |     |                  |                  |                               |                        |                      |
| Musculo-skeletal<br>(Conventional)       |                   |   |     |     |                  |                  |                               |                        |                      |
| Musculo-skeletal<br>(Superficial)        |                   |   |     |     |                  |                  |                               |                        |                      |
| Intravascular                            |                   |   |     |     |                  |                  |                               |                        |                      |
| Cardiac Adult                            | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| Cardiac Pediatric                        |                   |   |     |     |                  |                  |                               |                        |                      |
| Intravascular (Cardiac)                  |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-esoph. (Cardiac)                   |                   |   |     |     |                  |                  |                               |                        |                      |
| Intra-cardiac                            |                   |   |     |     |                  |                  |                               |                        |                      |
| Peripheral vessel                        |                   |   |     |     |                  |                  |                               |                        |                      |
| Urology (including prostate)             |                   |   |     |     |                  |                  |                               |                        |                      |

N = new indication: P = previously cleared by FDA; E = added under appendix

* Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In-Vitro Diagnostic De

| Prescription User (Per 21 CFR 801.109) |  |
|----------------------------------------|--|
|----------------------------------------|--|

(Division Sigh-Off)
Division of Radiological Devices
510k K121789

ALPINION MEDICAL SYSTEMS Co., Ltd.

E-5

{9}------------------------------------------------

Diagnostic Ultrasound Indications for Use Form

## E-CUBE 7 with EN3-10 Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application                                                        | Mode of Operation |   |     |     |         |         |                     |           |           |
|-----------------------------------------------------------------------------|-------------------|---|-----|-----|---------|---------|---------------------|-----------|-----------|
|                                                                             | B                 | M | PWD | CWD | Color   | Power   | Tissue              | Combined* | Other**   |
|                                                                             |                   |   |     |     | Doppler | Doppler | Harmonic<br>Imaging | (Specify) | (Specify) |
| Ophthalmic                                                                  |                   |   |     |     |         |         |                     |           |           |
| Fetal                                                                       |                   |   |     |     |         |         |                     |           |           |
| Abdominal                                                                   |                   |   |     |     |         |         |                     |           |           |
| Intra-operative (Specify)                                                   |                   |   |     |     |         |         |                     |           |           |
| Intra-operative (Neuro)                                                     |                   |   |     |     |         |         |                     |           |           |
| Laparoscopic                                                                |                   |   |     |     |         |         |                     |           |           |
| Pediatric                                                                   |                   |   |     |     |         |         |                     |           |           |
| Small Organ<br>(breast, testes, thyroid)                                    |                   |   |     |     |         |         |                     |           |           |
| Neonatal Cephalic                                                           |                   |   |     |     |         |         |                     |           |           |
| Adult Cephalic                                                              |                   |   |     |     |         |         |                     |           |           |
| Trans-rectal                                                                | P                 | P | P   |     | P       | P       |                     | P         |           |
| Trans-vaginal                                                               | P                 | P | P   |     | P       | P       |                     | P         |           |
| Trans-urethral                                                              |                   |   |     |     |         |         |                     |           |           |
| Trans-esoph. (non-Card.)                                                    |                   |   |     |     |         |         |                     |           |           |
| Musculo-skeletal<br>(Conventional)                                          |                   |   |     |     |         |         |                     |           |           |
| Musculo-skeletal<br>(Superficial)                                           |                   |   |     |     |         |         |                     |           |           |
| Intravascular                                                               |                   |   |     |     |         |         |                     |           |           |
| Cardiac Adult                                                               |                   |   |     |     |         |         |                     |           |           |
| Cardiac Pediatric                                                           |                   |   |     |     |         |         |                     |           |           |
| Intravascular (Cardiac)                                                     |                   |   |     |     |         |         |                     |           |           |
| Trans-esoph. (Cardiac)                                                      |                   |   |     |     |         |         |                     |           |           |
| Intra-cardiac                                                               |                   |   |     |     |         |         |                     |           |           |
| Peripheral vessel                                                           |                   |   |     |     |         |         |                     |           |           |
| Urology (including prostate)                                                | P                 | P | P   |     | P       | P       |                     | P         |           |
| N = new indication; P = previously cleared by FDA; E = added under appendix |                   |   |     |     |         |         |                     |           |           |

* Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANQTHER FAGE IF NEEDED)

510k

Concurrence of CDRH, Office of In-Vitro Digg

Prescription User (Per 21 CFR 801.109)

#### ALPINION MEDICAL SYSTEMS Co., Ltd.

Radiological Devices

{10}------------------------------------------------

Diagnostic Ultrasound Indications for Use Form

### E-CUBE 7 with E3-10 Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application         | Mode of Operation |   |     |     |                  |                  |                               |                        |                      |
|------------------------------|-------------------|---|-----|-----|------------------|------------------|-------------------------------|------------------------|----------------------|
|                              | B                 | M | PWD | CWD | Color<br>Doppler | Power<br>Doppler | Tissue<br>Harmonic<br>Imaging | Combined*<br>(Specify) | Other**<br>(Specify) |
| Ophthalmic                   |                   |   |     |     |                  |                  |                               |                        |                      |
| Fetal                        |                   |   |     |     |                  |                  |                               |                        |                      |
| Abdominal                    |                   |   |     |     |                  |                  |                               |                        |                      |
| Intra-operative (Specify)    |                   |   |     |     |                  |                  |                               |                        |                      |
| Intra-operative (Neuro)      |                   |   |     |     |                  |                  |                               |                        |                      |
| Laparoscopic                 |                   |   |     |     |                  |                  |                               |                        |                      |
| Pediatric                    |                   |   |     |     |                  |                  |                               |                        |                      |
| Small Organ                  |                   |   |     |     |                  |                  |                               |                        |                      |
| (breast, testes, thyroid)    |                   |   |     |     |                  |                  |                               |                        |                      |
| Neonatal Cephalic            |                   |   |     |     |                  |                  |                               |                        |                      |
| Adult Cephalic               |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-rectal                 | P                 | P | P   |     | P                | P                |                               | P                      |                      |
| Trans-vaginal ·              | P                 | P | P   |     | P                | P                |                               | P                      |                      |
| Trans-urethral               |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-esoph. (non-Card.)     |                   |   |     |     |                  |                  |                               |                        |                      |
| Musculo-skeletal             |                   |   |     |     |                  |                  |                               |                        |                      |
| (Conventional)               |                   |   |     |     |                  |                  |                               |                        |                      |
| Musculo-skeletal             |                   |   |     |     |                  |                  |                               |                        |                      |
| (Superficial)                |                   |   |     |     |                  |                  |                               |                        |                      |
| Intravascular                |                   |   |     |     |                  |                  |                               |                        |                      |
| Cardiac Adult                |                   |   |     |     |                  |                  |                               |                        |                      |
| Cardiac Pediatric            |                   |   |     |     |                  |                  |                               |                        |                      |
| Intravascular (Cardiac)      |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-esoph. (Cardiac)       |                   |   |     |     |                  |                  |                               |                        |                      |
| Intra-cardiac                |                   |   |     |     |                  |                  |                               |                        |                      |
| Peripheral vessel            |                   |   |     |     |                  |                  |                               |                        |                      |
| Urology (including prostate) | P                 | P | P   |     | P                | P                |                               | P                      |                      |

N = new indication; P = previously cleared by FDA; E = added under appendix

* Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

510k

Concurrence of CDRH, Office of In-Vitro *Diagnostic Devices*

Prescription User (Per 21 CFR 801.109)

E-7

Division Sign-Off of Radiological Devices

---

**Source:** [https://fda-staging.innolitics.com/device/K121729](https://fda-staging.innolitics.com/device/K121729)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).