The New Device is intended for clinical practice of general dentistry, as an aid in the diagnosis of pit and fissure caries, as an aid to highlight dental plaque and gingival inflammations (restricted to gingival inflammations which lead to bleeding upon probing) and as intra-oral camera to visualize anatomical details invisible to the naked eye or with a mirror (thanks to its magnification).
Device Story
SOPROCARE is an intra-oral video camera for general dentistry. It uses white and blue (450nm) LEDs to illuminate the oral cavity. The device operates in three modes: CARIO (blue light, fluorescence for pit/fissure caries detection), DAYLIGHT (white light, standard magnification), and PERIO (yellow light, chromatic amplification for dental plaque and gingival inflammation). A 1/4" CCD sensor captures images, which are transmitted to a monitor for clinician review. The device aids practitioners in diagnosis, patient education, and tracking treatment progress. It is used in clinical settings by dentists or hygienists. The device provides visual feedback to assist in identifying conditions that might be missed by the naked eye, potentially improving diagnostic accuracy and patient motivation.
Clinical Evidence
Clinical study (n=20, ages 20-41) evaluated diagnostic aid capabilities against gold standards (SOPROLIFE for caries, GC Tri Plaque ID Gel for plaque, bleeding upon probing for inflammation). 120 teeth analyzed for plaque/inflammation, 128 for caries. SOPROCARE achieved 87% accuracy, 91% sensitivity, and 90% specificity compared to gold standards. Practitioner visual inspection alone achieved 66% accuracy, 25% sensitivity, and 97% specificity. No adverse events reported.
Technological Characteristics
Intra-oral camera with 4 white LEDs and 3 blue LEDs (450nm). 1/4" high-sensitivity CCD sensor. Resolution: 752x582 (PAL) or 768x494 (NTSC). Definition: 470 lines. Sensitivity: 2 lux. Housing: self-extinguishing material (UL94-V0). Connectivity: docking station (USB2). Sterilization: uses disposable tips and protection sheaths. Operates in 10-40°C.
Indications for Use
Indicated for general dentistry patients as an aid in diagnosing pit and fissure caries, highlighting dental plaque, and identifying gingival inflammations (specifically those leading to bleeding upon probing). Also used as an intra-oral camera for visualizing anatomical details.
Regulatory Classification
Identification
A laser fluorescence caries detection device is a laser, a fluorescence detector housed in a dental handpiece, and a control console that performs device calibration, as well as variable tone emitting and fluorescence measurement functions. The intended use of the device is to aid in the detection of tooth decay by measuring increased laser induced fluorescence.
Special Controls
*Classification.* Class II, subject to the following special controls:(1) Sale, distribution, and use of this device are restricted to prescription use in accordance with § 801.109 of this chapter;
(2) Premarket notifications must include clinical studies, or other relevant information, that demonstrates that the device aids in the detection of tooth decay by measuring increased laser induced fluorescence; and
(3) The labeling must include detailed use instructions with precautions that urge users to:
(i) Read and understand all directions before using the device,
(ii) Store probe tips under proper conditions,
(iii) Properly sterilize the emitter-detector handpick before each use, and
(iv) Properly maintain and handle the instrument in the specified manner and condition.
1. restrictions on the sale, distribution, and use of the device to prescription use in accordance with 21 CFR 801.109;
2. clinical studies, or other relevant information, that demonstrates that the device aids in the detection of tooth decay by measuring increased laser induced fluorescence; and
3. labeling that includes detailed use instructions with precautions that urge users to (i) read and understand all directions before using the device, (ii) store probe tips under proper conditions, (iii) properly sterilize the emitter-detector handpiece before each use, and (iv) properly maintain and handle the instrument in the specified manner and condition.
{0}------------------------------------------------
**Section 05 – 510(k) Summary**
K12+685
# MAR 1 2 2013
This Summary of Safety and Effectiveness is submitted in accordance with 21 CFR 807.92.c.
### 1 – Administrative Information
#### Type of 510(k) submission: 1- a.
These documents constitute a Traditional 510(k) Submission.
1- b. Submission date: June 1, 2012
510(k) Submitter: 1-с.
SOPRO Acteongroup Avenue des Genevriers ZAC Athelia IV 13705 La Ciotat Cedex France Telephone: 33 (0) 442 98 01 01 Fax: 33 (0) 442 71 76 90
1-d. Contact Person:
Rick Rosati SOPRO clo ACTEON, Inc. 124 Gaither Drive, Suite 140 Mt. Laurel, NJ 08054 Telephone: 800-289-6367 Ext. 390 Fax: 856-222-4726 E-Mail: rick.rosati@us.acteongroup.com
1- ė. Establishment registration number:
9615782
Pre-Market Notification 510(k) Submission for SOPROCARE By SOPRO Version 04
{1}------------------------------------------------
2 - Device Information
| 2-a. | Common Name of device: | Laser fluorescence caries detection device |
|------|----------------------------|--------------------------------------------|
| 2-b. | Trade Name of device: | SOPROCARE |
| 2-c. | Classification regulation: | 21 CFR 872.1745 |
| 2-d. | Medical Device Class: | II |
| 2-e. | Panel: | Dental |
| 2-f. | Product code: | NBL |
| 2-g | Subsequent Product Codes | ΝΥΗ, ΕΙΧ |
3 - Identification of legally marketed device(s)
The Substantial Equivalence (SE) of the New Device is based on the Predicates Devices identified in the Table 01.
| Trade Name | Manufacturer | Product<br>Code | 510(k) number | Date Cleared |
|-----------------------|----------------|-----------------|---------------|---------------------|
| SOPROLIFE | SOPRO | NBL | K092583 | January 13,<br>2010 |
| GC Tri Plaque ID Gel | GC CORPS | NYH | Not required | NA |
| PROBE,<br>PERIODONTIC | AESCULAP<br>AG | EIX | Exempted | NA |
Table 01 - Identification of legally marketed devices
SOPRO SOPROLIFE (K092583) device is used as Predicate for pit and fissure caries and intra-oral camera claims for the New Device.
GC CORPS Tri Plaque ID Gel (510K not required) is used as Predicate for dental plaque highlight claims for the New Device.
AESCULAP AG PROBE (510K Exempted) is used as Predicate for gingival inflammations highlight claims for the New Device (restricted to gingival inflammations which lead to bleeding upon probing).
Pre-Market Notification 510(k) Submission for SOPROCARE By SOPRO Version 04
{2}------------------------------------------------
### 4 - Description of the Device .
The New Device is intended for clinical practice of general dentistry, as an aid in the diagnosis of pit and fissure caries, as an aid to highlight dental plaque and gingival inflammations (restricted to gingival inflammations which lead to bleeding upon probing) and as intra-oral camera to visualize anatomical details invisible to the naked eye or with a mirror (thanks to its magnification).
It provides the following benefits:
- Aids in the detection of pit and fissure caries l
- Information about patient dental hygiene -
- Highlight dental plaque -
-
- Highlight gingival inflammations (restricted to gingival inflammations which lead to bleeding upon probing)
- . Show the difference between "before" and "after" care (follow-up).
In CARIO mode (blue mode), the camera helps the dental practitioner to highlight carious warning on pits and fissures of the occlusal side of the teeth.
In DAYLIGHT mode (white mode), the camera enables to visualize anatomical details invisible to the naked eye or with a mirror (thanks to its magnification).
In PERIO mode (yellow mode), the camera helps the dental practitioner to see the presence of dental plaque but also to highlight gingival inflammations (restricted to gingival inflammations which lead to bleeding upon probing). This mode offers the dentist and/or hygienist a tool for improved communication, motivation and education of his/her patients, who will then become aware of their oral health condition.
The New Device is an intra-oral video camera equipped with LEDs to light the inspection site. Thanks to its technology based on fluorescence phenomenon (given by its LED lamps) and chromatic amplification, this camera will permit a global care of the patient. The optics and the charge coupled device (CCD) sensor in the New Device picks up the images containing this fluorescence highlight site and converts them to a video signal that is sent to a video monitor or computer monitor. Dental practitioner and/or hygienist, as an aid for diagnosis, can use the results of this image.
The handpiece can be connected with different SOPRO docking station such as: DOCK M USB2, DOCK MU_USB2, DOCK USB2 .....
### Pre-Market Notification 510(k) Submission for SOPROCARE By SOPRO Version 04
{3}------------------------------------------------
### TABLE 2
### Main Design Characteristics of New Device
| Light Source | 4 LEDs White<br>3 LEDs Blue |
|--------------------------------------------|----------------------------------------------------------------------------------------|
| Wavelength (blue led)<br>excitation signal | 450nm |
| Returned light | Standard image and Fluorescence |
| CCD | ¼" High sensitivity |
| Resolution | 752 x 582 PAL<br>768 x 494 NTSC |
| Definition | 470 lines |
| Sensitivity | 2 lux |
| Adjustment | 4 presets positions :<br>Extra oral<br>Intra oral<br>Care = fluorescence mode<br>Macro |
| System of image<br>capture | Image capture through SoproTouch or footswitch<br>(optional) |
| Angle of view | 70° |
| Cable length | 2,5 m |
| Hand piece<br>dimension | L = 200 mm x W = 28 mm x H = 24 mm |
| Hand piece weight | 78g |
| Working temperature | 10-40°C ( 50-104°F) |
| Power supply | Depending docking station |
| User interface | LCD Screen or Computer |
#### 5 - Intended Use
The New Device is intended for clinical practice of general dentistry, as an aid in the diagnosis of pit and fissure caries, as an aid to highlight dental plaque and gingival inflammations (restricted to gingival inflammations which lead to bleeding upon probing) and as intra-oral camera to visualize anatomical details invisible to the naked eye or with a mirror (thanks to its magnification).
### Pre-Market Notification 510(k) Submission for SOPROCARE By SOPRO Version 04
{4}------------------------------------------------
### 6 - Technological characteristics of the New Device compared to the Predicates Devices
6.1 Technological characteristics of the New Device compared to the Predicate Device for pit and fissure caries detection and intra-oral camera.
The Predicate Device is SOPROLIFE (K092583, January 13, 2010)
#### Technological perspective:
The New Device and the Predicate Device use the same technology (CCD sensor, Blue and white LEDs) and the same technical concept (fluorescence phenomenon).
#### Material perspective:
The New Device and the Predicate Device are extremely similar because they both share the same external housing (shape and material, only color is different).
#### Design perspective:
The New Device and the Predicate Device use the same:
- Electronic board (except led board). -
- External Housing (except shape color). -
- Optical design. -
- Accessories. -
-
#### Eneray source perspective:
The New Device and the Predicate Device:
- Use the same input energy source (electric power supply). ı
- Deliver the same output energy source (light emission).
#### Material in contact with the patient:
Because the New Device and the Predicate Device use the same tips and protection sheath, the materials in contact with the patient are the same.
Displayed image for aid to pit and fissure caries detection
- The New Device and the Predicate Device:
- Use an anatomic view of the tooth ।
- Highlight the suspicious pit and fissure caries in red
The Predicate Device displays healthy dentine fluorescence in green.
The New Device displays the area outside pit and fissure caries in black and white.
### Pre-Market Notification 510(k) Submission for SOPROCARE By SOPRO Version 04
{5}------------------------------------------------
6.2 Technological characteristics of the New Device compared to the Predicate Device for highlight dental plaque.
The Predicate Device is GC Tri Plaque ID Gel (510k exempt)
#### Technological perspective:
The New Device uses a different technology (CCD sensor, Blue and white LEDs) compared to a chemical solution. The New Device uses fluorescence of bacteria present in dental plaque whereas Predicate Device is a coloration of dental plaque.
#### Material perspective:
The New Device is made with plastic and electronic. The Predicate Device is a solution that needs to be applied everywhere in the patient's mouth.
#### Design perspective:
The New Device is made up of plastic and electronic.
The Predicate Device is a solution that needs to be applied everywhere in the patient's mouth.
#### Energy source perspective:
The New Device uses electric power supply. The Predicate Device does not require energy source.
#### Material in contact with the patient:
The New Device materials in contact with the patient are the tips and protection sheath, the contact is quite brief and punctual.
The Predicate Device needs to be applied everywhere in the mouth and coloration remains several hours after examination.
{6}------------------------------------------------
6.3 Technological characteristics of the New Device compared to the Predicate Device for highlighting gingival inflammations (restricted to gingival inflammations, which lead to bleeding upon probing)
The Predicate Device is AESCULAP AG Periodontic probe (510k Exempt)
The gingival inflammation is mainly detected by a visual inspection of the gum, if a more effective diagnosis is to be placed, practitioner can use a more invasive technique: bleeding upon probing test.
#### Technological perspective:
The New Device uses a different technology, chromatic amplification (using CCD sensor, Blue and white LEDs) compared to bleeding upon probing (which is a visual inspection of site after probing).
#### Material perspective:
The New Device is made with plastic and electronic. Other cited techniques use human eye and a probe.
#### Design perspective:
The New Device is made with plastic and electronic. Other cited techniques use human eye and a probe.
#### Energy source perspective:
The New Device uses electric power supply. Other cited techniques don't require energy source.
#### Material in contact with the patient:
The New Device materials in contact are the tips and protection sheath, the contact is quite brief and punctual.
For bleeding after probing techniques, the probe is firmly pressed in the gum in several areas.
### Pre-Market Notification 510(k) Submission for SOPROCARE By SOPRO Version 04
{7}------------------------------------------------
.
#### 7 - Determination of substantial equivalence
The New Device is an intra-oral video camera equipped with LEDs to light the inspection site. The Predicate Device SOPROLIFE is the same kind of device. Indications for use are similar for pit and fissure caries.
The New Device is the same as the Soprolife (K092583, January 13, 2010) Predicate Device in terms of functioning principle. Both devices use the fluorescence phenomenon. Also the Devices use the same housing, dental barrier and the same tips.
From internal structure perspective, Devices are extremely similar, all components used are the same for the New Device and the Predicate Device (with exception of the led board).
From external structure perspective, the devices are extremely similar because the casings are made in self-extinguishing material (UL94-V0) only color of housing has changed.
Moreover, the materials in contact with the patient are exactly the same.
Regarding dental plaque highlight, New Device and Predicate Device (GC Tri Plaque ID Gel) have the same intended use. The technology is different but the new characteristics do not affect safety or effectiveness.
Regarding gingival inflammations highlight (restricted to gingival inflammations which lead to bleeding upon probing), New Device and Predicate Device (Periodontic probe) have the same intended use. The technology is different but the new characteristics don't affect safety or effectiveness.
### Discussion of the non-clinical Tests:
The aim of the evaluation was to demonstrate the Substantial Equivalence between New Device and the selected Predicate Device SOPROLIFE in terms of Performances.
The evaluated Performances were:
- Image focus position ।
- Image luminosity -
- Led illumination -
After tests, the obtained results for the New Device have been directly compared to the Predicate Device Performances. The results of the comparison show that the Performances of the New Device and the Predicate Device are similar.
#### Discussion of the safety Tests:
From a safety point of view: The device has been tested to the following standards:
The Electromagnetic Compatibility has been performed according to:
Pre-Market Notification 510(k) Submission for SOPROCARE By SOPRO Version 04
{8}------------------------------------------------
- AAMI / ANSI / IEC 60601-1-2:2007, Medical electrical equipment Part 1-2: -General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests (Edition 3). (General) (Recognition Number: 5-54).
The Electrical Safety has been performed according to:
- The IEC 60601-2-18: Edition 3.0 2009-08, Medical electrical equipment Part 2-18: Particular requirements for the basic safety and essential performance of endoscopic equipment. (Dental/ENT) Recognition Number 4-187
Discussion about clinical data:
Clinical data has been collected during the following study:
### "Diagnosis aid capabilities of Soprocare camera" January 6 to February 7 2012 at Marseille
Definition: In medicine and statistics, gold standard test refers to a diagnostic test or benchmark that is the best available under reasonable conditions. The gold standard is the reference and it is supposed to never be wrong and so have sensitivity and specificity of 100%.
The aims of this study were to determine if:
- New Device has a comparable effectiveness as Predicate Device SOPROLIFE . (used as gold standard for pit and fissure caries detection).
- New Device has a comparable effectiveness as GC Tri Plaque ID Gel Predicate e Device (used as gold standard for dental plaque highlight).
- New Device has a comparable effectiveness as bleeding upon probing technique . (used as gold standard) regarding gingival inflammations (restricted to gingival inflammations which lead to bleeding upon probing).
- New Device would improve diagnosis capability compared to practitioner using . only visual inspection
20 people have been enrolled in this study (9 women and 11 men) aged from 20 to 41 years.
- A total of 120 teeth have been studied for dental plaque and inflammation analysis .
- A total of 128 teeth have been studied for pit and fissure caries detection .
In terms of safety the study did not show any particular risk associated with New Device, it showed same safety as Predicates Devices. No adverse events or complications occurred.
Pre-Market Notification 510(k) Submission for SOPROCARE By SOPRO Version 04
{9}------------------------------------------------
During this study, New Device has demonstrated its effectiveness by achieving:
- High accuracy (87 % in average) : capacity to get the same results as the gold ● standard
- High Sensitivity (91 % in average) : capacity to highlight a trouble detected by the . gold standard
- High specificity (90 % in average) : capacity to confirm the absence of a trouble . stated by the gold standard
The practitioner eyes achieved:
- Moderate accuracy (66 % in average) : capacity to get the same results as the ● gold standard
- Low Sensitivity (25 % in average) : capacity to highlight a trouble detected by the . gold standard
- High specificity (97 % in average) : capacity to confirm the absence of a trouble . stated by the gold standard
### 8 - Conclusion
The New Device is the same as the SOPROLIFE identified Predicate Device in terms of indication for use for aid to pit and fissure caries detection.
Because of the used technologies, characteristics and performances are similar to the Predicate Device; the characteristics of the SOPROCARE do not affect the Safety of the patients or of the operator. Moreover, the Effectiveness is the same as of the Predicate Device.
The New Device is the same as the GC Tri Plaque ID Gel identified Predicate Device in terms of indication for use for dental plaque highlight. Even though technology used is different, it doesn't affect safety and effectiveness as confirmed by performance data provided.
The New Device is the same as the periodontic probe identified Predicate Device in terms of indication for use for gingival inflammations highlight (restricted to gingival inflammations which lead to bleeding upon probing). Even though technology used is different, it doesn't affect safety and effectiveness as confirmed by performance data provided.
The New Device is as safe and effective as the predicates for the intended uses.
The New Device is substantially equivalent to the SOPROLIFE Predicate Device (K092583. January 13, 2010) as an aid to pit and fissure caries detection.
The New Device is substantially equivalent to GC Tri Plaque ID Gel Predicate Device as an aid to highlight dental plaque.
The New Device is substantially equivalent to periodontic probe Predicate Device as an aid to highlight gingival inflammations (restricted to gingival inflammations which lead to bleeding upon probing).
Pre-Market Notification 510(k) Submission for SOPROCARE By SOPRO Version 04
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized human figure with three arms or lines extending upwards and to the right. The figure is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the left side of the logo.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 12, 2013
SOPRO Acteongroup C/O Mr. Rick Rosati Quality Manager Acteon, Incorporated 124 Gaither Drive, Suite 140 MOUNT LAUREL NJ 08054
Re: K121685
Trade/Device Name: SOPROCARE Regulation Number: 21 CFR 872.1745 Regulation Name: Laser Fluorescence Caries Detection Device Regulatory Class: II Product Code: NBL, NYH Dated: February 14, 2013 Received: February 21, 2013
Dear Mr. Rosati:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{11}------------------------------------------------
Page 2 - Mr. Rosati
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Kwame O. Ulmer for
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{12}------------------------------------------------
### Section 04 - Indication for Use
#### Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name:
SOPROCARE
Indications for Use:
The New Device is intended for clinical practice of general dentistry, as an aid in the diagnosis of pit and fissure cares, as an aid to highlight dental plaque and gingival inflammations (restricted to gingival inflammations which lead to bleeding upon probing) and as intra-oral camera to visualize anatomical details invisible to the naked eye or with a mirror (thanks to its magnification).
× Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE ~ CONTINUE ON ANOTHER PAGE IF NEEDED)
AND/OR
## Concurrence of CDRH, Office of Device Evaluation (ODE)
Mary S. Runner -S .. Aus DOS MA 2013.03.11 - 15:30:11 -04'00'
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Page 1 of 1
| 510(k) Number: | K121685 |
|----------------|---------|
|----------------|---------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.