K121678 · Gyrus Acmi, Inc. · ERL · Aug 30, 2012 · Ear, Nose, Throat
Device Facts
Record ID
K121678
Device Name
DTAD
Applicant
Gyrus Acmi, Inc.
Product Code
ERL · Ear, Nose, Throat
Decision Date
Aug 30, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The DTAD is intended for cutting, coagulation, and removal of tissue in general ENT procedures. Specific indications include: Tonsillectomy and Adenoidectomy.
Device Story
Disposable, battery-operated microdebrider for ENT surgery; performs tissue cutting, coagulation, and removal. Input: manual activation via handpiece button for blade oscillation; electrosurgical generator footswitch for monopolar coagulation. Operation: internal motor oscillates inner blade gear at ~3500 rpm; monopolar lead connects to outer blade for coagulation. Features: flexible blade (0°-50° bend), nosecone for 90° tip rotation, suction port for fluid/tissue removal. Used in OR by surgeons. Output: mechanical resection of tissue and monopolar RF coagulation. Benefits: single-use convenience, integrated suction/cutting/coagulation, adjustable tip geometry for tonsillectomy/adenoidectomy access.
Clinical Evidence
No clinical testing conducted. Evidence provided via non-clinical bench testing (electrical safety per IEC 60601-1/2-2, mechanical/functional performance, stability/shelf-life) and preclinical evaluations (ex vivo bovine tissue for thermal impact/coagulation; in vivo porcine models for ergonomics, usability, and tissue removal).
Technological Characteristics
Single-use battery-operated microdebrider. Materials: patient-contacting parts identical to K034004; silicone RF insulator same as K041285. Energy: battery-powered motor (oscillation) and external electrosurgical generator (monopolar RF). Dimensions: flexible blade (0°-50°). Connectivity: monopolar cable to standard OR generator. Sterilization: Ethylene Oxide (EtO) per ISO 11135. Standards: IEC 60601-1, IEC 60601-2-2, ISO 10993-1.
Indications for Use
Indicated for patients undergoing general ENT procedures, specifically tonsillectomy and adenoidectomy.
Regulatory Classification
Identification
An ear, nose, and throat electric or pneumatic surgical drill is a rotating drilling device, including the handpiece, that is intended to drive various accessories, such as an ear, nose, and throat bur (§ 874.4140), for the controlled incision or removal of bone in the ear, nose, and throat area.
Predicate Devices
Gyrus ACMI Inc. Diego® RF Powered Dissector & Drill System (K034004)
{0}------------------------------------------------
Gyrus ACMI DTAD Gyrus ACMI, Inc.
K121678
Traditional 510(k) Notification June 2012
Page 1 of 6
# 510(k) Summary of Safety and Effectiveness Gyrus ACMI, Inc. DTAD
AUG 30 2012
## General Information
Manufacturer:
Gyrus ENT, L.L.C. (subsidiary of Gyrus ACMI, Inc.) 2925 Appling Road Bartlett , TN 38113 Phone: 1-800-262-3540 Fax: 1-901-373-0237
Establishment Registration Number:
510(k) Submitter:
Establishment Registration Number:
Contact Person:
Date Prepared:
## Device Description
Classification Name:
1037007
Gyrus ACMI, Inc. 136 Turnpike Rd. Southborough, MA 01772-2104
3003790304
Dolan Mills, RAC Specialist, Regulatory Affairs
June 6, 2012
Electrosurgical Cutting & Coagulation Device and Accessories, ENT Electric or Pneumatic Surgical Drill Class 2
21 CFR 878.4400 GEI General and Plastic Surgery Panel
21 CFR 874.4250 ERL Ear, Nose, Throat
Gyrus ACMI DTAD
Project Name:
ﻏ
{1}------------------------------------------------
Gyrus ACMI DTAD Gyrus ACMI, Inc. 121678
Traditional 510(k) Notification June 2012
Page 2 of 6
Trade Name(s):
Generic/Common Name:
Disposable Tonsil and Adenoid Device (DTAD)
Electrical Surgical Drill / Shaver or Electrosurgical cutting and coagulation device and accessories
# Predicate Devices
| Gyrus ACMI Inc. Diego® RF Powered Dissector & Drill System: | K034004 |
|-------------------------------------------------------------|---------|
| Covidien / ValleyLab Suction Coagulator: | K091223 |
| Bovie SEER: | K082568 |
| Peak TnA Plasmablade: | K083415 |
## Device Description and Technological Characteristics
The DTAD (Disposable Tonsil and Adenoid Device) is a disposable microdebrider blade with a built-in motor that will plug into the users electrosurgical generator. The sole energy application mode is Monopolar coagulation. The monopolar use will require a return electrode for the generator. The cutting performance is the same as the current Gyrus ACMI Diego® tonsil and adenoid blades cleared under K034004, with the addition of monopolar coagulation. The monopolar effect is completely dependent upon the generator used and its settings, which is the same as the other monopolar predicates.
The device will be a single-use battery operated microdebrider with monopolar capability. A button will activate the blade oscillation and the standard OR electrosurgical unit footswitch will power the monopolar effect. A nosecone will allow the tip to rotate 90° left or right. The blade will be provided at a 40° bend for adenoids with the capability of being reduced to 15° for tonsillectomies. The blade angle is flexible, and the design allows the blade to be bent between 0° and 50°,
The microdebrider device will include a monopolar cable that connects to a separate standard operating room electrosurgical generator. A standard suction tube will be attached to the suction port and a clip will attach the tubing to the cable. Two batteries, provided with a three year shelf life from the manufacturer, will be provided in the DTAD cable to power the blade oscillation. The battery leads connect internally to the PCB in the housing which connects to the oscillation switch and motor. When the handpiece button is pressed, power is sent to the motor which oscillates a gear which in turn oscillates the inner blade gear at approximately 3500 rpm. The monopolar lead from the cable is connected directly to the outer blade and is powered by the electrosurgical generator footswitch.
For user convenience the package will also include a disposable stylet for declogging if needed. The stylet is the same as the one provided in the current Diego® tubeset
{2}------------------------------------------------
Gyrus ACMI DTAD Gyrus ACMI, Inc. K121678
Traditional 510(k) Notification June 2012
Page 3 of 6
package (K034004). The device packaging is similar to the predicate Diego blades and disposables, and the devices are not labeled as non-pyrogenic.
# Material
The DTAD uses patient-contacting materials that are utilized in the predicate devices. as well as other legally marketed devices manufactured by Gyrus ACMI.
The composition of the DTAD is similar to the devices within the previously cleared predicate, Diego® RF Powered Dissector and Drill System, K034004. The handle and RF cable are not patient-contacting materials, and therefore no biocompatibility testing is warranted for those materials. The composition of the patient contacting shaver blade portion is similar to the devices cleared under K034004. The silicone material used as an RF insulator is the same material used in the disposable Gyrus ACMI Plasmacision devices tied to K041285. These devices are used in a similar manner, for the same procedures, manufactured in the same value stream, and are subjected to the same EtO sterilization cycle and process as the DTAD.
The DTAD is classified in accordance with ISO 10993-1, as an external communicating, tissue/bone/dentin device for limited exposure (<24hrs.). ISO10993-1 and FDA Blue Book memo #G95-1 guidelines recommend that these direct patient contact parts have supporting data for cytotoxicity, sensitization and irritation. Full biocompatibility testing (Cytotoxicity, sensitization, and irritation) to ISO10993-1 is available for the predicate devices containing the same patient contacting materials in similar quantities.
Since Gyrus ACMI does currently utilize the patient contacting materials in other devices with the same intended use no biocompatibility testing is required.
## Intended Use / Indications
The DTAD is intended for cutting, coagulation, and removal of tissue in general ENT procedures.
Specific indications include: Tonsillectomy and Adenoidectomy.
#### Compliance to Standards
The following standards were used during the design and validation of the DTAD:
IEC 60601-1: 2005 IEC 60601-2-2: 2009 ISO 10993-1: 2009 ISO 10993-7: 2008
12
{3}------------------------------------------------
Gvrus ACMI DTAD Gyrus ACMI, Inc.
K121678
Traditional 510(k) Notification June 2012
June 2012
Page 4 of 6
ISO 11135-1: 2007 ISO 11607-1: 2006 ISO 11607-2: 2006 ISO 15223-1: 2008 ASTM F88 - 09 ASTM F1886-98: 2004
The risk analysis was carried out in accordance with established in-house acceptance criteria based on ISO 14971:2007. The design verification tests were identified and performed as a result of risk analysis assessment.
## Summary of Sterilization and Shelf Life Discussion
The DTAD will be shipped sterile, for single use only. Each device is sterilized using Ethylene Oxide, using an over-kill cycle validated in accordance with ISO 11135 to provide a sterility assurance level of 106. The maximum allowable levels of residues of ethylene oxide, and ethylene chlorohydrin are established. The residual levels are less than those specified in ISO 10993-7.
The product will be released with a one year shelf life, which will later be extended to 3 years after functional accelerated age testing is complete. The Shelf Life period for the DTAD was determined through an analysis of the shelf-life stability of the materials used in the design and packaging of the DTAD, which are the same as those used in the predicate Diego® devices, as well as an analysis of the packaging materials and processes used with other Gyrus ACMI devices. Accelerated shelf-life functionality studies were conducted to support an initial one year shelf life, with real time testing in process to confirm an initial one year expiration date. Functional shelflife is also dependent upon the useful life of the batteries. Additional functionality studies (accelerated and real time) are in process to support a three-year expiration date in the future.
## Summary of Performance Testing
The following non-clinical, preclinical tests and usability studies were conducted:
#### Non-Clinical / Preclinical Performance
Evidence of safety and effectiveness was obtained from two primary areas:
- 1) non-clinical (electrical, mechanical, functional, stability) performance testing
- 2) preclinical (bench tissue, animal) evaluations and testing
Non-clinical: Basic safety and performance testing was performed in accordance with IEC 60601-1 and IEC 60601-2-2. In addition, verification and comparison bench studies were conducted to evaluate the mechanical and functional performance. Specifically, test results in the following areas were provided: functionality, RF
{4}------------------------------------------------
Gyrus ACMI DTAD Gyrus ACMI, Inc.
K121678
Traditional 510(k) Notification June 2012
Page 5 of 6
isolation, tip vibration, fluid ingress, battery and blade endurance, cable and connector pull force, monopolar effect, strain relief strength, suction and clog performance, blade flex capability, ability of the cutting edge to resist damage, blade torque, generator compatibility, and operational temperature.
Stability: Representative samples were subjected to accelerated and real time ageing to confirm that the device maintains functionality and continues to meet their specifications over time. The results of the accelerated age testing demonstrate that the device will be stable for the stated shelf-life. In addition, real time age testing will confirm the results of the accelerated age testing. Representative samples were also subjected to ship testing.
Preclinical: Evidence obtained from preclinical bench tissue (ex vivo) and animal (in vivo) studies demonstrate that the DTAD performs substantially equivalent to the predicate devices in relevant aspects associated with usability, cutting, coagulation, and removal of tissue.
Bench tissue - evaluated ex vivo using bovine tissue:
- Thermal margin -
- -Thermal impact
- Visual comparison of coagulation ・
Animal - evaluated in vivo using porcine models:
- Cut, coagulation, and suction -
- Ergonomics ー
- Usability aspects such as device setup, tip rotation, and tip malleability -
- -Overall design confidence
Testing demonstrated that the device is as safe, as effective, and performs as well as or better than the predicate devices.
No clinical testing was conducted. The use of Electrosurgical Cutting and Coagulation Devices / Microdebriders has been documented in published literature and indicates safe and effective use for the target procedures and patient populations.
## Substantial Equivalence to Predicates
The DTAD utilizes features incorporated into the following legally marketed predicate devices:
The DTAD utilizes high frequency electrosurgical energy to cut and coagulate tissue in a similar manner as the predicate devices.
With the exception of monopolar capability and the built-in battery powered motor, the DTAD cuts and removes tissue in the same manner as the predicate Diego® RF
{5}------------------------------------------------
Gyrus ACMI DTAD Gyrus ACMI, Inc. K121678 Traditional 510(k) Notification June 2012
Page 6 of 6
Dissector and Drill System. The DTAD blade offers monopolar coagulation at the cutting tip only when connected to a monopolar electrosurgical generator. The DTAD blade consists of a rotating inner member with a serrated tip within a stationary outer tube with a cutting window at the tip. The DTAD inner blade has a double helix laser cut that allows the inner blade to be flexible. The outer is spiral cut under an annealed aluminum tube and shrink tubing that allows the blade to be bent between 0° and 50°. The DTAD inner blade oscillates at a high rate of speed and as tissue is pulled into the cutting window it is resected and removed. The oscillating portion of the DTAD blade is driven by an electric motor inside the handle portion of the device that incorporates a gear transmission to provide the desired range of speeds and torque. The speed is limited by the motor within the handle. The hollow inner blade is attached to a suction supply to facilitate removal of tissue and fluid from the surgical site.
The Covidien / Valleylab suction monopolar coagulators are intended for general and ENT surgical procedures where the coagulation of tissue and suction of fluids are desired. The Bovie resection device is intended for cutting and monopolar coagulation of soft tissue using radiofrequency. The Peak Plasmablade TnA Tonsil and Adenoid Tissue Dissection device is intended to be used for the cutting and coagulation of soft tissue during otolaryngology (ENT) surgery including adenoidectomy and tonsillectomy.
#### Conclusion:
The cutting and coagulation performance of the DTAD was compared against performance requirements and the predicate systems listed above. Testing demonstrated that the performance requirements were met, and that the Gyrus ACMI DTAD exhibited comparable performance characteristics to the predicates.
In summary, the Gyrus ACMI DTAD is substantially equivalent to the predicate devices and presents no new questions of safety or efficacy.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Gyrus ACMI, Incorporated % Mr. Dolan Mills, RAC Specialist, Regulatory Affairs 136 Turnpike Road Southborough, Massachusetts 01772
AUG 300 2012
Re: K121678
Trade/Device Name: DTAD Regulation Number: 21 CFR 874.4250 Regulation Name: Ear, nose, and throat electric or pneumatic surgical drill Regulatory Class: II Product Code: ERL Dated: August 17, 2012 Received: August 20, 2012
Dear Mr. Mills:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{7}------------------------------------------------
## Page 2- Mr. Dolan Mills, RAC
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Erind Keith
Image /page/7/Picture/7 description: The image shows a stylized letter 'f' on the left, followed by the letters 'N' and 'R' on the right. The 'f' is written in a cursive style with a loop at the top and a curved tail. The letters 'N' and 'R' are in a simple, sans-serif font and are positioned close to the 'f'.
Mark Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
Traditional 510(k) Notification
June 2012
Page 1 of 1
# Indications for Use
510(k) Number: K121678
Device Name: DTAD
# Indications for Use:
. :
The DTAD is intended for cutting, coagulation, and removal of tissue in general ENT procedures.
Specific indications include: Tonsillectomy and Adenoidectomy.
Prescription Use: X (Per 21 CFR 801.109)
AND/OR
Over-the-Counter Use: (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
E.C. Keith
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K121678
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.