K121676 · Ge Medical Systems, LLC · LNH · Sep 20, 2012 · Radiology
Device Facts
Record ID
K121676
Device Name
1.5T SIGNA HDXT 3.0T SIGNA HDXT SIGNA VIBRANT
Applicant
Ge Medical Systems, LLC
Product Code
LNH · Radiology
Decision Date
Sep 20, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The 1.5T Signa HDx family and 3.0T Signa HDx family are a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used. The images produced by the 1.5T Signa HDx family and 3.0T Signa HDx family reflects the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
Device Story
Whole body MRI system (1.5T and 3.0T field strengths) for diagnostic imaging. Inputs: time-varying magnetic fields (gradients) and RF transmissions to detect spatial distribution/molecular environment of nuclei. Outputs: high-resolution axial, sagittal, coronal, oblique images, spectroscopic data, parametric maps, and dynamic images. Used in clinical settings by trained physicians/technicians. System supports up to 32 independent receive channels. Output interpreted by physicians to assist in clinical diagnosis. Benefits: high signal-to-noise ratio, short scan times, and comprehensive anatomical/functional visualization.
Clinical Evidence
No clinical studies required. Evidence based on bench testing, module verification, system integration testing, performance verification, safety testing, and simulated use validation. Sample clinical images included in submission.
Technological Characteristics
Whole body MRI system; 1.5T and 3.0T field strengths; up to 32 independent receive channels. Complies with IEC60601-1, IEC60601-2-33, IEC60601-1-1, IEC60601-1-2, IEC60601-1-4, IEC60601-1-6, ISO14971, ISO10993-1, and IEC62304.
Indications for Use
Indicated for diagnostic imaging of the entire body (head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, musculoskeletal) in patients requiring axial, sagittal, coronal, oblique images, spectroscopic data, parametric maps, or dynamic images. Contrast agents may be used.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows handwritten text. The top line reads "K121676". The bottom line reads "Page 1 of 3". The text appears to be a page number or identifier within a larger document.
SEP 20 2012
## 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Date: | 25-May-2012 |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | GE Healthcare. (GE Medical Systems LLC)<br>3200 N Grandview Blvd.<br>Waukesha, WI 53188 USA |
| Primary Contact Person: | Shashidhar C S<br>Regulatory Affairs Leader - MR<br>GE Healthcare,<br>Phone: +91-80-40883613 |
| Secondary Contact Person: | Glen Sabin<br>Regulatory Affairs Director - MR<br>GE Healthcare,<br>Phone: (262) 521-6848<br>Fax: (262) 364-2785 |
| Device /<br>Trade Name: | 1.5T Signa HDx family and 3.0T Signa HDx family |
| Common /Usual Name: | Magnetic Resonance Imaging System |
| Classification Names: | Magnetic resonance diagnostic device |
| Product Code: | LNH |
| Predicate Device(s): | 1.5T and 3.0T Signa HDx MR System (K052293) |
| | Discovery MR750w 3.0T (K103327) |
| Device Description: | The 1.5T Signa HDx family and 3.0T Signa HDx family<br>systems are a whole body magnetic resonance system<br>designed to support high resolution, high signal-to-noise<br>ratio, and short scan times. The Signa HDx family of<br>scanners is available in two different field strengths of<br>1.5T and 3.0T. The system uses a combination of time-<br>varying magnetic fields (gradients) and RF transmissions<br>to obtain information regarding the density and position of<br>elements exhibiting magnetic resonance. The data<br>acquisition system accommodates up to 32 independent<br>receive channels in various increments and multiple |
{1}------------------------------------------------
K121676
page 2 of 3
independent coil elements per channel during a single acquisition series. The System can image axial, sagittal, coronal, and oblique anatomical images, spectroscopic data, parametric maps, or dynamic images of the structures or functions of the entire body.
- Intended Use: The 1.5T Signa HDx family and 3.0T Signa HDx family are a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to. head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used. The images produced by the 1.5T Signa HDx family and 3.0T Signa HDx family reflects the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
- Technology: The 1.5T Signa HDx family and 3.0T Signa HDx family employs the same fundamental scientific technology as its predicate devices.
Determination of Substantial Equivalence:
## Summary of Non-Clinical Tests:
The 1.5T Signa HDx family and 3.0T Signa HDx family and its applications comply with voluntary standards, including IEC60601-1, IEC60601-2-33, IEC60601-1-1, IEC60601-1-2, IEC60601-1-4, IEC60601-1-6, ISO14971, ISO10993-1 and IEC62304.
The following quality assurance measures were applied to the development of the system:
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the text "K121676 Page 3 of 3". The text appears to be handwritten. The phrase "Page 3 of 3" indicates that this is the last page of a document.
- . Risk Analysis
- Requirements Reviews .
- Design Reviews .
- Testing on unit level (Module verification) .
- Integration testing (System verification) .
- Performance testing (Verification) .
- . Safety testing (Verification)
- Simulated use testing (Validation) .
## Summary of Clinical Tests:
The subject of this premarket submission, 1.5T Signa HDx family and 3.0T Signa HDx family did not require clinical studies to support substantial external equivalence. Internal scans were conducted as part of validation for workflow and image quality and that sample clinical images are included in the submission.
Conclusion:
GE Healthcare considers the 1.5T Signa HDx family and 3.0T Signa HDx family to be as safe; as effective, and performance is substantially equivalent to the predicate device(s).
Image /page/2/Picture/14 description: The image shows the words "Section 5" in a simple, sans-serif font. The word "Section" is capitalized, and the number "5" is written out as a digit. The text is black against a white background.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
SEP 20 2012
Mr. C.S. Shashidhar Regulatory Affairs Leader GE Medical Systems LLC 3200 N. Grandview Blvd WAUKESHA WI 53188
Re: K121676
Trade/Device Name: 1.5T Signa HDx family and 3.0T Signa HDx family Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH, MOS, and LNI Dated: September 7, 2012 Received: September 11, 2012
Dear Mr. Shashidhar:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your Scellon 910(t) promised is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the Citilosure) to regarry management date of the Medical Device American County commerce phor to May 26, 1976, the encordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The You may, therefore, market the device, sourcements for annual registration, listing of general controls provisions of the rist labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such II your device is Classified (see above) into in affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please oe auvised that I Dri s issuance of a read on a complies with other requirements of the Act
that FDA has made a determination that your device complies with other requ that IDA has made a dolorimmentin as administered by other Federal agencies. You must or any Federal Statutes and regulations and limited to: registration and listing (21)
comply with all the Act's requirements, including, but not limited to: researching of comply with an the Act 3 requirements, and 809); medical device reporting (reporting of
{4}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice medical device related adverse ovality systems (QS) regulation (21 CFR Part 820). This letter requirements as set form in the quality -Jevice as described in your Section 510(k) premarket with anow you to begal mading of your device of your device to a legally marketed noticate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 601) 70 11 you desire specific ad rice for your estic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to under the MDK regulation (27 Of RP Latv 900), Portal Problem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Tou thay obtain of Scherers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### K121676 510(k) Number (if known):
Device Name: 1.5T Signa HDx family and 3.0T Signa HDx family
## Indications for Use:
The 1.5T Signa HDx family and 3.0T Signa HDx family are a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used.
The images produced by the 1.5T Signa HDx family and 3.0T Signa HDx family reflects the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
Prescription Use X _ (Part 21 CFR 801 Subpart D) Over-The-Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
AND/OR
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Mashad D. Othman
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) Kia1676
Section 4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.