The Gemini is indicated for the fragmentation of urinary tract stones, i.e., renal calyceal stones, renal pelvic stones, and upper ureteral stones.
Device Story
Gemini is a modular urological workstation for extracorporeal shock wave lithotripsy (ESWL) and diagnostic/therapeutic urological procedures. System modules include a basic unit with integrated X-ray C-arm and therapy arm, patient table, control desk, and ultrasound unit. Device uses an electromagnetic shock wave emitter (EMSE) to fragment stones. Clinicians operate the system via the control desk, utilizing integrated X-ray and ultrasound for stone localization. The system aligns the shock wave focus area with the stone to perform fragmentation. Modifications from predicate devices include updated patient table specifications (weight limit, travel, speed), optional larger X-ray image intensifier, alternate X-ray generator, alternate ultrasound unit, and alternate image storage. Benefits include non-invasive stone fragmentation. Used in clinical settings by urologists.
Clinical Evidence
Bench testing only. Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-1-6, IEC 60601-2-7, IEC 60601-2-28, IEC 60601-2-36, IEC 60601-2-32, ISO 13485, and IEC 61846. Design verification confirmed anti-collision system performance and accuracy of stone localization at the shock wave focus area.
Technological Characteristics
Modular urological workstation. Electromagnetic shock wave emitter (EMSE) 140f. Integrated X-ray C-arm, patient table, control desk, and ultrasound unit. Standards: IEC 60601 series (safety, EMC, radiation, usability), ISO 13485 (quality management), IEC 61846 (ultrasonics). Connectivity includes image storage system. Software-controlled for system operation and stone localization.
Indications for Use
Indicated for fragmentation of urinary tract stones, including renal calyceal, renal pelvic, and upper ureteral stones in adult patients.
Regulatory Classification
Identification
An extracorporeal shock wave lithotripter is a device that focuses ultrasonic shock waves into the body to noninvasively fragment urinary calculi within the kidney or ureter. The primary components of the device are a shock wave generator, high voltage generator, control console, imaging/localization system, and patient table. Prior to treatment, the urinary stone is targeted using either an integral or stand-alone localization/imaging system. Shock waves are typically generated using electrostatic spark discharge (spark gap), electromagnetically repelled membranes, or piezoelectric crystal arrays, and focused onto the stone with either a specially designed reflector, dish, or acoustic lens. The shock waves are created under water within the shock wave generator, and are transferred to the patient's body using an appropriate acoustic interface. After the stone has been fragmented by the focused shock waves, the fragments pass out of the body with the patient's urine.
Special Controls
*Classification.* Class II (special controls) (FDA guidance document: “Guidance for the Content of Premarket Notifications (510(k)'s) for Extracorporeal Shock Wave Lithotripters Indicated for the Fragmentation of Kidney and Ureteral Calculi.”)
{0}------------------------------------------------
2121656 PE. I DF
# 510(k) SUMMARY
#### Gemini Lithotripter
SEP 24 2012
# Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Dornier MedTech America 1155 Roberts Blvd. Kennesaw, GA 30144 Phone: 770-514-6163 Fax: 770-514-6291
Contact Person: John Hoffer
Date Prepared: September 4, 2012
#### Device Trade Name
Gemini Lithotripter
# Classification Name
Extracorporeal Shock Wave Lithotripters (21 C.F.R. § 876.5990)
### Predicate Devices
Dornier Lithotripter (Doli) {P840008, S062)
Dornier Lithotripter 140 (Doli 140) (K011773)
#### Intended Use / Indications for Use
The Gemini is indicated for the fragmentation of urinary tract stones, i.e., renal calyceal stones, renal pelvic stones, and upper ureteral stones.
#### Purpose of the Special 510(k) Notice
The Gemini is a modification to Dornier's Lithotripter (P840008, S062) and the Doli 140 (K011773). Specifically, the Gemini is the same device as the cleared predicate Doli products with the exception of the modifications outlined below:
- Minor changes to the patient table, including weight limit, travel range and speed; o
- Minor changes to the X-ray system to include an optional larger Image Intensifier and 0 use of an alternate X-ray generator;
- O Use of an alternate ultrasound unit; and
- Use of an alternate image storage system. O
# Technological Characteristics/Principles of Operation
The Gemini is a modular urological work station designed for extracorporeal shock wave lithotripsy ("ESWL") and for diagnostic and therapeutic procedures usual in Urology. The Gemini is composed of the following modules: (1) basic unit with integrated X-ray C-arm and
{1}------------------------------------------------
therapy arm for shockwave treatment; (2) patient table; (3) control desk – user interface; and (4) ultrasound unit.
As indicate above, the Gemini is an extracorporeal shock wave lithotripter used for fragmentation of urinary tract stones, i.e., renal calyceal stones, renal pelvic stones, and upper ureteral stones. The technology to perform this function involves use of an electromagnetic shock wave emitter ("EMSE"). In the case of the subject device and the predicate devices, the identical shock wave source is used. This source, the 140f EMSE, has a well-established record of performance and also is utilized in other Lithotripers that have been cleared by FDA (e.g., Dornier Compact Alpha (K002929) and Dornier Compact Delta (P840008, S066)).
The other elements of the Gemini, i.e., the patient table and the X-ray unit, are also are similar to that of other cleared stationary Lithotripters, the Doli 140 (K011773) and the Dornier Lithotriper (P840008, S062). They perform the same function and operate in the same manner during the procedures involved in the fragmenting of urological stones.
#### Performance Data
The company has complied with all of the requirements described in FDA's Guidance for the Content of Premarket Notifications (510k's) for Extracorporeal Shock Wave Lithotrioters Indicated for the Fragmentation of Kidney and Ureteral Calculi.
Device testing was performed to confirm compliance with the following standards:
| o IEC 60601-1:2007 | Electrical safety of medical devices |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------|
| o IEC60601-1-2:2007 | Electromagnetic compatibility |
| o IEC 60601-1-3:2008 | Radiation protection |
| o IEC 60601-1-6:2008 | Usability |
| o IEC 60601-2-7 | Safety of high-voltage generators of diagnostic X-ray<br>generator |
| o IEC 60601-2-28 | Particular requirements for the safety of X-ray<br>source assemblies and X-ray tube assemblies for<br>medical diagnosis |
| o IEC 60601-2-36:1997 | Extracorporeally induced Lithotripsy |
| o IEC 60601-2-32:1995 | Safety of X-ray equipment |
| o ISO 13485:2003+AC:2007 | Quality management system |
| o IEC 61846 | Ultrasonics - Pressure pulse lithotripters<br>characteristic of fields |
In all instances, the Gemini Lithotripter functioned as intended and results observed were as expected.
In addition, the system and its software was tested and validated during the company's design control process and the system met all acceptance criteria. The anti-collision system of the subject device was verified to perform as designed to mitigate the risk of injury. The methodology for accurately locating the stone was tested during the design verification testing and the assessment demonstrated that the system is capable of locating the shock wave focus area with sufficient accuracy. Lastly, the device manual was reviewed and approved as part of the design control process. It contains all necessary warnings, cautions and instructions to mitigate potential injuries.
{2}------------------------------------------------
K121656 PG.3 VF- 3
# Substantial Equivalence
The Gemini Lithotripter is as safe and effective as the identified predicate devices listed above. The Gemini Lithotripter has the same intended uses / indications for use, technological characteristics, and principles of operation as its predicate devices. Performance data demonstrate that the Gemini Lithotripter is as safe and effective as the predicate devices. Thus, the Gemini Lithotripter is substantially equivalent.
{3}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image is a circular seal for the Department of Health & Human Services - USA. The seal features the department's name in a circular arrangement around the edge. In the center of the seal is a stylized image of an eagle or other bird-like figure, with its wings spread and head turned to the side. The design is simple and monochromatic.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
SEP 24 2012
Mr. John Hoffer VP Quality, Regulatory, Clinical Dornier MedTech America 1155 Roberts Blvd. KENNESAW GA 30144
Re: K121656
Trade/Device Name: Gemini Regulation Number: 21 CFR§ 876.5990 Regulation Name: Extracorporeal shock wave lithotripter Regulatory Class: II Product Code: LNS Dated: September 4, 2012 Received: September 5, 2012
Dear Mr. Hoffer:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becaon 3.10(t) premier on substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the enclosure) to regard management date of the Medical Device Amendments, or to commerce prior to May 26, 1776, the enaounters with the provisions of the Federal Food, Drug, devices that have been reclassinou in access approval of a premarket approval application (PMA). and Cosmelle Act (Act) that do not require apple of the general controls provisions of the Act. The You may, dielefore, market the device, basics to the ments for annual registration, listing of general controls provisions of the rict labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability additeration. Thease note: ODICT acce how care labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), If your device is classified (SCC above) into othing major regulations affecting your device can be
it may be subject to additional controls. Existing major regulations affe It may be subject to additional conseiler. This 21, Parts 800 to 898. In addition, FDA may found in the Code of I cacal resguing your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA s issuation of a backer with other requirements of the Act
that FDA has made a determination that your device complies with of the Act that FDA has made a decemination that your accessed by other Federal agencies. You must of ally Federal statutes and roganatelle , including, but not limited to: registration and listing comply with an the Act 3 requirements, mondaing, Frinately, Frances
(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
Page 2-
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set device-related adverse events) (21 CFR Part 820); and if applicable, the electronic forul in the quality systems (QD) rogalications (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please IT you desire specific advice for your do need on the corress.
go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/com/15809.html/sear go to million.rwww.ida.gov/rtloudio.org/w/index.php?tice of Compliance. Also, please the Concertion Borroos and Ramorogious of reference to premarket notification" (21CFR Patt note the regulation onlined, "informating of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the I ou thay of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Sincerely yours,
Benjamin K. Twitchell
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use Statement
K121656 510(k) Number (if known):
Device Name: Gemini
Indications for Use:
The Gemini is indicated for the fragmentation of urinary tract stones, i.e. renal calyceal stones, renal pelvic stones, and upper ureteral stones.
Prescription Use _ × (Per 21 C.F.R. 801.109) AND/OR
Over-The-Counter Use_ (Per 21 C.F.R. 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
```
(Division Sign-Off)
Division of Reproductive, Gastro-Renal, and
Urological Devices
510(k) Number K121656
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.