These devices are implantable bone plates and bone screws for neuro procedures including: 1. Fractures 2. Osteotomies 3. Reconstructive procedures 4. Revision procedures where other treatments or devices have failed
Device Story
Biomet Microfixation Neuro Plating System consists of titanium plates and screws for cranial skeletal alignment and stabilization. Plates feature various geometries (straight, angle, mesh, etc.) and screws range from 1.0mm to 2.3mm diameter. Used by surgeons in clinical settings for neurosurgical procedures. Devices are provided non-sterile and require steam sterilization by the user. Implants provide mechanical fixation to support bone healing; failure to follow postoperative instructions or presence of contraindications may lead to complications like nonunion, migration, or infection.
Clinical Evidence
Bench testing only. Testing included validation of cleaning protocols and steam sterilization efficacy per AAMI TIR 12 and ANSI/AAMI/ISO 17665-1:2006. Worst-case sample components demonstrated effective sterilization achieving a 10^-6 Sterility Assurance Level (SAL).
Technological Characteristics
Titanium implantable bone plates and screws. Various geometries including straight, angle, mesh, and matrix. Screw diameters 1.0mm-2.3mm; lengths 3.0mm-6.0mm. Provided non-sterile; steam sterilization required. Mechanical fixation principle.
Indications for Use
Indicated for patients requiring cranial skeletal alignment and stabilization during neuro procedures, including fractures, osteotomies, reconstructive procedures, and revisions. Contraindicated in patients with active infection, foreign body sensitivity, limited blood supply, insufficient bone quantity, latent infection, or those unable to follow postoperative care.
Regulatory Classification
Identification
A preformed alterable cranioplasty plate is a device that is implanted into a patient to repair a skull defect. It is constructed of a material, e.g., tantalum, that can be altered or reshaped at the time of surgery without changing the chemical behavior of the material.
Predicate Devices
Wuerzburg Titanium Mini Bone Plates and Bone Screws (K854886)
Lorenz 1.0mm, 1.5mm, 2.0mm Plating system (K953385)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Biomet Microfixation. The logo is in black and white and features the word "BIOMET" in large, bold letters on top. Below that, the word "MICROFIXATION" is written in smaller, thinner letters. The logo appears to be for a medical or surgical company.
#### 510(k) Summary
510(k) Number: Date Prepared: Submitter:
K121624 October 2, 2012 Biomet Microfixation 1520 Tradeport Drive Jacksonville, FL 32218-2480
Contact:
#### Sheryl Malmberg, Global Regulatory Affairs Specialist 904-741-4400 fax 904-741-9425
| Common or Usual Name: | Neuro Plating System | | | | Biomet Microfixation<br>Neuro Plating System | Walter Lorenz<br>Surgical K854886 | Walter Lorenz<br>Surgical K922741 | Walter Lorenz<br>Surgical K953385 |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification Name: | Preformed Alterable<br>Cranioplasty Plate | Burr Hole Cover | Cranioplasty Plate<br>Fastener | Material | Titanium | Titanium | Titanium | Titanium |
| Device Classification: | II (882.5320) | II (882.5250) | HBW (882.5360) | Indications<br>for Use | These devices are<br>implantable bone plates<br>and bone screws for<br>cranial procedures<br>including:<br>1. Fractures<br>2. Osteotomies<br>3. Reconstructive<br>procedures<br>4. Revision procedures<br>where other treatments<br>or devices have failed. | These devices are<br>implantable bone<br>plates and bone<br>screws for oral,<br>cranio-maxillofacial<br>procedures including:<br>1. Fractures, 2.<br>Osteotomies,<br>including<br>orthognathic<br>procedures, 3.<br>Reconstructive<br>procedures, and 4.<br>Revision procedures<br>where other<br>treatments or devices<br>have failed. | These devices are<br>implantable bone<br>plates and bone<br>screws for oral,<br>cranio-maxillofacial<br>procedures including:<br>1. Fractures, 2.<br>Osteotomies,<br>including<br>orthognathic<br>procedures, 3.<br>Reconstructive<br>procedures, and 4.<br>Revision procedures<br>where other<br>treatments or devices<br>have failed. | These devices are<br>implantable bone<br>plates and bone<br>screws for oral,<br>cranio-maxillofacial<br>procedures including:<br>1. Fractures, 2.<br>Osteotomies,<br>including<br>orthognathic<br>procedures, 3..<br>Reconstructive<br>procedures, and 4.<br>Revision procedures<br>where other<br>treatments or devices<br>have failed. |
| Device Product Code: | GWO | GXR | HBW | Use | Single | Single | Single | Single |
| Sterility | Provided non-sterile;<br>steam sterilization<br>validation on file | Provided non-sterile;<br>steam sterilization<br>validation on file | Provided non-sterile;<br>steam sterilization<br>validation on file | Provided non-sterile;<br>steam sterilization<br>validation on file | | | | |
| Shape of<br>Plates | New shape includes pre-<br>contoured mesh.<br>Additional system<br>plates include variations<br>of straight, curved,<br>angle, double angle, L-,<br>T-, double T-, Z-, X-,<br>Y-, double Y-shape,<br>burr hole, triangle,<br>rectangle, square,<br>matrix, and mesh<br>options with various<br>lengths and thicknesses. | Variations of straight,<br>curved, L-, T-, Y-<br>shape, matrix, plate<br>options. | Variations of straight,<br>square, rectangular,<br>and mesh plate<br>options | Variations of straight,<br>angle, curved, L-, T-,<br>double T-, Z-, X-, Y-,<br>double Y-, H-shape,<br>triangle, square,<br>rectangle, matrix,<br>mesh, orbital floor,<br>and LeFort plate<br>options. | | | | |
| Screw Sizes | No new screws in this<br>submission. Additional<br>system screws range in<br>diameters of 1.0mm to<br>2.3 and lengths from<br>2.0mm to 7.0mm. Head<br>geometry includes<br>cross-drive and center<br>drive, with HT (high<br>torque) and SD (self-<br>drilling) options. | Screws range in<br>diameters of 1.0mm<br>to 2.3mm and lengths<br>from 2.0mm to<br>29.0mm<br>Head geometry<br>includes cross-drive<br>and center drive, with<br>HT (high torque) and<br>SD (self-drilling)<br>options | Not Applicable | Not Applicable | | | | |
Device Name: Biomet Microfixation Neuro Plating System
#### Intended Use:
These devices are implantable bone plates and bone screws for neuro procedures including:
- Fractures l .
- 2. Osteotomies
- 3. Reconstructive procedures
- Revision procedures where other treatments or devices have failed 4.
#### Contraindications:
- Active infection 1.
- 2. Foreign body sensitivity. Where material sensitivity is to be completed prior to implantation.
- 3. Patients with limited blood supply, insufficient quantity of bone, or latent infection.
- 4. Patients with mental or neurologic conditions who are unwilling or incapable of following postoperative care instructions.
Description: Biomet Microfixation manufactures and distributes a variety of internal fixation devices intended to aid in the alignment and stabilization of bone in the cranial skeletal system. Instrumentation has been designed specifically for use with each system of implants.
The Biomet Microfixation Cranio/Neuro Plaing System is comprised of a variety of titanium plates and screws with shapes and sizes intended to aid in the alignment and stabilization of bone in the cranial skeletal system.
The plates include variations of straight, angle, double angle, L-shape, Z-shape, X-shape, burr hole, triangle, square, matrix, mesh, and crescent options with various lengths and thickness. The screws range in diameters of 1.0mm to 2.3mm and lengths from 3.0mm to 6.0mm.
Sterility Information: The plates and screws will be marketed as non-sterile, single use devices.
#### Possible risks:
- Poor bone formation, Osteolysis, Osteolysis, Osteomyelitis, inhibited revascularization, or infection can 1. cause loosening, bending, cracking or fracture of the device.
- 2. Nonunion or delayed union, which may lead to breakage of the implant.
- 3. Migration, bending, fracture or loosening of the implant.
- Metal sensitivity, or allergic reaction to a foreign body. 4.
- న్. Decrease in bone density due to stress shielding.
- 6. Pain, discomfort, abnormal sensation, or palpability due to the presence of the device.
- 7. Increased fibrous tissue response around the fracture site and/or the implant.
- 8. Necrosis of bone.
- 9. Inadequate healing.
OCT 1 2 2012
{1}------------------------------------------------
1421624
Substantial Equivalence Biomet Microfixation considers the Neuro Plating System modifications equivalent to the Biomet Microfixation (formally Walter Lorenz Surgical) Wuerzburg Titanium Mini Bone Plates and Bone Screws cleared under K854886, Lorenz 1.0 MM System cleared under K922741, and Lorenz 1.0mm, 1.5mm, 2.0mm Plating system cleared under K953385. Subject pre-contoured mesh are similar to the predicate devices in terms of material, indications, use, sterility, and design.
{2}------------------------------------------------
Nonclinical testing: Referenced testing demonstrates that the device can be cleaned and sterilized within the required parameters, demonstrating equivalence to the predicate devices. A worst case scenario of instrumentation was compromised and underwent testing resulting in achieving the required cleaning results as outlined in the testing. The cleaning testing was performed per an FDA approved protocol. Sterility testing, (Reference AAMI TIR 12 and ANSI/AAMI/ISO 17665-1:2006) provided worst case testing by inoculating worst case sample components that then demonstrated results to have been effectively sterilized per required parameters.
Conclusion: This testing data supports the determination of substantial equivalence by demonstrating worst case scenarios for both cleaning and sterilization that fall within acceptable parameters providing a 10 * Sterility Assurance Level (SAL). No new safety and efficacy issues were raised.
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three wavy lines representing the feathers. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
OCT 12 2012
Biomet Microfixation % Ms. Sheryl Malmberg Global Regulatory Affairs Specialist 1520 Tradeport Drive Jacksonville, FL 32218
Re: K121624
Trade/Device Name: Biomet Microfixation Neuro Plating System Regulation Number: 21 CFR 882.5320 Regulation Name: Preformed alterable cranioplasty plate Regulatory Class: Class II Product Code: GWO, GXR, HBW Dated: September 11, 2012 Received: September 12, 2012
Dear Ms. Malmberg:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
## Page 2 - Ms. Sheryl Malmberg
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Kesia Alexander
Image /page/4/Picture/6 description: The image shows a handwritten word, "for", in cursive script. The "f" is elongated with a large loop extending below the baseline, and the "o" and "r" are connected with a smooth, flowing line. The writing appears to be done with a pen or marker, and the overall style is elegant and fluid.
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
· Enclosure
{5}------------------------------------------------
# Indications for Use
K121624 510(k) Number (if known):
Device Name: Biomet Microfixation Neuro Plating System
### Indications For Use:
These devices are implantable bone plates and bone screws for neuro procedures including:
- 1. Fractures
- 2. Osteotomies
- 3. Reconstructive procedures
- 4. Revision procedures where other treatments or devices have failed
AND/OR Prescription Use XX (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number K121624
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.