K121607 · Terumo (Philippines) Corporation · FMI · Nov 8, 2012 · General Hospital
Device Facts
Record ID
K121607
Device Name
TERUMO(R) NEEDLE
Applicant
Terumo (Philippines) Corporation
Product Code
FMI · General Hospital
Decision Date
Nov 8, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The TERUMO® Needle is a hypodermic single lumen needle intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin.
Device Story
Sterile, single-lumen hypodermic needle; stainless steel tube sharpened at one end; attached to female luer connector (hub) at other end. Designed for connection to male nozzle of piston syringe or intravascular administration set. Operated manually by clinicians to inject or withdraw fluids. Biocompatible materials; limited duration of contact (≤24hr).
Clinical Evidence
Bench testing only. Performance testing included protector fit, adhesive hold, conical fitting, and seal strength. Biocompatibility testing performed per ISO 10993.
Technological Characteristics
Stainless steel needle; polypropylene hub; adhesive attachment. Manual operation. Various gauges (18G-30G) and lengths (1/2"-1 1/2"). Sterilized via radiation per ANSI/AAMI/ISO 11137 (SAL 10^-5).
Indications for Use
Indicated for injection of fluids into or withdrawal of fluids from parts of the body below the skin surface. No specific patient population, age, or gender restrictions stated.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
# 510(k) SUMMARY
1121607
NOV
8 2012
: TERUMO (PHILIPPINES) CORPORATION Prepared for 124 East Main Avenue, Laguna Technopark, Binan, Laguna,Philippines
Prepared by
: Ma. Cristina Faderagao - Supervisor, Regulatory Affairs Tel. (+63) 49 541 2111 Fax (+63) 49 541 2121
Date prepared : February 10, 2012
Contact Person : Sandi Hartka
Regulatory Affairs Manager Terumo Medical Corporation 950 Elkton Blvd Elkton, MD 21921 Phone#: (410) 392-7243 Fax#: (410) 398-6079 E-mail: sandi.hartka@terumomedical.com
#### A. DEVICE NAME
Proprietary Name TERUMO® NEEDLE
Classification Name Hypoderinic Single Lumen Needle (880.5570)
Product Code: FMI
Classification: Class II
Common Name: Hypodermic Needle
{1}------------------------------------------------
# B. INTENDED USE
The TERUMO® Needle is a hypodermic single lumen needle intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin.
# C. DEVICE DESCRIPTION
The TERUMO® NEEDLE is a hypodermic sterile single lumen needle, consisting of a stainless tube that is sharpened at one end and at the other end joined to a female luer connector (hub) designed to be connected with a male connector (nozzle) of a piston syringe or an intravascular administration set.
| Product<br>Code | Hub<br>Color | Needle Ø<br>(Gauge and<br>mm) | Needle<br>Length<br>(inch - mm) | Wall<br>thickness | Needle<br>Bevel |
|-----------------|--------------|-------------------------------|---------------------------------|--------------------|------------------|
| NN-1825R | Pink | 18G - 1.20mm | 1" – 25mm | Thin Wall | Regular<br>Bevel |
| NN-1838R | Pink | 18G - 1.20mm | 1 ½" – 38mm | Thin Wall | Regular<br>Bevel |
| NN-1925R | Cream | 19G - 1.10mm | 1" – 25mm | Thin Wall | Regular<br>Bevel |
| NN-1938R | Cream | 19G - 1.10mm | 1 ½" – 38mm | Thin Wall | Regular<br>Bevel |
| NN-2025R | Yellow | 20G - 0.90mm | 1" – 25mm | Ultra Thin<br>Wall | Regular<br>Bevel |
| NN-2025V | Yellow | 20G - 0.90mm | 1" – 25mm | Regular<br>Wall | Regular<br>Bevel |
| NN-2038R | Yellow | 20G - 0.90mm | 1 ½" – 38mm | Ultra Thin<br>Wall | Regular<br>Bevel |
| NN-2125R | Green | 21G - 0.80mm | 1" – 25mm | Ultra Thin<br>Wall | Regular<br>Bevel |
| NN-2138R | Green | 21G - 0.80mm | 1 ½" – 38mm | Ultra Thin<br>Wall | Regular<br>Bevel |
| NN-2225R | Black | 22G - 0.70mm | 1" – 25mm | Ultra Thin<br>Wall | Regular<br>Bevel |
| NN-2238R | Black | 22G - 0.65mm | 1 ½" – 38mm | Ultra Thin<br>Wall | Regular<br>Bevel |
00013
{2}------------------------------------------------
| Product<br>Code | Hub<br>Color | Needle Ø<br>(Gauge and<br>mm) | Needle<br>Length<br>(inch - mm) | Wall<br>thickness | Needle<br>Bevel |
|-----------------|--------------|-------------------------------|---------------------------------|--------------------|------------------|
| NN-2325R | Blue | 23G - 0.65mm | 1" - 25mm | Ultra Thin<br>Wall | Regular<br>Bevel |
| NN-2338R | Blue | 23G - 0.65mm | 1 ½" - 38mm | Ultra Thin<br>Wall | Regular<br>Bevel |
| NN-2516R | Orange | 25G - 0.50mm | 5/8" - 16mm | Ultra Thin<br>Wall | Regular<br>Bevel |
| NN-2522R | Orange | 25G-0.50mm | 7/8" - 22mm | Ultra Thin<br>Wall | Regular<br>Bevel |
| NN-2525R | Orange | 25G - 0.50mm | 1" - 25mm | Ultra Thin<br>Wall | Regular<br>Bevel |
| NN-2538R | Orange | 25G - 0.50mm | 1 ½" - 38mm | Regular<br>Wall | Regular<br>Bevel |
| NN-2613R | Brown | 26G - 0.45mm | ½" - 13mm | Regular<br>Wall | Regular<br>Bevel |
| NN-2713R | Gray | 27G - 0.40mm | ½" - 13mm | Regular<br>Wall | Regular<br>Bevel |
| NN-2732R | Gray | 27G - 0.40mm | 1 ¼" - 32mm | Regular<br>Wall | Regular<br>Bevel |
| NN-3025R | Yellow | 30G -0.30mm | 1" - 25mm | Regular<br>Wall | Regular<br>Bevel |
| NN-3013R | Yellow | 30G-0.30mm | ½" - 13mm | Regular<br>Wall | Regular<br>Bevel |
# D. SUBSTANTIAL EQUIVALENCE
The TERUMO® NEEDLE manufactured by Terumo Philippines Corporation is substantially equivalent in its intended use, technology/principal of operation, materials, and performance to the following cleared devices:
- 1. K001572 TERUMO® NEOLUS® Necdle manufactured by Terumo Europe, Leuven, Belgium
- 2. K771203 TERUMO® Hypodermic Needle manufactured by Tcrumo Medical Corporation, Elkton, Maryland
- K012646 (30G) TERUMO® Needle manufactured by Terumo Medical Corp, Elkton, MD mi
{3}------------------------------------------------
### Intended Use
The TERUMO® Needle manufactured from Terumo (Philippines) Corporation and predicate devices cleared under K001572, K771203 and K012646 (30G) are hypodermic single lumen needle intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin.
# Principles of Operation / Technology
The TERUMO® Needle manufactured by Terumo (Philippines) Corporation and the predicate devices cleared under K001572 manufactured by Terumo Europe N.V. Belgium, and K771573 and K012646 (30G) manufactured by Terumo Medical Corporation, USA are all operated manually.
## Design and Materials
The TERUMO® Needle manufactured by Terumo (Philippines) Corporation and the predicate devices cleared under K001572 manufactured by Terumo Europe N.V. Belgium, and K771573 and K012646 (30G) manufactured by Terumo Medical Corporation, USA are made up of stainless steel needle that is sharpened at one end and at the other end was attached to a plastic polypropylene hub by means of adhesive.
### Performance
The following performance tests were performed on the TERUMO® Needle manufactured by Terumo (Philippines) Corporation:
- . Protector Pit
- Adhesive Hold .
- Conical Fitting .
- . Seal Strength
None of the data raises any new issues on safety and effectiveness.
{4}------------------------------------------------
### E. Additional Safety Information
Manufacturing controls include visual, functional and sterility tests.
The sterility of the device is assured using a sterilization method validated in accordance with ANSUAAMI/ISO 11137 - Medical Devices - Validation and Routine Control of Radiation Sterilization. The TERUMO® Needle is sterilized to provide a Sterility Assurance Level (SAL) of 10-5.
The TERUMO® Needle is classified as Externally Communicating Device, Circulating Blood. Limited Duration of Contact (≤24hr.). The device's blood contacting materials were tested in accordance with the tests recommended in the FDA General Program Memorandum #G95-1 (5/1/95): Use of International Standard ISO - 10993, "Biological Evaluation of Medical Devices part 1: Evaluation and Testing". Results of the testing demonstrate that the blood contacting materials are biocompatible.
### F. Conclusion
In summary, the TERUMO® Needle manufactured by Terumo Philippines Corporation is substantially equivalent in its intended use, technology/principal of operation, materials, and performance to the following cleared devices:
- 1. K001572 TERUMO® NEOLUS® Needle manufactured by Terumo Europe, Leuven, Belgium
- 2. K771203 TERUMO® Hypodermic Needle manufactured by Terumo Medical Corporation, Elkton, Maryland
- 3. K012646 (30G) TERUMO® Needle manufactured by Terumo Medical Corporation, Elkton, Maryland
There is no significant change that raises any new issues on safety and effectiveness.
Terumo's statement of substantial equivalence is done solely to comply with the requirements of the Federal Food, Drug and Cosmetic Act and is not intended whatsoever to be the basis for a patent infringement action.
{5}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged around the caduceus in a circular fashion.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
# Letter Date: November 8, 2012
Terumo Medical Corporation Ms. Sandi Hartka Manager, Regulatory Affairs 950 Elkton Boulevard Elkton, Maryland 21921
Re: K121607
Trade/Device Name: Terumo® Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: FMI Dated: November 2, 2012 Received: November 5, 2012
### Dear Ms. Hartka:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register.
{6}------------------------------------------------
Page 2 - Ms. Hartka
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Digitally signed by Anthony D. Watson DN: c=US. o=U.S. Government ou=HiHS, ou= Anthony D. Watson eople, cn=Anthony D. Watson, 342.19200300.100.1.1=1300092402 2012.11.08 11:16:45 -05'00
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Respiratory, Infection Control and
Dental Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{7}------------------------------------------------
Indications for Use
510(k) Number (if known): K121607
Device Name: TERUMO® Needle
Indications for Use:
The TERUMO® Needle is a hypodermic single lumen needle intended to inject The TERUMO Necule is a nypousiness and the body below the surface of the skin.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page __ of
N
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
**510(k) Number:** K121607
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.