← Product Code [NKB](/productcode/NKB) · K121568

# DALI SPINAL FIXATION SYSTEM (K121568)

_Accel Spine · NKB · Sep 21, 2012 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K121568

## Device Facts

- **Applicant:** Accel Spine
- **Product Code:** [NKB](/productcode/NKB.md)
- **Decision Date:** Sep 21, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Dali Spinal Fixation System is non-cervical spinal fixation devices intended for use as posterior pedicle fixation systems (T1-S2/ilium), or as an anterolateral fixation system (T8-L5). All components in the system are limited to skeletally mature patients. System components are to be used for immobilization and stabilization of the spine as an adjunct to fusion. These devices are indicated for all of the following indications regardless of the intended use: - degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies); - spondylolisthesis; - trauma (i.e. fracture or dislocation); - deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis); - tumor; - stenosis, and; - failed previous fusion (pseudoarthrosis)

## Device Story

Dali Spinal Fixation System is a posterior pedicle (T1-S2/ilium) or anterolateral (T8-L5) fixation system. Comprised of titanium alloy or CoCrMo alloy screws, set screws, rods, crosslinks, and connectors. Used by surgeons in clinical settings to provide spinal immobilization and stabilization as an adjunct to fusion. Mechanical components are implanted to support the spine; no electronic or software components involved. Benefit is stabilization of the spinal column in patients with degenerative, traumatic, or neoplastic conditions.

## Clinical Evidence

No clinical data provided. Substantial equivalence established via bench testing.

## Technological Characteristics

Materials: Titanium-6Aluminum-4Vanadium ELI (ASTM F136) and Cobalt-28Chromium-6Molybdenum (ASTM F1537). Components: screws (4.0-8.5mm diameter, 20-150mm length), set screws, rods, crosslinks, connectors. Mechanical fixation system; no energy source or software.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- Venus Spinal Fixation System ([K120270](/device/K120270.md))
- Venus Basic Spinal Fixation System ([K100706](/device/K100706.md), [K103085](/device/K103085.md))
- Anyplus Spinal Fixation System ([K091717](/device/K091717.md))
- APEX Spine System ([K062513](/device/K062513.md), [K092825](/device/K092825.md), [K102488](/device/K102488.md))
- Xia III Spinal System ([K071373](/device/K071373.md), [K083393](/device/K083393.md), [K091291](/device/K091291.md))
- USS Polyaxial and Synthes USS Iliosacral System ([K022949](/device/K022949.md), [K082572](/device/K082572.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# CAccessPINE

SEP 21 2012

### 510(k) SUMMARY

The following 510(k) summary is being submitted as required by 21 CFR 807.92(a):

| Submitter:      | Accel SPINE                               |
|-----------------|-------------------------------------------|
|                 | 14850 Quorum Dr.Suite220 Dallas, TX 75254 |
| Contact Person: | Daniel Chon                               |
|                 | Telephone. 800.713.9489                   |
|                 | FAX .888.831.4892                         |
|                 | dchon(@AccelSpine LLC LLC.com             |
| Prepared Date   | May 22, 2012                              |

#### 2. Device Identification

1.

| Trade Name:             | Dali Spinal Fixation System                                       |
|-------------------------|-------------------------------------------------------------------|
| Common Name:            | Spinal Fixation Appliances                                        |
| Product Code:           | NKB, KWP, KWQ, MNH, MNI                                           |
| Class                   | Class III                                                         |
| Classification<br>Name: | l } Spinal Interlaminal Fixation Orthosis, 21 CFR §888.3050       |
|                         | 2) Spinal Intervertebral Body Fixation Orthosis, 21 CFR §888.3060 |
|                         | 3) Pedicle Screw Spinal System, 21 CFR §888.3070                  |

#### 3. Identification of Legally Marketed Devices

L&K BIOMED: Venus Spinal Fixation System (K120270) L&K BIOMED: Venus Basic Spinal Fixation System (K100706,K103085) GS Medical: Anyplus Spinal Fixation System (K091717) Spine Craft: APEX Spine System (K062513,K092825, K102488) Stryker Spine: Xia@III Spinal System (K071373,K083393,K091291) Synthes Spine: USS Polyaixal and Synthes USS Iliosacral System (K022949,K082572)

#### 4. Device Description

This system is comprised of screws, set screws, rods, crosslink and connectors. The components of this system are manufactured of Titanium alloy (Titanium-6Aluminum-4Vanadium ELI, per ASTM F136) and CoCrMo alloy (Cobalt-28Chromium-6Molybdenum, per ASTM F1537). The screws are available from 4.0 to 8.5mm diameters with lengths ranging from 20-150mm.

#### 5. Intended use / Indications for Use

The Dali Spinal Fixation System is non-cervical spinal fixation devices intended for use as posterior pedicle fixation systems (T1-S2/ilium), or as an anterolateral fixation system (T8-L5). All components in the system are limited to skeletally mature patients. System

{1}------------------------------------------------

AccelSPINE

components are to be used for immobilization and stabilization of the spine as an adjunct to fusion. These devices are indicated for all of the following indications regardless of the intended use:

- degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies);
- spondylolisthesis;
- trauma (i.e. fracture or dislocation);
- deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis);
- tumor;

.

- stenosis, and;
- failed previous fusion (pseudoarthrosis)

#### 6. Comparison of the Technology Characteristics

The Dali Spinal Fixation System shares technological characteristics similar to the predicate devices. These characteristics include similar design, the same materials, substantially equivalent performance characteristics and the same intended use.

#### 7. Performance Data

Non-Clinical Performance and Conclusions:

Bench testing results demonstrate that Dali Spinal Fixation System performs equivalently to the predicates in static compression bending, static torsion, static torsion, dynamic compression bending (in accordance with ASTM F1717-10) and gripping-push down (in accordance with ASTM F1798).

Clinical Performance and Conclusions:

Clinical data and conclusions were not needed for this device.

#### 8. Conclusion

The Dali Spinal Fixation System is substantially equivalent to the device referenced above and is therefore safe and effective for its intended use.

{2}------------------------------------------------

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Accel SPINE % L & K Biomed Company, Limited Ms. Ki Hyang Kim Vice President 1104-ho, 145, Gasandigital 1-ro, Geumcheon-gu Seoul, 153-787, Korea

Re: K121568

Trade/Device Name: Dali Spinal Fixation System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, KWP, KWQ, MNH, MNI Dated: August 18, 2012 Received: August 20, 2012

Dear Ms. Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical

SEP 21 2012

{3}------------------------------------------------

### Page 2 - Ms. Ki Hyang Kim

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use Statement

CAccessPINE

510(k) Number : K 121568 Device Name: Dali Spinal Fixation System Indications For Use:

The Dali Spinal Fixation System is non-cervical spinal fixation devices intended for use as posterior pedicle fixation systems (TI-S2/ilium), or as an anterolateral fixation system (T8-L5). All components in the system are limited to skeletally mature patients. System components are to be used for immobilization and stabilization of the spine as an adjunct to fusion. These devices are indicated for all of the following indications regardless of the intended use:

- degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies);
- spondy lolisthesis;
- trauma (i.e. fracture or dislocation);
- deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis);
- tumor:
- stenosis, and;
- failed previous fusion (pseudoarthrosis)

| Prescription Use           | √ |
|----------------------------|---|
| (Part 21 CER801 Subpart D) |   |
| AND/OR                     |   |
| Over-The-Counter Use       |   |
| (21 CER801 Subpart C)      |   |

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concyptence of CDRH, Office of Device Evaluation (OED)

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number_

K121568

Page 1/1

Dali Spinal Fixation System 510(K)

---

**Source:** [https://fda-staging.innolitics.com/device/K121568](https://fda-staging.innolitics.com/device/K121568)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).