A&J 5L POCA series Oxygen Concentrator is intended for use as an oxygen concentrator to provide supplemental low flow oxygen therapy at home, in nursing homes, at patient care facilities. A&J 5L POCA series Oxygen concentrator is available by prescription only under the supervision of a physician.
Device Story
AC-powered oxygen concentrator; draws ambient air via compressor; utilizes molecular sieve adsorbent to selectively retain nitrogen; releases nitrogen when pressure vents; cycle controlled by motorized valve; delivers high-purity supplemental oxygen via nasal cannula or mask; optional bubble humidifier. Used in homes, nursing homes, or patient care facilities; operated by patients or caregivers. Front panel includes status lights (power, oxygen purity, flow, service), flow meter, and monitor display for pressure/purity/hour meter. Provides supplemental oxygen therapy; does not sustain life.
Clinical Evidence
Bench testing only. Performance validated against design specifications and standards including ASTM F1464, ISO 8359, IEC 60601-1, and IEC 60601-1-2. Gas output tested for particulate matter, VOCs, CO, CO2, and ozone to ensure biocompatibility of gas pathway.
Technological Characteristics
AC-powered portable oxygen generator; Pressure Swing Adsorption (PSA) technology; molecular sieve adsorbent. Flow rate 0-5 LPM; oxygen purity 93% ± 2%; outlet pressure 8.5 psi. Type B equipment. Complies with IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), ASTM F1464, and ISO 8359.
Indications for Use
Indicated for patients requiring supplemental low flow oxygen therapy in home, nursing home, or patient care facility settings. Available by prescription only under physician supervision. Not intended to support or sustain life.
Regulatory Classification
Identification
A portable oxygen generator is a device that is intended to release oxygen for respiratory therapy by means of either a chemical reaction or physical means (e.g., a molecular sieve).
{0}------------------------------------------------
### Chapter 4. 510(k) Summary
1121531
JUL 1 7 2012
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA and 21 CFR 807.92.
### 1. Submitter Information
- � Establishment Registration Number: 3004443928
- � Company Name: Zhongshan A&J Medical Equipment Co., Ltd.
- Address: #3 Shenghui South Road, Nantou Town, Zhongshan City, Guangdong, China �
- � Phone: +86-760-2313 0562
- Fax: +86-760-2313 0754 �
- � Contact Person (Title): Ms. Hebe An (Management Representative)
- E-mail: xa1218@126.com �
#### 2. Application Correspondent
- � Company Name: MEDLAB (Shenzhen) Information Service Co., Ltd.
- Address: Room 2706, Block A, ZhongFang JinYuan Buiding, Xinwen Road, Shenzhen, � Guangdong, P.R. China, 518034
- Phone: +86-755-8308 9699 �
- � Fax: +86-755-8632 9134
- � Contact Person (Title): Ms. Sabrina Wei (Project Manager)
- � E-mail: sabrinawei@hotmail.com
#### Subject Device Information: 3.
- � Product Code: CAW
- Regulation Number: 868.5440 �
- Class: 2 �
- Review Panel: Anesthesiology �
- Classification Name: Generator, oxygen, portable �
- Trade Name: A&J 5L POCA series Oxygen Concentrator �
- Model: POCA01B, POCA03, POCA04, POCA05, POCA06 ◆
{1}------------------------------------------------
Zhongshan A&J Medical Equipment Co., Ltd. Sponsor: Subject Device: A&J 5L POCA series Oxygen Concentrator File No .: 510(k) submission report (V1.0), Chapter 4
#### 4. Predicate Device Information:
- � 510(k) Number: K071608
- Sponsor: Zhongshan A&J Medical Equipment Co., Ltd. �
- Classification Name: Generator, oxygen, portable �
- Trade Name: A&J 5L Oxygen concentrator �
- Product Code: CAW �
- Model: POCA01 �
#### Device Description 5.
A&J 5L POCA series Oxygen concentrator is AC power electrically operated. The unit separates oxygen from room air (ambient air) which allows high-purity supplemental oxygen to be delivered through the oxygen outlet, although the concentrator filters the oxygen in a room, it will not affect the normal amount of oxygen in your room. Air is drawn into the device with a compressor and exposed to molecular sieve adsorbent that selectively retains nitrogen and other components until they are released when the pressure is vented to the atmosphere. This cycle is controlled by a motorized value and protected from over pressurization by the compressor's pressure relief value.
Oxygen provided by the A&J 5L POCA series Oxygen concentrator is delivered to the user by means of standard oxygen supply tubing and a standard nasal cannula or mask. A standard bubble humidifier may be used, if physician has prescribed an oxygen humidifier as part of therapy.
The front panel of the A&J 5L POCA series Oxygen concentrator contains the controls and indicators. These include the status lights (included power light, normal oxygen light, fow oxygen light and service required light), standard power switch, flow meter knob, a circuit breaker which could reset the device after electrical overload shutdown, an oxygen outlet which oxygen is dispersed through, a monitor display which indicates the condition of system status (included pressure status, oxygen purity status and electric hour meter, etc.). The user could operate the device conveniently according to the instructions.
#### 6. Intended Use
A&J 5L POCA series Oxyqen Concentrator is intended for use as an oxygen concentrator to provide supplemental low flow oxygen therapy at home, in nursing homes, at patient care facilities. A&J 5L POCA series Oxygen concentrator is available by prescription only under the supervision of a physician.
Report by Zhongshan A&J Medical Equipment Co., Ltd.
{2}------------------------------------------------
Sponsor: Zhongshan A&J Medical Equipment Co., Ltd. Subject Device: A&J 5L POCA series Oxygen Concentrator 510(k) submission report (V1.0), Chapter 4 File No.:
### 7. Contraindication
The device is not intended to support or sustain life.
#### 8. Performance Summary
Technologies utilized by the A&J 5L POCA series Oxygen Concentrator conduces no questions of safety and effectiveness. The same technologies are being used on the identified predicate device. Bench performance testing has demonstrated that the A&J 5L POCA series Oxygen Concentrator is substantially equivalent to the predicate device.
### 9. Testing
.
Laboratory testing was conducted to validate and verify that the A&J 5L POCA series Oxygen Concentrator met all design specifications and was substantially equivalent to the predicate device. The testing consisted of all environmental testing identified in the FDA's "Reviewer Guidance for Nebulizers, Metered Dose Inhalers, Spacers and Actuators (October 1, 1993)". Testing was also performed to demonstrate the sponsored device series achieves the standards' requirements of ASTM F1464 and ISO 8359.
Hazard analysis of the system and its software was performed and testing was conducted to validate overall operation of the system. The A&J 5L POCA series Oxygen Concentrator has also been tested to assure its compliance to the requirements of various standards of IEC60601-1, IEC60601-1-2.
### 10. Comparison to Predicate Device
Compare the subject devices to the predicate device: they are same in design principle, intended use, functions, performance, material and applicable standards. The main difference between subject devices and the predicate device is enclosure style and the following list notes. These differences do not raise any new issues of safety or effectiveness.
| Elements of<br>Comparison | Subject Device | Predicate Device | Judgment | | |
|-----------------------------------|--------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|--------------|--|--|
| Device Name | A&J 5L POCA series Oxygen<br>Concentrator, Models:<br>POCA01B, POCA03, POCA04,<br>POCA05, POCA06 | A&J 5L Oxygen Concentrator,<br>Model: POCA01 | | | |
| Operation and Storage Environment | | | | | |
| Operating<br>Conditions | Temperature: 10-40 ℃<br>Humidity: 30-70 %<br>Atmospheric Pressure: 50~106<br>kPa (7.3-15.4 psi) | Temperature: 10-35 ℃<br>Humidity: 30~70 %<br>Atmospheric Pressure: 50~106<br>kPa (7.3-15.4 psi) | SE<br>Note 1 | | |
{3}------------------------------------------------
#### Zhongshan A&J Medical Equipment Co., Ltd. Sponsor: Subject Device: A&J 5L POCA series Oxygen Concentrator
510(k) submission report (V1.0), Chapter 4 File No.:
| Elements of<br>Comparison | Subject Device | Predicate Device | Judgment |
|---------------------------------------------------|-----------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Storage<br>Conditions | Temperature: -40~70 ℃<br>Humidity: 10~100 %<br>Atmospheric Pressure: 50~106 kPa (7.3~15.4 psi) | Temperature: -40~70 ℃<br>Humidity: 10~100 %<br>Atmospheric Pressure: 50~106 kPa (7.3~15.4 psi) | SE |
| Safety Factor | | | |
| Power Input | 115V, 60Hz, 3.1A | 115 V, 60 Hz, 1.8A (Max.) | SE<br>Note 2 |
| Degree of<br>Protection Against<br>Electric Shock | Type B Equipment | Type B Equipment | SE |
| Electrical Safety | Compliance with IEC 60601-1 | Compliance with IEC 60601-1 | SE |
| EMC | Compliance with IEC 60601-1-<br>2 | Compliance with IEC 60601-1-<br>2 | SE |
| Performance and Specification | | | |
| Delivery Rate | 0 to 5 LPM | 0 to 5 LPM | SE |
| Oxygen<br>Percentage | 93% ± 2% | 93% ± 2% | SE |
| Outlet Pressure | 8.5 psi (58.6 kPa) | 8.5 psi (58.6 kPa) | SE |
| Sound Level | < 45 dbA (Overall Average) | 45-47 dbA (Overall Average) | SE<br>Note 3 |
| Operating<br>System | Time Cycle /<br>Pressure Swing Adsorption | Time Cycle /<br>Pressure Swing Adsorption | SE |
| Mode of<br>Operation | Continuous | Continuous | SE |
| Others | | | |
| Biocompatibility | All the patient contacting<br>material and gas pathway<br>material are the same with<br>predicate device. | Conduct the following output<br>gases tests to replace the<br>biocompatibility testing for the<br>materials in the gas pathway:<br>- Particulate Matter per EPA<br>PM 2.5;<br>- Volatile Organic Compounds<br>per ASTM D5466 or equivalent;<br>- Measurement of output gas<br>for carbon monoxide;<br>- Measurement of output gas<br>for carbon dioxide;<br>- Measurement of output gas<br>for ozone. | SE |
| Elements of<br>Comparison | Subject Device | Predicate Device | Judgment |
| Accessories | All the specified accessories<br>are the same with predicate<br>device. | All the following accessories<br>are not included in the<br>package, and need to use the<br>legitimate marketing products<br>in the United States:<br>- Oxygen Mark<br>- Nasal Cannula<br>- Bubble Humidifier<br>- Oxygen Outlet Connector<br>- Oxygen Tubing | SE |
Report by Zhongshan A&J Medical Equipment Co., Ltd.
{4}------------------------------------------------
Sponsor: Zhongshan A&J Medical Equipment Co., Ltd. Subject Device: A&J 5L POCA series Oxygen Concentrator File No.: 510(k) submission report (V1.0), Chapter 4
### Note:
- 1) The operating temperature range of subject device is wider than predicate device. This can seem to be SE.
- 2) Although the power input of subject device is different from predicate device, both of them are complied with IEC 60601-1, the difference will not affect the SE.
- 3) The sound level of subject device is lower than predicate device, and both results are complied with ISO 8359. The difference will not affect the SE.
### 11. Conclusion
As a summary to the above testing results and performance specifications, A&J 5L POCA series Oxygen Concentrator is as same safe and effective as the predicate device, therefore, is Substantial Equivalent to the predicate device.
## 12. Summary Prepared Date: 2012-06-19
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure with outstretched wings.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Zhongshan A&J Medical Equipment Company, Limited C/O Mr. Ned Devine Responsible Third Party Official Underwrites Laboratories, Incorporated 333 Pfingsten Road Northbrook, Illinois 60062
JUL 1 7 2012
Re: K121531
Trade/Device Name: A&J 5L POCA series Oxygen Concentrator, models: POCA01B, POCA03, POCA04, POCA05 & POCA06 Regulation Number: 21 CFR 868.5440 Regulation Name: Portable Oxygen Generator Regulatory Class: II Product Code: CAW Dated: June 26, 2012 Received: July 2, 2012
Dear Mr. Devine:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register.
{6}------------------------------------------------
### Page 2- Mr. Devine
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
hh f..
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{7}------------------------------------------------
Zhongshan A&J Medical Equipment Co., Ltd. Sponsor: A&J 5LOxygen Concentrator POCA series Subject Device: 510(k) submission report (V1.0), Chapter 3 File No .:
Chapter 3. Indications for Use
## Indications for Use
510(k) Number (if known):
Device Name: A&J 5L POCA series Oxygen Concentrator, models: POCA01B, POCA03, POCA04, POCA05, POCA06
Indications for Use:
A&J 5L POCA series Oxygen Concentrator is intended for use as an oxygen concentrator to provide supplemental low flow oxygen therapy at home, in nursing homes, at patient care facilities. A&J 5L POCA series is available by prescription only under the supervision of a physician.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Y Shulttin
Page 1 of 1
(Division Sign-Off) (Division of Anesthesiology. General Hospital Infection Control, Dental Devices
510(k) Number: k/ 2153/
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.