K121513 · Siemens Medical Solution · KPR · Aug 10, 2012 · Radiology
Device Facts
Record ID
K121513
Device Name
MULTIX FUSION
Applicant
Siemens Medical Solution
Product Code
KPR · Radiology
Decision Date
Aug 10, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1680
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Multix Fusion system is a radiographic system used in hospitals, clinics, and medical practices. Multix Fusion enables radiographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and bariatric patients. It can also be used for intravenous, small interventions (like biopsy, punctures, etc.) and emergency (trauma, critical ill) applications. Exposures may be taken with the patient sitting, standing, or in the prone position. The Multix Fusion system is not meant for mammography. The Multix Fusion uses a mobile (wired) or portable (wireless) digital detector for generating diagnostic images by converting x-rays into electronic signals. The Multix Fusion is also designed to be used with conventional film/screen or Computed Radiography (CR) cassettes.
Device Story
Multix Fusion is a stationary X-ray system for diagnostic imaging; configurations include analog, digital (wired/wireless flat panel detector), or DR-upgrade. System comprises radiologic table, X-ray generator, X-ray tube, ceiling-mounted support, and Bucky wall stand. Operated by healthcare professionals in hospitals/clinics. Input: X-ray radiation; output: diagnostic images via digital detector or film/CR cassettes. Semi-motorized tube movement available. System includes Automatic Cassette Size Sensing (ACSS) collimator. Healthcare providers use images for clinical decision-making in trauma, biopsy, and general radiography. Benefits include versatile patient positioning (sitting, standing, prone) and diagnostic flexibility across diverse patient populations.
Clinical Evidence
No clinical data. Evidence consists of non-clinical verification and validation testing, including phantom testing and risk analysis, confirming software and hardware specifications met acceptance criteria.
Technological Characteristics
Stationary X-ray system; 80kW/65kW/55kW generator; ceiling-mounted tube support; flat panel detector (wired/wireless); ACSS collimator. Supports digital imaging (Fluorospot Compact) and conventional film/screen/CR. System includes visual/audible safety warnings and continuous error monitoring.
Indications for Use
Indicated for radiographic imaging of whole body (skull, chest, abdomen, extremities) in pediatric, adult, and bariatric patients. Suitable for clinical, hospital, and emergency/trauma settings. Supports intravenous and small interventional procedures (biopsy, punctures). Contraindicated for mammography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows handwritten text on a white background. The text at the top reads "K121513", which appears to be a code or identifier. Below this, the text "Page 1 of 4" indicates that this is the first page of a four-page document. The handwriting is clear and legible, with a simple, unadorned style.
AUG 1 0 2012
# 510(k) Summary: Multix Fusion X-ray System
Company:
: - -
Siemens Medical Systems, Inc. 1 Valley Stream Parkway Malvern, PA 19355
July 26, 2012 Date Prepared:
Date Frepared. In our offectiveness information is being submitted in This order) - Sammants of SMDA 1990 and 21 CFR § 807.92.
### General Information: 1.
Importer / Distributor: Siemens Medical Solutions, Inc. 51 Valley Stream Parkway, E-50 Malvern, PA 19355 Establishment Registration Number: 2240869
## Location of Manufacturing Site:
Siemens AG Medical Solutions X-Ray Products Henkestrasse 127 DE-91052 Erlangen Establishment Registration Number: 3002808157
## Manufacturer:
Siemens Shanghai Medical Equipment Ltd. 278 Zhou Zhu Road, Shanghai 201318, China Headquarters: Siemens AG Wittelsbacherplatz 2 D-80333 Munich 2, Germany Establishment Registration Number: 3003202425
#### 2. Contact Person:
Ms. Patricia D Jones Technical Specialist, Regulatory Submissions Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway D-02 Malvern, PA 19355 Fax: (610) 448-1787 Phone: (610) 448 -3536
Email: patricia.d.jones@siemens.com
- Device Name and Classification: 3. Trade Name: Classification Name: Classification Panel:
Multix Fusion Stationary X-Ray System Radiology
{1}------------------------------------------------
Classification Regulation: Device Class: Product Code:
21 CFR §892.1680 Class II 90 KPR
- 4. Legally Marketed Predicate Device Trade Name: Ysio 510(k) #: K0817227 Clearance Date: August 22, 2008 Classification Name: Solid State X-ray Imager (SSXI) Classification Panel: Radiology CFR Section: 21 CFR §892.1680 Device Class: Class II Product Code: 90 MQB
KIZISI3
age 2 of H
#### 5. Device Description:
The Multix Fusion is a modification of Siemens' Ysio X-ray system cleared under Premarket Notification K081722 on 08/22/2008. The Multix Fusion offers the following system configurations:
- An analog radiography system or; .
- A digital radiography system with a mobile (wired) and/or a portable . (wireless) flat panel detector or;
- An optional Multix DR-Upgrade for the analog version of the Multix Fusion . and other tables within the Multix and Vertix analog family.
This modification does not affect the general intended use of the device nor does it alter its fundamental scientific technology.
The Multix Fusion system consists of a radiologic table, x-ray generator, x-ray tube, flat panel detector (wired or wireless) and Bucky-wall stand. It will be marketed as an analog system, a digital radiographic system with a mobile (wired) and/or portable (wireless) flat panel detector or an optional Multix DR- Upgrade for the analog version of the Multix Fusion and other tables within the Multix and Vertix analog family.
The key components are a patient table, a ceiling mounted support and a Bucky wall stand which are available in different configurations. A semi-motorized movement in the ceiling-mounted X-ray tube is available. The optional Multix DR-Upgrade for the analog version of the Multix Fusion consists of a mobile (wired) and/or portable (wireless) flat panel detector and an imaging system. Components used to upgrade (Multix DR-Upgrade) the analog version of the Multix Fusion and other tables within the Multix and Vertix analog family are the same or similar components cleared in the Ysio 510(k) K0817227 on August 22, 2008.
The Multix Fusion may be combined into different configurations to provide specialized customer requirements.
Similar to the cleared Ysio stationary x-ray system, the Multix Fusion and the Mutix DR-Upgrade have the same or similar comparable components. The subject device the Multix Fusion does not affect the indication for use nor the intended use of the device nor does it alter its fundamental scientific technology.
{2}------------------------------------------------
K121513
Page 3 of 4
### 6. Indication for Use:
The Multix Fusion system is a radiographic system used in hospitals, clinics, and medical practices. Multix Fusion enables radiographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and bariatric patients. It can also be used for intravenous, small interventions (like biopsy, punctures, etc.) and emergency (trauma, critical ill) applications. Exposures may be taken with the patient sitting, standing, or in the prone position. The Multix Fusion system is not meant for mammography.
The Multix Fusion uses a mobile (wired) or portable (wireless) digital detector for generating diagnostic images by converting x-rays into electronic signals. The Multix Fusion is also designed to be used with conventional film/screen or Computed Radiography (CR) cassettes.
### 7. Substantial Equivalence:
The Multix Fusion is substantially equivalent to the commercially available Siemens Ysio radiographic x-ray system. The Ysio was described in premarket notification K081722 which received FDA Clearance on August 22, 2008. Ysio is indicated for tomographic procedures, The Multix Fusion is not indicated for tomographic procedures. The technical differences between the Multix Fusion and the Ysio are listed below:
| Comparable<br>Properties | Subject Device<br>Multix Fusion | Predicate Device<br>Ysio (K081722) | Comparison<br>Results |
|--------------------------------------------------------|---------------------------------------------------------|---------------------------------------------------------|----------------------------|
| X-ray tube | 80 kW Two-focus | 80 kW Two-focus | Same |
| Flat panel detector<br>portable | 43cm x 36 cm<br>Type: 4336X | 43cm x 36 cm<br>Type: 3543pR | Same imaging<br>technology |
| Digital Imaging system | Fluorospot Compact<br>High Res Digital<br>Imaging | Fluorospot Compact<br>High Res Digital<br>Imaging | Similar<br>Functionality |
| Collimator | ACSS collimator<br>(Automatic Cassette<br>Size Sensing) | ACSS collimator<br>(Automatic Cassette<br>Size Sensing) | Similar<br>Functionality |
| X-ray Generator | 80 kW<br>65 kW<br>55 kW | 80 kW<br>65 kW | Similar |
| Conventional<br>film/screen systems<br>or CR cassettes | Film/Screen or CR<br>Cassettes | Film/Screen or CR<br>Cassettes | Same |
Table 1: Subject and Predicate Device Comparable Properties
## 8. Summary of Technological Characteristics of the Subject Device as Compared with the Predicate Device:
The Multix Fusion does not have significant changes in materials, energy source, or technological characteristics when compared to the predicate device. The intended use and fundamental scientific technology are similar to the predicate device. The Multix Fusion is designed based on the Ysio radiography x-ray system. It uses the same or similar components cleared in the Ysio (e.g. tube, generator, ceilingmounted tube support, table, Bucky wall stand and imaging system). The Multix Fusion is similar to the predicated system with respect to technical characteristics,
{3}------------------------------------------------
K121513
Page 4 of 4
performance and intended use. The subject device the Multix Fusion does not affect the indication for use nor the intended use of the device nor does it alter its fundamental scientific technology
### 9. Non-clinical Testing:
Non-clinical testing was conducted for the Multix Fusion during product development. This includes verification and validation testing as well as phantom testing. The risk analysis was combation and risk controls implemented to mitigate identified hazards. The testing results supports that all the software specificating have met the acceptance criteria.
Testing for verification and validation of the device was found acceptable to support the claims of substantial eqivalence.
## 10. General Safety and Effectiveness Concerns:
Instructions for use are included within the device labeling, and the information provided will enable the user the device in a safe and effective manner. Several safety features including visual and audible warnings are incorporated into the system design. In addition the Multix Fusion is continually monitored, and if an error occurs , the system the mank I asion is continually montored, and If an
displayed displayed.
Furthermore the operators are health care professionals familiar with and responsible for the evaluating and post processionals familiar will and performed. To mininize electrical, mechanical and radiation hazards, Siemens adheres to recognized and established industry practice, and all equipment is subject to final performance testing.
## 11. Conclusion as to Substantial Equivalence:
The Multix Fusion is intended for the same use as Ysio (with the exception of tomography). It uses the same or similar components cleared for the Ysio (e.g. tube, generator, ceiling-mounted tube support, table, Bucky wall stand and imaging system). It is Siemens opinion that the Multix Fusion is substantially equivalent to the cleared predicate device the Ysio radiographic x-ray system.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle symbol on the right and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged in a circular fashion around the left side. The eagle symbol is composed of three stylized lines that suggest the wings and body of an eagle.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service .
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
AUG 10 2012
Ms. Patricia D. Jones Technical Specialist, Regulatory Submissions Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway, E-50 MALVERN PA 19355
Re: K121513
Trade/Device Name: Multix Fusion Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: KPR Dated: July 26, 2012 Received: July 27, 2012
Dear Ms. Jones:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{5}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice mource no real forth in the quality systems (QS) regulation (21 CFR Part 820). This letter requirements as bet form in the quar device as described in your Section 510(k) premarket witi anow you to begal finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 11 you desire specifice of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Special 510(k) Submission: Multix Fusion
# Indications for Use Statement
510(k) Number (if known):
Device Name: Multix Fusion
Indications for Use:
The Multix Fusion system is a radiographic system used in hospitals, clinics, and medical practices. Multix Fusion enables radiographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and bariatric patients. It can also be used for intravenous, small interventions (like biopsy, punctures, etc.) and emergency (trauma, critical ill) applications. Exposures may be taken with the patient sitting, standing, or in the prone position. The Multix Fusion system is not meant for mammography.
The Multix Fusion uses a mobile (wired) or portable (wireless) digital detector for generating diagnostic images by converting x-rays into electronic signals. The Multix Fusion is also designed to be used with conventional film/screen or Computed Radiography (CR) cassettes.
Prescription Use X (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
AND/OR
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
hnd D'O'hm
Division Sian-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K121513
Page 1 of
Special 510(k) Multix Fusion
Siemens Medical Systems, Inc.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.