DISPOSABLE LINEAR STAPLER, DISPOSABLE LINEAR CUTTER
K121474 · Changzhou Kangdi Medical Stapler Co., Ltd. · GDW · Jul 16, 2012 · General, Plastic Surgery
Device Facts
Record ID
K121474
Device Name
DISPOSABLE LINEAR STAPLER, DISPOSABLE LINEAR CUTTER
Applicant
Changzhou Kangdi Medical Stapler Co., Ltd.
Product Code
GDW · General, Plastic Surgery
Decision Date
Jul 16, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
Disposable Linear Stapler, Model KYFB-30 / 60 / 90 The Disposable Linear Stapler has application throughout the alimentary tract and in thoracic surgery for transaction and resection of internal tissues. Disposable Linear Cutter, Model KYQII-60 / 80 / 100 The Disposable Linear Cutter has applications in abdominal, gynecological, pediatric and thoracic surgical procedures for resection, transection and creation of anastomosis.
Device Story
Disposable linear staplers and cutters deliver two staggered rows of titanium staples to approximate internal tissues. The stapler is available in 30, 60, and 90 mm sizes; the cutter in 60, 80, and 100 mm sizes. Staple height is adjustable to accommodate varying tissue thicknesses. Used in OR settings by surgeons for resection, transection, and anastomosis. The device functions mechanically to deliver and secure staples, providing tissue approximation and closure. Benefits include reliable tissue management during surgical procedures.
Clinical Evidence
Bench testing only. Performance testing confirmed energy requirements for staple delivery and removal were consistent with predicate devices. Biocompatibility testing passed ISO 10993-1, -3, -5, -6, -10, and -11. Sterilization and shelf-life validation performed per ISO 11137, ISO 11737, ISO 11607, and ASTM standards.
Technological Characteristics
Constructed of unalloyed titanium (ISO 5832-2:1999, ASTM F67-06). Mechanical stapling/cutting principle. Available in multiple lengths (30-100 mm). Sterilized via validated methodology (ISO 11137). No software or electronic components.
Indications for Use
Indicated for patients undergoing alimentary tract, thoracic, abdominal, or gynecological surgical procedures, including pediatric patients, requiring resection, transection, or creation of anastomosis of internal tissues.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
Predicate Devices
PROXIMATE Reloadable Linear Stapler, Model TL30, TL60, TL90, TLH30, TLH60, TLH90 (K020779)
DST Series GIA: Auto Suture GIA Reloadable Stapler, GIA60, GIA80, GIA100 (K111825)
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: April 22, 2012
JUL 16. 2012
Company making the submission: Name - Changzhou Kangdi Medical Stapler Co., Ltd Address - No. 167-5, East Road, Changzhou, Jiangsu, China 213022 Telephone - 86-519-85162780 Fax - 86-519-85139853 Contact - Mr. Fleming Jiang Email - jiangmingfang@kanghui-china.com
Correspondent: Name: Charles Mack Address: 77325 Joyce Way; Echo, Oregon 97826 Telephone: 931-625-4938 Email: charliemack@irc-us.com
### Device:
Trade/proprietary names:
Disposable Linear Stapler, Model KYFB Disposable Linear Cutter, Model KYQII
Common Name: Implantable Staple Classification Name: Staple, implantable Predicate Devices:
# Linear Stapler:
Ethicon Endo-Surgery, Inc. K020779 The PROXIMATE Reloadable Linear Stapler, Model TL30, TL60, TL90,
TLH30, TLH60, TLH90
#### Linear Cutter:
Surgical Devices, a global business unit of Tyco Healthcare Group LP (d/b/a Covidien) K111825 DST Series™ GIA™: Auto Suture™ GIA™ Reloadable Stapler, GIA60, GIA80, GIA100
Classifications Names & Citations: 21CFR 878.4520, GDW, Polymer, ENT synthetic-polyamide (mesh or foil material), Class II
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows handwritten text. The first line reads "<12147.4". The second line reads "page 2/3". The text appears to be written in black ink on a white background.
# Description: General Disposable Linear Stapler
The Disposable Linear Stapler delivers two staggered rows of titanium staples to approximate internal tissues. The instrument is available in three sizes to accommodate different tissues. Staple height is adjustable to compensate for various tissue thickness .The 30 mm instruments create a 30 mm staple line, the 60 mm instruments create a 60 mm staple line, the 90 mm instruments create a 90 mm staple line.
# Disposable Linear Cutter
The disposable linear cutter and the cartridge for disposable linear cutter have applications in abdominal gynecological, pediatric and thoracic surgical procedures for resection, transection and creation of anastomosis.
# Indications For Use
Disposable Linear Stapler, Model KYFB-30 / 60 / 90
The Disposable Linear Stapler has application throughout the alimentary tract and in thoracic surgery for transaction and resection of internal tissues.
Disposable Linear Cutter, Model KYQII-60 / 80 / 100
The Disposable Linear Cutter has applications in abdominal, gynecological, pediatric and thoracic surgical procedures for resection, transection and creation of anastomosis.
### Technical Characteristics:
The Disposable Linear cutter and Disposable Linear cutter have not changed in basic design. The only difference in the current submitted device is the size of the staples in the submitted device The stable differences are not significant and testing of the staples during insertion and extraction showed that there are no appreciable differences in the energy needed to insert or extract the staples.
### Performance data:
In order to establish substantial equivalence to the identified predicate device's, we performed the following tests:
Bench testing was performed to demonstrate the energy to deliver staples and also to remove staples was consistent with the predicate device. ISO 5832-2:1999 Implants for surgery - Metallic materials -Part 2: Unalloved Titanium and ASTM F67-06 Standard Specification for Unalloyed Titanium for Surgical Implant Applications (UNS R50250, UNS R50400, UNS R50550, UNS R50700).
Biocompatibility testing was performed to ensure the stapler and cutters were constructed of material which passes the ISO 10993 benchmarks. The devices successfully passed tests for ISO 10993-1, ISO 10993-3, ISO 10993-5, ISO 10993-6, ISO 10993-10 and ISO 10993-11.
As these products are delivered sterility testing was performed to ensure the sterilization methodology was sufficient and that the packaging kept the for the stated 3 year shelf
{2}------------------------------------------------
life. Testing was performed to ISO 11137-1:2006, ISO 11137-2:2006, ISO 11737-1:2006 and ISO 11737-2:2009, ISO 11607-1:2006, ASTM F1929-98 (2004), ASTM F1980-01 and ASTM D3078-02 (2008) with the products successfully passing all tests.
K121474 page 3/3
#### Conclusions:
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification Changzhou Kangdi Medical Stapler Co., Ltd. concludes that Disposable Linear Stapler, Model KYFB and Disposable Linear Cutter, Model KYQII are substantially equivalent to predicate devices as described herein.
END
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a seal for the Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" around the edge. In the center of the seal is an image of an eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Changzhou Kangdi Medical Stapler Company, LTD % International Regulatory Consultants, LLC Mr. Charles Mack 77325 Joyce Way Echo, Oregon 97826
JUL 16 2012
Re: K121474
Trade/Device Name: Disposable Linear Stapler, Model KYFB-30/60/90, Disposable Linear Cutter, Model K YQII-60/80/100
Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable staple Regulatory Class: II Product Code: GDW Dated: May 13, 2012 Received: May 18, 2012
Dear Mr. Mack
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{4}------------------------------------------------
### Page 2 - Mr. Charles Mack
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific.advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions.regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Erin Keith
Fo Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Indications for Use
510(k) Number (if known): 510(k) Number (if known):
K121474
Device Name:
Disposable Linear Stapler, Model KYFB-30 / 60 / 90 Disposable Linear Cutter, Model KYQII-60 / 80 / 100
Indications For Use:
# Disposable Linear Stapler, Model KYFB-30 / 60 / 90
The Disposable Linear Stapler has application throughout the alimentary tract and in thoracic surgery for transaction and resection of internal tissues.
# Disposable Linear Cutter, Model KYQH-60 / 80 / 100
The Disposable Linear Cutter has applications in abdominal, gynecological, pediatric and thoracic surgical procedures for resection, transection and creation of anastomosis.
AND/OR Prescription Use X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Raul D. Kane by MM
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number
Page 1 of
12
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.