KD-972 Fully Automatic Wireless Blood Pressure Wrist Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to 13.5cm-22cm.
Device Story
The KD-972 is a non-invasive, automated wrist-worn blood pressure monitor. It utilizes an inflatable cuff and silicon-integrated pressure sensor technology to perform oscillometric measurements. Unlike traditional monitors, the device lacks an onboard display; it pairs via Bluetooth with an iPhone, iPod touch, or iPad running support software to function as a complete system. The device measures blood pressure during the cuff inflation phase. It provides systolic/diastolic pressure and pulse rate, detects irregular heartbeats, and calculates measurement averages. Data is transmitted wirelessly to the mobile device for display and user review. Used by clinicians or patients at home, the system enables remote monitoring and tracking of blood pressure trends, facilitating clinical decision-making through accessible, digitized health data.
Clinical Evidence
Bench testing only. The device conforms to IEC 60601-1, EN 60601-1-2, and IEC 80601-2-30 standards for safety, electromagnetic compatibility, and performance of automated non-invasive sphygmomanometers.
Technological Characteristics
Oscillometric measurement principle using silicon-integrated pressure sensors. Wrist cuff (13.5cm-22cm). Bluetooth connectivity for data transmission to mobile devices (iPhone/iPod/iPad). No onboard LCD. Powered by internal energy source. Complies with IEC 80601-2-30, IEC 60601-1, and EN 60601-1-2.
Indications for Use
Indicated for adult individuals requiring non-invasive measurement of systolic/diastolic blood pressure and pulse rate via wrist cuff (13.5cm-22cm circumference). Suitable for use by medical professionals or home users.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{0}------------------------------------------------
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of SMDA and 21 CFR 807.92.
JUN 1 4 2012
K121470
\$1/3
### 1.0 submitter's information
| Name: | Andon Health Co., Ltd. |
|----------------------|------------------------------------------------------------|
| Address: | No 3, Jinping Street Ya An Road, Nankai District, Tianjin, |
| | P.R. China |
| Phone number: | 86-22-6052 6161 Ext 8060 |
| Fax number: | 86-22-6052 6162 |
| Contact: | Liu Yi |
| Date of Application: | 05/08/2012 |
### 2.0 Device information
| Trade name: | iHealth BP7-Wireless Blood Pressure Wrist Monitor |
|----------------------|-----------------------------------------------------------------|
| Device name: | KD-972 Fully Automatic Wireless Blood Pressure Wrist<br>Monitor |
| Classification name: | Noninvasive blood pressure measurement system |
### 3.0 Classification
Production code: DXN- Noninvasive blood pressure measurement system. Regulation number: 870.1130 Classification: III Panel: Cardiovascular
### 4.0 Predicate device information
Manufacturer: Andon Health Co., Ltd. Device: KD-7964 Fully Automatic Electronic Blood Pressure Monitor 510(k) number: K102906
### 5.0 Device description
KD-972 Wireless Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system
{1}------------------------------------------------
intended to measure the diastolic and systolic blood pressures and pulse rate K121470 of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to 13.5cm-22cm.
833
It is designed and manufactured according to IEC 80601-2-30- Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers.
The operational principle is based on oscillometric and silicon integrates pressure sensor technology. it can calculate the systolic and diastolic blood pressure, the measurement results can also be classified by the function of blood pressure classification indicator. If any irregular heartbeat is detected, it can be shown to the user. More over, it also obtains the function of averaging the measurement results.
KD-972 Fully Automatic Wireless Blood Pressure Wrist Monitor achieves its function by integrate the device with an iPhone, iPod touch or iPad. For it does not contain an LCD or other display components, so It's necessary for the new device to connect to an iPhone, iPod touch or iPad containing a support software to constitute a complete blood pressure measurement system. And the new device connect iPhone, iPod or iPad through bluetooth.
#### 6.0 Intended use
KD-972 Fully Automatic Wireless Blood Pressure Wrist Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to 13.5cm-22cm.
The intended use and the indication for use of KD-972, as described in the labeling are the same as its predicate device KD-7964.
### 7.0 Summary comparing technological characteristics with predicate device
| Technological Characteristics | Comparison result |
|-------------------------------|-------------------|
| Design principle | Identical |
| Appearance | Similar |
| Patients contact Materials | Identical |
| Performance | Similar |
2- 3
{2}------------------------------------------------
| Biocompatibility | Identical |
|-------------------|-----------|
| Mechanical safety | Identical |
| Energy source | Identical |
| Standards met | Identical |
| Electrical safety | Identical |
| EMC | Identical |
| Function | Similar |
121470 p3/3
### 8.0 Performance summary
KD-972 Fully Automatic Wireless Blood Pressure Wrist Monitor conforms to the following standards:
- · IEC 60601-1, Medical Electrical Equipment Part 1: General Requirements for Safety, 1988; Amendment 1, 1991-11, Amendment 2, 1995.
- · EN 60601-1-2, Medical Electrical Equipment Part 1-2: General Requirements for Safety - Collateral standard: Electromagnetic Compatibility - Requirements and Tests, 2007.
- · IEC 80601-2-30, Medical electrical equipment-Part 2-30:Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanmeters,2009.
### 9.0 Comparison to the predicate device and the conclusion
Our device KD-972 Fully Automatic Wireless Blood Pressure Wrist Monitor is substantially equivalent to the Fully Automatic Electronic Blood Pressure Monitor KD-7964 whose 510(k) number is K102906.
KD-972 and KD-7964 are very similar in the intended use, the design principle, the performance and the applicable standards. Only their appearance, the memory time, the average function are different. The measure process is also changed, that is the new device will get the measurement results when the device is inflating, while KD -7964 gets the result during the deflating period. The data transfer method of the new device KD-972 is changed to Bluetooth and the data displayed on iPhone, while KD-7964 transfer the data to PC while receive available command.
However, the test in this submission provides demonstration that these small differences do not raise any new questions of safety and effectiveness
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in all caps. The text is in a bold, sans-serif font. The text is black and the background is white.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JUN 1 4 2012
Andon Health Co., Ltd. c/o Ms. Liu Yi President No. 3 Jin Ping Street, Ya An Road, Nankai District Tianjin China 300190
Re: K121470
> Trade/Device Name: Fully Automatic Wireless Blood Pressure Wrist Monitor, KD-972 Regulatory Number: 21 CFR 870.1130 Regulation Name: Non-invasive Blood Pressure Measurement System Regulatory Class: II (two) Product Code: DXN Dated: May 14, 2012 Received: May 17, 2012
Dear Ms. Yi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be
{4}------------------------------------------------
Page 2 - Ms. Liu Yi
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's.Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Statement of Indications for Use
### 510(k) Number :
Device name:
KD-972 Fully Automatic Wireless Blood Pressure Wrist Monitor
#### Indications for use:
KD-972 Fully Automatic Wireless Blood Pressure Wrist Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to 13.5cm-22cm.
Prescription use Part 21 CFR 801 Subpart D)
Over-The-Counter Use YES AND/OR (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-COUNTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) (Division Sign-on)
Division of Cardiovascular Devices
Page I of
510(k) Number K121470
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.