K121444 · Canon, Inc. - Medical Equpment Group · OWB · Jun 7, 2012 · Radiology
Device Facts
Record ID
K121444
Device Name
FLAT PANEL DETECTOR
Applicant
Canon, Inc. - Medical Equpment Group
Product Code
OWB · Radiology
Decision Date
Jun 7, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1650
Device Class
Class 2
Indications for Use
The flat panel detector CSX-20 is designed to provide fluoroscopic and spot radiographic images of human anatomy during diagnostic, surgical and interventional procedures. Examples of clinical application may include angiography, endoscopy, urologic, orthopedic, neurologic; vascular, cardiac, critical-care and emergency room procedures or other imaging applications at the physician's discretion. The device is intended to replace spot-film devices. The device is also intended to replace fluoroscopic images obtained through image intensifier technology. Not intended for mammography applications.
Device Story
CSX-20 is a digital radiography flat panel detector component for X-ray systems; captures X-ray images and converts them into high-resolution digital signals. Used in clinical settings (OR, clinic, ER) by physicians/technicians during diagnostic, surgical, and interventional procedures. Replaces traditional spot-film devices and image intensifier technology. Features 16-bit A/D conversion (upgraded from 14-bit in predicate) to improve resolution, signal-to-noise ratio, and tonal precision; supports higher frame rates (up to 240 fps). Output is a digital image displayed for clinician review to support diagnostic and interventional decision-making. Benefits include improved image quality and workflow efficiency compared to legacy intensifier systems.
Clinical Evidence
Bench testing only. No clinical data provided. Evidence includes performance testing, software validation, electrical safety, and electromagnetic compatibility testing. Device complies with US Performance Standards for radiographic equipment.
Technological Characteristics
Digital flat panel X-ray detector. 16-bit A/D conversion. Grayscale depth: 16384 gradations. Max frame rate: 240 fps. Component of an X-ray system. Not intended for mammography.
Indications for Use
Indicated for human patients requiring fluoroscopic and spot radiographic imaging during diagnostic, surgical, and interventional procedures, including angiography, endoscopy, urologic, orthopedic, neurologic, vascular, cardiac, critical-care, and emergency room applications. Not intended for mammography.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
**Canon**
## 5. 510(k) SUMMARY
# JUN - 7 2012
K1214444
.
| 5. 510(k) SUMMARY | JUN - 7 20 | |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Submitter: | Canon, Inc. - Medical Equipment Group<br>30-2 Shimomaruko, 3-chrome<br>Ohta-ku, Tokyo 146-8501 Japan | |
| Contact Person: | Mr. Naoyasu Asaka<br>Staff Manager<br>TEL: 81-3-3758-2111<br>FAX: 81-3-5482-3960<br>asaka.naoyasu@canon.co.jp | |
| Date Prepared: | April 24, 2012 (Revised June 7, 2012) | |
| Trade Name: | CSX-20 Flat Panel Detector | |
| Common Name: | Solid State X-Ray Imager (Flat Panel/Digital Imager) | |
| Classification Name: | OWB, JAA 892.1650 Image-intensified fluoroscopic x-ray system | |
| Predicate Device: | K111824 Canon CSX-10 (MQB 892.1650) | |
| Device Description: | The Canon CSX-20 is a digital radiography flat panel detector that can take<br>fluoroscopic and spot radiographic images of any part of the body. It directly<br>converts the X-ray images captured by the sensor into high-resolution digital<br>images. The instrument is a component of an x-ray system and as such cannot be<br>used outside of such a system. This unit converts the X-rays into digital signals. Not<br>intended for mammography applications. | |
| Statement of<br>Intended Use: | The flat panel detector CSX-20 is designed to provide fluoroscopic and spot<br>radiographic images of human anatomy during diagnostic, surgical and<br>interventional procedures. Examples of clinical application may include<br>angiography, endoscopy, urologic, orthopedic, neurologic; vascular, cardiac,<br>critical-care and emergency room procedures or other imaging applications at the<br>physician's discretion. The device is intended to replace spot-film devices. The<br>device is also intended to replace fluoroscopic images obtained through image<br>intensifier technology. Not intended for mammography applications. | |
| Summary of<br>Technological<br>Characteristics: | Comparison with the predicate shows the technological characteristics of the<br>CSX-20 are substantially equivalent to the predicate device. The flat panel detector<br>units are functionally identical.<br>The A/D Conversion was modified from 14-bit precision to 16-bit precision to<br>improve the resolution and S/N. This modification allows for improved tonal<br>precision.<br>Grayscale is a function of the A/D conversion. As a result of the A/D conversion<br>modification, the grayscale gradations have quadrupled from the 4096 of the CSX-<br>10 to the 16384 gradations now available with the CSX-20.<br>By modifying the A/D conversion to 16-bits, a maximum frame rate of the 240 fps<br>has been improved.<br>The modifications described above are all related to the change in A/D conversion<br>from 14-bit to 16-bit precision. These modifications have been incorporated into the<br>CSX-20 in an effort to improve the product performance. | |
. . . . .
{1}------------------------------------------------
## Canon
### Summary of Non-Clinical /Test Data:
Tests were performed on the device which demonstrated that the device is safe and effective, performs comparably to and is substantially equivalent to the predicate device. Tests include: Performance testing and Software Validation. Electrical safety and Electromagnetic Compatibility testing has been performed. The unit complies with the US Performance Standard for radiographic equipment. .
Conclusion:
Canon, Inc. - Medical Equipment Group considers the CSX-20 to be substantially equivalent to the predicate device listed above. This conclusion is based on the similarities in primary intended use, principles of operation, functional design, and established medical use for the flat panel detector.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. In the center of the circle is a stylized symbol that resembles a caduceus, a traditional symbol of medicine.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
JUN - 7 2012
Canon, Inc. - Medical Equipment Group % Ms. Diane Rutherford Regulatory Engineer Ken Block Consulting 1201 Richardson Drive, Suite 280 RICHARDSON TX 75080
Re: K121444
Trade/Device Name: CSX-20 Flat Panel Detector Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB and JAA Dated: April 24, 2012 Received: May 15, 2012
#### Dear Ms. Rutherford:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA): You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{3}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### INDICATIONS FOR USE
#### 510(k) Number:
Device Name: CSX-20 Flat Panel Detector
#### Indications for Use:
The flat panel detector CSX-20 is designed to provide fluoroscopic and spot radiographic images of human anatomy during diagnostic, surgical and interventional procedures. Examples of clinical application may include angiography, endoscopy, urologic, orthopedic, neurologic, vascular, cardiac, critical-care and emergency room procedures or other-imaging applications at the physician's discretion. The device is intended to replace the spot-film devices. The device is also intended to replace fluoroscopic images obtained through image intensifier technology. Not intended for mammography applications.
Prescription Use _ X (21 CFR 801 Subpart D)
Over-the-Counter Use AND/OR (21. CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDHR, Office of Device Evaluation (ODE)
Jahmal DD
(Division Sign-Off)
(Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
510KLA21444
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.