NOVATION CROWN CUP LINERS; NEUTRAL, LIPPED, +5MM LATERALIZED, +5/10 DEGREE
K121392 · Exactech, Inc. · LZO · Jan 16, 2013 · Orthopedic
Device Facts
Record ID
K121392
Device Name
NOVATION CROWN CUP LINERS; NEUTRAL, LIPPED, +5MM LATERALIZED, +5/10 DEGREE
Applicant
Exactech, Inc.
Product Code
LZO · Orthopedic
Decision Date
Jan 16, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3353
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
All Exactech Hip Systems are indicated for use in skeletally mature individuals undergoing primary surgery for hip replacement due to osteoarthritis, rheumatoid arthritis, osteonecrosis, post-traumatic degenerative problems of the hip, and for treatment of proximal femoral fractures where prosthetic replacement is determined by the surgeon as the preferred treatment. Components of Exactech Hip Systems are also potentially indicated for ankylosing spondylitis, congenital hip dysplasia. revision of failed previous reconstructions where sufficient bone stock is present, and to restore mobility resulting from previous fusion. Cemented femoral stems and cemented acetabular cups are intended for . cemented fixation only. Press-fit femoral stems and acetabular cups are intended for press-fit . fixation. . Femoral heads and endoprostheses are intended for use in cemented and press-fit applications.
Device Story
The device consists of modified acetabular liners and femoral heads for total hip arthroplasty. The liners feature a minimum polyethylene thickness of 4mm to accommodate 32mm, 36mm, and 40mm femoral heads; the femoral heads include a new 40mm diameter option. These components are used by orthopedic surgeons in clinical settings to replace damaged hip joints. The device functions as a mechanical bearing surface within a hip prosthesis system. The modification aims to provide additional sizing options for surgeons to better match patient anatomy, potentially improving joint stability and range of motion. The device is implanted during surgery; clinical decision-making is based on the surgeon's assessment of bone stock and patient pathology.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing and engineering analyses, including wear, fatigue, impingement, and mechanical strength testing (ASTM and ISO standards).
Technological Characteristics
Components include UHMWPE acetabular liners, Cobalt-Chromium (CoCr) femoral heads, and Bioloxdelta/Biolox Option ceramic femoral heads. Design features include 4mm minimum polyethylene thickness and 40mm head diameters. Mechanical testing performed per ASTM F1714 (wear), ASTM F1820 (push-out), and ISO 7206-10 (fatigue/burst).
Indications for Use
Indicated for skeletally mature individuals undergoing primary hip replacement or revision due to osteoarthritis, rheumatoid arthritis, osteonecrosis, post-traumatic degeneration, proximal femoral fractures, ankylosing spondylitis, or congenital hip dysplasia; intended to restore mobility in patients with previous fusion.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
Exactech Bioloxdelta® and Biolox Option Femoral Heads and Adapters (K103012)
Submission Summary (Full Text)
{0}------------------------------------------------
## Exactech® Femoral Heads and Novation® Crown Cup® Acetabular Liners Special 510(k) - 510(k) Summary of Safety and Effectiveness
Sponsor: Exactech. Inc. 2320 N.W. 66th Court Gainesville, FL 32653
> Phone: (352) 377-1140 (352) 378-2617 Fax:
FDA Establishment Number 1038671
Contact: Nicole Kassner QA/RA Compliance, Project Manager
Date: January 11, 2013
## Trade of Proprietary or Model Name(s):
Exactech® Novation® Crown Cup® GXL UHMWPE Liner Exactech 12/14 CoCr Femoral Head Exactech 12/14 Bioloxdelta® Femoral Head Exactech 12/14 Biolox Option Femoral Head Exactech Novation Crown Cup 40mm Liner Trials Exactech 12/14 40mm Head Trials
#### Common Name:
Total Hip Arthroplasty Prosthesis - Acetabular Liners and Femoral Heads
#### Classification Name:
Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented (CFR 888.3353, Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis, Class II, Product Code LZO)
Prosthesis, hip, semi-constrained, metal/polymer, cemented (CFR 888.3350. Hip joint metal/polymer semi-constrained cemented, Class II, Product Code JDI)
Prosthesis, hip, semi-constrained, metal/polymer, uncemented (CFR 888.3360, Hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis, Class II, Product Code LWJ)
Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calciumphosphate (CFR 888.3353, Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis, Class II, Product Code MEH)
Prosthesis, hip, semi-constrained, metal/polymer, porous uncemented (CFR 888.3358, Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis, Class II, Product Code LPH)
## JAN 1 6 2013
ાજિ
K121392
{1}------------------------------------------------
## Exactech® Femoral Heads and Novation® Crown Cup® Acetabular Liners Special 510(k) - 510(k) Summary of Safety and Effectiveness
#### Information on devices to which substantial equivalence is claimed:
| 510(k) Number | Trade of Proprietary Model Name | Manufacturer |
|---------------|-----------------------------------------------------------------------|----------------|
| K070479 | Exactech Novation® Crown Cup® Liners | Exactech, Inc. |
| K041906 | Exactech 12/14 CoCr Femoral Heads | Exactech, Inc. |
| K103012 | Exactech Bioloxdelta® and Biolox Option<br>Femoral Heads and Adapters | Exactech, Inc. |
#### Indications for Use:
All Exactech Hip Systems are indicated for use in skeletally mature individuals undergoing primary surgery for hip replacement due to osteoarthritis, rheumatoid arthritis, osteonecrosis, post-traumatic degenerative problems of the hip, and for treatment of proximal femoral fractures where prosthetic replacement is determined by the surgeon as the preferred treatment. Components of Exactech Hip Systems are also potentially indicated for ankylosing spondylitis, congenital hip dysplasia. revision of failed previous reconstructions where sufficient bone stock is present, and to restore mobility resulting from previous fusion.
- Cemented femoral stems and cemented acetabular cups are intended for . cemented fixation only.
- Press-fit femoral stems and acetabular cups are intended for press-fit . fixation.
- . Femoral heads and endoprostheses are intended for use in cemented and press-fit applications.
#### Device Description:
The proposed Novation Crown Cup Acetabular Liners and associated Liner Trials are a modification to the Novation Crown Cup Acetabular Liners and associated Liner Trials cleared through premarket notification #K070479.
The proposed 12/14 Cobalt Chromium, Bioloxdelta and Biolox Option Femoral Heads and associated Head Trials are a modification to the 12/14 Cobalt Chromium, Bioloxdelta and Biolox Option Femoral Heads and associated Head Trials cleared through premarket notifications #K041906 and #K103012.
The predicate and proposed devices have the same intended use and basic fundamental scientific technology.
The modified devices share the following similarities with the predicate devices:
- . Indications for use
- . Design features
- Material ●
- Shelf life
{2}------------------------------------------------
## Exactech Femoral Heads and Novation® Crown Cup® Acetabular Liners Special 510(k) - 510(k) Summary of Safety and Effectiveness
<121392
3/3
- . Packaging and sterilization materials and processes
This submission proposes the following design change:
- The proposed acetabular liners have a minimum polyethylene thickness of 4mm . for compatibility with 32mm, 36mm and 40mm femoral heads.
- The proposed femoral heads have an outer diameter of 40mm. .
#### Substantial Equivalence Conclusion:
The following engineering analyses were conducted to demonstrate substantial equivalence of the proposed devices to the predicate devices:
- Wear testing per ASTM F1714 to determine that the proposed devices have ● equivalent or better wear results than the predicate devices.
- . Lever-out testing per Tradonsky, et al. 1993, Axial push-out per ASTM F1820, Torque testing was conducted to determine that the proposed acetabular liner devices are substantially equivalent to the predicate devices.
- . Impingement fatigue testing was conducted to determine that the proposed acetabular liner devices are substantially equivalent to the predicate devices.
- . Engineering justification to support the applicability of the predicate device testing to the proposed devices:
- Burst Testing per ISO 7206-10
- 트 Fatigue per ISO 7206-10
- 트 Post Fatigue per ISO 7206-10
- 트 Pull-Off per Ceramtec specifications, FDA Guidance and ASTM F2009
- Torque per Ceramtec specifications, FDA Guidance 이
- D Range of Motion per ISO 21535
The results of engineering analyses demonstrate the proposed devices are substantially equivalent to the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling a stylized caduceus or a representation of the human form.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 16, 2013
Exactech, Incorporated % Ms. Nicole Kassner Quality and Regulatory Compliance Project Manager 2320 Northwest 66th Court Gainesville, Florida 32653
Re: K121392
Trade/Device Name: Exactech® Femoral Heads and Novation® Crown Cup® Acetabular Liners
Regulation Number: 21 CFR 888.3353
Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis
Regulatory Class: Class II Product Code: LZO, JDI, LWJ, MEH, LPH Dated: December 21. 2012 Received: December 26, 2012
Dear Ms. Kassner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{4}------------------------------------------------
#### Page 2 - Ms. Nicole Kassner
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
# Erin Dieith
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Exactech® Femoral Heads and Novation® Crown Cup® Acetabular Liners Special 510(k) - Indications for Use
510(k) Number: _
Device Name: Exactech® Femoral Heads and Novation® Crown Cup® Acetabular Liners
## INDICATIONS
All Exactech Hip Systems are indicated for use in skeletally mature individuals undergoing primary surgery for hip replacement due to osteoarthritis, rheumatoid arthritis, osteonecrosis, post-traumatic degenerative problems of the hip, and for treatment of proximal fractures where prosthetic replacement is determined by the surgeon as the preferred treatment. Components of Exactech Hip Systems are also potentially indicated for ankylosing spondylitis, congenital hip dysplasia, revision of failed previous reconstructions where sufficient bone stock is present, and to restore mobility resulting from previous fusion.
- Cemented femoral stems and cemented acetabular cups are intended for . cemented fixation only.
- Press-fit femoral stems and acetabular cups are intended for press-fit fixation. .
- Femoral heads and endoprostheses are intended for use in cemented and � press-fit applications.
Prescription Use_ X (Part 21 CFR 801 Subpart D) and/or
Over-The-Counter Use_ (21 CFR 807 Subpart C)
Please do not write below this line - use another page if needed.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Krishna Asundi, PhD Division of Orthopedic Devices
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.