← Product Code [MBI](/productcode/MBI) · K121356

# MICRO/MINI N-PK(H) BONE ANCHOR (K121356)

_Anulex Technologies, Inc. · MBI · Nov 9, 2012 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K121356

## Device Facts

- **Applicant:** Anulex Technologies, Inc.
- **Product Code:** [MBI](/productcode/MBI.md)
- **Decision Date:** Nov 9, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Micro/Mini N-PK Bone Anchor is intended for the reattachment of soft tissue to bone for the following indications: Elbow, Hand and Wrist Biceps tendon reattachment Ulnar or lateral collateral ligament reconstruction . Scapholunate ligament reconstruction . Lateral epicondylitis repair

## Device Story

Micro/Mini N-PK(H) Bone Anchor is a two-part orthopedic fixation system comprising a nitinol sleeve and a PEEK core. Device includes an adjustable loop of non-absorbable suture. Nitinol sleeve expands upon delivery to provide pullout resistance. Provided sterile and preloaded on a disposable delivery instrument. Used by surgeons in orthopedic procedures to reattach soft tissue to bone in the elbow, hand, and wrist. Output is mechanical fixation of soft tissue to bone, facilitating healing and structural repair.

## Clinical Evidence

Bench testing only. Biocompatibility testing per ISO 10993-1; corrosion testing per ASTM F2129; tensile testing per USP suture requirements; fixation/static disassembly strength and cyclic fatigue performance testing conducted.

## Technological Characteristics

Materials: Polyetheretheretherketone (PEEK) and nickel-titanium (nitinol). Sensing/Actuation: Mechanical expansion of nitinol sleeve. Form factor: Bone anchor with adjustable non-absorbable suture loop. Sterilization: Sterile. Standards: ISO 10993-1 (biocompatibility), ASTM F2129 (corrosion).

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Bioraptor PK Suture Anchor ([K071586](/device/K071586.md))
- DePuy Mitek Mini QuickAnchor Plus ([K071257](/device/K071257.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

510(k) Summary
K121356

Page 1/2

.

As Required by 21 CFR 807.92

|                      | NOV 9 2012                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                      | Anulex Technologies, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Submitter:           | 5600 Rowland Road, Suite 280                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                      | Minnetonka, MN 55343                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Contact Person:      | Rachel Kennedy                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|                      | Director, Regulatory Affairs and Quality Systems                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                      | Anulex Technologies, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                      | Direct: 952.224.4034<br>Fax: 952.224.5417<br>E-Mail: rkennedy@anulex.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Date Prepared:       | October 25, 2012                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Trade Name:          | Micro/Mini N-PK(H) Bone Anchor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Classification:      | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Product Code:        | MBI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|                      | 21 CFR 888.3040                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Predicate Device(s): | The subject device is substantially equivalent to the<br>following predicate devices:                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|                      | Smith & Nephew, Inc., Bioraptor PK Suture Anchor<br>(K071586 cleared August 17, 2007) DePuy Mitek Mini QuickAnchor Plus (K071257<br>cleared June 29, 2007)                                                                                                                                                                                                                                                                                                                                                                                           |
| Device Description:  | The Micro/Mini N-PK(H) Bone Anchor consists of an<br>adjustable loop of non-absorbable suture (Sizes 2, 0 and 2-<br>0) with one (1) or two (2) attached anchors. The bone<br>anchor is a two part system, one part nickel-titanium<br>(nitinol) and the second part polyetheretherketone (PEEK).<br>The PEEK part resides within the nitinol sleeve and retains<br>the suture. The nitinol part expands upon delivery<br>providing resistance to pullout.<br>The construct is provided sterile and preloaded on a<br>disposable delivery instrument. |
| Indications for Use: | The Micro/Mini N-PK Bone Anchor is intended for the<br>reattachment of soft tissue to bone for the following<br>indications:<br>Elbow, Hand and Wrist  Biceps tendon reattachment Ulnar or lateral collateral ligament                                                                                                                                                                                                                                                                                                                               |

Anulex Technologies, Inc.

.

# K121356A2

{1}------------------------------------------------

Page 2/2

#### reconstruction

●

- Scapholunate ligament reconstruction
- Lateral epicondylitis repair ●

## Functional and Safety Testing:

#### Comparison to Predicate:

#### Conclusion:

Biocompatibility testing was conducted in accordance with ISO 10993-1 standards. Corrosion testing was conducted in accordance with ASTM F2129. Tensile testing was performed to verify compliance with USP suture requirements and comparison testing to support the safety and effectiveness of the Micro/Mini N-PK(H) Bone Anchor included evaluation of fixation/static disassembly strength and cyclic fatigue performance.

The Micro/Mini N-PK(H) Bone Anchor that is the subject of this submission has the same materials and similar technological characteristics in comparison to the predicate suture anchors. The intended use is the same as the predicates, fixation of soft tissue to bone for a variety of orthopedic procedures in the elbow, hand and wrist while the predicate anchors include procedures in additional anatomical locations.

Substantial equivalence is demonstrated through the detailed device description, performance testing and conformance with voluntary performance standards.

Anulex Technologies, Inc.

#### K121356A2

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

#### Letter Dated: November 9, 2012

Anulex Technologies, Incorporated % Ms. Rachel Kennedy Director, Regulatory Affairs and Quality Assurance 5600 Rowland Road, Suite 280 Minnetonka, Minnesota 55343

Re: K121356

Trade/Device Name: Micro/Mini N-(PK)(H) Bone Anchor Regulation Number: 21 CFR 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener. Regulatory Class: Class II Product Code: MBI Dated: October 26, 2012 Received: October 31, 2012

Dear Ms. Kennedy:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Ms. Rachel Kennedy

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address · http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Mark N. Melkerson

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Appendix 4: Indications for Use Form

KI2i356

Device Name: Micro/Mini N-PK(H) Bone Anchor

#### Indications for Use:

The Micro/Mini N-PK Bone Anchor is intended for the reattachment of soft tissue to bone for the following indications:

#### Elbow, Hand and Wrist

- Biceps tendon reattachment
- Ulnar or lateral collateral ligament reconstruction .
- ◆ Scapholunate ligament reconstruction
- . Lateral epicondylitis repair

Prescription Use _________X (Part 21 CFR 801 Subpart D AND/OR

Over-The-Counter Use _ (21 CFR 801 Subpart C)

### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

ASh

(Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K171356

Micro/Mini N-PK(H) Bone Anchor Anulex Technologies, Inc. Confidential

Traditional 510(k) Premarket Notification

Page 52

---

**Source:** [https://fda-staging.innolitics.com/device/K121356](https://fda-staging.innolitics.com/device/K121356)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com