K121327 · Biolase Technology, Inc. · GEX · Feb 1, 2013 · General, Plastic Surgery
Device Facts
Record ID
K121327
Device Name
DIOLASE 10S
Applicant
Biolase Technology, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Feb 1, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Hemostasis, incision, excision, ablation, coagulation, and vaporization of tissues from the ear, nose, throat and adjacent areas, including soft tissue in the oral cavity; joint tissues during arthroscopic surgery; tissue in the upper and lower gastrointestinal tracts via endoscopy; soft tissue, abdominal, rectal, skin, fat or muscle tissue, and dermabrasion; abdominal and rectal tissues; tissue in gynecology; tissue in neurosurgery; tissue in ophthalmology; tissue in pulmonary surgery; cardiac tissue; thoracic surgery; and tissues in urology. Photocoagulation of vascular & dermatological lesions of the face and extremities; telangiectasia, venulectasia of the legs and face; reticular veins and branch varicosities; pyrogenic granuloma, lymphangioma and lymphangiomatosis disease, angiofibromas; superficial benign vascular lesions; dermatological surgery; seborrheic keratosis; mixoid cyst; papillary varix; and acne treatment.
Device Story
Diolase™ 10S is an InGaAsP solid-state diode laser system; emits infrared laser energy for surgical tissue interaction. System comprises console with touch screen/keypad and delivery system (fiber optic cable and surgical handpiece). Visible aiming beam identifies treatment site. Operated by clinicians in various surgical settings (OR, clinic). Laser energy performs hemostasis, incision, excision, ablation, coagulation, vaporization, and photocoagulation. Clinical utility spans broad surgical specialties; enables precise tissue management; reduces bleeding; facilitates minimally invasive procedures. Clinician directs energy to target site; visual feedback via aiming beam; clinical decision-making based on patient pathology and surgical requirements.
Clinical Evidence
No clinical data presented. Substantial equivalence supported by comparison to previously cleared devices and demonstration of identical technological characteristics.
Technological Characteristics
InGaAsP solid-state laser diode; infrared energy output; fiber optic delivery; console with touch screen/keypad interface; visible aiming beam; Class II device; 21 CFR 878.4810.
Indications for Use
Indicated for patients requiring surgical intervention across multiple specialties including ENT, oral surgery, arthroscopy, gastroenterology, general/plastic/podiatric surgery, GI/GU, gynecology, neurosurgery, ophthalmology, pulmonary, cardiac, thoracic, urology, and dermatology/aesthetics. Contraindicated for patients with specific medical conditions or medications that may contraindicate use of laser sources, including local anesthetic allergies, heart/lung disease, malignancies, bleeding disorders, sleep apnea, or immune deficiency.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
K.121327
Biolase Technology, Inc.
Diolase™ Expanded Indication for Use
# FEB 0 1 2013
# 510(k) Summary for Diolase™ 10S by Biolase Technology, Inc. (As required by 21CFR 807.92)
# 1. GENERAL INFORMATION
Date Prepared:
January 30, 2013
Company:
Biolase Technology, Inc. 4 Cromwell Irvine, CA 92618 Tel: (949) 361-1200 Fax: (949) 273-6687
Submitter:
Marcia VanValen Director of Business Development Tel: (949) 226-8159 e-mail: mvanvalen@biolase.net
Contact:
Colleen Boswell VP of Regulatory Affairs Tel: (949) 226-8470 e-mail: cboswell@biolase.net
### 2. NAMES / REGULATIONS
| Trade/Device Name: | Diolase™ 10S |
|----------------------|----------------------------------------------------------------------------------------|
| Common Name: | Diode Laser |
| Regulation Number: | 21CFR 878.4810 |
| Classification name: | Laser surgical instrument for use in general and plastic<br>surgery and in dermatology |
| Regulatory Class: | II |
| Product Code: | GEX |
{1}------------------------------------------------
# 3 ... PREDICATE DEVICES
ﯾ ﺩﺭ ﺍﯾ
- (K100558) Quanta System QUANTA Diode Laser Family by Quanta System . SpA
- (K110375) Blueshine GOLD Series by Blueshine srl .
- Reference Device (K061898) ezlase™ by Biolase Technology, Inc. .
- Reference Device (K083069) ezlase™ 10W by Biolase Technology, Inc. .
## 4. DEVICE DESCRIPTION
The Diolase™ 10S system uses an Indium Gallium Arsenide Phosphorous (InGaAsP) solid state laser diode to emit infrared laser energy which is transmitted via a flexible fiber optic cable to a hand piece that emits the energy to the target site. A visible light is emitted at the same time to visually identify the treatment location. The Diolase™ 10S laser consists of two permanently connected components: Console, and Delivery System. The Console has a Control Panel (Touch Screen and Keypad) in front and a detachable base attached at the bottom rear of the Console. The Diolase™10S Laser Delivery System consists of the following: Fiber Optic Assembly and Surgical Handpiece
### 5. INDICATIONS FOR USE
### Expanded Indications for Use:
### Ear, Nose and Throat and Oral Surgery:
Hemostasis, incision, excision, ablation, coagulation, and vaporization of tissues from the ear, nose, throat and adjacent areas, including soft tissue in the oral cavity, such as:
- Removal of benign lesions from ear, nose and throat .
- Excision and vaporization of vocal cord nodules and polyps
- Incision and excision of carcinoma in-situ .
- Ablation and vaporization of hyperkeratosis .
- Laryngeal papillomectomy .
- Excision and vaporization of herpes simplex 1 and II
- Neck dissection
### Arthroscopy:
Hemostasis, incision, excision, coagulation, vaporization, and ablation of joint tissues during arthroscopic surgery, such as:
- Menisectomy .
- Syovectomy .
- Chondromalacia
{2}------------------------------------------------
### Gastroenterology:
Hemostasis, incision, excision, ablation, coagulation, and vaporization of tissue in the upper and lower gastrointestinal tracts via endoscopy, such as:
- Hemostasis of upper and lower GI bleeding .
- Excision and vaporization of colorectal carcinoma .
- Excision of polyps .
- Hemostasis of colonoscopy ●
- Hemostasis of esophageal varices ●
### Orthopedics
- . Dissect and coagulate
### General Surgery, Dematology & Plastic Surgery, and Podiatry:
Excision, ablation, vaporization, and photocoagulation of skin lesions, hemostasis, incision, excision, vaporization, ablation, and debulking of soft tissue, abdominal, rectal, skin, fat or muscle tissue, and dermabrasion, such as:
- Matrixectomy .
- Excision of neuromas .
- Excision of periungual and subungual warts ●
- Excision of plantar warts .
- Excision of Keloids
- Excision of cutaneous lesions ●
- Hemorrhoidectomy ●
- Appendectomy .
- Debridement of decubitus ulcer ●
- Dermabrasion .
- Vaporization & hemostasis of capillary hemangioma �
- Excision, vaporization & hemostasis of abdominal tumors .
- Excision, vaporization & hemostasis of rectal pathology ●
- Pilonidal cystectomy ●
- Herniorraphy ●
- Adhesiolysis .
- Parathyroidectomy �
- Laparoscopic cholecystecomy
- Thyroidectomy
- Resection of organs
### GI/GU:
Excision, vaporization, and hemostasis of abdominal and rectal tissues, such as:
- Hemorrhoidectomy .
- . Excision, vaporization, and hemostasis of rectal pathology
- Excision, vaporization, and hemostasis of abdominal tumors .
{3}------------------------------------------------
Biolase Technology, Inc.
Diolase™ 10S Expanded Indication for Use
# Gynecology:
CP-DEN 123
Ablation, excision, incision, coagulation, hemostasis, and vaporization of tissue, such as:
- Excision or vaporization of condylomata acuminata .
- Vaporization of CIN (cervical intraepithelial neoplasia) ●
- Cervical conization
- Menorrhagia .
- Myomectomy
- Ovarian cystectomy
# Neurosurgerv:
Vaporization, coagulation, excision, incision, ablation and hemostasis of tissue, such as:
- Hemostasis in conjunction with meningiomas .
- . Percutaneous Disc Decompression (PLDD)
### Ophthalmology:
- Dacryocystorhinostomy transcanalicular �
- Open DCR .
- Tumor Excision
- Blepharoplasty
## Pulmonary Surgery:
Hemostasis, vaporization, coagulation, incision, ablation, and excision of tissue, such as:
- Tracheobronchial malignancy or stricture .
- Benign and malignant pulmonary obstruction .
### Cardiac Surgery:
- . Coagulation and hemostasis of cardiac tissue
### Thoracic Surgery:
- . Thoracotomy
- Pulmonary resection .
- Hemostasis t
- Pericardiectomy ●
- Adhesiolysis
- Coagulation of blebs and bullae .
### Urology:
Hemostasis, vaporization, incision, coagulation, ablation, and excision of tissues, such as:
- Vaporization of urethral tumors ●
- . Release of urethral stricture
- Removal of bladder neck obstruction .
- Excision and vaporization of condyloma ●
- Lesions of external genitalia
- Circumcision
- Vaporization of the prostate to treat benign prostate hyperplasia (BPH)
{4}------------------------------------------------
# Dermatology/Aesthetics:
- Photocoagulation of vascular & dermatological lesions of the face and . extremities
- Photocoagulation of telangiectasia, venulectasia of the legs and face .
- Treatment of reticular veins and branch varicosities .
- . Pyrogenic granuloma, lymphangioma and lymphangiomatosis disease, angiofibromas
- . Superficial benign vascular lesions including Telangiectasias, hemangioma, Port wine stains, angiokeratoma, and benign epidermal pigment lesions as lentigines. Epidermal nevi, spider nevi.
- Dermatological surgery: Condyloma acuminate, warts, small non-malignant . skin tumors, small semi-malignant tumors as basalomas, Bowe, Kaposi
- sarcoma. Warty leucoplasty and ulcers debridement.
- Seborrheic keratosis ●
- Mixoid cyst �
- Papillary varix .
- Acne treatment ●
## Vascular Surgery:
- Photocoagulation of vascular & dermatological lesions of the face and ● extremities
- Photocoagulation of telangiectasia, venulectasia of the legs and face .
- . Treatment of reticular veins and branch varicosities
# 6. DEVICE TECHNOLOGICAL CHARACTERISTICS
The Diolase "10 10S system has the same technological characteristics (i.e., design, material, chemical composition, energy source) as the predicate devices; a summary of the technological characteristics of this device in comparison to those of the predicate devices is included in the body of the 510(k) submission.
# 7. PERFORMANCE ASSESSMENT
Non-clinical performance data is not presented. Comparison with previously cleared devices is included in the body of the 510(k) submission for the demonstration of safety and effectiveness of the new indication of this device and to support substantial equivalence to legally marketed devices.
# 8. CONTRAINDICATIONS
The contraindications are identical to that of the previously cleared extase™ system by Biolase Technology, Inc.
{5}------------------------------------------------
Biolase Technology, Inc. Diolase™ 10S Expanded Indication for Use
All clinical procedures performed with Diolase™ 10S must be subjected to the same clinical judgment and care as with traditional techniques. Patient risk must always be considered and fully understood before clinical treatment. The clinician must completely understand the patient's medical history prior to treatment. Exercise caution for general medical conditions that might contraindicate a local procedure. Such conditions may include allergy to local or topical anesthetics, heart disease, lung disease, malignancies, bleeding disorders, sleep apnea, immune system deficiency, or any medical conditions or medications that may contraindicate use of certain light/laser type sources associated with this device. Medical clearance from the patient's physician is advisable when doubt exists regarding treatment.
### 9. SUBSTANTIAL EQUIVALENCE
The purpose of this 510(k) submission is to add indications for use to the previously cleared ezlase™ diode laser system and market it as Diolase™ 10S. The requested additional indications have been cleared by FDA for several equivalent medical devices, including the following: K100558 (Quanta System QUANTA Diode Laser Family by Quanta System SpA) and K110375 (Blueshine GOLD Series by Blueshine srl). Based on comparison, the Diolase™ 10S system is substantially equivalent to previously cleared devices.
### 10. CONCLUSION
The indications requested by this 510(k) submission is the same as those previously cleared by FDA for other predicate devices. Substantial Equivalence for the Diolase™ 10S system has been determined through comparison to previous cleared devices. This summary demonstrates that the Diolase™ 10S system is as safe, as effective, and will perform as well as the predicate devices.
{6}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized eagle with three tail feathers, representing the three levels of government: federal, state, and local. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Biolase Technology, Incorporated % Ms. Marcia Van Valen Director of Business Development 4 Cromwell Irvine, California 92618-1816
February 1, 2013
Re: K121327
Trade/Device Name: Diolase 10S Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: January 02, 2013
Received: January 25, 2013
Dear Ms. Valen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{7}------------------------------------------------
Page 2 - Ms. Marcia Van Valen
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
For
# Peter D.,Rumm -S
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Indications for Use
# 510(k) Number: K121327
Device Name: Diolase 10S
### Indications for Use:
### Ear, Nose and Throat and Oral Surgery:
Hemostasis, incision, excision, ablation, and vaporization of tissues from the ear, nose, throat and adjacent areas, including soft tissue in the oral cavity, such as:
- . Removal of benign lesions from ear, nose and throat
- Excision and vaporization of vocal cord nodules and polyps .
- Incision and excision of carcinoma in-situ
- Ablation and vaporization of hyperkeratosis .
- Laryngeal papillectomy ◆
- Excision and vaporization of herpes simplex I and II
- Neck dissection
### Arthroscopy:
Hemostasis, incision, excision, vaporization, and ablation of joint tissues during arthroscopic surgery, such as:
- Meniscectomy .
- Synovectomy ●
- Chondromalacia
### Gastroenterology:
Hemostasis, incision, excision, and vaporization of tissue in the upper and lower gastrointestinal tracts via endoscopy, such as:
- Hemostasis of upper and lower GI bleeding •
- . Excision and vaporization of colorectal carcinoma
- ◆ Excision of polyps
- Hemostasis of colonoscopy .
- Hemostasis of esophageal varices
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-the-Counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
### Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil R Ogden 2013.02.01 10:17:59 -05'00'
(Division Sign-Off) Division of Surgical Devices 510(k) Number K121327
{9}------------------------------------------------
### Orthopedics
● Dissect and coagulate
# General Surgery, Dematology & Plastic Surgery, and Podiatry:
Excision, ablation, vaporization, and photocoagulation of skin lesions, hemostasis, incision, vaporization, ablation, and debulking of soft tissue, abdominal, rectal, skin, fat or muscle tissue, and dermabrasion, such as:
- Matrixectomy .
- Excision of neuromas .
- . Excision of periungual and subungual warts
- . Excision of plantar warts
- . Excision of Keloids
- . Excision of cutaneous lesions
- . Hemorrhoidectomy
- Appendectomy .
- . Debridement of decubitus ulcer
- ◆ Hepatobiliary
- . Mastectomy
- . Dermabrasion
- . Vaporization & hemostasis of capillary hemangioma
- . Excision, vaporization & hemostasis of abdominal tumors
- Excision, vaporization & hemostasis of rectal pathology
- . Pilonidal cystectomy
- ◆ Herniorrhaphy
- . Adhesiolysis
- . Parathyroidectomy
- . Laparoscopic cholecystecomy
- Thyroidectomy
- Resection of organs
Prescription Use - X AND/OR Over-the-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil R Ogden 2013.02.01 ,10:19:05 -05'00'
(Division Sign-Off) Division of Surgical Devices 510(k) Number K121327
{10}------------------------------------------------
### GI/GU:
Excision, vaporization, and hemostasis of abdominal and rectal tissues. such as:
- . Hemorrhoidectomy
- Excision, vaporization, and hemostasis of rectal pathology .
- Excision, vaporization, and hemostasis of abdominal tumors .
### Gynecology:
Ablation, excision, hemostasis, and vaporization of tissue, such as:
- . Excision or vaporization of condylomata acuminata
- . Vaporization of CIN (cervical intraepithelial neoplasia)
- . Cervical conization
- Menorrhagia .
- Ovarian cystectomy
### Neurosurgerv:
Vaporization, coagulation, excision, incision, ablation and hemostasis of tissue, such as:
- Hemostasis in conjunction with meningiomas .
- . Percutaneous Disc Decompression (PLDD)
### Ophthalmology:
- Dacryocystorhinostomy transcanalicular .
- . Open DCR
- Tumor Excision .
- Blepharoplasty .
# Pulmonary Surgerv:
Hemostasis, vaporization, and excision of tissue, such as:
- . Tracheobronchial malignancy or stricture
- Benign and malignant pulmonary obstruction .
### Cardiac Surgery:
- . Coagulation and hemostasis of cardiac tissue
Prescription Use X AND/OR Over-the-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
### Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil R Ogden 2013.02.01 10:19:56 -05'00'
(Division Sign-Off) Division of Surgical Devices 510(k) Number K121327
{11}------------------------------------------------
# Thoracic Surgery:
- . Thoracotomy
- Pulmonary resection .
- . Hemostasis
- Pericardiectomy .
- . Adhesiolysis
- Coagulation of blebs and bullae .
# Urology:
Hemostasis, vaporization, incision, coagulation, and excision of tissues, such as:
- . Vaporization of urethral tumors
- . Release of urethral stricture
- . Removal of bladder neck obstruction
- . Excision and vaporization of condyloma
- Lesions of external genitalia .
- Circumcision .
- Vaporization of the prostate to treat benign prostate hyperplasia (BPH) .
# Dermatology/Aesthetics:
- Photocoagulation of vascular & dermatological lesions of the face and extremities .
- Photocoagulation of telangiectasia, venulectasia of the legs and face ◆
- . Treatment of reticular veins and branch varicosities
- Pyrogenic granuloma, lymphangioma and lymphangiomatosis disease, . angiofibromas
- Superficial benign vascular lesions including Telangiectasias, hemangioma, Port . wine stains, angiokeratoma, and benign epidermal pigment lesions as lentigines. Epidermal nevi, spider nevi.
- . Dermatological surgery: Condyloma acuminate, warts, small non-malignant skin tumors, small semi-malignant tumors as basaliomas, Bowe, Kaposi sarcoma. Warty leucoplasty and ulcers debridement.
- Seborrheic keratosis .
- . Mixoid cyst
- Papillary varix .
- . Acne treatment
| Prescription Use | X | AND/OR | Over-the-Counter Use |
|-----------------------------|---|--------|------------------------|
| (Part 21 CFR 801 Subpart D) | | | (21 CFR 801 Subpart C) |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil R Ogden 2013.02.01 10:20:55 -05'00'
(Division Sign-Off) Division of Surgical Devices 510(k) Number _ K121327
{12}------------------------------------------------
# Vascular Surgery:
- Photocoagulation of vascular & dermatological lesions of the face and extremities •
- Photocoagulation of telangiectasia, venulectasia of the legs and face .
- Treatment of reticular veins and branch varicosities .
Prescription Use X Over-the-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil R Ogden 2013.02.01 ,10:22:04 -05'00'
(Division Sign-Off) Division of Surgical Devices 510(k) Number _ K121327
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.