Mesothelium Dental Membrane is indicated for: simultaneous use of GBR-membrane and implants. augmentation around implants placed in immediate extraction sockets. augmentation around implants placed in delayed extraction sockets. localized ridge augmentation for later implantation. alveolar ridge reconstruction for prosthetic treatment. filling of bone defects after root resection, cystectomy, removal of retained teeth. guided bone regeneration in dehiscence defects. guided tissue regeneration procedures in periodontal defects.
Device Story
Resorbable porcine-mesothelium-derived extracellular matrix membrane; supplied as lyophilized sheet. Used in dental surgery for guided tissue and bone regeneration. Clinician trims membrane to fit defect site; placed over bone/implant site to act as cell-occlusive barrier. Prevents soft tissue ingrowth into bone defect; promotes bone regeneration. Sterile, single-use device.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Resorbable porcine-mesothelium-derived extracellular matrix. Lyophilized sheet form factor. Cell-occlusive. Sterile, double-layer packaging. Biocompatible and non-pyrogenic.
Indications for Use
Indicated for dental patients requiring guided tissue and bone regeneration, including ridge augmentation, alveolar reconstruction, and defect filling following extractions, resections, or cystectomies. For prescription use only.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
K121310
Image /page/0/Picture/1 description: The image shows the logo for DSM, a global science-based company. The logo consists of a stylized circular graphic on the left, followed by the letters "DSM" in bold, sans-serif font. Below the letters, the tagline "BRIGHT SCIENCE. BRIGHTER LIVING." is printed in a smaller font size. The overall design is clean and modern, reflecting the company's focus on innovation and improving lives through science.
#### Mesothelium Dental Membrane
# JUL 1 1 2013
#### 510(k) Summary
DSM Biomedical Submitted by: 735 Pennsylvania Drive Exton, PA 19341 USA
Lori Burns, MS, RAC Contact Person: Manager Regulatory Affairs Ph: (484) 713-2186 Fax: (484) 713-2903
July 2, 2013 Date Prepared: K121310 510(K) #:
Device:
| 'Trade Name: | Mesothelium Dental Membrane |
|--------------------------|-------------------------------------------------------------|
| Common/Usual Name: | Resorbable Membrane for Guided Tissue and Bone Regeneration |
| Classification Name: | Barrier, Animal Source, Dental |
| Proposed Classification: | 21 CFR 872.3930, Class II, NPL |
Device Description:
Mesothelium Dental Membrane is a barrier membrane for guided tissue and bone regencration in dental applications. It is a resorbable porcine-mesothelium-derived extracellular matrix available as a lyophilized sheet. Mesothelium Dental Membrane is available in a variety of shapes and sizes. Furthermore, the device can be casily trimmed or shaped to the appropriate size to fit the defect to be treated. The device is packaged sterile in a double-layer package.
{1}------------------------------------------------
## Intended Use:
Mesothelium Dental Membrane is indicated for:
- . simultaneous use of GBR-membrane and implants.
- augmentation around implants placed in immediate extraction sockets. .
- augmentation around implants placed in delayed extraction sockets. .
- . localized ridge augmentation for later implantation.
- alveolar ridge reconstruction for prosthetic treatment. .
- filling of bone defects after root resection, cystectomy, removal of retained teeth. .
- guided bone regeneration in dehiscence defects. .
- guided tissue regencration procedures in pcriodontal defects. .
## Predicate Devices:
| Manufacturer | Device | 510(k) |
|-----------------------------------------------------|---------|------------------------------|
| Ed. Geistlich Soehne Ag Fuer<br>Chemische Industrie | K050446 | Bio-Gide® |
| RTI Biologics | K073097 | CopiOs™ Pericardium Membrane |
## Technological Characteristics:
The intended use, product design, material, function and target population of Mesothelium Dental Membrancs are substantially equivalent to the FDA cleared and legally marketed predicate devices Bio-Gide (K050446) and CopiOs Pericardium Membrane (K073097).
・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・
…
{2}------------------------------------------------
| Medical Device Name | Mesothelium Dental<br>Membrane | Bio-Gide | CopiOs Pericardium<br>Membrane |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Origin | Porcine | Porcine | Bovine |
| Device Characteristics | Extracellular matrix | Extracellular Matrix | Extracellular Matrix |
| | Cell-occlusive | Cell-occlusive | Cell-occlusive |
| | Implantable | Implantable | Implantable |
| | Resorbable | Resorbable | Resorbable |
| Intended Use | Mesothelium Dental<br>Membrane is<br>indicated for<br>simultaneous use of<br>GBR-membrane and<br>implants;<br>augmentation around<br>implants placed in<br>immediate extractions<br>sockets, augmentation<br>around implants<br>placed in delayed<br>extraction sockets,<br>localized ridge<br>augmentation for later<br>implantation, alveolar<br>ridge reconstruction<br>for prosthetic<br>treatment, filling of<br>bone defects after root<br>resection, cystectomy<br>removal of retained<br>teeth, guided bone<br>regeneration in<br>dehiscence defects,<br>guided tissue<br>regeneration<br>procedures in<br>periodontal defects. | Bio-Gide is indicated<br>for simultaneous use<br>of GBR-membrane<br>and implants;<br>augmentation around<br>implants placed in<br>immediate extractions<br>sockets, augmentation<br>around implants<br>placed in delayed<br>extraction sockets,<br>localized ridge<br>augmentation for later<br>implantation, alveolar<br>ridge reconstruction<br>for prosthetic<br>treatment, filling of<br>bone defects after root<br>resection, cystectomy<br>removal of retained<br>teeth, guided bone<br>regeneration in<br>dehiscence defects,<br>guided tissue<br>regeneration<br>procedures in<br>periodontal defects. | CopiOs is intended<br>for use in oral surgical<br>procedures as a<br>resorbable material<br>for augmentation<br>around implants<br>placed in immediate<br>extraction sockets,<br>augmentation around<br>implants placed in<br>delayed extraction<br>sockets; localized<br>ridge augmentation<br>for later implantation;<br>alveolar ridge<br>reconstruction for<br>prosthetic treatment;<br>filling of bone defects<br>after root resection,<br>cystectomy, removal<br>of retained teeth;<br>guided bone<br>regeneration in<br>dehiscence defects<br>and guided tissue<br>regeneration<br>procedures in<br>periodontal defects. |
| Biocompatibility | Yes | Yes | Yes |
| Non pyrogenic | Yes | Yes | Yes |
| Resorbable | Yes | Yes | Yes |
. . .
. .
יי
. .
· :
: :
"","t":"","title":"","title":"","title":"","title":"","title":"","title":"","title":"","title":"","title":"","title":"","title":"","title":"","title":"","title":"","title":"
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which features three abstract human figures connected by flowing lines. The symbol is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". The text is arranged in a circular fashion around the symbol.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - W(066-G609 Silver Spring, MD 20993-0002
July 11, 2013
DSM Biomedical C/O Ms. Lori Burns, MS. RAC Manager, Regulatory Affairs 735 Pennsylvania Drive EXTON PA 19341
Re: K121310
Trade/Device Name: Mesothelium Dental Membrane Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: II Product Code: NPL Dated: July 2, 2013 Received: July 5, 2013
Dear Ms. Burns:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Burns
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFF Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mary S. Runner -S
Kwame Ulmer, M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### Indications for Use
510(k) Number (if known): _K121310
Device Name: Mesothelium Dental Membrane
Indications for Use: Mesothelium Dental Membrane is indicated for: simultaneous use of GBRmembrane and implants, augmentation around implants placed in immediate extraction sockets, augmentation around implants placed in delayed extraction sockets, localized ridge augmentation for later implantation, alveolar ridge reconstruction for prosthetic treatment, filling of bone defects after root resection, cystectomy, removal of retained teeth, guided bone regeneration in dehiscence defects and guided tissue regeneration procedures in periodontal defects.
The device is intended for one time use.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
Mary S. Runner -S
Susan Run DDS, MA 2013.07.11
11:29:45-04'00'
Page __ of
(Division Sign-Off) (Division Sign-On)
Division of Anesthestology, General Hospital Unfection Control, Dental Devices
s10(k) Number: K121510
Page 60
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.