The OSTEOMED Cranial Distraction System intended for use in the treatment of cranial conditions such as syndromic craniosynostosis and congenital deficiencies in which osteotomies and gradual bone distraction are indicated. This device is intended to provide temporary stabilization and gradual lengthening of the cranial bones. This device is intended to be removed after consolidation. The OSTEOMED Cranial Distraction System is intended for single patient use only.
Device Story
Internal cranial distraction system; comprises distractors, spacers, screws, and instrumentation. Anchored between osteotomy sites using 1.2mm screws. Activated via distraction tool and preassembled distraction rod. Used by surgeons in clinical settings to facilitate gradual bone elongation. Non-threaded portion of rod removable during consolidation; entire assembly removed post-consolidation. Provides temporary stabilization and bone lengthening; benefits patients with congenital cranial deficiencies or syndromic craniosynostosis.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing and mechanical verification comparisons against the predicate device.
Technological Characteristics
Materials: Titanium (ASTM F-67), Titanium Alloy (ASTM F-136), surgical grade stainless steel, anodized aluminum, medical grade plastic. Principle: Internal mechanical distraction via threaded rod. Components: Distractors, spacers, screws, instrumentation (drills, forceps, cutters, screwdrivers). Single-patient use. Sterilization: Not specified.
Indications for Use
Indicated for patients with syndromic craniosynostosis or congenital cranial deficiencies requiring osteotomies and gradual bone distraction for temporary stabilization and lengthening of cranial bones.
Regulatory Classification
Identification
A preformed nonalterable cranioplasty plate is a device that is implanted in a patient to repair a skull defect and is constructed of a material, e.g., stainless steel or vitallium, that cannot be altered or reshaped at the time of surgery without changing the chemical behavior of the material.
{0}------------------------------------------------
K 12 1304
Image /page/0/Picture/1 description: The image shows the logo for Osteomed. The logo is in black and white and features the word "OSTEOMED" in a bold, sans-serif font. The word is enclosed in an oval shape. The logo has a modern and professional look.
# 510(k) Summary
AUG 222 2012
Submitter Information:
OsteoMed 3885 Arapaho Road Addison, Texas 75001 Phone: (972) 677-4600 Fax: (972) 677-4601
Contact Person:
Mrs. Piedad Peña
Date Prepared:
August 20, 2012
Device Information:
Proprietary/Trade Name: Common Name:
OsteoMed Cranial Distraction System Cranial Distraction System
Classification Name:
- · Regulation Number: 21 CFR 882.5330
- · Regulation Name: Preformed nonalterable cranioplasty plate
- · Product Code:
- PBJ o
Device Class: II
#### Predicate Devices:
KLS-Martin, K003883 Classification Name: Bone Plate (21CFR 872.4760, Product Code JEY) Device Class: II
#### Device Description:
The OsteoMed Cranial Distraction System is comprised of distractors, spacers, screws and instrumentation. The distractor is an internal distraction device for bone elongation of cranial bones. The distractor is anchored between the osteotomy with 1.2mm standard or 1.2mm AutoDrive screws previously cleared under K924138 and K023260, respectively. The distractor is activated using a distractor tool via the distraction rod. The distraction rod is preassembled within the distractor device. The non-threaded portion of the distractor rod can be removed during the consolidation period. The distractors, spacers, and screws are removed after consolidation.
The instruments include distractor tool, drills, plate bending forceps, plate cutters and screwdrivers to facilitate the placement of screws and modification of the distractor assembly plates. A distractor rod removal tool is also available to partially remove the distractor rod leaving the distracted device in place.
OsteoMed 3885 Arapaho Road Addison, Texas 75001 (972) 677-4600 FAX: (972) 677-4601 Customer Service: (800) 456-7779
Image /page/0/Picture/25 description: The image shows a logo with the word "Colson" in a stylized font inside a circle. To the right of the circle, the text "A COLSON ASSOCIATE" is written in a smaller, sans-serif font. The logo appears to be for a company or organization named Colson and its associates.
{1}------------------------------------------------
K121304
Image /page/1/Picture/1 description: The image shows the logo for OsteoMed. The logo is in black and white and features the word "OsteoMed" in bold, sans-serif font. The word is enclosed in an oval shape with a black stripe above the word. The oval shape is outlined in black.
The distractor assembly is made from Titanium (ASTM F-67) and Titanium Allov (ASTM F-136), and the spacer is made from Titanium (ASTM F-67). The screws are made from Titanium Alloy (ASTM F-136). The instrumentation is made from various grades of surgical grade stainless steel, anodized aluminum, and/or medical grade plastic.
#### Intended Use:
The OSTEOMED Cranial Distraction System intended for use in the treatment of cranial conditions such as syndromic craniosynostosis and congenital deficiencies in which osteotomies and gradual bone distraction are indicated. This device is intended to provide temporary stabilization and gradual lengthening of the cranial bones. This device is intended to be removed after consolidation. The OSTEOMED Cranial Distraction System is intended for single patient use only.
Intended use is equivalent to the KLS and OsteoMed predicate devices.
### Technological Characteristics:
The KLS predicate device is an internal distractor for bone elongation, which distracts via a threaded rod. OsteoMed internal distractor also uses a threaded rod for distraction and is anchored using bone screws. Both use a distraction tool for distraction.
Materials used for the distractor are the same as the predicate devices, Titanium/Titanium Alloy. These materials are biocompatible.
### Performance / Clinical Data:
The OSTEOMED Cranial Distraction System performed equivalent as the KLS predicate device based on verification testing. Verification testing consisted of mechanical testing and comparisons against the KLS predicate device. The intended use if the device is the same as the KLS Predicate Device.
Clinical Testing is not required to support substantial equivalence.
In conclusion, the device is safe and effective and performs as well as the predicate device.
#### Substantial Equivalence:
The basis of substantial equivalence for this device is based on similarities in intended use, material, function, performance, sterilization design, technology and operational principle to the predicate device, KLS-Martin (K003883) Distraction System, based on their promotional materials, labeling and clearance letter.
Due to the similarity of materials and design to the predicate device, OsteoMed believes that the OsteoMed Cranial Distraction System does not raise any new safety or effectiveness issues.
OsteoMed 3885 Arapaho Road Addison, Texas 75001 (972) 677-4600 FAX: (972) 677-4601 Customer Service: (800) 456-7779
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
OsteoMed c/o Mrs. Piedad Peña Sr. Regulatory Affairs Specialist 3885 Arapaho Road Addison, TX 75001
AUG 222 2012
Re: K121304
OsteoMed Cranial Distraction System Regulation Number: 21 CFR 885.5330 Regulation Name: Preformed nonalterable cranioplasty plate Regulatory Class: Class II Product Code: PBJ Dated: July 29, 2012 Received: July 31, 2012
Dear Mrs. Peña:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drig, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not maching.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements not in Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kesia Alexander
Malvina B. Eydelman, M.D. Director Division of Ophthalmic. Neurological. and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known): K121304
Device Name: OsteoMed Cranial Distraction System
. Indications for Use:
The OsteoMed Cranial Distraction System is intended for use in the treatment of cranial conditions such as syndromic craniosynostosis and congenital deficiencies in which osteotomies and gradual bone distraction are indicated. This device is intended to provide temporary stabilization and gradual lengthening of the cranial bones. This device is intended to be removed after consolidation.
The OsteoMed Cranial Distraction System implants are intended for single use only.
Prescription Use X · (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign Off,
Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number K121304
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.