XPRESSWAY RX CATHETER (6F LD-VERISON)

K121301 · Kaneka Pharma America, LLC · QEZ · Jul 30, 2012 · Cardiovascular

Device Facts

Record IDK121301
Device NameXPRESSWAY RX CATHETER (6F LD-VERISON)
ApplicantKaneka Pharma America, LLC
Product CodeQEZ · Cardiovascular
Decision DateJul 30, 2012
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.5150
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Xpress-Way™ RX Extraction Catheter is indicated for the removal of fresh, soft emboli and thrombi from vessels in the coronary and peripheral vasculature. The Xpress-Way™ RX Extraction Catheter is not intended for use in the cerebral vasculature.

Device Story

Xpress-Way™ RX Extraction Catheter is a mechanical device used by physicians in clinical settings to remove fresh, soft emboli and thrombi from coronary and peripheral vessels. It functions as an extraction catheter, facilitating the physical removal of obstructive material to restore blood flow. It is not intended for use in cerebral vasculature. The device is operated by a physician during interventional procedures.

Clinical Evidence

Bench testing only.

Technological Characteristics

Mechanical extraction catheter. Designed for coronary and peripheral vasculature. Class II device (21 CFR 870.5150).

Indications for Use

Indicated for removal of fresh, soft emboli and thrombi from coronary and peripheral vasculature. Contraindicated for use in cerebral vasculature.

Regulatory Classification

Identification

An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.

Submission Summary (Full Text)

{0}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the bird-like figure. ## Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 ## JUL 3 0 2012 Kaneka Pharma America LLC c/o Mr. Joseph DePaolo Regulatory Affairs Atrium Medical Corporation 5 Wentworth Drive Hudson, NH 03051 Re: K121301 Trade Name: XpressWay™ RX Catheter (6F LD-Version) Regulation Number: 21 CFR 870.5150 Regulation Name: Catheter, embolectomy Regulatory Class: Class II Product Code: DXE Dated: June 27, 2012 Received: June 29, 2012 Dear Mr. DePaolo: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must {1}------------------------------------------------ Page 2 – Mr. Joseph DePaolo comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, M.g. Whillchen Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use Page 1 of 1 510(k) Number (if known): צוג 2/3 0\ Device Name: Xpress-Way™ RX (6F LD-Version) Extraction Catheter Indications for Use: The Xpress-Way™ RX Extraction Catheter is indicated for the removal of fresh, soft emboli and thrombi from vessels in the coronary and peripheral vasculature. The Xpress-Way™ RX Extraction Catheter is not intended for use in the cerebral vasculature. Prescription Use _ X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF ---NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) CM.J. Keller Division Sign-Off) Division of Cardiovascular Devices 510(k) Number_ K121301
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...