K121292 · Siemens Medical Solutions USA, Inc. · LLZ · Jul 30, 2012 · Radiology
Device Facts
Record ID
K121292
Device Name
DYNAPBV BODY SOFTWARE
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
LLZ · Radiology
Decision Date
Jul 30, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Indications for Use
syngo DynaPBV Body is an extended software application to the InSpace 3D software option which allows the reconstruction of two-dimensional images acquired with a standard angiographic C-arm device into a three-dimensional image format. syngo DynaPBV Body is intended for imaging primarily soft tissue for diagnosis, surgical planning, interventional procedures and treatment follow-up. It is design for the visualization of contrast enhanced blood distribution in the body using color coded relative values for diagnosis. This software is designed to visually assist physicians in the diagnosis and treatment of vessel malformations (i.e. Aneurysms, AVM's and Stenoses).
Device Story
syngo DynaPBV Body is an add-on software extension for InSpace 3D; operates on angiographic C-arm systems. Input: 2D X-ray images. Transformation: reconstructs 2D images into 3D format; generates color-coded maps representing relative contrast-enhanced blood distribution. Output: 3D visualization of blood flow/vessel structures. Used in clinical settings (e.g., interventional suites) by physicians. Output assists in diagnosing/treating vessel malformations (aneurysms, AVMs, stenoses) and surgical planning. Benefits include improved visualization of vascular perfusion to guide clinical decision-making.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing, software verification, and validation of the software extension within the existing InSpace 3D framework.
Technological Characteristics
Software-only extension to InSpace 3D. Interfaces with angiographic C-arm hardware. Features include 3D image reconstruction, color-coded relative value mapping for blood distribution, and integration into existing InSpace 3D task cards. Operates as a non-standalone application.
Indications for Use
Indicated for physicians to assist in diagnosis, surgical planning, interventional procedures, and treatment follow-up of vessel malformations including aneurysms, AVMs, and stenoses in soft tissue, specifically for visualizing contrast-enhanced blood distribution in the body (thorax/abdomen).
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
SIEMENS
syngo DynaPBV Body 510(k) K121292 Al Response
# 510(k) Summary: syngo DynaPBV Body Software
JUL 30 2012
Siemens Medical Systems, Inc. Company: 51 Valley Stream Parkway Malvern, PA 19355
Date Prepared: July 25, 2012
This 510(k) summary of safety and effectiveness information is being submitted in i accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.
## 1. General Information:
lmporter / Distributor: Siemens Medical Systems, Inc. 51 Valley Stream Parkway Malvern, PA 19355
Establishment Registration Number: 2240869 Manufacturing Site: SIEMENS AG Sector Healthcare Siemensstraße 1 D-91301 Forchheim, Germany
Establishment Registration Number: 3004977335
## 2. Contact Person:
Ms. Patricia D Jones Technical Specialist, Regulatory Submissions Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway D-02 Malvern, PA 19355 Fax: (610) 448-1787 Phone: (610) 448 -3536 Email: patricia.d.jones@siemens.com
# Device Name and Classification: 3.
syngo DynaPBV Body Trade Name: Classification Name: System, Image Processing Radiological Classification Panel: Radiology Classification Regulation: 21 CFR §892.2050 Class II Device Class: LLZ Product Code:
Legally Marketed Predicate Device 4. InSpace 3D Software Option Trade Name: K011447 510(k) #: Clearance Date: August 3, 2001
{1}------------------------------------------------
# SIEMENS
Classification Name: Classification Panel: CFR Section: Device Class: Product Code:
Accessory to Angiographic X-Ray System Radiology 21 CFR §892.1600 Class II JAA
Legally Marketed Predicate Device Trade Name: syngo Neuro PBV K111052 510(k) #: May 20, 2011 Clearance Date: Classification Name: System, Image Processing Radiological Classification Panel: Radiology 21 CFR §892.2050 CFR Section: Class II Device Class: Product Code: LLZ
## 5. Device Description:
The syngo DynaPBV Body software is an optional extension to the Inspace 3D application originally cleared under Premarket Notification K011447 on 08/03/2001. It is also similar to the cleared syngo Neuro PBV IR (K111052, May 20, 2011) which was designed for the visualization of contrast enhanced blood distribution in the arterial and venous vessels in the head.
Similar to syngo Neuro PBV IR the synqo DynaPBV Body is an add-on software option used for the visualization of contrast enhanced blood distribution in the body (e.g. thorax and abdomen) using color coded relative values for diagnosis.
This software modification does not affect the intended use of the device nor does it alter its fundamental scientific technology.
## 6. Indication for Use:
The syngo DynaPBV Body is an extended software application to the InSpace 3D software option which allows the reconstruction of two-dimentional images acquired with a standard angiographic C-arm device into a three-dimentional image format.
The syngo DynaPBV Body is intended for imaging primarily soft tissue for diagnosis, surgical planning, interventional procedures and treatment follow-up. It is design for the visualization of contrast enhanced blood distribution in the body using color coded relative values for diagnosis.
This software is designed to visually assist physicians in the diagnosis and treatment of vessel malformations (i.e. Aneurysms, AVM's and Stenoses).
## Substantial Equivalence: 7.
The syngo DynaPBV Body software application is substantially equivalent to the commercially available Siemens software application, syngo Neuro PBV IR which is an option to the Siemens InSpace 3D software. The InSpace 3D software option was described in premarket notification K011447 which received 510(k) clearance
{2}------------------------------------------------
on August 03, 2001. The syngo Neuro PBV IR software option was described in premarket notification K111052 and received 510(k) clearance on May 20, 2011.
The syngo DynaPBV Body software is an optional extension to InSpace 3D and uses the same hardware and software components as the inSpace 3D software.
# Summary of Technological Characteristics of the Subject Device as Compared 8. with the Predicate Device:
The syngo DynaPBV Body is a software extension to InSpace 3D. The syngo DynaPBV Body features the same post processing software, user interface, archiving and communication as the predicate InSpace 3D. The syngo DynaPBV Body software is not a stand-alone software. It interfaces with InSpace 3D. The syngo DynaPBV Body user function keys are integrated into the InSpace 3D task card. The user function is similar to InSpace 3D task card except for an additional activation button for the syngo DynaPBV Body software features.
## General Safety and Effectiveness Concerns: 9.
Instructions for use are included within the device labeling, and the information provided will enable the user to operate the device in a safe and effective manner.
Risk management is ensured via a hazard analysis, which is used to identify potential hazards. These potential hazards are controlled via software development, verification and validation testing. To minimize electrical, mechanical and radiation hazards. Siemens adheres to recognized and established industry practice, and all equipment is subject to final performance testing. Furthermore, the operators are health care professionals familiar with and responsible for the evaluating and post processing of X-ray images.
## 10. Conclusion as to Substantial Equivalence:
syngo DynaPBV Body software is intended for similar indications as cleared in the predicate InSpace 3D. The syngo DynaPBV Body Software add-on application is designed for use with the InSpace 3D (K011447) for the visualization of contrast enhanced blood distribution in the body using color coded relative values for diagnosis.
The functionality of syngo DynaPBV Body software is similar to the predicate device. It is Siemens opinion, that the syngo DynaPBV Body add-on software is substantially equivalent to the InSpace 3D software (K011447) and the syngo Neuro PBV IR software (K111052).
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circular fashion around the symbol. The text is in all caps and appears to be in a sans-serif font. The logo is black and white.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
# Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
JUL 30 2012
Ms. Patricia D. Jones Technical Specialist. Regulatory Submissions Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway D-02 MALVERN PA 19355
Re: K121292
Trade/Device Name: Syngo DynaPBV Body Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: July 25, 2012 Received: July 26, 2012
# Dear Ms. Jones:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean i Toure of act note a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I vatual the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
- for argument in
{4}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
SIEMENS
Special 510(k) Submission: syngo DynaPBV Body
# Indications for Use Statement
510(k) Number (if known):
Device Name: syngo DynaPBV Body
Indications for Use:
syngo DynaPBV Body is an extended software application to the InSpace 3D software option which allows the reconstruction of two-dimensional images acquired with a standard angiographic C-arm device into a three-dimensional image format.
syngo DynaPBV Body is intended for imaging primarily soft tissue for diagnosis, surgical planning, interventional procedures and treatment follow-up. It is design for the visualization of contrast enhanced blood distribution in the body using color coded relative values for diagnosis.
This software is designed to visually assist physicians in the diagnosis and treatment of vessel malformations (i.e. Aneurysms, AVM's and Stenoses)
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C)
(Please do not write below this line - continue on another page if needed)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Martin D. Thoma
Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K121292
Page 1 of
Special 510(k) syngo Liver PBV-IR Software
Siemens Medical Systems, Inc.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.