The VBOX Trooper oxygen concentrator device is used on a prescriptive basis by patients requiring supplemental oxygen. It supplies a high concentration of oxygen and is used with a nasal cannula to channel oxygen from the concentrator to the patient. The VBOX Trooper may be used in home, institution, vehicle and various mobile environments.
Device Story
Trooper Oxygen Concentrator uses molecular sieve and differential pressure swing adsorption to extract oxygen from ambient air; produces 87-94% pure oxygen. Device senses patient inhalation via pressure change (trigger sensitivity ≤ 0.13 cm water); releases oxygen pulse upon detection. Between breaths, device regenerates oxygen pulse. Front panel features LEDs, ON/OFF button, flow rate controls, and battery/flow displays. Used in home, institution, vehicle, or mobile settings by patients. System includes concentrator unit, Li-Ion batteries, AC power supply, battery charger, carrying case, and nasal cannula. Healthcare providers prescribe device; patients operate it to receive supplemental oxygen therapy, potentially improving mobility and oxygenation.
Clinical Evidence
Bench testing only. Evaluated output gas composition, display/LED functions, flow rate, oxygen concentration, electromagnetic compatibility, and electrical safety. Functional side-by-side comparison with predicates assessed trigger sensitivity, pulse timing (time, delay, total delivery), pulse volume, oxygen purity, and FIO2 across various breath rates.
Technological Characteristics
Molecular sieve (synthetic zeolite) with differential pressure swing adsorption. 5 flow settings (1-5 LPM equivalent). Pulsed delivery. Li-Ion battery or AC power. Analog/digital electronics with microprocessor. Dimensions: 6 x 2.5 x 6.25 inches; weight: 3.2 lbs. Standards: ASTM F1464-93:2005, ISO 8359:1996, IEC 60601-1:2003, ISO 10993-1:2009, EN 60601-1-2:2007.
Indications for Use
Indicated for patients requiring supplemental oxygen on a prescriptive basis. Used in home, institutional, vehicle, and mobile environments via nasal cannula.
Regulatory Classification
Identification
A portable oxygen generator is a device that is intended to release oxygen for respiratory therapy by means of either a chemical reaction or physical means (e.g., a molecular sieve).
{0}------------------------------------------------
#### 510(k) Summary 12/260 : 5.0
VBOX. Inc.
Submitter:
WH
TI
Contact Person:
2340 East County Road J White Bear Lake, MN 55110 Theodore Jagger Director, Regulatory & Test
Telephone: 651-407-6206 Fax: 651-407-6206 Email: jagger@vboxinc.com
Trooper Oxygen Concentrator
Portable Oxygen Concentrator
September 19, 2012
Date Prepared: Trade Name: Common Name: Classification: Product Code:
Class II, 21 CFR 868.5440 CAW
Predicate Devices:
The subject device is substantially equivalent to the following devices:
SEP 2 1 2012
- · Inogen One Oxygen Concentrator (K032818)
- · EverGo Portable Oxygen Concentrator (K043615)
Device Description:
The Trooper oxygen concentrator utilizes a molecular sieve and differential pressure swing adsorption to separate the gases in ambient air. The device takes the room air and concentrates the oxygen portion to produce a pulse of oxygen between 87-94 % in purity. When the patient inhales, the device senses the pressure change and is triggered to release the oxygen pulse. In between breaths, the device regenerates an oxygen pulse and waits for the next inhalation breath before dispensing it.
The front panel of the Trooper contains controls and indicators such as, device status indicator LEDs, an ON/OFF button, oxygen flow rate controls, and flow rate and battery status displays. The oxygen outlet is also located on the front panel of the device.
The VBOX Trooper oxygen concentrator system will be provided under a single model number, A-1000, which includes the following items:
- One (1) Trooper Oxygen Concentrator Unit .
- Two (2) Li-Ion Batteries (only one connected to the device . at a time)
- One (1) Auxiliary AC Power Supply .
- One (1) Battery Charger ●
- One (1) Carrying case .
- One (1) nasal cannula .
{1}------------------------------------------------
- User manual .
# per Device Specifications:
| Trooper Device Specifications: | |
|----------------------------------------|----------------------------------------------------------------------|
| Dimensions (LxWxH) | 6 x 2.5 x 6.25 inches |
| Weight | 3.2 lbs (includes battery) |
| Materials | |
| Sieve Bed | Synthetic zeolite |
| Nasal Cannula | PVC (standard cannula supplied<br>by Salter Labs) |
| Battery | Li-Ion |
| Performance Specifications | |
| Method of oxygen<br>concentration | Molecular sieve (mechanical) |
| Process by which Oxygen is<br>released | Differential pressure swing<br>adsorption |
| Flow Rate | 5 settings: 1 to 5 (flow rates<br>equivalent to 1 LPM to 5 LPM) |
| Duration of flow | Pulsed |
| Trigger Sensitivity | $ \leq $ 0.13 cm water ( $\leq$ 12.7 Pa) |
| Oxygen concentration | 87-94% at all settings |
| Software/ Hardware | Analog and digital electronics<br>with microprocessor |
| Rechargeable Battery | Yes |
| Power Options | Battery, AC |
### Intended Use:
The VBOX Trooper oxygen concentrator device is used on a prescriptive basis by patients requiring supplemental oxygen. It supplies a high concentration of oxygen and is used with a nasal cannula to channel oxygen from the concentrator to the patient. The VBOX Trooper may be used in home, institution, vehicle and various mobile environments.
## Summary Comparison to Predicate Devices:
| | VBOX Trooper<br>Oxygen Concentrator<br>(Subject Device) | Inogen One Oxygen<br>Concentrator<br>(Predicate Device) | EverGo Portable<br>Oxygen Concentrator<br>(Predicate Device) |
|-------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | VBOX, Inc. | Inogen | Philips Respironics |
| Classification | Class II | Class II | Class II |
| Product Code | CAW | CAW | CAW |
| Indications for<br>Use | The VBOX Trooper oxygen concentrator<br>device is used on a<br>prescriptive basis by<br>patients requiring<br>supplemental oxygen. It<br>supplies a high<br>concentration of oxygen<br>and is used with a nasal<br>cannula to channel<br>oxygen from the<br>concentrator to the<br>patient. The VBOX<br>Trooper may be used in<br>home, institution,<br>vehicle and various<br>mobile environments. | The Inogen One<br>Oxygen Concentrator is<br>used on a prescriptive<br>basis by patients<br>requiring supplemental<br>oxygen. It supplies a<br>high concentration of<br>oxygen and is used<br>with a nasal cannula to<br>channel oxygen from<br>the concentrator to the<br>patient. The Inogen<br>One may be used in<br>home, institution,<br>vehicle and various<br>mobile environments. | The EverGo Portable<br>Oxygen Concentrator is<br>intended for prescription<br>use by patients requiring<br>high concentrations of<br>oxygen on a<br>supplemental basis. It is<br>small, portable and is<br>capable of continuous<br>use in home,<br>institutional, and travel /<br>mobile environments. |
| Prescription<br>Required | Yes | Yes | Yes |
| Patient<br>Interface | Standard nasal cannula | Standard nasal cannula | Standard nasal cannula |
| Dimensions<br>(LxWxH) | 6 x 2.5 x 6.25 inches | 11.6 x 6.0 x 10.7 inches | 12 x 6 x 8.5 inches |
| Weight | 3.2 lbs<br>(includes battery) | 9.8 lbs<br>(includes battery) | 10 lbs<br>(includes batteries) |
| Materials | | | |
| Sieve Bed | Synthetic zeolite | Synthetic zeolite | Synthetic zeolite |
| Nasal Cannula | PVC (standard cannula<br>supplied by Salter Labs) | PVC (standard cannula<br>supplied by Salter<br>Labs) | Not supplied with the<br>device |
| Battery | Li-Ion | Li-Ion | Li-Ion |
| Performance Specifications | | | |
| Method of<br>oxygen<br>concentration | Molecular sieve<br>(mechanical) | Molecular sieve<br>(mechanical) | Molecular sieve<br>(mechanical) |
| Process by<br>which Oxygen is<br>released | Differential pressure<br>swing adsorption | Differential pressure<br>swing adsorption | Differential pressure<br>swing adsorption |
| Flow Rate | 5 settings: 1 to 5 (flow<br>rates equivalent to 1<br>LPM to 5 LPM) | 5 settings: 1 to 5 (flow<br>rates equivalent to 1<br>LPM to 5 LPM) and<br>one setting of<br>"Satellite" | 6 settings: 1 to 6 (flow<br>rates equivalent to 1<br>LPM to 6 LPM) |
| Duration of<br>flow | Pulsed | Pulsed | Pulsed |
| Trigger<br>Sensitivity | ≤ 0.13 cm water<br>(≤ 12.7 Pa) | 0.12 cm water (12 Pa) | 0.16 cm water (16 Pa) |
| Oxygen<br>concentration | 87-94% at all settings | 87-93% at all settings | 86-92% at all settings |
| Software/<br>Hardware | Analog and digital<br>electronics with<br>microprocessor | Analog and digital<br>electronics with<br>microprocessor | Analog and digital<br>electronics with<br>microprocessor |
| Rechargeable<br>Battery | Yes | Yes | Yes |
| Power Options | Battery, AC | Battery, AC | Battery, AC |
{2}------------------------------------------------
Pre-Market Notification for the VBOX Trooper Oxygen Concentrator
{3}------------------------------------------------
Functional and Safety Testing:
Applicable portions of the following standards were applied during development and testing of the Trooper Oxygen Concentrator:
- ASTM F1464-93:2005 Oxygen Concentrators for . Domiciliary Use
- ISO 8359:1996 Oxygen Concentrators for medical use -. Safety Requirements
- EN 60601-1-2:2007 Medical Electrical Equipment-Part 1-. 2: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests
- . EN 55011:2007 (including Amendment A2:2007) - Limits and Methods of Measurement of Radio Interference Characteristics of Industrial, Scientific and Medical (ISM) Equipment
- Federal Communications Commission (FCC) Part 15 . Subpart B
- IEC 60601-1:2003 Medical Electrical Equipment Part 1: . General Requirements for Safety
- ISO 10993-1:2009 Biological evaluation of medical devices .
Bench testing was performed to provide assurance that the proposed device conforms to the requirements for its intended use. This included the following testing:
- Output gas composition (e.g. VOCs, particulate matter, � ozone/carbon monoxide/carbon dioxide content)
- User display and LED functions .
- Oxygen flow rate and concentration ●
- Electromagnetic compatibility and electrical safety ●
- Functional performance (e.g. trigger sensitivity and delay, . pulse volume and duration)
{4}------------------------------------------------
- Output gas temperature .
In addition, functional side-by-side comparison testing was performed to demonstrate substantial equivalence of the proposed device to each of the predicate devices. The following parameters were evaluated across all breath rates:
- Trigger Sensitivity .
- . Oxygen Pulse Timing
- o Pulse Time
- o Pulse Delay
- o Total Time to Deliver Pulse
- Pulse Volume �
.
- · Oxygen Purity
- Relative Fraction of Inspired Oxygen (FIO2) .
The pulse delivery waveforms (liters per minute flow over time) were also assessed for the proposed device and each of the predicate devices.
The similarities between the Trooper (proposed device) and the Conclusion: predicate devices referenced above with respect to the principles of operation, technology, materials, indications for use, and functional performance clearly support a conclusion of substantial equivalence.
{5}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle with three curved lines forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -- WO66-G609 Silver Spring, MD 20993-0002
Vbox, Incorporated C/O Mr. Mark Job Regulatory Technology Services, Limited Liability Company 1394 25TH Street, North West Buffalo, Minnesota 53313
SEP
2 1 2012
Re: K121260
Trade/Device Name: Trooper Oxygen Concentrator Regulation Number: 21 CFR 868.5440 Regulation Name: Portable Oxygen Generator Regulatory Class: II Product Code: CAW Dated: September 6, 2012 Received: September 7, 2012
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2- Mr. Job
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Nh for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
#### 4.0 Indications for Use Statement
Device Name: VBOX Trooper Oxygen Concentrator
### Indications for Use:
The VBOX Trooper oxygen concentrator device is used on a prescriptive basis by patients requiring supplemental oxygen. It supplies a high concentration of oxygen and is used with a nasal cannula to channel oxygen from the concentrator to the patient. The VBOX Trooper may be used in home, institution, vehicle and various mobile environments.
Prescription Use X (21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
L. Schultze
(Division Division of Anesthesiology, General Hospitai Intection Control, Dental Device
510(k) Number;
K L L B
VBOX, Inc.
Pre-Market Notification for the VBOX Trooper Oxygen Concentrator Page 14 of 57
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.