K121128 · Irvine Scientific Sales Co., Inc. · MQL · Jun 6, 2012 · Obstetrics/Gynecology
Device Facts
Record ID
K121128
Device Name
MULTIPURPOSE HANDLING MEDIUM
Applicant
Irvine Scientific Sales Co., Inc.
Product Code
MQL · Obstetrics/Gynecology
Decision Date
Jun 6, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.6180
Device Class
Class 2
Indications for Use
Multipurpose Handling Medium™ (MHM™) is intended for use in assisted reproductive procedures which involve the manipulation of gametes or embryos. Specifically, Multipurpose Handling Medium™ (MHM™) is indicated for use as an oocyte retrieval medium during ovarian follicle aspiration procedures (not for flushing ovarian follicles), washing sperm prior to IVF and ICSI fertilization procedures, and for transport of the embryo to the uterus during embryo transfer procedures.
Device Story
Multipurpose Handling Medium™ (MHM™) is a ready-to-use liquid medium for assisted reproductive technology (ART) procedures. It serves as an oocyte retrieval medium, sperm-washing medium, and embryo transport medium. The medium utilizes a HEPES, MOPS, and Sodium Bicarbonate buffering system, allowing for pH maintenance in ambient atmosphere without a CO2 incubator. It is used by clinicians in a laboratory or clinical setting during IVF/ICSI procedures. The device supports gametes and embryos during manipulation, washing, and transport, facilitating successful fertilization and embryo transfer. It is supplied in 100 mL and 500 mL bottles.
Clinical Evidence
No clinical data provided. Evidence consists of bench testing: 1-cell Mouse Embryo Assay (MEA) and Human Sperm Survival Assay (HSSA). MEA results showed ≥ 80% development. HSSA demonstrated ≥ 70% of original motility at 24 hours. Testing confirmed the device is non-toxic and functional for its intended use, with performance comparable to the predicate mHTF.
Technological Characteristics
Liquid medium containing HEPES, MOPS, and Sodium Bicarbonate buffering system. Ingredients include WFI water, salts (NaCl, KCl, CaCl2), amino acids, pyruvate, lactate, and gentamicin sulfate. Formulated for ambient atmosphere use. Sterility, endotoxin (≤ 0.25 EU/mL), pH (7.25-7.54), and osmolality (275-295 Osm/KgH2O) specifications are controlled. Non-pyrogenic, sterile-filtered.
Indications for Use
Indicated for use in assisted reproductive technology (ART) procedures involving gamete or embryo manipulation, specifically oocyte retrieval (excluding follicle flushing), sperm washing prior to IVF/ICSI, and embryo transport for uterine transfer.
Regulatory Classification
Identification
Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
{0}------------------------------------------------
# K121128
### JUN - 6 2012
#### 510(k) SUMMARY AS REQUIRED BY 21 CR § 807.87(h)
Submitter's Name and Address:
Manufacturing Site:
Contact Person:
Irvine Scientific Sales Co., Inc. 2511 Daimler Street Santa Ana, CA 92705-5588
2511 Daimler Street Santa Ana, CA 92705-5588
Jayme Yamaquchi-Owens Requlatory Affairs Manager Irvine Scientific 2511 Daimler Street Santa Ana, CA 92705-5588 Tel: 949-261-7800 ext 254 949-261-6522 Fax: Email: jfy@irvinesci.com
Date of Application:
April 9, 2012
Establishment Registration Number: 2022379
510(k):
K121128
Trade or Proprietary Name:
Device Name:
Device Classification:
Device Regulation:
Device Classification:
Product Code:
Predicate Device:
Performance Standards:
R121120
Multipurpose Handling Medium™ (MHM TM)
Handling Medium
Reproductive Media and Supplements
21 CFR § 884.6180
Class II (Special Controls)
85 MQL
Modified HTF (mHTF), K983586 G-MOPS™/G-MOPS™ PLUS, K081115
None established
{1}------------------------------------------------
#### Purpose:
Purpose of this application is to obtain 510(k) marketing clearance for the Multipurpose Handling Medium™ (MHM™).
#### Indication for Use:
Multipurpose Handling Medium™ (MHM™) is intended for use in assisted reproductive procedures which involve the manipulation of gametes or embryos. Specifically, Multipurpose Handling Medium™) is indicated for use as an oocyte retrieval medium during ovarian follicle aspiration procedures (not for flushing ovarian follicles), washing sperm prior to IVF and ICSI fertilization procedures, and for transport of the embryo to the uterus during embryo transfer procedures.
#### Description of the Device:
The Multipurpose Handling Medium™ (MHM™) is based upon the mHTF (K983586) formulation. The formula for the Multipurpose Handling Medium™ (MHM™) is based upon the composition of modified HTF, K983586. The medium uses a buffering system composed of HEPES (N-2-Hydroxyethylpiperazine-N1 - - 2 - - ethanesulfonic acid). MOPS . (3 Morpholinopropane - 1- sulfonic acid) and Sodium Bicarbonate. This buffering system provides pH maintenance over the physiologic range and does not require the use of a CO2 Incubator. - The product also contains 10 ug/mL gentamicin.
Multipurpose Handling Medium™ (MHM™) has utility as an oocyte retrieval medium, in procedures that aspirate oocytes from the patient's ovarian follicles. Multipurpose Handling Medium™ (MHM™) is used only in the oocyte collection vessel and is not for use in flushing oocytes from ovarian follicles. Once the oocyte has been retrieved, it is placed into a culture dish with an appropriate amount of a culture medium (does not include MHM™), as specified by the user's internal procedures, and fertilization is allowed to occur. Multipurpose Handling Medium™ (MHM™) can also be used to wash sperm in preparation for
{2}------------------------------------------------
fertilization by conventional IVF or ICSI procedures. The fertilized oocyte (zygote) is allowed to grow in the culture dish, supported by a culture medium (does not include MHM™), as specified by the user's internal procedures, and an appropriate protein supplement, in a carbon dioxide incubator at 37°C until the desired stage of development is achieved. The embryo is removed from the culture dish, placed into an appropriate amount of Multipurpose Handling Medium™ (MHM™) for transfer into the uterus.
Multipurpose Handling Medium™ (MHM™) is supplied as a ready to use liquid, in 100 and 500 mL bottles.
### Technological Characteristics:
Multipurpose Handling Medium™ (MHM™) can be used as an oocyte collection medium during in procedures that aspirate oocytes from the patient's ovarian follicles. Multipurpose Handling Medium™ (MHM™) is used only in the oocyte collection vessel and is not for use in flushing oocytes from ovarian follicles. Once the oocyte has been retrieved, it is placed into a culture dish with an appropriate amount of a culture medium, as specified by the user's internal procedures, and fertilized. After fertilization, the embryo is allowed to develop in a culture medium, as specified by the user's internal procedures, until an appropriate developmental stage is reached. After incubation of the embryo, to the appropriate developmental stage, the embryo is removed from the incubation dish and placed into a suitable amount of Multipurpose Handling Medium™ (MHM™) for transport and transfer into the patient. Multipurpose Handling Medium™ (MHM™) can be used for washing sperm from its surrounding seminal fluid. Once semen has liquefied it is placed into a conical tube with a volume of room temperature Handling Medium™ (MHM™) and centrifuged to separate the sperm from the seminal fluid prior to IVF and ICSI fertilization procedures. Multipurpose Handling Medium™ (MHM™) is therefore intended for use as an oocyte retrieval medium, as a transport and storage medium and as a spermwashing medium.
{3}------------------------------------------------
Multipurpose Handling Medium™ (MHM™) is similar to mHTF (K983586) it is intended for use in assisted reproductive technology procedures that involve the manipulation of gametes or embryos. mHTF is also intended for use as a spermprocessing medium in washing procedures, as an oocyte retrieval medium, for transport of the embryo, and as a support medium for transfer of the embryo. mHTF is intended to simulate the substances found in the human, female reproductive system.
Multipurpose Handling Medium™ (MHM™) is similar to G-MOPS and G-MOPS" PLUS (K081115) in that is intended for oocyte collection and for handling and manipulating oocytes and embryos in ambient atmosphere.
#### Performance Data:
Multipurpose Handling Medium™ is assayed by one (1) -cell mouse embryo assay (MEA) prior to release to the market. This assay assures that the product is both functional for its intended use and no embryo-toxic components are present in the formulation.
Multipurpose Handling Medium™ (MHM™) is also assayed human sperm survival assay was also performed prior to release to the market. This assay also assures that the product functions in accordance with its intended use with regards to sperm wash procedure and that no sperm-toxic components are present in the formulation.
#### Nonclinical Tests:
One (1) -cell MEA was performed on the Multipurpose Handling Medium™ (MHM™) as part of the design validation. Testing was performed at three (3) different test facilities after 2 hours exposure to the Multipurpose Handling Medium™ (MHM™).
In addition, human sperm survival assay was also performed on donor specimens that were initially processed by gradient separation and resulting motile specimens were equally divided and the average % motility after 24 hours
{4}------------------------------------------------
at 37°C in ambient air was determined. The field evaluations demonstrate that the Multipurpose Handling Medium™ (MHM™) was "equal" to the proven control medium of mHTF (K983586).
#### Additional Information:
Endotoxin, pH, osmolality, mouse embryo assay, human sperm survival assay and sterility tests will be performed as a condition of release for Multipurpose Handling Medium™ (MHM™). Results of all release assays performed will be reported on a lot-specific certificate of analysis, and will be indicated on the labeling.
#### Conclusion:
The conclusion from the performance testing as well as the nonclinical and clinical data demonstrates that the Multipurpose Handling Medium™ (MHM™) is suitable for its intended use, and meets the criteria in the comparison to predicate devices in which substantial equivalence has been demonstrated and meets the criteria outlined in the Notice of Final Rule, 63 FR 48428, Docket Number 97N-0335.
A comparison of the Predicate Device Intended Use, Product Formula and Product Specifications are summarized below that demonstrate substantial equivalence.
| Product Name | K# | Retrieval /<br>Collection | Culture | Transfer | Ambient |
|------------------------------------------|---------|---------------------------|---------|----------|---------|
| Modified HTF<br>(mHTF) Medium –<br>HEPES | K983586 | + | + | + | + |
| G-MOPS™/G-<br>MOPS™ PLUS | K081115 | + | + | + | + |
#### Table 1: Predicate Intended Use Comparison
{5}------------------------------------------------
| Raw Materials | Multipurpose<br>Handling<br>Medium™<br>Catalog #90163 | Modified HTF<br>(K983586) | G-MOPS™<br>(K081115) | G-MOPS™<br>PLUS<br>(K081115) |
|------------------------------------------------------------------------------------------------|---------------------------------------------------------|---------------------------|-----------------------|-------------------------------|
| WFI Quality Water | + | + | + | + |
| Phenol Red, Na Salt | + | + | - | - |
| Sodium Bicarbonate | + | + | + | + |
| Sodium Chloride | + | + | + | + |
| Potassium Chloride | + | + | + | + |
| Dextrose | + | + | + | + |
| Magnesium Sulfate | + | + | + | + |
| Potassium Phosphate | + | + | - | - |
| Pyruvic Acid, sodium<br>salt | + | + | - | - |
| DL-Lactic Acid, sodium | + | + | + | + |
| Glycine | + | - | + | + |
| Taurine | + | - | + | + |
| Calcium Chloride,<br>anhydrous | + | + | + | + |
| HEPES ½ sodium salt<br>[4-(2-hydroxyethyl)-1-<br>piperazineethanesulfonic acid, ½ sodium salt] | + | + | - | - |
| HEPES Sodium salt<br>[4-(2-hydroxyethyl)-1-<br>piperazineethanesulfonic acid, sodium salt] | - | + | - | - |
| MOPS, sodium salt | + | - | + | + |
| Gentamicin Sulfate | + | + | + | + |
| Alanine | + | + | + | + |
| Alanyl-glutamine | + | + | + | + |
| Asparagine | + | + | + | + |
| Aspartate | + | + | + | + |
| Raw Materials | Multipurpose<br>Handling<br>MediumTM,<br>Catalog #90163 | Modified HTF<br>(K983586) | G-MOPSTM<br>(K081115) | G-MOPSTM<br>PLUS<br>(K081115) |
| Glutamate | + | + | + | + |
| Glycine | + | + | + | + |
| Lipoic Acid | + | + | + | + |
| Methionine | + | + | + | + |
| Proline | + | + | + | + |
| Serine | + | + | + | + |
| Sodium citrate | + | + | + | + |
| Sodium Hydrogen<br>Phosphate | + | + | + | + |
| Sodium Pyruvate | + | + | + | + |
| Human serum albumin | -1 | + | + | + |
# Table 2: Product Formulation Comparison
.
·
:
·
{6}------------------------------------------------
Table 3: Predicate Device Product Specification Comparison
| Specification | Multipurpose<br>Handling Medium™ | Modified HTF<br>(K983586) | G-MOPS™/<br>G-MOPS™ PLUS<br>(K081115) |
|---------------|-------------------------------------------|---------------------------|---------------------------------------|
| pH | 7.25 - 7.54 | 7.25 - 7.54 | Unknown |
| Osmolality | 275 - 295 Osm/KgH2O | 272 - 288 Osm/KgH2O | Unknown |
| Sterility | Pass | Pass | Pass |
| Endotoxin | ≤ 0.25 EU/mL | ≤ 0.25 EU/mL | ≤ 0.25 EU/mL |
| MEA | ≥ 80% | ≥ 70% | ≥ 80% |
| HSSA | ≥ 70% of original<br>motility at 24 hours | NT2 | NT |
.
1 Protein supplementation is to be performed prior to use
.
² NT — Not Tested
{7}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, symbolizing protection and care. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Jayme Yamaguchi-Owens Regulatory Affairs Manager Irvine Scientific Sales Co., Inc. 2511 Daimler Street SANTA ANA CA 92705
JUN - 6 2012
Re: K121128
> Trade/Device Name: Multipurpose Handling Medium™ (MHMTM) Regulation Number: 21 CFR§ 884.6180 Regulation Name: Reproductive media and supplements Regulatory Class: II Product Code: MQL Dated: April 9, 2012 Received: April 13, 2012
Dear Ms. Yamaguchi-Owens:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate 10 abo career in to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it your device to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act that 1 17 has interes and regulations administered by other Federal agencies. You must of any Feather statest is requirements, including, but not limited to: registration and listing eoulpry .mart 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{8}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Benjamin K. Evans
Benjamin R. Fisher, Ph.D. Director Division of Reproductive. Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
Irvine Scientific
.
#### INDICATIONS FOR USE STATEMENT (page 1 of 1) র্ব
# 510(K) Number: K121128
Multipurpose Handling Medium™ (MHM™) Device Name:
#### Indications for Use:
Multipurpose Handling Medium™ (MHM™) is intended for use in assisted reproductive procedures which involve the manipulation of gametes or embryos. Specifically, Multipurpose Handling Medium™ (MHM™) is indicated for use as an oocyte retrieval medium during ovarian follicle aspiration procedures (not for flushing ovarian follicles), washing sperm prior to IVF and ICSI fertilization procedures, and for transport of the embryo to the uterus during embryo transfer procedures.
Prescription Use (Part 21 CFR § 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR § 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of _
Benjamin R. Folds 06 June 2012
(Division Sign-Off)
astro-Renal, and
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.