BD PosiFlush™ Normal Saline Flush Syringes are intended to be used only for the flushing of indwelling vascular access devices. 10ml BD PosiFlush™ Normal Saline. Flush Svringes are generally compatible for use with syringe pumps.
Device Story
Three-piece, sterile, single-use syringe pre-filled with 0.9% Sodium Chloride Injection, USP; sealed with tip cap. Modified from predicate by changing stopper material and adding extra thread to plunger rod/stopper insert. Used in clinical settings for flushing indwelling vascular access devices (VADs) to maintain patency. 10ml configuration compatible with syringe pumps. Operated by clinicians. Provides mechanical flushing action to clear VADs; prevents occlusion; maintains catheter function.
Clinical Evidence
Bench testing only. No clinical data. Performance verified through functional testing (container closure integrity, break loose/out/sustaining forces, pump force, dead space/expelled volume, syringe-induced reflux) and chemical/biocompatibility testing (USP <85>, <788>, <11>, <231>, <241>, <851>, <791>; ISO 10993-3, -4, -5, -10, -11). Results demonstrate equivalence to predicate.
Technological Characteristics
Three-piece syringe with 6% Luer connector. Materials identical to predicate. Sterilized via moist heat (SAL 10^-6). Complies with USP 34<11> for saline and ISO 10993 for biocompatibility. Mechanical operation; no software or electronic components.
Indications for Use
Indicated for flushing indwelling vascular access devices in patients requiring maintenance of catheter patency. Prescription use only.
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
{0}------------------------------------------------
KI2050
## 510(k) Summarv
FEB 7 2013
#### 】. Submitted Bv:
Avital Merl Staff Regulatory Affairs Specialist BD Medical - Medical Surgical Systems I Becton Drive Franklin Lakes. NJ 07417 201 847 4739 Tel: 201 847 5307 Fax:
#### ri Device Name: .
Trade Name: 1) 0.9% Sodium Chloride Injection, USP, BD PosiFlush Flush Syringe 2) 0.9% Sodium Chloride Injection, USP, BD PosiFlush SF Flush Syringe
0.9% Sodium Chloride Injection Flush Syringe Common Name: Classification Name: Saline, Vascular Access Flush Class II, 21 CFR § 880.5200 Classificiation:
#### 3. Predicate Device:
Trade Name: 0.9% Sodium Chloride Injection, USP, BD Pre-Filled Flush Syringe Manufacturer: Becton, Dickinson and Company 510(k) Number: K003553
Trade Name: 0.9% Sodium Chloride Injection, USP, BD PosiFlush SF Flush Syringe Becton. Dickinson and Company Manufacturer: 510(k) Number: K042061
#### 4. Device Description:
The modified BD PosiFlush Flush Syringe is a three-piece, sterile, single use syringe with a 6% (Luer) connector pre-filled with 0.9% Sodium Chloride Injection, USP and sealed with a tip cap. BD has altered the modified device from the predicate by changing the stopper material and adding an extra thread to the plunger rod, threaded stopper insert. All performance characteristics are equivalent to the predicate device. The 0.9% Sodium Chloride Injection, USP BD PosiFlush Flush Syringes are intended for use in maintaining patency of vascular access devices (VAD's).
This submission represents two 2 devices: the BD PosiFlush Flush Syringe, which is provided with a sterile fluid path; and the BD PosiFlush Flush SF Syringe, which is provided externally sterile for use on a sterile field. Both configurations are sterilized via moist heat.
{1}------------------------------------------------
#### Intended Use: ri
BD PosiFlush™ Normal Saline Flush Syringes are intended to be used only for the flushing of indwelling vascular access devices. 10ml BD PosiFlush™ Normal Saline. Flush Svringes are generally compatible for use with syringe pumps.
#### Technological Characteristics: 6.
The principal device of this 510(k) premarket notification is the result of a design change to the predicate device (K003553 and K042061) conducted in accordance with Quality System Regulations, 21 CFR § 820. The BD PosiFlush Flush Syringe is Substantially Equivalent to the predicate device, given that:
- a) The BD PosiFlush Flush Syringe has the same intended use as the predicate device
- b) The BD PosiFlush Flush Syringe operates under the same operating principle as the predicate device
- c) The BD PosiFlush Flush Syringe barrel and stopper use a design identical to the predicate device.
- d) The BD PosiFlush Flush Syringe barrel and plunger rod use materials identical to the predicate device.
- e) The BD PosiFlush Flush Syringe pre-filled 0.9% Sodium Chloride Injection solution meets the requirements of USP 34<1 1>: Sodium Chloride Injection, as does the predicate device.
- The BD PosiFlush Flush Syringe meets the requirements of ISO 10993 as applicable f) to the intended use of the device, as does the predicate device.
- g) The BD PosiFlush Flush Syringe are sterilized to an SAL of 106 (sterile fluid path or externally sterile) via a Moist Heat sterilization process, as does the predicate device.
- h) The BD PosiFlush Flush Syringe demonstrated equivalent performance to the predicate device during design verification testing.
#### Performance: 6.
Design Verification testing included the following:
| Performance Characteristic | Acceptance Criteria |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| Functional Testing | |
| Container Closure Integrity | No Dye in Solution;<br>No Leakage in the luer well or tip threads;<br>No Leakage Past the Stopper Ribs;<br>No Dye Between Stopper Ribs |
| Break Loose Force | Equivalence to Predicate |
| Break Out Force | Equivalence to Predicate |
| Sustaining Force | Equivalence to Predicate |
| Pump Force | 10ml/hr – 20N<br>1ml/hr - 13N |
# Table VI - Modified Device Performance
{2}------------------------------------------------
.
| | 0.1ml/hr - 9N |
|----------------------------------------|----------------------------------------------------|
| Dead Space / Expelled Volume | Equivalent to labeled volume (3ml. 5ml or 10ml) |
| Syringe Induced Reflux | 0 average reflux when connected to a 4 Fr catheter |
| Sodium Chloride Injection, USP Testing | |
| Bacterial Endotoxin | Per USP <85> |
| Particulate Matter | Per USP <788> |
| Assay of NaCl | Per USP <11> |
| Heavy Metals | Per USP <231> |
| Iron | Per USP <241> |
| UV/vis | Per USP <851> |
| pH | Per USP <791> |
| Biocompatibility Testing | |
| Test | Method/Acceptance |
| Cytotoxicity | Per ISO10993-5:1999, Non-Toxic |
| Hemolysis | Per ISO10993-4:2002/A:2006, Non-Toxic |
| Acute Systemic Toxicity | Per ISO10993-11:2006, Non-Toxic |
| Intracutaneous Reactivity | Per ISO10993-10:2002/A1:2006, Non-Irritant |
| Sensitization | Per ISO10993-10:2002/A1:2006, Non-Sensitizer |
| Bacterial Mutagenicity | Per ISO10993-3, Non-Mutagenic |
| In Vitro Mouse Lymphoma | Per ISO10993-3, Non-Mutagenic |
| Mouse Embryo Assay | Per ISO10993-3, Non-Mutagenic |
| Occular Irritation | Per ISO10993-10:2002/A1:2006, Non-Irritant |
| Rabbit Pyrogen | Per ISO10993-11:2006, Non-Pyrogenic |
| Subchronic Intracutaneous Toxicity | Per ISO10993-11:2006, Non-Toxic. |
| Chemical Extractables Analysis | LC/DAD/MS & GC/MS. No significant extractables |
Design Verification tests were performed based on the risk analysis conducted, and the results of these tests demonstrate that the BD PosiFlush Flush Syringe performed in an equivalent manner to the predicate device and is safe and effective when used as intended.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's commitment to health, human services, and well-being. The eagle is positioned within a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged around the circle's perimeter.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
### February 7, 2013
Ms. Avital Merl Staff Regulatory Affairs Specialist Becton, Dickinson and Company One Becton Drive FRANKLIN LAKES NJ 07417
Re: K121050
Trade/Device Name: 0.9% Sodium Chloride Injection, USP BD PosiFlush Flush Syringe 0.9% Sodium Chloride Injection, USP BD PosiFlush SF Flush
Svringe Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: II Product Code: NGT Dated: February 1, 2013
Received: February 4, 2013
Dear Ms. Merl:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Merl
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Anthony Lo m
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use Statement
# 510(k) Number (if known): K121050
1
Device Name:
0.9% Sodium Chloride Injection, USP BD PosiFlush Flush Syringe
| Model Number | Syringe Size/Fill Volume |
|--------------|----------------------------------------------------------|
| 306544 | 3mL BD PosiFlush™ SP |
| 306545 | 5mL BD PosiFlush™ SP |
| 306546 | 10mL BD PosiFlush™ SP with Regular Plunger Rod |
| 306547 | 10mL BD PosiFlush™ SP |
| 306553 | 10mL BD PosiFlush™ SF with Regular Plunger Rod |
| 306549 | 3mL BD PosiFlush™ SP with BPC (Blunt Plastic<br>Cannula) |
| 306550 | 5mL BD PosiFlush™ SP with BPC |
| 306551 | 10mL BD PosiFlush™ SP (Regular Plunger Rod) with<br>BPC |
Over-The-Counter Use
(21 CFR 801 Subpart C)
Indications for Use:
Prescription Use
(Part 21 CFR 801 Subpart D)
X
BD PosiFlush™ Normal Saline Flush Syringes are intended to be used only for the flushing of indwelling vascular access devices. Catalog No. 306547 10ml BD PosiFlush™ Normal Saline Flush Syringes are generally compatible for use with syringe pumps.
AND/OR
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE OF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) |
|--------------------------------------------------------|
|--------------------------------------------------------|
| | Elizabeth F. Claverie |
|----------------|----------------------------------------------|
| | 2013.02.07 11:32:28 -05'00' |
| Page __ of __ | (Division Sign-Off) |
| | Division of Anesthesiology, General Hospital |
| | Infection Control, Dental Devices |
| 510(k) Number: | |
Page 10 of 249
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.