ILLUMIGENE GROUP B STREPTOCOCCUS CARROT BROTH CLAIM
K121044 · Meridian Bioscience, Inc. · NJR · May 1, 2012 · Microbiology
Device Facts
Record ID
K121044
Device Name
ILLUMIGENE GROUP B STREPTOCOCCUS CARROT BROTH CLAIM
Applicant
Meridian Bioscience, Inc.
Product Code
NJR · Microbiology
Decision Date
May 1, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3740
Device Class
Class 1
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K121044 · May 1, 2012
ILLUMIGENE GROUP B STREPTOCOCCUS CARROT BROTH CLAIM
Meridian Bioscience, Inc.
Clinical patient samples (vaginal/rectal swabs) from routine clinical practice
Clinical performance (sensitivity and specificity) was established by comparing the device results against standard GBS bacterial culture results from patient samples collected at multiple clinical sites.
Antepartum women (ages 15-48); Sample Size: 600; Number of Sites: 3
GBS bacterial culture
Clinical sensitivity and specificity
Indications for Use
The illumigene Group B Streptococcus (GBS) assay, performed on the illumipro-10™, is a qualitative in vitro diagnostic for the detection of Streptococcus agalactiae in enriched cultures obtained from vaginal/rectal swab specimens from antepartum women. Enriched cultures are obtained by 18-24 hour incubation of vaginal/rectal swab specimens in selective broth medium, either Lim Broth, TransVag Broth or Carrot Broth. The illumigene GBS assay utilizes loop-mediated isothermal DNA amplification (LAMP) technology to detect Streptococcus agalactiae by targeting a segment of the Streptococcus agalactiae genome. Results from the illumigene GBS assay can be used as an aid in establishing the GBS colonization status of antepartum women. This assay is not intended to diagnose or monitor treatment for GBS infections. The illumigene GBS assay does not provide susceptibility results. Culture isolates are needed for performing susceptibility testing as recommended for penicillin-allergic women. illumigene Group B Streptococcus is intended for use in hospital, reference or state laboratory settings. The device is not intended for point-of-care use.
Device Story
The illumigene GBS assay is a molecular diagnostic test for Streptococcus agalactiae colonization in antepartum women. Input consists of vaginal/rectal swab specimens enriched in selective broth (Lim, TransVag, or Carrot). The process involves manual sample preparation followed by automated isothermal amplification and detection on the illumipro-10 system. The system uses Loop-mediated isothermal DNA amplification (LAMP) to target a specific genome segment; magnesium pyrophosphate precipitate forms during amplification, changing light transmission through the reaction mixture. The illumipro-10 measures this change to report positive or negative results. Used in hospital, reference, or state laboratories by trained personnel. The output aids clinicians in determining GBS colonization status to guide intrapartum antibiotic prophylaxis; it does not monitor treatment or provide susceptibility data.
Clinical Evidence
Clinical performance established via two studies comparing illumigene GBS to bacterial culture. Study 1 (Lim/TransVag Broth, n=826) reported 97.4% sensitivity and 92.3% specificity. Study 2 (Carrot Broth, n=600) reported 100% sensitivity and 94.6% specificity. Combined data (n=1426) showed 98.6% sensitivity and 93.2% specificity. Discrepant analysis using alternate molecular methods supported the results.
Technological Characteristics
Qualitative in vitro diagnostic using Loop-mediated isothermal DNA amplification (LAMP). Components: illumigene GBS Test Device, Heat Treatment Tube, Control Reagent, Reaction Buffer. Instrumentation: illumipro-10 Automated Isothermal Amplification and Detection System. Detection via visible light transmission change caused by magnesium pyrophosphate precipitation. No calibration required.
Indications for Use
Indicated for qualitative detection of Streptococcus agalactiae in enriched vaginal/rectal swab cultures from antepartum women. Not for point-of-care use; requires prescription. Does not provide susceptibility results.
Regulatory Classification
Identification
Streptococcus spp. serological reagents are devices that consist of antigens and antisera (excluding streptococcal exoenzyme reagents made from enzymes secreted by streptococci) used in serological tests to identify Streptococcus spp. from cultured isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by bacteria belonging to the genus Streptococcus and provides epidemiological information on these diseases. Pathogenic streptococci are associated with infections, such as sore throat, impetigo (an infection characterized by small pustules on the skin), urinary tract infections, rheumatic fever, and kidney disease.
Predicate Devices
illumigene Group B Streptococcus (GBS) Assay (K112125)
Submission Summary (Full Text)
{0}------------------------------------------------
| Image: Meridian Bioscience, Inc. logo | illumigene® Group B Streptococcus (GBS) DNA Amplification Assay | |
|---------------------------------------|-----------------------------------------------------------------|----------------|
| | Application Number: | K121044 |
| | Application Section 2: | 510(k) Summary |
## 510(k) Substantial Equivalence Determination Decision Summary
MAY - 1 2002
## A. 510(k) number: K121044
- B. Purpose for Submission:
To determine substantial equivalence for the illumigene® Group B Streptococcus (GBS) DNA Amplification Assay used for the qualitative detection of Streptococcus agalactiae.
#### C. Measurand:
Segment of the Streptococcus agalactiae genome
#### D. Type of Test:
Qualitative in vitro diagnostic using Loop-mediated isothermal DNA amplification (LAMP) technology
#### ﻧﻨﺎ Applicant:
Meridian Bioscience, Inc.
#### F. Propriety and Established Names:
illumigene® Group B Streptococcus (GBS) DNA Amplification Assay
#### G. Regulatory Information:
| Product Code | Classification | Regulation Section | Panel |
|-------------------------------------------------------------------------------------------|----------------|--------------------|-------------------|
| NJR, Nucleic acid amplification assay system,<br>Streptococcus spp., serological reagents | Class I | 21 CFR § 866.3740 | Microbiology (83) |
#### H. Intended Use:
#### 1. Intended use(s):
The illumigene Group B Streptococcus (GBS) assay, performed on the illumipro-10, is a qualitative in vitro diagnostic for the detection of Streptococcus agalactiae in enriched cultures obtained from vaginal/rectal swab specimens from antepartum women. Enriched cultures are obtained by 18-24 hour incubation of vaginal/rectal swab specimens in selective broth medium, Lim Broth, TransVag Broth or Carrot Broth.
The illumigene GBS assay utilizes loop-mediated isothermal DNA amplification (LAMP) technology to detect Streptococcus agalactiae by targeting a segment of the Streptococcus agalactive genome. Results from the illumigene GBS assay can be used as an aid in establishing the GBS colonization status of antepartum women. This assay does not diagnose or monitor treatment for GBS infections. The illumigene GBS assay does not provide susceptibility results. Culture isolates are needed for performing susceptibility testing as recommended for penicillin-allergic women.
{1}------------------------------------------------
| Image: Meridian<br>Bioscience, Inc. | illumigene® Group B Streptococcus (GBS) DNA Amplification Assay | |
|-------------------------------------|-----------------------------------------------------------------|--|
| Application Number: | K121044 | |
| Application Section 2: | 510(k) Summary | |
illumigene Group B Streptococcus is intended for use in hospital, reference or state laboratory settings. The device is not intended for point-of-care use.
#### 2. Indication(s) for use:
The illumiaene Group B Streptococcus (GBS) assay, performed on the illumipro-10, is a qualitative in vitro diagnostic for the detection of Streptococcus agalactiae in enriched cultures obtained from vaginal/rectal swab specimens from antepartum women. Enriched cultures are obtained by 18-24 hour incubation of vaginal/rectal swab specimens in selective broth medium, Lim Broth, TransVag Broth or Carrot Broth.
The illumigene GBS assay utilizes loop-mediated isothermal DNA amplification (LAMP) technology to detect Streptococcus agalactioe by targeting a segment of the Streptococcus agalactiae genome. Results from the illumigene GBS assay can be used as an aid in establishing the GBS colonization status of antepartum women. This assay does not diagnose or monitor treatment for GBS infections. The illumigene GBS assay does not provide susceptibility results. Culture isolates are needed for performing susceptibility testing as recommended for penicillin-allergic women.
illumigene Group B Streptococcus is intended for use in hospital, reference or state laboratory settings. The device is not intended for point-of-care use.
#### 3. Special Conditions for use statement(s):
- For Prescription Use Only .
- The device is not intended for point-of-care use ●
#### 4. Special instrument requirements:
illumioro-10™ Automated Isothermal Amplification and Detection System
#### l. Device Description:
The illumigene Molecular Diagnostic Test System is comprised of the illumigene® Group B Streptococcus (GBS) DNA Amplification Test Kit, the illumigene Group B Streptococus (GBS) External Control Kit and the illumipro-10™ Automated Isothermal Amplification and Detection System.
The illumigene Group B Streptococcus (GBS) DNA amplification assay utilizes loop-mediated isothermal amplification (LAMP) technology to detect the presence of Streptococcus agalactiae in enriched cultures obtained from vaginal/rectal swab specimens taken from antepartum women. Each illumigene GBS assay is completed using illumigene Control Reagent, illumigene Reaction Buffer, an illumigene GBS Test Device and an illumigene Heat Treatment Tube. Samples are diluted with the illumigene Control Reagent, target DNA is made available for isothermal amplification via heat-treatment in the illumigene Heat Treatment Tube and DNA amplification occurs in the illumigene GBS Test Device.
The illumipro-10 heats each illumigene GBS Test Device containing prepared samples and Control Reagent, facilitating amplification of target DNA. When S. agalactiae is present in the enriched culture sample, a conserved sequence of the S. ggdractiae is amplified and magnesium pyrophosphate is formed. Magnesium pyrophosphate forms a precipitate in the reaction mixture. The illumipro-10 detects the change in light transmission through the reaction mixture created by the precipitating magnesium pyrophosphate. Sample results are reported as Positive or Negative based on the detected change in light transmission.
{2}------------------------------------------------
| Image: Meridian Bioscience, Inc. logo | illumigene ® Group B Streptococcus (GBS) DNA Amplification Assay |
|---------------------------------------|-------------------------------------------------------------------|
| | Application Number: K121044 |
| | Application Section 2: 510(k) Summary |
The illumigene Group B Streptococcus (GBS) External Control Kit consists of a Positive Control Reagent and a Negative Control Reagent. External Control reagents are provided to aid the user in detection of reagent deterioration, adverse environmental or test conditions, or variance in operator performance that may lead to test errors. The illumigene Group B Streptococcus External Control Kit is required for routine Quality Control.
#### J. Substantial Equivalence Information:
- 1. Predicate device name(s): illumigene Group B Streptococcus (GBS) Assay
- 2. Predicate 510(k) numbers: K112125
- 3. Comparison with predicates:
| Similarities | | |
|---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------|
| Item | DEVICE<br><i>illumigene</i> ® GBS<br>K121044 | PREDICATE<br><i>illumigene</i> ® GBS<br>K112125 |
| Intended Use | Qualitative | Same |
| Test Principle | DNA Amplification Assay | Same |
| DNA Amplification<br>Technology | Loop-Mediated Isothermal Amplification (LAMP) | Same |
| Target Sequences<br>Detected | 213 base pair (bp) sequence residing in the 593-805<br>bp region of <i>S. agalactiae</i> genome Segment 3 | Same |
| Specimen Types | Vaginal/Rectal Swab Specimen Enriched in Lim Broth Vaginal/Rectal Swab Specimen Enriched in TransVag Broth | Same |
| Reagents/Components | <i>illumigene</i> Control Reagent <i>illumigene</i> Reaction Buffer <i>illumigene</i> GBS Test Device <i>illumigene</i> Heat Treatment Tubes | Same |
| Amplification | Self contained and automated | Same |
| Detection | Self contained and automated | Same |
| Testing Time | 60 minutes | Same |
| Instrumentation | illumipro-10™ Automated Isothermal Amplification<br>and Detection System | Same |
| Reading Method | Visible Light Transmission | Same |
| Calibration | Not required | Same |
{3}------------------------------------------------
| Image: Meridian Bioscience, Inc. logo | illumigene® Group B Streptococcus (GBS) DNA Amplification Assay |
|---------------------------------------|-----------------------------------------------------------------|
| | Application Number: K121044 |
| | Application Section 2: 510(k) Summary |
| Differences | | |
|----------------------------------------|----------------------------------------------------------------|----------------------------------------------------------------|
| Item | DEVICE<br><i>illumigene</i> ® GBS<br>K121044 | PREDICATE<br><i>illumigene</i> ® GBS<br>K112125 |
| Specimen Types | Vaginal/Rectal Swab Specimen Enriched in Carrot<br>Broth | N/A |
| Performance<br>Sensitivity Specificity | 98.6% [95% CI: 96.5% - 99.5%]<br>93.2% [95% CI: 91.6% - 94.5%] | 97.4% [95% CI: 91.9% - 99.0%]<br>92.3% [95% CI: 90.0% - 94.1%] |
#### Standard/Guidance Document Referenced (if applicable): K.
- Clinical and Laboratory Standards Institute. 2008. User Protocol for Evaluation of Qualitative Test Performance; . Approved Guideline- Second Edition (EP12-A)
- Clinical and Laboratory Standards Institute. 2005. User Verification of Performance for Precision and Trueness; . Approved Guideline- Second Edition (EP15-A2)
- Clinical and Laboratory Standards Institute. 2005. Interference Testing in Clinical Chemistry; Approved Guideline-. Second Edition (EP7-A2)
### L. Test Principle:
The illumigene Group B Streptococcus assay is based on loop-mediated isothermal amplification technology (LAMP). Loop-mediated amplification of DNA is accomplished by the use of specially designed primers that provide specific and continuous isothermal amplification. Magnesium-pyrophosphate is a by-product of LAMP amplification. The magnesium-pyrophosphate forms a white precion solution, giving the reaction solution a turbid appearance. Change in sample absorbance created by precipitation of magnesium pyrophosphate indicates the presence of target DNA and is considered a positive reaction. The absence of target DNA results in no detectable change in sample absorbance and is considered a negative reaction.
### M. Performance Characteristics (if/when applicable):
- 1. Analytical Performance:
- a. Precision/Reproducibility:
Reproducibility studies were completed with K12225. Blind coded panels of 10 samples were supplied to three independent laboratories. Samples were randomly sorted within each panel to mask sample identities. The panels included contrived samples manufactured as low positive samples (i.e. limit of detection, n = 3) and high negative samples (n = 3). The panels also included contrived positive (n = 3) samples and natural negative samples (n = 1). Testing was performed by different operators at each site on the same day (intra-assay variability) for five days (interassay variability). Three lots of illumipro-10 instruments were used in this study. The results are given in the table below:
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Meridian Bioscience, Inc. The logo consists of a globe-like graphic on the left, followed by the word "Meridian" in a bold, stylized font. Below "Meridian" is the text "Bioscience, Inc." in a smaller, also bold font. The overall design is simple and professional.
| illumigene® Group B Streptococcus (GBS) DNA Amplification Assay | |
|-----------------------------------------------------------------|----------------|
| Application Number: | K121044 |
| Application Section 2: | 510(k) Summary |
| | Site 1 | | Site 2 | | Site 4 | | Total | |
|---------------|--------|-------------------|--------|-------------------|--------|-------------------|-------|-------------------|
| Sample Type | | Percent agreement | | Percent agreement | | Percent agreement | | Percent agreement |
| Negative | 10/10 | 100% | 10/10 | 100% | 10/10 | 100% | 30/30 | 100% |
| High Negative | 30/30 | 100% | 30/30 | 100% | 30/30 | 100% | 90/90 | 100% |
| Low Positive | 30/30 | 100% | 30/30 | 100% | 30/30 | 100% | 90/90 | 100% |
| Positive | 30/30 | 100% | 30/30 | 100% | 30/30 | 100% | 90/90 | 100% |
#### Linearity/assay reportable range: b.
Not applicable as this assay is qualitative
#### Traceability, Stability, Expected values (controls, calibrators, or methods): ﻥ
#### Stability:
Stability studies were completed with K112125 and supplemented for K121044. Sample storage and hold time studies were performed to characterize illumiqene® Group B Streptococcus (GBS) assay ranges. Validation studies performed at Meridian indicated that Lim Broth represented the worst-case matrix; therefore studies were completed using Lim Broth. Assay ranges were confirmed for the remaining TransVag Broth and Carrot Broth by testing one high negative and one LoD positive sample prepared from each broth.
Study results demonstrated that undiluted samples can be held at 19 -27 C for up to 6 hours or at 2-8 C for up to 7 days after culture enrichment and prior to testing. Samples diluted with illumigene® Control Reagent can be held at 21 -27 C for up to 15 minutes prior to heat-treated samples may be held at 19-29 C for up to 45 minutes prior to further processing.
Sample freeze/thaw studies were also performed. Samples were enriched 18-24 hours and frozen prior to initiation of testing. Samples were subject to multiple freeze/thaw cycles. Study results showed that samples should not be freeze/thawed.
Final testing demonstrated that Lim Broth, in combination with Dacron/polyester, flocked nylon and foam swab types perform acceptably with the illumigene® GBS assay. Amies Clear and Calcium Alginate Swabs produced invalid results and are not considered validated for use with illumigene® GBS. Swab types with demonstrated performance in the illumigene® GBS assay appear in the Package Insert.
#### d. Detection limit:
Detection limit studies were completed with K12225. Sensitivity studies were designed to determine the analytical limit of detection of S. agalactiae in Lim Broth. Six common strains of S.agalactiae, representing six serotypes, were evaluated with the illumigene® GBS assay. Each strain was spiked into negative Lim Broth and diluted serially. A minimum of twenty replicates for each dilution were individually processed and tested to establish LoD. Testing was performed using three production lots of illumigene® GBS assay and six illumipro-10™ instruments. External Positive and Negative Controls were tested each day throughout the study.
The LoD of the assay ranged from 60 CFU/Test to 1280 CFU/Test. The table below shows the lowest concentration for each serotype that produced positive results for 95% of the replicates tested.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Meridian Bioscience, Inc. The logo features a globe-like icon on the left, followed by the word "Meridian" in a bold, italicized font. Below "Meridian" is the text "Bioscience, Inc." in a smaller, bold font. A horizontal line separates "Meridian" from "Bioscience, Inc."
| illumigene® Group B Streptococcus (GBS) DNA Amplification Assay | |
|-----------------------------------------------------------------|----------------|
| Application Number: | K121044 |
| Application Section 2: | 510(k) Summary |
| Serotype | Streptococcus agalactiae Strain<br>Description | CFU/Test |
|----------|------------------------------------------------|----------|
| Ia | NCTC 11248 | 60 |
| Ib | ATCC 12401 | 80 |
| Ic | NCTC 11253 | 640 |
| II | II/2 | 320 |
| III | ATCC 12403 | 160 |
| V | ATCC BAA-611 | 1280 |
Additional S. agoloctioe strains were tested and produced positive reactions at 1280 CFU/test with Strains and serotypes were tested as follows: Serotype IV: NCTC 11930, hemolytic; illumiaene GBS. Serotype VIa: NCTC 08188, non-hemolytic; Serotype VII: VII/2, hemolytic; Serotype VIII: VIII/2, hemolytic; Serotype X: NCTC 11249, hemolytic; Unknown Serotype: ATCC 13813, non-hemolytic; and ATCC12386, hemolytic.
Limit of Detection studies are acceptable.
#### e. Analytical specificity:
#### Interference Testing:
Interference testing was completed with K112125. Selected substances that might be expected to be present in vaginal/rectal swab samples taken from antepartum women were added to a negative Lim broth sample and two contrived positive Lim broth sample was prepared by pooling confirmed negative Lim Broth samples while the contrived positive samples were prepared by spiking a pooled, confirmed negative Lim sample with either Streptococcus agalactiae, strain 11248 Serotype la (123 CFU/test) or Streptococcus agalactiae, strain 12401 Serotype Ib (80 CFU/test). Potentially interfering substances were added to Lim broth samples at final concentrations of 2.5% v/v or greater when the substances could be pipetted. Substances that could not be pipetted were coated onto cotton swabs, immersed in the negative/positive Lim broth samples and tested. Dilution Controls were prepared by adding a phosphate buffered saline solution in place of the potentially cross-reactive organisms. Each inoculated sample was tested in triplicate.
The following substances, at the specified saturated solvent/diluents concentrations, do not interfere with illumigene Group B Streptococcus test results: Amniotic fluid (10% v/v), Human DNA (100 ng/Test), Urine (30%v/v), Whole Blood (2.5% v/v). The following substances do not interfere with test results: Meconium, Stool, Hemorrhoidal cream (30.65 mg/100mg), Miconazole (fungicide), Mucin (0.5-1.5%), Spermicidal gel (nonoxynol 9) (4 mg/100mg). Lubricating gel produced False Negative Results in 1 of 11 replicates tested. Body Powder produced False Negative Results in 1 of 10 replicates tested. Whole Blood at concentrations greater than 2.5% v/v interferes with the illumigene GBS assay.
Interference studies are acceptable.
#### Cross-Reactivity Study:
Cross-reactivity studies were completed with K12125. Potentially cross-reacting microorganisms expected to be present in vaginal/rectal swab specimens were added to negative and contrived positive Lim Broth The negative sample was prepared by pooling confirmed negative Lim Broth samples. The samples. contrived positive sample was prepared by spiking a confirmed negative matrix with Streptococcus
{6}------------------------------------------------
| Image: Meridian Bioscience, Inc. logo | illumigene® Group B Streptococcus (GBS) DNA Amplification Assay |
|---------------------------------------|-----------------------------------------------------------------|
| | Application Number: K121044 |
| | Application Section 2: 510(k) Summary |
agoloctioe, strain 12401, at 122 CFU/test, near the limit of detection for this strain. Potentially crossreactive microorganisms were added at concentrations of 1.2 x 10° CFU/mL (bacteria and fungi) or virus at a minimum of 1 x 10 TCIDso/mL (viruses). Dilution controls for each sample were prepared by adding sterile saline solution in place of the potentially cross-reactive microorganisms. Each sample was tested in triplicate. Cross-reactivity with Enterococcus dispar was observed in one of seven replicates tested.
The following microorganisms at the indicated concentrations do not interfere with illumigene GBS: Aeromonas hydrophila, Alcaligenes faecalis, Bacillus subtilis, Bacteroides fragilis, Campylobacter coli, Campylobacter fetus, Campylobacter jejuni, Candida albicans, Candida glabrata, Candida krusei, Candida parapsilosis, Citrobacter freundii, Clostridium bifermentans, Clostridium butyricum, Clostridium difficile, Clostridium histolyticum, Clostridium novyi, Clostridium perfringens, Clostridium septicum, Clostridium sordelli, Clostridium sporogenes, Clostridium tetani, Corynebacterium genitalium, Corynebacterium urealyticum, Corynebacterium xerosis, Enterobacter aerogenes, Enterobacter cloocae, Enterococcus avium, Enterococcus durans, Enterococcus faccalis, Enterococus faecium, Escherichia coli, Escherichia coli 0157:H7, Escherichia fergusonii, Escherichia hermannii, Gardnerella vagindiis ducreyi, Helicobacter pylori, Klebsiella pneumoniae, Lactobacillus acidophilus, Lactobacillus brevis, Lactobacillus delbruekii subspecies Iactis, Lactobacillus jensenii, Lactococcus lactis, Legionella pneumophila, Listeria monocytogenes, Moraxella osloensis, Morganella morganii, Neisseria gonorrhoeae, Peptostreptococcus anaerobius, Plesiomonas shigelloides, Porphyromonas asaccharolytica, Prevotella melaninogenica, Propionibacterium acnes, Proteus mirabilis, Proteus vulgaris, Pseudomonas aeruginosa, Pseudomonus fluorescens, Salmonella Group B, Salmonella Group C, Salmonella Group D, Salmonella Group E, Serratia marcescens, Shigella boydii. Shigella flexneri. Staphylococcus aureus, Staphylococcus aureus (Cowan), Staphylococcus epidermidis, Staphylococcus saprophyticus, Streptococus anginosus, Streptococcus bovis, Streptococcus dysgalactiae equisimilis, Streptococcus intermedius, Streptococcus mitis, Streptococcus ordis, Streptococcus pneumoniae, Streptocacus pyogenes, Streptococcus salivarius, Streptococcus sanguinis, Vibrio parahaemolyticus, Yersinia enterocolitica, Adenovirus 41, BK virus, Coxsackievirus, Echovirus, Epstein Barr virus, Herpes Simplex Virus-1, Herpes Simplex Virus-2, Rotavirus.
Mycoplasma genitalium, Mycoplasma hominis and Ureaplasma urealyticum were tested at final concentrations ranging between 1.6 x 10 and 9.9 x 10 CFU/mL with no reaction with the illumigene GBS assay.
Cross-reactivity studies are acceptable.
- f. Assay cut-off:
The illumigene Group B Streptococcus Assay has a fixed cut-off based on the measured change in light transmission at the assay endpoint. The illumipro-10 measures transmission of light through the Test and the Control reactions at the start of the Assay Run (Signalmon) and at the end of the Assay Run (Signalma). The illumipro-10 calculates that change in transmission between the Signalmar and compares the result to a fixed cut-off value. Test results are reported as Positive based on comparison to the assay cut-off. Fixed cut-off values were based on well characterized clinical specimens.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Meridian Bioscience, Inc. The logo consists of a globe-like graphic on the left, followed by the word "Meridian" in a bold, sans-serif font. Below "Meridian" is the phrase "Bioscience, Inc." in a smaller, also bolded, sans-serif font. A horizontal line underlines the word "Meridian".
#### illumigene® Group B Streptococcus (GBS) DNA Amplification Assay
| Application Number: | K121044 | | |
|-------------------------|----------------|--|--|
| Application Section 2: | 510(k) Summary | | |
#### 2. Comparison Studies:
- Method comparison with predicate device: 0.
Not applicable
#### b. Matrix comparison:
Not applicable
#### 3. Clinical Studies:
- Clinical Sensitivity: a.
Clinical trials for the illumigene Group B Streptococcus (GBS) assay, including the illumipro-10 Automated Isothermal amplification and detection system, were conducted in 2012. Performance characteristics of the illumigene GBS assay were determined by comparison to GBS bacterial culture in two separate studies: (1) Lim and TransVag Broth Enrichment, reference K112125; and (2) Carrot Broth Enrichment. Combined performance data for all studies and all enrichment broth types is shown in Table 1.
| Sample Type | Positive Samples | | | Negative Samples | | |
|----------------|-----------------------------|---------------|---------------|-----------------------------|---------------|--------------|
| | illumigene /<br>GBS Culture | Sensitivity % | 95% CI | illumigene /<br>GBS Culture | Specificity % | 95% CI |
| Total | 285/289 | 98.6% | 96.5 - 99.5% | 1045/1121 | 93.2% | 91.6 - 94.5% |
| Lim Broth | 82/84 | 97.6% | 91.7 - 99.3% | 296/315 | 94.0% | 90.8 - 96.1% |
| TransVag Broth | 68/70 | 97.1% | 90.2 - 99.2% | 314/346 | 90.8% | 87.2 - 93.4% |
| Carrot Broth | 135/135 | 100% | 97.2 - 100.0% | 435/460 | 94.6% | 92.1 - 96.3% |
Table 1: illumigene Group B Streptococcus assay performance characteristics
(1) Lim and TransVag Broth Enrichment: illumigene GBS assay performance using Lim Broth and TransVag Broth enriched specimens was evaluated in 2011 by four independent clinical test sites located in the Midwestern and Southern regions of the United States. A total of 826 qualified patient samples were evaluated. Samples were obtained according to established guidelines for the collection of clinical specimens for culture of Group B Streptococcus and enriched for 18-24 hours in either Lim Broth or TransVag Broth prior to illumigene testing. Four hundred three (403, 48.8%) specimens were enriched with Lim Broth and 423 (51.2%) specimens were enriched with TransVag Broth prior to testing. The age groups of patients tested ranged from 15 years of age to 44 years of age, with age unknown for 3 (0.4%) of the patient population. No differences in test performance were observed based on enrichment medium or patient age. Overall assay Sensitivity was reported as 97.4% [95% Cl: 91.9% - 99.0%] where Specificity was 92.3% [95% Cl: 90.0% - 94.1%]. Table 2 shows overall assay performance reported for the study; Table 3 summarizes assay performance by Clinical Site.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Meridian Bioscience, Inc. The logo features a globe-like icon on the left, followed by the word "Meridian" in a bold, sans-serif font. Below "Meridian" is the text "Bioscience, Inc." in a smaller, also bolded, sans-serif font. A horizontal line separates "Meridian" from "Bioscience, Inc."
| illumigene® Group B Streptococcus (GBS) DNA Amplification Assay | |
|-----------------------------------------------------------------|----------------|
| Application Number: | K121044 |
| Application Section 2: | 510(k) Summary |
| Group B Streptococcus | illumigene Group B Streptococcus (GBS) | | | | |
|-----------------------|----------------------------------------|------------|--------------|--|--|
| Culture | Positive | ా Negative | Total | | |
| Positive | 150 · | য | 154 | | |
| Negative | 51 | eio | 661 | | |
| Total | 201 | 614 | 845 | | |
| | | | વેટી વિદર્ભ | | |
| Sensitivity | 150/154 | 97.4% | 91.9 - 99.0% | | |
| Specificity | 610/661 | 92.3% | 90.0 - 94.1% | | |
| Correlation | 760/815 | 93.3% | 91.3 - 94.8% | | |
#### I im and TransVas enrichment performance data
Forty-eight of the 51 false positive results were positive by another molecular method. Invalid results were obtained for 11/826 samples tested or 1.3%. Two of the 11 samples remained invalid after repeat testing of the original sample.
| Table 3: Performance characteristics summary: Lim and TransVag Broth Enrichment | | |
|---------------------------------------------------------------------------------|--|--|
| | | |
| Site<br>Identification | Enrichment<br>Broth | Positive Samples | | | Negative Samples | | |
|------------------------|---------------------|-----------------------------------|------------------|--------------|-----------------------------------|------------------|--------------|
| | | <i>illumigene</i><br>/GBS Culture | %<br>Sensitivity | 95% CI | <i>illumigene</i><br>/GBS Culture | %<br>Specificity | 95% CI |
| Total | N/A | 150/154 | 97.4% | 91.9 - 99.0% | 610/661 | 92.3% | 90.0 - 94.1% |
| Site 1 | TransVag | 32/33 | 97.0% | 84.7 - 99.5% | 197/199 | 99.0% | 96.4 - 99.7% |
| Site 3 | TransVag | 36/37 | 97.3% | 86.2 - 99.5% | 117/147 | 79.6% | 72.4 - 85.3% |
| Site 2 | Lim | 38/39 | 97.4% | 86.8 - 99.5% | 162/168 | 96.4% | 92.4 - 98.4% |
| Site 4 | Lim | 44/45 | 97.8% | 88.4 - 99.6% | 134/147 | 97.8% | 85.5 - 94.8% |
Specimens that generated discrepant results were further evaluated by independent testing laboratories using FDA cleared or laboratory validated molecular assays. Sixteen of nineteen Lim Broth False Positive results were positive by an alternate molecular method. All thirty-two TransVag Broth False Positive results were positive by an alternate molecular method. In addition to discrepant sample analysis, a selection of concordant samples was tested with non-illumigene molecular methodologies. Concordant result testing showed a combined correlation between molecular methods of 97.7%.
(2) Carrot Broth Enrichment: Performance characteristics specific to Carrot Broth Enrichment were established by studies involving three independent clinical test sites located in the Midwestern and Southern regions of the United States. Independent clinical test sites located in the Midwestern and Southern regions of the United States evaluated a total of 600 qualified patient samples were obtained according to established guidelines for the collection of clinical specimens for culture of Group B Streptococcus and enriched in Carrot Broth prior to illumigene testing. The age groups of patients tested ranged from 15 years of age to 48 years of age; no differences in test performance were observed based on patient age. Table 4 shows assay performance reported for the study; Table 5 summarizes assay performance by Clinical Site.
{9}------------------------------------------------
| illumigene® Group B Streptococcus (GBS) DNA Amplification Assay | |
|-----------------------------------------------------------------|----------------|
| Application Number: | K121044 |
| Application Section 2: | 510(k) Summary |
| Group B Streptococcus Culture | <i>illumigene</i> Group B Streptococcus (GBS) | | |
|-------------------------------|-----------------------------------------------|----------|---------------|
| | Positive | Negative | Total |
| Positive | 135 | 0 | 135 |
| Negative | 25 | 435 | 460 |
| Total | 160 | 435 | 595 |
| | | | 95% CI |
| Sensitivity | 135/135 | 100.0% | 97.2 - 100.0% |
| Specificity | 435/460 | 94.6% | 92.1 - 96.3% |
| Correlation | 570/595 | 95.8% | 93.9 - 97.1% |
Invalid results were obtained for 5/600 (0.8%) samples. One of the 5 samples remained invalid after repeat testing. Repeat test results are not included in assay performance calculations. Twenty four of the 25 illumigene Positive: GBS Culture Negative specimens were further analyzed by an independent laboratory developed GBS Molecular Assay. Sixteen of the 24 discrepant samples were positive by the alternate molecular amplification assay.
#### Table 5: illumigene Group B Streptococcus assay performance by Site
| Site<br>Identification | Positive Samples | | | Negative Samples | | | Invalid Samples | |
|------------------------|-----------------------------|------------------|---------------|----------------------------|------------------|--------------|-------------------|--------------|
| | illumigene /<br>GBS Culture | %<br>Sensitivity | 95% CI | illumigene<br>/GBS Culture | %<br>Specificity | 95% CI | Invalid<br>/Total | %<br>Invalid |
| Total | 135/135 | 100.0% | 97.2 - 100.0% | 435/460 | 94.6% | 92.1 - 96.3% | 5/600 | 0.8% |
| Site 1 | 58/58 | 100.0% | 93.8 - 100.0% | 92/93 | 98.9% | 94.2 - 99.8% | 0/151 | 0.0% |
| Site 2 | 36/36 | 100.0% | 90.4 - 100.0% | 145/149 | 97.3% | 93.3 - 99.0% | 1/186 | 0.5% |
| Site 3 | 41/41 | 100.0% | 91.4 - 100.0% | 198/218 | 90.8% | 86.3 - 94.0% | 4/263 | 1.5% |
Enriched carrot broth samples were visually inspected for color at the enrichment incubation and prior to initiation of illumigene testing. Table 3 summarizes performance data and observed broth color. Thirty four of the 135 (25.1%) illumigene/GBS Culture positive specimens were reported as showing no visual change in color at the end of enrichment incubation.
Table 6: Performance Characteristics by Carrot Broth Results
| | Positive Samples | | | Negative Samples | | |
|-----------------|-----------------------------|---------------|---------------|-----------------------------|---------------|---------------|
| Sample Type | illumigene /<br>GBS Culture | Sensitivity % | 95% CI | illumigene /<br>GBS Culture | Specificity % | 95% CI |
| Color Change | 100/100 | 100.0% | 96.3 - 100.0% | 2/2 | 100.0% | 34.2 - 100.0% |
| No Color Change | 34/34 | 100.0% | 89.8 - 100.0% | 433/458 | 94.5% | 92.1 - 96.3% |
| Unspecified | 1/1 | 100.0% | 20.7 - 100.0% | 0/0 | 0.0 | 0.0 - 0.0% |
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Meridian Bioscience, Inc. The logo is black and white and features a globe-like image on the left. To the right of the image is the company name, with "Meridian" in a larger, bold font and "Bioscience, Inc." underneath in a smaller font. There is also a table with the text "illumigene Group B", "Application Number", and "Application Section".
| illumigene® Group B Streptococcus (GBS) DNA Amplification Assay | |
|-----------------------------------------------------------------|----------------|
| Application Number: | K121044 |
| Application Section 2: | 510(k) Summary |
- Clinical Specificity: b.
See Section 3a
- Other Clinical Supportive Data C.
Not applicable
- 4. Clinical cut-off:
Not applicable
### 5. Expected values/Reference range:
Approximately 10-30% of antepartum women are colonized with Group B Streptocccus in the vagina or rectum: Clinical performance of the illumigene GBS Assay was established during two clinical studies completed in 2011 (Lim Broth and TransVag Broth) and 2012 (Carrot Broth). The overall incidence of GBS colonization in antepartum women tested during the 2011 study was 24.3% (201 of 826). Incidence of GBS colonization for enrichment performed using Lim Broth was found to be 25.1% (101 of 403); while incidence for specimens enriched by TransVag Broth was found to be 23.6% (100 of 423). The overall incidence of GBS colonization in antepartum women tested during the 2012 Carrot Broth study was found to be 22.5% (135 of 600).
### N. Other Supportive Device and Instrument Information:
Instrument: illumipro-10™
### O. System Descriptions:
System Description was reviewed in previous submission, K100818, K110012 and K12225. No system or software changes were made.
### 1. Modes of Operation:
The illumipro-10™ heats each illumigene® GBS Test Device containing prepared samples and Control Reagent, facilitating amplification of target DNA. When S. agoloction is present in the enriched culture sample, a conserved sequence of the S. agalaction is amplified and magnesium pyrophosphate is formed. Magnesium pyrophosphate forms a precipitate in the reaction mixture. The illumipro-10 detects the change in light transmission through the reaction mixture created by the precipitating magnesium pyrophosphate. Sample results are reported as Positive or Negative based on the detected change in light transmission.
- 2. Software:
FDA has reviewed applicant's Hazard Analysis and software development processes for this line of product types:
Yes Yes No
- 3. Specimen Identification:
The illumipro-10™ utilizes software to automate incubation of illumigene Molecular Diagnostic in vitro diagnostic test reactions. The illumipro-10™ reports sample results as INVALID, POSITIVE or NEGATIVE.
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for Meridian Bioscience, Inc. The logo features a globe-like icon on the left, followed by the word "Meridian" in a bold, sans-serif font. Below "Meridian" is the text "Bioscience, Inc." in a smaller, also bolded, font. The overall design is simple and professional.
| illumigene® Group B Streptococcus (GBS) DNA Amplification Assay | |
|-----------------------------------------------------------------|----------------|
| Application Number: | K121044 |
| Application Section 2: | 510(k) Summary |
### 4. Specimen Sampling and Handling:
Specimens are prepared manually. Incubation and detection are automated using the illumipro-10™.
#### 5. Calibration:
.
Calibration of the illumipro-10™ is not required.
### 6. Quality Control:
The illumigene® Group B Streptococcus External Control Kit consists of a Positive Control Reagent and a Negative Control Reagent. External Control reagents are provided to aid the user in detection of reagent deterioration, adverse environmental or test conditions, or variance in operator performance that may lead to test errors. The illumigene® Group B Streptococcus External Control Kit is required for routine Quality Control.
{12}------------------------------------------------
Statement for the Record, K 121044 Meridian BioScience, Inc. Illumigene Group B Streptococcus (GBS) DNA Amplification Assay
This 510(k) was reviewed under OIVD's Pilot Triage Program. This program represents an internal workload management tool intended to reduce internal FDA review resources for 510(k) applications that are of good quality upon receipt by FDA.
The information in the 510(k) is complete and supports a substantial equivalence (SE) determination. Please refer to the applicant's 510(k) summary for a summary of the information that supports this SE determination.
{13}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/13/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized human figure with three arms reaching upwards, next to a circular seal. The seal contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter.
#### Food and Drug Administration
10903 New Hampshire Avenue Silver Spring, MD 20993
MAY - 1 2012
Meridian Bioscience, Inc. C/o Michelle L. Smith Senior Director, RA/DA 3471 River Hills Drive Cincinnati, Ohio 45244
Re: k121044
Trade/Device Name: illumigene® Group B Streptococcus (GBS) DNA Amplification Assay Regulation Number: 21 CFR 888.3740 Regulation Name: Streptococcus spp. serological reagents Regulatory Class: Class I Product Code: NJR Dated: April 5, 2012 Received: April 6, 2012
Dear Ms. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act that 1 Driveral statutes and regulations administered by other Federal agencies. You must of any I cural statutios and registanents, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{14}------------------------------------------------
## Page 2 - Ms. Michelle L. Smith
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Sall artgins
Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics Device Evaluation and Safety Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{15}------------------------------------------------
| | illumigene® Group B Streptococcus (GBS) DNA Amplification Assay | | |
|--------------|-----------------------------------------------------------------|----------------------------|--|
| OFFICE TO | Application Number: | K121044 | |
| cience. Inc. | Application Section 1: | Indication(s) for Use Form | |
## Indication(s) for Use Form
### 510(k) Number: K121044
# Device Name: illumigene Group B Streptococcus (GBS) DNA Amplification Assay
### Indications for Use:
The illumigene Group B Streptococus (GBS) assay, performed on the illumipro-10™, is a qualitative in vitro diagnostic for the detection of Streptococcus agalactiae in enriched cultures obtained from vaginal/rectal swab specimens from antepartum women. Enriched cultures are obtained by 18-24 hour incubation of vaginal/rectal swab specimens in selective broth medium, either Lim Broth, TransVag Broth or Carrot Broth.
The illumigene GBS assay utilizes loop-mediated isothermal DNA amplification (LAMP) technology to detect Streptococus agalactiae by targeting a segment of the Streptococcus agalactiae genome. Results from the illumigene GBS assay can be used as an aid in establishing the GBS colonization status of antepartum women. This assay is not intended to diagnose or monitor treatment for GBS infections.
The illumigene GBS assay does not provide susceptibility results. Culture isolates are needed for performing susceptibility testing as recommended for penicillin-allergic women.
illumigene Group B Streptococcus is intended for use in hospital, reference or state laboratory settings. The device is not intended for point-of-care use.
Prescription Use _ AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Freddie Lee Pool, MS
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K120044 (Sol)
K121044
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.