eRAD PACS/ eRAD RIS/PACS is a PACS software product used to receive DICOM images, scheduling information and textual report, organize and store them in an internal format, and to make that information available across a network via web and customized user interfaces. The eRAD PACS/ eRAD RIS/PACS viewer software is intended for use as a primary diagnostic and analysis tool for diagnostic images. eRAD PACS/ eRAD RIS/PACS is for hospitals, imaging centers, radiologists, reading practices and any user who requires and is granted access to patient image, demographic and report information. The eRAD PACS/ eRAD RIS/PACS viewer displays images from CT, computed radiography, MRI, mammography, nuclear medicine, PET, secondary capture, ultrasound, x-ray angiography, x-ray fluoroscopy and visible light modalities. Lossy compressed mammography images and digitized film screen mammography images must not be reviewed for primary image interpretations. Mammography images may only be interpreted using an FDA approved monitor that offers at least 5 mega-pixel resolution and meets other technical specifications reviewed and accepted by FDA.
Device Story
eRAD PACS/RIS is a software-based Picture Archiving and Communication System. It receives DICOM images, scheduling data, and reports; organizes and stores them; and distributes them across networks via web/customized interfaces. Components include acquisition modules, a central systems manager, diagnostic viewers, and archiving storage. The system retrieves prior studies, registers images, and manages worklists. Radiologists use diagnostic workstations to view images and dictate reports; the system supports playback and transcription. Reports are then accessible to referring physicians or exported to other information systems. The device is used in clinical settings (hospitals, imaging centers) by radiologists and authorized personnel. It facilitates clinical decision-making by providing centralized access to patient imaging and diagnostic reports, improving workflow efficiency and diagnostic capability.
Clinical Evidence
Bench testing only. System verification and validation testing were conducted in accordance with international standards and internal design requirements to confirm performance and substantial equivalence to the predicate device.
Technological Characteristics
Software-only PACS system. Supports DICOM standard for image acquisition and communication. Web-based and customized user interfaces for image viewing and report management. Compatible with various imaging modalities (CT, MRI, PET, etc.). Requires FDA-approved 5MP monitors for primary mammography interpretation. Operates as a networked system for image/data storage and retrieval.
Indications for Use
Indicated for use by hospitals, imaging centers, radiologists, and reading practices as a primary diagnostic and analysis tool for viewing and managing medical images (CT, CR, MRI, mammography, nuclear medicine, PET, ultrasound, angiography, fluoroscopy, visible light). Contraindicated for primary interpretation of lossy compressed or digitized film-screen mammography images.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
K120995
Image /page/0/Picture/1 description: The image shows a logo for "CeRAD". The logo features a stylized lowercase "e" with curved lines extending from the top and bottom, resembling a crescent shape. To the right of the "e" is the word "RAD" in bold, uppercase letters, with a trademark symbol next to it. The overall design is simple and modern, with a focus on the company's name.
# 510(k) Summary
eRAD, Inc.
C 0 3 2012
### Submitter:
Section 5:
9 Pilgrim Road Greenville, SC 29607
# Contact Person:
Jim Connors Vice President, Product Management eRAD, Inc. O: 864-640-8664 F: 864-234-7412
# Official Correspondent:
Jillian M. Reed Regulatory and Clinical Affairs Consultant Reed Technical Associates, LLC 23 Griswold Lane Winsted, CT 06098 O: 860-238-7734 F: 203-538-3960 jillianreed@charter.net
| Date Prepared: | October 5, 2012 |
|----------------------|----------------------------------------|
| Classification Name: | System, Image Processing, Radiological |
| Common Name: | PACS |
| Proprietary Name: | eRAD PACS and eRAD RIS/PACS |
| Predicate Devices: | eRAD PACS- #K061421(July 25, 2006) |
# Device Description:
eRAD PACS/ eRAD RIS/PACS Software Product is a PACS system, comprised of acquisition components, a central systems manager component, diagnostic viewing components, and an archiving component. The data flow is such that patient and procedure information is optionally delivered to the central system manager, followed by the acquisition of the image objects directly from the image sources or by one of the acquisition components. After receiving the procedure information or after receiving image objects, the central system manager searches for and retrieves relevant prior procedure data from the archiving a component. When the central system manager registers the acquired image objects and the retrieved prior procedure data, a user can access the information by selecting the item from the operator worklist. The image data is
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for eRAD, a medical imaging software company. The logo features a stylized lowercase "e" with three curved lines emanating from it, resembling radio waves. To the right of the "e" is the word "RAD" in bold, sans-serif font. The logo is simple, modern, and easily recognizable.
transmitted to and rendered on the user's workstation using the diagnostic viewing components. After using the workstation to view the images, the user optionally dictates a report into the system, after which, a user can play back the diction and transcribe it to text. Once eRAD PACS's central system manager registers a report, the report is available for access by the referring physician, or it can be exported into an information system. At some configured point in time, the image data and the report information is delivered to the archiving component for backup and long-term storage.
### Intended Use: . .
eRAD PACS/ eRAD RIS/PACS is a PACS software product used to receive DICOM images, scheduling information and textual report, organize and store them in an internal format, and to make that information available across a network via web and customized user interfaces.
The eRAD PACS/ eRAD RIS/PACS viewer software is intended for use as a primary diagnostic and analysis tool for diagnostic images. eRAD PACS/ eRAD RIS/PACS is for hospitals, imaging centers, radiologists, reading practices and any user who requires and is granted access to patient image, demographic and report information.
The eRAD PACS/ eRAD RIS/PACS viewer displays images from CT, computed radiography, MRI, mammography, nuclear medicine, PET, secondary capture, ultrasound, x-ray angiography, x-ray fluoroscopy and visible light modalities.
Lossy compressed mammography images and digitized film screen mammography images must not be reviewed for primary image interpretations. Mammography images may only be interpreted using an FDA approved monitor that offers at least 5 mega-pixel resolution and meets other technical specifications reviewed and accepted by FDA.
#### Substantial Equivalence:
The modifications to the eRAD PACS/ eRAD RIS/PACS Software do not alter the fundamental scientific technology of the device. The only modification made to the predicate device is making the software available to the user in a software-only option.
# Discussion of Non-Clinical Testing Performed:
Thorough non-clinical system verification and validation testing was conducted in accordance with applicable international standards and internal design requirement to verify that the eRAD PACS/ eRAD RIS/PACS Software Product meet user needs and indications for use. Testing demonstrated that the eRAD PACS/ eRAD RIS/PACS Software Product were substantial equivalent to their predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a logo with the letters 'eRAD' in a stylized font. The 'e' is formed by three curved lines that resemble a stylized letter 'C' facing right. The 'RAD' is in bold, sans-serif font, with a small circle containing a registered trademark symbol to the right of the 'D'. The overall design is simple and modern.
# Conclusions:
1. 14.
1. 11.1
:
. .
:
The information provided in this premarket notification submission has shown that the eRAD PACS/ eRAD RIS/PACS Software Product is substantially equivalent to the predicate devices and are safe and effective for its intended use.
. . .
1.1
...
.
September 1998 - 1998
:
: : :
. 1.11 ..............................................................................................................................................................................
:
... . . . . :
: : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : :
:
: : :
:
. .
:
ټ
.
・・・
:
: ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・
..
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines representing its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
eRAD, Inc. % Ms. Jillian M Reed Regulatory and Clinical Affairs Consultant Reed Technical Associates, LLC 23 Griswold Lane WINSTED CT 06098
December 3, 2012
Re: K120995
Trade/Device Name: eRAD PACS/eRAD RIS/PACS Software Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: October 8, 2012 Received: October 9, 2012
Dear Ms. Reed:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2-Ms. Reed
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D. O'Hara
Janine M. Morris. M.S. Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### Indications for Use
510(k) Number (if known): #K120995
Device Name: eRAD PACS/ eRAD RIS/PACS Software
Indications for Use:
eRAD PACS/ eRAD RIS/PACS is a PACS software product used to receive DICOM images, scheduling information and textual report, organize and store them in an internal format, and to make that information available across a network via web and customized user interfaces.
The eRAD PACS/ eRAD RIS/PACS viewer software is intended for use as a primary diagnostic and analysis tool for diagnostic images. eRAD PACS/ eRAD RIS/PACS is for hospitals, imaging centers, radiologists, reading practices and any user who requires and is granted access to patient image, demographic and report information.
The eRAD PACS/ eRAD RIS/PACS viewer displays images from CT, computed radiography, MRI, mammography, nuclear medicine, PET, secondary capture, ultrasound, x-ray angiography, x-ray fluoroscopy and visible light modalities.
Lossy compressed mammography images and digitized film screen mammography images must not be reviewed for primary image interpretations. Mammography images may only be interpreted using an FDA approved monitor that offers at least 5 mega-pixel resolution and meets other technical specifications reviewed and accepted by FDA.
Prescription Use XX (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of in-vitro Diagnostics and Radiological Health (OIR)
Michael D'Okon
(Division Sign Off)
Division of Radiological Health
Ollice of In Vitro Diagnostics and Radiological Health
K120995
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.