← Product Code [OAS](/productcode/OAS) · K120948

# PREXION 3D (K120948)

_The Yoshida Dental Mfg. Co., Ltd. · OAS · Dec 21, 2012 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K120948

## Device Facts

- **Applicant:** The Yoshida Dental Mfg. Co., Ltd.
- **Product Code:** [OAS](/productcode/OAS.md)
- **Decision Date:** Dec 21, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1750
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

PreXion 3D is intended to produce two-dimensional and three-dimensional digital X-ray images of the dento-maxillo-facial regions at the direction of healthcare professionals as diagnostic support.

## Device Story

PreXion 3D is a dental computed tomography (CT) and panoramic X-ray system. It uses an X-ray tube head and flat panel detector to capture X-ray absorption data from multiple directions. The system includes a scanner, a console computer, and software for image processing and analysis. It generates 2D panoramic images and 2D/3D CT images of hard tissues (bones, teeth) in the jaw and mouth cavity. Operated by healthcare professionals in a clinical setting, the device provides diagnostic support through cross-section views and image editing. Clinicians use these images to visualize anatomical structures, aiding in diagnosis and treatment planning. The system benefits patients by providing high-resolution imaging for dental and maxillofacial assessment.

## Clinical Evidence

Bench testing only. No clinical data provided. Testing included Modulation-Transfer Function (MTF) per IEC 61223-3-5 to evaluate spatial resolution, metal artifact analysis compared to the FINECUBE predicate, and electrical/radiation safety testing per IEC 60601 series standards.

## Technological Characteristics

Computed tomography X-ray system; flat panel detector (FPD); X-ray tube head; console computer. Connectivity: standalone/networked. Software: console and viewer software for image reconstruction and analysis. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-7, IEC 60601-2-28, IEC 60601-2-32, IEC 61223-3-5, ISO 14971.

## Regulatory Identification

A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- FINECUBE ([K063622](/device/K063622.md))
- CS 9300 ([K103659](/device/K103659.md))

## Reference Devices

- PANOURA 18S ([K111231](/device/K111231.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# DEC 2 1 2012

## 510(k) Summary

- a. Owner/Company name, address THE YOSHIDA DENTAL MFG CO., LTD. 1-3-6, Kotobashi, Sumida-ku Tokyo 130-8516, Japan
Michizo Yamanaka President

· Contact person Hidenori Watanabe Regulatory Affairs Phone: 011-81-3-3631-2165 Fax: 011-81- 3-3633-9420 Email: hi-watanabe@yoshida-net.co.jp

## b. Contact/Application Correspondent

Fumiaki Kanai, Ph.D. President and CEO MIC International 4-1-17 Hongo, Bunkyo-ku Tokyo, 113-0033, Japan

011-81-3-3818-8577 Phone: 011-81-3-3818-8573 Fax: Email: kanaif(@mici.co.jp

- c. Date prepared March 28, 2012
## d. Name of device

PREXION 3D Trade Name: Computed tomography x-ray system Common Name: Classification Name: X-ray, tomography, computed, dental Classification Regulation: 21 CFR 892.1750

YOSHIDA DENTAL MFG CO., LTD. NOVEMBER 9, 2012

ATTACHMENT 1 A 1-12 OF A 1-17

{1}------------------------------------------------

## e. Predicate devices

The PREXION 3D is substantially equivalent to the following legally marketed device: .

| 510(k) Number | Trade name | Product code |
|---------------|------------|--------------|
| K063622       | FINECUBE   | OAS          |
| K103659       | CS 9300    | OAS          |

The predicate devices are hereinafter called "the FINECUBE (k063622)" or "CS 9300 (k103659)" in this application.

## f. Description of the device

The PREXION 3D consists of scanner and two software including Console software, and Viewer software used for the Image Analysis System and Data processing. A qualified computer named Console computer is distributed with the PREXION 3D.

The PREXION 3D uses the Image Analysis System and the processed data acquired by the scanner to analyze 2D and 3D images, perform image edition, such as creating cross-section views, and output results to a printer or other output device.

During scanning, X-rays are generated from the X-ray tube head mounted in the arm of the scanner and the X-rays passing through a patient are then detected by the flat panel detector of the scanner under the control of the firmware and the Console software installed on the qualified computer. The detected X-ray absorption data is processed by the Console software and viewer software on a computer to reconstruct images. Scanning is performed using X-ray penetration signals of a patient taken from multiple directions for the diagnosis of hard tissue including bones and teeth of the jaw and mouth cavity. The PREXION 3D performs two types of scanning including CT scan generating two (2) and three (3) dimensional images and Panoramic scan generating two (2) dimensional images.

## g. Indications for Use

The PREXION 3D is intended to produce two-dimensional and three-dimensional digital X-ray images of the dento-maxillo-facial region at the direction of healthcare professionals as diagnostic support.

### h. Statement of substantial equivalence

The PREXION 3D was developed from the FINECUBE (k063622) by adding panoramic scan function. Accordingly, the characteristics of the PREXION 3D are identical or similar to those of the FINECUBE (k063622) regarding X-ray Generation Device characteristics including tube voltage, tube current, and Focal Spot Size, X-ray image capturing device including type of detector, Pixel size, pixel number. (See Table 6-1 Comparison Table).

The only difference of the PREXION 3D from the FINECUBE (k063622) is addition of panoramic scanning function. The panoramic scanning function of the PREXION 3D is similar to CS 9300 (k103659). Both the PREXION 3D and CS 9300 (k103659) have the panoramic scanning function in addition to CT scan function. The PREXION 3D has the similar characteristics regarding intended use, X-ray Generation Device characteristics

YOSHIDA DENTAL MFG CO., LTD. NOVEMBER 9, 2012

ATTACHMENT 1 A1-13 OF A1-17

{2}------------------------------------------------

including tube voltage, tube current, and Focal Spot Size, X-ray image capturing device including type of detector, pixel number as CS 9300 (k103659).

By considering above device characteristics, the intended use of the PREXION 3D is identical to part of intended use of the CS 9300 (k103659), and similar to the intended use of the FINECUBE (k063622).

In order to ensure same performance characteristics as predicate devices, software verification/validation, performance testing, and risk analysis were performed. Such test results and risk analysis indicate that the PREXION 3D meets the requirements of the recognized consensus or voluntary standard. Based on the information presented above we conclude that the PREXION 3D is substantially equivalent to the predicate devices and does not raise any new questions regarding safety or effectiveness.

#### Comparison table i.

Table 6-1 compares the characteristics between the PREXION 3D and the predicates.

{3}------------------------------------------------

ON 3D (K120948)
ONSE TO FDA LETTER DATED MAY

x 149 mm max (Panoram x 1536 pixels (Panorar PD (TFT ર0 - 90k પ 2 - 15m/ 0.7mm l 4 bits .6mm x 123.2m 608 x 610 0.2mm 200μm l 2bits 90k V 4mA FPD Xion sional pisal managalad to provinsi
Adimensional manakal manakaran para na mana mana mana marka marin
sa of the villaga marka mana mana mana mana mana mana mana mana man 123.2mm x 121.6mm (CT)
mm x 131.6mm (Panoram 616 x 608 (CT)
216 x 72 (Panorami 200μm (CT)
0μm (Panorami 0.2mm 90k V 4bits 4mA FPD Tube Current
Tocal Spot Size Number of B ube Voltag ixel numbe Size of Area
receiving receiving
X-ray le 6-1. Comparison Ts ixel size Detector ications for l X ray Generatic
Device K ray image
capturing devi

SHIDA DENTAL MFG CO., 
VEMBER 9, 2012

ATTACHMENT
A I-15 OF A I-1

{4}------------------------------------------------

XION 3D (K120948)
K... SPONSE TO FDA LETTER DATED MAY 16, 2012

|                               | PREXION 3D                                                                 | FINECUBE (k063622)                                                                                | CS 9300 (k103659)                                                                     |                                                                                                                                                                                                                                                                                                    |
|-------------------------------|----------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Scanner                       | SID/SOD                                                                    |                                                                                                   |                                                                                       |                                                                                                                                                                                                                                                                                                    |
|                               | 700mm/468mm<br>(Hi-Def/Ultra Hi-Def/Rapid Scan)<br>700mm/322mm<br>(Hi-Res) | 700mm/322mm(Magnification<br>imaging)<br>700mm/468mm(Wide area<br>imaging)                        | 615mm (SID)                                                                           |                                                                                                                                                                                                                                                                                                    |
|                               | Dimension<br>(W×D×H)                                                       | 1170mm×1530mm×1930mm                                                                              | 1158mm×1595mm×2378mm                                                                  |                                                                                                                                                                                                                                                                                                    |
|                               | Weight                                                                     | 390kg                                                                                             | 390kg                                                                                 | 160kg                                                                                                                                                                                                                                                                                              |
| Scan mode                     |                                                                            | CT Scan and Panoramic Scan                                                                        | CT Scan                                                                               | CT Scan and Panoramic Scan                                                                                                                                                                                                                                                                         |
| Panoramic scan<br>Performance | Scan time                                                                  | 16.2 sec                                                                                          |                                                                                       | 4 - 16 sec                                                                                                                                                                                                                                                                                         |
|                               | Scan time                                                                  | 8.6 sec<br>(Rapid Scan)<br>16.8 sec<br>(Hi-Def/Hi-Res)<br>33.5 sec<br>(Ultra Hi-Def/Ultra Hi-Res) | 19sec. (Standard mode)<br>37sec. (High density mode)                                  | 12 - 20 sec<br>28 sec                                                                                                                                                                                                                                                                              |
| CT scan<br>Performance        |                                                                            | φ81mm/H74mm<br>(Hi-Def/Ultra Hi-Def/Rapid Scan)<br><br>φ56mm/H51mm<br>(Hi-Res/Ultra Hi-Res)       | Wide area imaging<br>φ82.0mm H75.1mm<br><br>Magnification imaging<br>φ56.5mm, H51.7mm | φ170mm/H135mm<br>(0.090mm - 0.500mm)<br><br>φ170mm/H110mm<br>(0.090mm-0.500mm)<br><br>φ170mm/H60mm<br>(0.090mm - 0.500mm)<br><br>φ100mm/H100mm<br>(0.090mm - 0.500mm)<br><br>φ80mm/H80mm<br>(0.090mm - 0.500mm)<br><br>φ100mm/H50mm<br>(0.090mm - 0.500mm)<br><br>φ50mm/H50mm<br>(0.090mm-0.500mm) |
|                               | FOV<br>(voxel size)                                                        |                                                                                                   |                                                                                       |                                                                                                                                                                                                                                                                                                    |

ATTACHMENT ]
A1-16 OF A1-17

YOSHIDA DENTAL MFG CO., LTD
NOVEMBER 9, 2012

{5}------------------------------------------------

#### j. Risk Analysis

The PREXION3D was evaluated in accordance with ISO14971:2007. The risk management of the device was deemed satisfactory.

### k. Bench Testing

THE YOSHIDA DENTAL MFG CO., LTD has performed bench tests to ensure safety and effectiveness as follows;

## 1. Laser safety

The laser system of the PREXION3D is identical to that of the PANOURA 18S (K111231). Therefore, the test report for IEC 60825-1 for the PANOURA 18S (K111231) is used as the test report for the PREXION3D.

## 2. Modulation-Transfer Function

In order to evaluate the spatial resolution of the PreXion3D, we measured the MTF in accordance with IEC 61223-3-5. The spatial resolution of all scan modes met the acceptance criteria.

## 3. Artifact Analysis

In order to evaluate the artifact of the image of the PREXION 3D, the images of all scan mode of the PREXION 3D were compared to those of the FineCube (K063622). There was no difference of pattern and strength of the metal artifact between the PREXION 3D and the FineCube(K063622).

The software of the PREXION3D has been validated according to "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."

EMC, Electric safety, and X-ray radiation safety are confirmed in accordance with IEC60601-1, IEC 60601-1-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-7, IEC 60601-2-28, and IEC 60601-2-32.

#### I. Conclusion

The PREXION 3D has similar intended use and technical characteristics as the predicate devices including the FINECUBE (k063622) and CS 9300 (k103659). A number of test results and risk analysis indicate that the PREXION 3D meets the requirements of the recognized consensus or voluntary standard. Based on those information, we conclude that the PREXION 3D is substantially equivalent to the predicate devices and does not raise any new questions regarding safety or effectiveness.

YOSHIDA DENTAL MFG CO., LTD. NOVEMBER 9, 2012

· ATTACHMENT 1 A1-17 OF A1-17

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure or a caduceus, rendered in black.

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

December 21, 2012

The Yoshida Dental MFG Co., LTD. % Dr. Fumiaki Kanai President and CEO MIC international 4-1-17 Hongo, Bunkyo-ku TOKYO, 113-0033, JAPAN

Re: K120948

Trade/Device Name:PREXION 3D Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-ray System Regulatory Class: Class II Product Code: OAS Dated: November, 09, 2012 Received: November 13, 2012

Dear Dr. Kanai:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{7}------------------------------------------------

Page 2 - Dr. Kanai

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostics and Radiological Health at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

## Janine M. Morris -S

Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

# Indications for Use

510(k) Number (if known): K120948

Device Name: PreXion 3d

Indications For Use:

PreXion 3D is intended to produce two-dimensional and three-dimensional digital X-ray images of the dento-maxillo-facial regions at the direction of healthcare professionals as diagnostic support.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Janine M. Morris -S 2012.12.21 10:48:57 -05'00'

Page 1 of

---

**Source:** [https://fda-staging.innolitics.com/device/K120948](https://fda-staging.innolitics.com/device/K120948)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
