K120946 · Immuno-Mycologics, Inc. · MDU · Dec 17, 2012 · Microbiology
Device Facts
Record ID
K120946
Device Name
ALPHA CRYPTOCOCCAL ANTIGEN EIA
Applicant
Immuno-Mycologics, Inc.
Product Code
MDU · Microbiology
Decision Date
Dec 17, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3165
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K120946 · Dec 17, 2012
ALPHA CRYPTOCOCCAL ANTIGEN EIA
Immuno-Mycologics, Inc.
Clinical specimens submitted for routine cryptococcal antigen testing at three sites (manufacturer and two national reference laboratories)
The clinical specimens were used to perform a split-sample method comparison study to evaluate the performance (sensitivity/specificity) of the new device against a predicate device and a lateral flow assay.
Three-site split-specimen comparison study; Retrospective method comparison study
Patients with serum and CSF specimens submitted for cryptococcal antigen testing; Sample Size: 1782 specimens (426 CSF, 1356 serum); Number of Sites: 3
Percent agreement positive, percent agreement negative, and overall percent agreement
Indications for Use
The ALPHA Cryptococcal Antigen enzyme immunoassay (CrAg EIA) is a qualitative or semi-quantitative (titration) test system for the detection of capsular polysaccharide antigens of Cryptococcus species complex (Cryptococcus neoformans and Cryptococcus gattii) in serum and cerebrospinal fluid (CSF). The ALPHA Cryptococcal Antigen Enzyme Immunoassay is an assay which can be used as an aid in the diagnosis of cryptococcosis. Test results are to be used in conjunction with information available from the patient clinical evaluation and other diagnostic procedures.
Device Story
Direct immunoenzymatic sandwich microplate assay; detects Cryptococcus capsular polysaccharide antigens in serum/CSF. Microwells coated with capture antibodies; patient specimen added/incubated; unbound material washed; HRP-conjugated detection antibodies added; TMB substrate added; blue color development proportional to antigen concentration. Reaction stopped; optical density (OD) measured via spectrophotometer at 450nm (reference 630nm). Used in clinical laboratories; operated by trained personnel. Output is qualitative (positive/negative) or semi-quantitative (titration). Results aid clinicians in diagnosing cryptococcosis in conjunction with patient evaluation.
Clinical Evidence
Bench testing only. Precision evaluated across 3 sites with 5 operators over 5 days. Analytical sensitivity (LoD) established at 5.3 ng/ml (serum) and 1.7 ng/ml (CSF). Specificity tested against 118 serum specimens and 15 fungal culture filtrates (3.9% cross-reactivity). Method comparison study (n=1782) against commercial EIA showed high agreement (Serum: 98.5% positive, 97.7% negative; CSF: 93.5% positive, 99.2% negative). ROC analysis (n=995) against IMMY LFA established 0.265 OD cutoff (97.4% sensitivity, 99.9% specificity).
Technological Characteristics
Direct immunoenzymatic sandwich microplate assay. Components: antibody-coated 96-well microplate, wash buffer, positive control, HRP-enzyme conjugate, TMB substrate, stop solution, sample diluent. Detection via microplate reader (450 nm). No automated software algorithm; manual/instrument-based optical density measurement.
Indications for Use
Indicated for qualitative or semi-quantitative detection of Cryptococcus neoformans and Cryptococcus gattii capsular polysaccharide antigens in serum and CSF as an aid in diagnosing cryptococcosis. For prescription use.
Regulatory Classification
Identification
Cryptococcus neoformans serological reagents are devices that consist of antigens used in serological tests to identify antibodies to Cryptococcus neoformans in serum. Additionally, some of these reagents consist of antisera conjugated with a fluorescent dye (immunofluorescent reagents) and are used to identify Cryptococcus neoformans directly from clinical specimens or from cultured isolates derived from clinical specimens. The identification aids in the diagnosis of cryptococcosis and provides epidemiological information on this type of disease. Cryptococcosis infections are found most often as chronic meningitis (inflammation of brain membranes) and, if not treated, are usually fatal.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 866.9.
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
K120946
B. Purpose for Submission:
To obtain substantial equivalence for an original 510(k) for a device which detects Cryptococcal antigen.
C. Measurand:
Capsular polysaccharide antigens of *Cryptococcus* species complex (*Cryptococcus neoformans* and *Cryptococcus gattii*)
D. Type of Test:
Qualitative and semi-quantitative enzyme immunoassay
E. Applicant:
Immuno-Mycologics, Inc.
F. Proprietary and Established Names:
ALPHA Cryptococcal Antigen EIA
G. Regulatory Information:
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| GMD | II | 866.3165 | 83-Microbiology |
H. Intended Use:
1. Intended use:
The ALPHA Cryptococcal Antigen enzyme immunoassay (CrAg EIA) is a qualitative or semi-quantitative (titration) test system for the detection of capsular polysaccharide antigens of *Cryptococcus* species complex (*Cryptococcus neoformans* and *Cryptococcus gattii*) in serum and cerebrospinal fluid (CSF). The ALPHA Cryptococcal Antigen Enzyme Immunoassay is an assay which can be used as an aid in the diagnosis of cryptococcosis. Test results are to be used in conjunction with information available from the patient clinical evaluation and other diagnostic procedures.
{1}
2. Indication for use:
The ALPHA Cryptococcal Antigen enzyme immunoassay (CrAg EIA) is a qualitative or semi-quantitative (titration) test system for the detection of capsular polysaccharide antigens of *Cryptococcus* species complex (*Cryptococcus neoformans* and *Cryptococcus gattii*) in serum and cerebrospinal fluid (CSF). The ALPHA Cryptococcal Antigen Enzyme Immunoassay is an assay which can be used as an aid in the diagnosis of cryptococcosis. Test results are to be used in conjunction with information available from the patient clinical evaluation and other diagnostic procedures.
3. Special conditions for use statement(s):
For prescription use only
4. Special instrument requirements:
Spectrophotometer microplate reader equipped with 450 nm and/or 450/630 nm filters
I. Device Description:
The ALPHA Cryptococcal Antigen Enzyme Immunoassay (EIA) is a direct immunoenzymatic sandwich microplate assay which detects Cryptococcus antigens in serum and CSF. Anti- Cryptococcus antibodies bound to microwell plates are used as capture antibodies, and horseradish peroxidase (HRP)-conjugated anti-Cryptococcus antibodies are used as detection antibodies. The positive control and standard curve material are composed of cryptococcal capsular polysaccharide antigen in a buffered protein solution with a preservative.
Either serum or CSF is added to the microwells coated with the capture antibodies and incubated. If the patient specimen contains cryptococcal antigens that are recognized by the capture antibodies, those antigens will become bound to the microwells. The microwells are washed to remove unbound patient material, and HRP-conjugated detect antibody is added to the wells. If Cryptococcus antigens are bound to the microwells by the capture antibodies, the detect antibody will also become bound to the microwells. The wells are then washed to remove any unbound detect antibody. Next, tetramethylbenzidine (TMB) substrate is added to the microwells, and in the presence of HRP, a blue color will develop. The reaction is stopped by the addition of a stop solution. The optical density (OD) is determined with a microplate reader at 450 nm with reference at 630 nm (reference is optional).
J. Substantial Equivalence Information:
1. Predicate device name(s):
Premier™ Cryptococcal Antigen EIA
{2}
2. Predicate 510(k) number(s): K904393
3. Comparison with predicate:
| SIMILARITIES | | |
| --- | --- | --- |
| Feature | Device: ALPHA Cryptococcal Antigen EIA | Predicate: Premier Cryptococcal Antigen EIA |
| Intended Use | Test for the qualitative or semi-quantitative detection of capsular polysaccharide antigens of Cryptococcus | Test for the qualitative or semi-quantitative detection of capsular polysaccharide antigens of Cryptococcus |
| Indication For Use | Prescription-use laboratory assay, which can aid in the diagnosis of cryptococcosis | Prescription-use laboratory assay, which can aid in the diagnosis of cryptococcosis |
| Assay Principle | Enzyme immunoassay | Enzyme immunoassay |
| Sample Matrix | Serum and CSF | Serum and CSF |
| Assay Components | Antibody coated 96-well microplate, wash buffer, positive control, enzyme conjugate, TMB substrate, stop solution, sample diluent | Antibody coated 96-well microplate, wash buffer, positive control, enzyme conjugate, TMB substrate, stop solution, sample diluent |
| Detection Chemistry | HRP + TMB | HRP + TMB |
| Instruments | Optical Density Microplate Reader | Optical Density Microplate Reader |
| Detection Antibody | Anti-cryptococcal monoclonal antibody | Anti-cryptococcal antibody |
| Specimen Pre-treatment | None | None |
| DIFFERENCES | | |
| --- | --- | --- |
| Feature | Device: ALPHA Cryptococcal Antigen EIA | Predicate: Premier Cryptococcal Antigen EIA |
| Output | Positive or Negative Only | Positive, Negative, or Indeterminate |
| Reagent Application | Pipette or equivalent | Reagent dropper |
| Optional Visual Read | No | Yes |
# K. Standard/Guidance Document Referenced (if applicable):
CLSI EP17-A2 Protocols for Determination of Limits of Detection and Limits of Quantitation; Approved Guideline.
# L. Test Principle:
In the qualitative procedure, specimens are analyzed undiluted. In the semi-
{3}
quantitative titration procedure, specimens are analyzed after serial dilution in specimen diluent. Either serum or CSF is added to the microwells coated with the capture antibodies and incubated. If the patient specimen contains cryptococcal antigens that are recognized by the capture antibodies, those antigens will become bound to the microwells. The microwells are washed to remove unbound patient material, and HRP-conjugated detect antibody is added to the wells. If Cryptococcus antigens are bound to the microwells by the capture antibodies, the detect antibody will also become bound to the microwells. The wells are then washed to remove any unbound detect antibody. Next, tetramethylbenzidine (TMB) substrate is added to the microwells, and in the presence of HRP, a blue color will develop. The reaction is stopped by the addition of a stop solution. The optical density (OD) is determined with a microplate reader at $450\mathrm{nm}$ with reference at $630\mathrm{nm}$ (reference is optional).
# M. Performance Characteristics (if/when applicable):
# 1. Analytical performance:
# a. Precision/Reproducibility:
The ALPHA Cryptococcal Antigen EIA was evaluated for reproducibility and precision by spiking serum and mock CSF with cryptococcal antigen to produce a panel consisting of a negative sample, a high negative (C5) sample, a low positive sample and a moderate positive sample. This panel was tested twice per day at three sites (Immuno-Mycologics, Inc., a national reference lab, and a clinical lab) with a total of five operators over a five-day period in order to determine both the inter-lab and the intra-lab reproducibility and precision of the assay.
The percent coefficient of variation (CV) for the positive CSF specimens ranged from $6.0\%$ to $9.4\%$ at the individual sites. The percent CV of the CSF results averaged over all sites ranged from $21.1\%$ to $23.3\%$ . The percent coefficient of variation for the positive serum specimens ranged from $7.7\%$ to $13.6\%$ at the individual sites. The percent CV of the serum results averaged over all sites ranged from $13.05\%$ to $14.16\%$ .
The study results are listed below.
| | | IMMY | | | National Reference Lab | | | Clinical Lab | | | Combined Data (3 Sites) | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Description | Type | Ave O.D. | Std Dev | % CV | Ave O.D. | Std Dev | % CV | Ave O.D. | Std Dev | % CV | Ave O.D. | Std. Dev. | % CV |
| Blank | Control | 0.066 | 0.012 | 17.8 | 0.000 | 0.0027 | 3686.2 | 0.075 | 0.005 | 7.0 | 0.046 | 0.034 | 74.67 |
| CRYPC1 | Control | 1.814 | 0.078 | 4.3 | 1.973 | 0.1083 | 5.5 | 2.473 | 0.158 | 6.4 | 2.059 | 0.295 | 14.34 |
| Negative | Serum | 0.022 | 0.011 | 48.7 | 0.012 | 0.0058 | 48.8 | 0.019 | 0.005 | 25.4 | 0.018 | 0.009 | 50.25 |
| High Negative | Serum | 0.040 | 0.008 | 19.2 | 0.028 | 0.0077 | 27.7 | 0.035 | 0.007 | 20.0 | 0.034 | 0.009 | 26.51 |
| Low Positive | Serum | 0.420 | 0.041 | 9.7 | 0.338 | 0.0436 | 12.9 | 0.372 | 0.038 | 10.3 | 0.377 | 0.053 | 14.16 |
| Moderate Positive | Serum | 1.683 | 0.229 | 13.6 | 1.666 | 0.1281 | 7.7 | 1.959 | 0.205 | 10.5 | 1.756 | 0.229 | 13.05 |
| Negative | CSF | 0.065 | 0.018 | 27.8 | 0.058 | 0.0087 | 15.1 | 0.099 | 0.010 | 10.0 | 0.072 | 0.022 | 29.85 |
| High Negative | CSF | 0.179 | 0.020 | 11.1 | 0.174 | 0.0190 | 10.9 | 0.298 | 0.027 | 9.2 | 0.211 | 0.059 | 28.09 |
{4}
The overall percent positive and percent negative result reproducibility was calculated by combining the data from all three sites. The reproducibility of results from low positive, moderate positive and negative samples was 100%.
| Serum | Site 1 % Positive | Site 2 % Positive | Site 3 % Positive | Overall % Positive |
| --- | --- | --- | --- | --- |
| Negative | 0% | 0% | 0% | 0% |
| High Negative | 0% | 0% | 0% | 0% |
| Low Positive | 100% | 100% | 100% | 100% |
| Moderate Positive | 100% | 100% | 100% | 100% |
| CSF | Site 1 % Positive | Site 2 % Positive | Site 3 % Positive | Overall % Positive |
| --- | --- | --- | --- | --- |
| Negative | 0% | 0% | 0% | 0% |
| High Negative | 0% | 0% | 83% | 24% |
| Low Positive | 100% | 100% | 100% | 100% |
| Moderate Positive | 100% | 100% | 100% | 100% |
b. Linearity/assay reportable range: Not applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Controls:
The Positive Control and Negative Control should be assayed with each batch of patient specimens to provide quality assurance of the reagents. The Positive and Negative controls are intended to monitor for substantial reagent failure.
| Reagent | Acceptable Blanked OD Parameters |
| --- | --- |
| Positive Control | > 0.265 |
| Negative Control | < 0.090 |
If the blanked OD of the Positive Control is not within these parameters, patient test results should be considered invalid and the assay should be repeated.
High Dose Hook Effect:
The effects of high doses were determined by spiking a serum specimen pool that was negative by the IMMY Latex-Cryptococcus Antigen Detection System and ALPHA Cryptococcal Antigen EIA, with cryptococcal antigen at 1 mg/ml and testing it in triplicate at IMMY on one lot of ALPHA Cryptococcal Antigen EIA, according to the package insert.
It was observed that at extremely high concentrations, the signal can be
{5}
reduced due to the high dose hook effect. Although the signal is reduced leading to lower than expected EIA values, the result is still positive. The semi-quantitative titration procedure can differentiate between a low EIA score due to the high dose hook effect and a low EIA score due to low concentrations of Cryptococcal Antigen.
## Sample stability:
Three CrAg-Negative human CSF specimens and four CrAg-Negative human serum specimens were spiked with concentrations of cryptococcal antigen near and slightly above the cutoff of the assay. The specimens were tested on the ALPHA CrAg EIA immediately after spiking and then they were frozen at $< -20^{\circ}\mathrm{C}$ for the Day 0 time point. The specimens were then divided into three aliquots and tested as follows:
- Aliquot 1 – Tested daily in triplicate, subjected to a freeze-thaw cycle each day
- Aliquot 2 – Frozen for one week, thawed and tested in triplicate
- Aliquot 3 – Frozen for two weeks, thawed and tested in triplicate
A reduction in OD values was observed after multiple freeze-thaw cycles and after storage at $< -20^{\circ}\mathrm{C}$ for one week. The OD value reductions were large enough to cause some very low positive samples (near 0.300 Blanked OD) to become negative when stored frozen for one week or subjected to more than two freeze-thaw cycles. Specimens should be tested as soon as possible after collection and multiple freeze-thaw cycles should be avoided.
## d. Detection limit/Analytical cut-off:
The limit of detection (LoD) was estimated at IMMY according to Clinical and Laboratory Standards Institute (CLSI) EP-17A. The limit of the blank (LoB) was estimated by running 80 replicates of normal human serum and 80 replicates of artificial CSF. The initial LoD was estimated by running 20 replicates of 4 concentrations near the LoB in both serum and CSF.
The LoD was established by testing 20 replicates of both serum and CSF spiked with Cryptococcal antigen at a range of concentrations near the initial LoD. Results were considered positive if they yielded a blanked OD that was greater than the cut-off.
Serum data:
| Concentration | % Positive |
| --- | --- |
| 4.9 ng/mL | 0% (0/24) |
| 5.0 ng/mL | 8% (2/24) |
| 5.1 ng/mL | 54% (13/24) |
| 5.2 ng/mL | 71% (17/24) |
| 5.3 ng/mL | 100% (24/24) |
{6}
CSF data:
| Concentration | % Positive |
| --- | --- |
| 1.3 ng/mL | 0% (0/24) |
| 1.4 ng/mL | 0% (0/24) |
| 1.5 ng/mL | 4% (1/24) |
| 1.6 ng/mL | 83% (20/24) |
| 1.7 ng/mL | 100% (24/24) |
Analysis shows that the LoD for serum is $5.3\mathrm{ng / ml}$ and the LoD for CSF is $1.7\mathrm{ng / ml}$ .
# e. Analytical specificity:
# Cross reactivity:
Analytical specificity for the ALPHA Cryptococcal Antigen EIA was determined by running specimens from potentially cross-reacting medical conditions unrelated to cryptococcosis. The following specimens were run on one lot of the ALPHA Cryptococcal Antigen EIA. A total of 118 serum specimen and 15 fungal culture filtrates were tested. Culture filtrates were tested at three different dilutions: undiluted, 1:10, and 1:100. Dilutions were made in 1X Specimen Diluent. Percent positive was determined for each condition.
| Pathology | Number of Specimens | Number of Positives | % Positive |
| --- | --- | --- | --- |
| HAMA | 5 | 0 | 0 |
| Syphilis | 10 | 0 | 0 |
| Rubella | 5 | 0 | 0 |
| Mycoplasma | 10 | 0 | 0 |
| Toxoplasmosis | 7 | 0 | 0 |
| CMV Infection | 10 | 0 | 0 |
| Rheumatoid factor | 10 | 0 | 0 |
| Penicilliosis | 5 | 0 | 0 |
| Sporotrichosis | 6 | 0 | 0 |
| Blastomycosis | 10 | 0 | 0 |
| Coccidioidomycosis | 10 | 0 | 0 |
| Histoplasmosis | 10 | 0 | 0 |
| Candidiasis | 10 | 0 | 0 |
| Aspergillus GM+ | 10 | 1 | 10 |
{7}
Culture filtrate results:
| Organism | Number of Specimens | Number of Positives | % Positive |
| --- | --- | --- | --- |
| Aspergillus terreus | 3 | 0 | 0 |
| Aspergillus niger | 3 | 0 | 0 |
| Aspergillus flavus | 3 | 0 | 0 |
| Aspergillus fumigatus | 3 | 0 | 0 |
| Paracoccidioides brasiliensis | 3 | 2 | 67 |
As expected, some fungal cross-reactivity was observed. One *Aspergillus galactomannan* positive specimen (positive by Bio-Rad’s Platelia™ Aspergillus EIA) was positive in the ALPHA Cryptococcal Antigen EIA and *Paracoccidioides brasiliensis* culture filtrates undiluted and 1:10 were positive as well. The total percent positive for fungal pathologies was 3.9%, well below the acceptance criteria of 10%. However, *Paracoccidioides brasiliensis* culture filtrate had a total percent positive of 67%.
Rheumatoid factor is known to cause false-positive results in the Cryptococcal latex agglutination method. Rheumatoid factor did not cause false positives in the ALPHA Cryptococcal Antigen EIA between the range of 112 IU/ml and 6479 IU/ml. The normal reference range for rheumatoid factor is 0-14 IU/ml.
## Interfering substances:
Interference testing was performed on five icteric, five hemolyzed, and five lipemic serum specimens. Each specimen was spiked with cryptococcal antigen at three times the C₉₅ concentration. All specimens were then tested at IMMY on one lot of the ALPHA Cryptococcal Antigen EIA in triplicate: spiked and unspiked. Percent positivity was determined for each condition. All of the unspiked specimens had negative results on the ALPHA Cryptococcal Antigen EIA. All spiked specimens were positive, thus, these types of serum specimens do not interfere with the ALPHA Cryptococcal Antigen EIA.
## f. Assay cut-off:
The assay cutoff was determined through Receiver Operator Curve (ROC) analysis using the IMMY CrAg Lateral Flow Assay (LFA) as the means of classifying a specimen as a true positive or negative. ROC Analysis was performed on a dataset consisting of 995 combined serum and CSF specimens
ROC Analysis suggested a blanked OD cutoff value of greater than 0.265 which yielded a positive percent agreement of 97.4% and a negative percent agreement of 99.9% when compared to the IMMY CrAg LFA. This cutoff
{8}
was confirmed through the method comparison comparing the results from the IMMY Cryptococcal Antigen EIA to the predicate device.
Assay cut-offs
| x ≤ 0.265 | Negative |
| --- | --- |
| x > 0.265 | Positive |
# 2. Comparison studies:
# a. Method comparison with predicate device:
A three-site (IMMY and two national reference laboratories) split specimen comparison study was performed on both serum and CSF specimens that had been submitted for cryptococcal antigen EIA testing. A total of 1782 specimens (CSF $n = 426$ ; serum $n = 1356$ ) were tested in both the ALPHA Cryptococcal Antigen EIA and a commercial Cryptococcal antigen EIA according to their respective package inserts. The resulting data from the split sample comparison is shown in the tables below.
Serum 2x2 Contingency Table
| | Other Manufacturer's EIA | | |
| --- | --- | --- | --- |
| | | Pos | Neg |
| IMMY ALPHA | Pos | 131 | 28 |
| CrAg EIA | Neg | 2 | 1195 |
Serum Data Analysis
| | Calculated | 95% CI |
| --- | --- | --- |
| % Positive Agreement | 98.5% (131/133) | 94.7 - 99.6% |
| % Negative Agreement | 97.7% (1195/1223) | 96.7 - 98.4% |
CSF 2x2 Contingency Table
| | Other Manufacturer's EIA | | |
| --- | --- | --- | --- |
| | | Pos | Neg |
| IMMY ALPHA | Pos | 29 | 3 |
| CrAg EIA | Neg | 2 | 392 |
CSF Data Analysis
| | Calculated | 95% CI |
| --- | --- | --- |
| % Positive Agreement | 93.5% (29/31) | 79.3 - 98.2% |
| % Negative Agreement | 99.2% (392/395) | 97.8 - 99.7% |
{9}
10
Supplemental studies:
A multi-site split-specimen comparison study was performed on both serum and CSF specimens that had been submitted for cryptococcal antigen EIA testing. A total of 1782 specimens (CSF n=426; serum n=1356) were tested in both the ALPHA Cryptococcal Antigen EIA and the IMMY Cryptococcal Antigen Lateral Flow Assay (LFA) according to their respective package inserts. Analysis of this data yielded percent positive agreements of 96.9% and 93.8% in serum and CSF, respectively. Analysis indicated percent negative agreements of 99.8% and 99.5% in serum and CSF, respectively.
b. Matrix comparison:
See sections M1a and M1d for comparisons of assay Precision, Reproducibility, and detection limits in serum and CSF matrices.
3. Clinical studies:
a. Clinical sensitivity: Not applicable
b. Clinical specificity: Not applicable
c. Other clinical supportive data (when a. and b. are not applicable): See M2a method comparison with predicate device
4. Clinical cut-off:
Not Applicable
5. Expected values/Reference range:
The frequency of cryptococcosis is dependent on several factors including: patient population, type of institution, and epidemiology. In these studies, the IMMY Alpha Cryptococcal Antigen EIA exhibited 97.6% agreement positive and 98.1% agreement negative with another manufacturer's EIA (serum and CSF combined).
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The information submitted in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.