The Constant Care LILAH Home Health Monitoring System is a non-emergency, delayed, remote patient monitoring device available by prescription. The Constant Care LILAH Home Health Monitoring System is designed to be used in the home of a patient or healthcare setting to provide remote monitoring for maintenance of chronic disease. The Constant Care LILAH Home Health Monitoring System provides guidance in operating medical sensor devices: reminders for medication compliance and connectivity to healthcare professionals through text messaging and realtime video conferencing technology. The Constant Care LILAH Home Health Monitoring System connectally available wireless and wired glucose meters, weight scales, blood pressure monitors, medication reminders, and pulse oximeters. The Constant Care LILAH Home Health Monitoring System stores and displays the information on a computer screen and transmits the information to the Caretaker Portal secure host server using connectivity including, but not limited to, FCC approved Wi-Fi, Cellular Wireless, Internet, or Ethernet. Healthcare professionals and remote caregivers can review the transmitted information utilizing the Caretaker Portal and set thresholds to trigger non-emergency alerts based on specific thresholds being exceeded. The Constant Care LILAH Home Health: Monitoring System is not intended for diagnosis or as a substitute for medical care, and it is not intended to provide real, time data: The data is made available to the patients when timecritical care is not required. The device is contraindicated for patients requiring direct medical or emergency intervention. The system is intended for patients who are willing and capable of managing its use. Judgment and experience by a caregiver or by a family member are required to check and interpret the information delivered.
Device Story
System consists of patient-side software (installed on touchscreen PC or tablet) and caregiver-side software (Caretaker Portal server). Inputs: data from wireless/wired medical devices (glucose meters, scales, BP monitors, pulse oximeters). Operation: system guides patient in sensor use, provides medication reminders, and facilitates text/video communication with caregivers. Output: patient data displayed on screen and transmitted to secure portal for remote review. Used in home or clinical settings by patients and healthcare professionals. Caregivers set thresholds for non-emergency alerts. Benefits: supports chronic disease maintenance via remote connectivity and compliance reminders; not for real-time or emergency care.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Software-based remote monitoring system. Patient-side hardware: Asus EeeTop PC ET1611PUT or Acer ICONIA Tab W500-C52G03iss. Connectivity: Wi-Fi, cellular, internet, Ethernet. Data transmission to secure host server. Threshold-based alert system.
Indications for Use
Indicated for patients with chronic disease requiring non-emergency, delayed remote monitoring. Contraindicated for patients requiring direct medical or emergency intervention. Intended for patients willing and capable of managing system use.
Regulatory Classification
Identification
A radiofrequency physiological signal transmitter and receiver is a device used to condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g., a central monitoring station. The received signal is reconditioned by the device into its original format so that it can be displayed.
Predicate Devices
Honeywell HomMed Genesis Touch Personal Health Device (K112858)
Submission Summary (Full Text)
{0}------------------------------------------------
K120941
## 510(k) Summary
JUL 3 2012
| Submitted by | Constant Care, LLC |
|-----------------|------------------------------------|
| Address: | 4333 Shreveport Highway |
| | Pineville, LA 71360 |
| Telephone: | (318) 861-0497 |
| Facsimile: | (866) 551-6030 |
| Contact Name: | Jodi Funderburk, VP of Development |
| Date Submitted: | March 27, 2012 |
| Trade Name: | Constant Care LILAH Home Health Monitoring System |
|--------------|---------------------------------------------------|
| Common Name: | Remote patient monitoring system |
Product Code / Regulation: DRG (21 C.F.R. 870.2910)
The Constant Care LILAH Home Health Monitoring System also supports the following medical device product codes by means of separate medical devices:
| Classification Regulation | Product Code | Classification Name | Device Class |
|---------------------------|--------------|--------------------------------------------------|--------------|
| 21 C.F.R. 870.1130 | DXN | Noninvasive Blood Pressure<br>Measurement System | II |
| 21 C.F.R. 880.2700 | FRI | Stand-on Patient Scale | I |
| 21 C.F.R. 870.2700 | DQA | Oximeter | II |
| 21 C.F.R. 862.1345 | NBW | Glucose Test System | II |
| 21 C.F.R. 890.5050 | NXQ | Daily Activity Assist Device | I |
The Constant Care LILAH Home Health Monitoring System consists of two Descrintion: components: patient-side software to be installed on either an Asus EeeTop PC ET1611PUT 15.6 inch TouchScreen Atom D425 All-in-One PC, or an Acer ICONIA Tab W500-C52G03iss 10.1' LED Tablet PC for use in performing remote patient monitoring, and sending data to the Caretaker Portal secure host server; and caregiver-side software for receiving patient-side data and transmitting that data to healthcare professionals and remote caregivers through the Caretaker Portal secure host server.
> The Constant Care LILAH Home Health Monitoring System is a remote patient monitoring device available by prescription designed to be used in the home or healthcare setting of a patient undergoing remote monitoring for maintenance of chronic disease. The Constant Care LILAH Home Health Monitoring System provides guidance in operating medical sensor devices, reminders for medication compliance and connectivity to healthcare professionals through text messaging and real-time video conferencing technology.
Intended Use: The Constant Care LILAH Home Health Monitoring System is intended for use in the home or healthcare setting to provide non-emergency, delayed, remote monitoring of a patient for maintenance of chronic disease; connecting to
{1}------------------------------------------------
commercially available wireless and wired medical devices such as glucose meters, weight scales, blood pressure monitors, medication reminders, and pulse oximeters; and providing guidance in operating medical sensor devices, reminders for medication compliance, and connectivity to healthcare professionals and remote caregivers through text messaging and real-time video conferencing technology.
## Substantial Equivalence:
The Constant Care LILAH Home Health Monitoring System is similar in intended use and technological characteristics to other Remote Patient Monitoring Devices, including the Honeywell HomMed Genesis Touch™ Personal Health Device (K112858).
Results of software validation and verification activities establish the device as safe and effective for its intended use, which is comparable to other predicate devices.
{2}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
> JUL 3 2012
Constant Care, LLC C/O Seth A. Mailhot Sheppard Mullin Richter & Hampton LLP 1300 I ST NW, 11th Floor East Washington, DC 20005
Re: K120941
Trade/Device Name: Constant Care LILAH Home Health Monitoring System Regulation Number: 21 CFR 870.2910 Regulation Name: Radiofrequency physiological signal transmitter and receiver Regulatory Class: Class II Product Codes: DRG, DXN, FRI, DQA, FLL, NBW, NXQ Dated: May 29, 2012 Received: May 30, 2012
Dear Mr. Mailhot:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{3}------------------------------------------------
#### Page 2 - Mr. Mailhot
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Bram Zuckerman, M.D.
Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known):
Device Name: Constant Care LILAH Home Health Monitoring System
#### Intended Use:
The Constant Care LILAH. Home Health Monitoring System is a non-emergency, delayed, remote patient monitoring device available by prescription. The Constant Care LILAH Home Health Monitoring System is designed to be used in the home of a patient or healthcare setting to provide remote monitoring for maintenance of chronic disease. The Constant Care LILAH Home Health Monitoring System provides guidance in operating medical sensor devices: reminders for medication compliance and connectivity to healthcare professionals through text messaging and realtime video conferencing technology.
The Constant Care LILAH Home Health Monitoring System connectally available wireless and wired glucose meters, weight scales, blood pressure monitors, medication reminders, and pulse oximeters. The Constant Care LILAH Home Health Monitoring System stores and displays the information on a computer screen and transmits the information to the Caretaker Portal secure host server using connectivity including, but not limited to, FCC approved Wi-Fi, Cellular Wireless, Internet, or Ethernet.
Healthcare professionals and remote caregivers can review the transmitted information utilizing the Caretaker Portal and set thresholds to trigger non-emergency alerts based on specific thresholds being exceeded.
The Constant Care LILAH Home Health: Monitoring System is not intended for diagnosis or as a substitute for medical care, and it is not intended to provide real, time data: The data is made available to the patients when timecritical care is not required. The device is contraindicated for patients requiring direct medical or emergency intervention. The system is intended for patients who are willing and capable of managing its use. Judgment and experience by a caregiver or by a family member are required to check and interpret the information delivered.
A list of devices that are compatible with the Constant Care LILAH Home Health Monitoring System will be available in the user's manual and the Constant Care website.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Cardiovascular Devices
Page 1 of 1
120941 510(k) Number
4-2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.