CHG ANTIBACTERIAL FOAM DRESSING, CHG ANTIBACTERIAL PATCH, CHG ANTIBACTERIAL FOAM PATCH, CHG PATCH, CHG DRESSING

K120930 · Medtrade Products , Ltd. · FRO · Oct 26, 2012 · SU

Device Facts

Record IDK120930
Device NameCHG ANTIBACTERIAL FOAM DRESSING, CHG ANTIBACTERIAL PATCH, CHG ANTIBACTERIAL FOAM PATCH, CHG PATCH, CHG DRESSING
ApplicantMedtrade Products , Ltd.
Product CodeFRO · SU
Decision DateOct 26, 2012
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic

Indications for Use

Under the supervision of a healthcare professional CHG Antibacterial Foam Patch is intended for use as: A hydrophilic foam patch that is used to absorb exudates and to cover the peri-wound area of a wound caused by the use of vascular and non-vascular percutaneous medical devices such as: IV catheters, central venous lines, arterial catheters, dialysis catheters, midline catheters, drains, chest tubes, externally placed orthopaedic pins, and epidural catheters.

Device Story

CHG Antibacterial Foam Patch is a hydrophilic polyurethane foam dressing containing chlorhexidine gluconate (CHG). Designed to absorb exudate and cover peri-wound areas around percutaneous medical devices (catheters, drains, pins). Applied by healthcare professionals in clinical settings. CHG agent inhibits/kills microorganisms on dressing surface. Benefits include microbial colonization reduction at insertion sites. Device is provided in various disk sizes with center holes and radial slits for secure placement around medical devices.

Clinical Evidence

No clinical data. Evidence consists of in-vitro antibacterial testing demonstrating Log 4 reduction of S. aureus (MRSA), E. coli, P. aeruginosa, and E. faecalis over 7 days, alongside biocompatibility testing per BS EN ISO 10993-1.

Technological Characteristics

Hydrophilic polyurethane foam with chlorhexidine gluconate (CHG) and polyurethane backing. Disk form factor with center hole and radial slit. Biocompatibility per BS EN ISO 10993-1. Risk analysis per BS EN ISO 14971. Sterilized per harmonized standards.

Indications for Use

Indicated for patients requiring exudate absorption and peri-wound coverage for sites of vascular and non-vascular percutaneous medical devices (e.g., IV, central venous, arterial, dialysis, midline, epidural catheters, drains, chest tubes, orthopaedic pins). Use under healthcare professional supervision.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Page 1 of 3 K120930 ## Section 5 – Traditional 510(k) Notification:- Summary This Traditional 510(k) notification is to provide substantial equivalence for the Medtrade Products Ltd CHG Antibacterial Foam Patch, which is substantially equivalent to currently marketed devices intended for wound care. OCT 2 6 2012 | Submitted by:- | Medtrade Product Ltd<br>Electra House, Crewe Business Park<br>Crewe, Cheshire<br>CW1 6GL<br>United Kingdom | |-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------| | Contact:- | Mrs Claire Ryan<br>Head of Regulatory & QA<br>Telephone: + 44(0)1270 500019<br>Fax:- + 44(0)1270 500045<br>Email: claire.ryan@medtrade.co.uk | | Date prepared:- | 26th March 2012 | | Proprietary Name: | CHG Antibacterial Foam Patch | | Common Name:- | CHG Antibacterial Foam Dressing<br>CHG Antibacterial Patch<br>CHG Antibacterial Dressing<br>CHG Patch<br>CHG Dressing | | Trade Names:- | Not yet defined | | Classification Name:- | Dressing, Wound, Drug | | Classification:- | Unclassified | | Product Code:- | FRO | Legally marketed device(s) for substantial equivalence comparison:-BIOPATCH, 510(k) # K003229, manufactured by Johnson and Johnson Wound Management. #### Device Description :- The Medtrade Product Ltd CHG Antibacterial Foam Patch consists of a hydrophilic Polyurethane absorbent foam with chlorhexidine gluconate (CHG) and a thin Polyurethane backing. The foam material absorbs up to eight times its own weight in fluid. The CHG present in the dressing inhibits or kills microorganisms on the surface of the dressing. {1}------------------------------------------------ posi 2 of 3 CHG Antibacterial Patch is available in the following sizes :-1" DISK (2.5cm) 4.0mm center hole with radial slit. %" DISK (1.9cm) 1.5mm center hole with radial slit. 1" DISK (2.5cm) 7.0mm center hole with radial slit. CHG Antibacterial Patches are packaged in individual pouches, then into shelf cartons with an instruction for use. #### Indications for use: Under the supervision of a healthcare professional CHG Antibacterial Foam Patch is intended for use as a hydrophilic foam patch that is used to absorb exudates and to cover the peri-wound area of a wound caused by the use of vascular and non-vascular percutaneous medical devices such as: IV catheters, central venous lines, arterial catheters, dialysis catheters, midline catheters, drains, chest tubes, externally placed orthopaedic pins, and epidural catheters. #### In-vitro Testing :- In-vitro Antibacterial data has shown that CHG Antibacterial Foam Patch reduces or inhibits microbial colonization in the dressing. In vitro CHG Antibacterial Foam Patch has demonstrated a Log 4 reduction and has been shown to kill the following organisms, over a 7 day period. Staphylococcus aureus (MRSA) NCTC 13142 – Gram Positive Bacterium Escherichia coli ATCC 8739 - Gram Negative Bacterium Pseudomonas aeruginosa ATCC 9027 - Gram Negative Bacterium Enterococcus faecalis NCTC 12201 – Gram Positive Bacterium Staphylococcus aureus ATCC 6538 - Gram Positive Bacterium #### Manufacturing :- CHG Antibacterial Foam Patch is manufactured according to the product specifications and under good manufacturing practices (GMP). A risk analysis has been performed in accordance with BS EN ISO 14971 to identify possible failure mode during manufacturing and design. Manufacturing controls have been developed and implemented to address the identified risk factors based on the criticality of the failure mode. Medtrade Product Ltd. CHG Antibacterial Foam Patch meets all the established specifications prior to release to ensure the device is safe, effective and correctly labeled for its intended use. {2}------------------------------------------------ #### Testing: - Performance data for the CHG Antibacterial Foam Patch have been established using antibacterial and bench testing. The biocompatibility of CHG Antibacterial Foam Patch has demonstrated to be in compliance with the requirements of BS EN ISO 10993-1 (Biological Evaluation of Medical Devices). Sterilisation has been performed in compliance with harmonised standards. The biocompatibility and performance testing for the CHG Antibacterial Foam Patch have demonstrated that the device is safe and effective for the indications of use. ### Statement of Substantial Equivalence:- The indication for use, performance testing and antibacterial activity for the CHG Antibacterial Foam Patch are substantially equivalent to the predicate device; BIOPATCH, 510(k) # K003229, cleared October 26, 2001, manufactured by Johnson & Johnson, as both products are absorbent polyurethane foams with the addition of CHG as an antibacterial. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Medtech Products, Limited % Ms. Clare Ryan Head of Regulatory and Quality Affairs Electra House, Crewe Business Cheshire, United Kingdom CWI 6GL Re: K120930 Trade/Device Name: CHG Antibacterial Foam Patch Regulatory Class: Unclassified Product Code: FRO Dated: October 12, 2012 Received: October 16, 2012 Dear Ms. Ryan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. OCT 26 2012 {4}------------------------------------------------ Page 2 - Ms. Clare Ryan This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Mark A. Millan Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ page 1 of 1 # Indications for Use 510(k) Number (if known): Device Name: CHG Antibacterial Foam Patch Indications for Use: Under the supervision of a healthcare professional CHG Antibacterial Foam Patch is intended for use as: A hydrophilic foam patch that is used to absorb exudates and to cover the peri-wound area of a wound caused by the use of vascular and non-vascular percutaneous medical devices such as: IV catheters, central venous lines, arterial catheters, dialysis catheters, midline catheters, drains, chest tubes, externally placed orthopaedic pins, and epidural catheters. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) । প্ (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) David Krone for MMM (Division Sign-Off) Division of Surgical. Orthoped and Restorative Devices 510(k) Number K120930
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...